K230660 · Cutera, Inc. · GEX · Jun 6, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230660
Device Name
AviClear Laser System
Applicant
Cutera, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The AviClear Laser System is indicated for the long-term treatment of mild to severe inflammatory acne vulgaris.
Device Story
The AviClear Laser System is an infrared diode laser (1726 nm) used in dermatology clinics by trained operators. The device targets sebum within sebaceous glands to reduce sebum production and improve acne. It features a console with a touchscreen interface and a handpiece with an integrated scanner, sapphire cooling window (0°C to 5°C), and skin-contact pressure sensors. The operator selects treatment parameters via the touchscreen; the system delivers laser energy in a pattern while the cooling window protects the epidermis and superficial dermis. Laser activation requires both footswitch depression and verified skin contact. The thermal injury to sebaceous glands provides a therapeutic effect for acne vulgaris. The device is a non-sterile, reusable system.
Clinical Evidence
Prospective, multi-center clinical study (n=104; 57% female, 43% male; age 16-40). Primary endpoint: Treatment success (≥50% reduction in inflammatory lesions) at 12 weeks post-final treatment. Long-term effectiveness: 91.5% of subjects achieved treatment success at 52 weeks (p<0.001). Secondary endpoints: 66.2% of subjects achieved 'Clear' or 'Almost Clear' skin at 52 weeks; non-inflammatory lesion counts reduced by 57.6% at 52 weeks (p<0.001). Safety: No SAEs or SUADEs reported. Common AEs included transient erythema, edema, mild acne purging, and dryness. Average discomfort score 5.23 ± 1.15.
Technological Characteristics
Infrared diode laser, 1726 nm wavelength. Quasi-CW output mode. Max fluence 30 J/cm² (single pulse) or 20 J/cm² (double pulse). Max pulse energy 5 J. Pulse duration up to 50 ms. Sapphire contact cooling window (0°C to 5°C). Optical fiber delivery. Touchscreen interface. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for long-term treatment of mild to severe inflammatory acne vulgaris in patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K251149 — AviClear Laser System · Cutera, Inc. · Jul 11, 2025
K213461 — AviClear Laser System · Cutera, Inc. · Mar 24, 2022
K242035 — Accure Laser System · Accure Acne, Inc. · Oct 11, 2024
K030834 — CANDELA SMOOTHBEAM LASER SYSTEM · Candela Corp. · Dec 3, 2003
K024206 — AURA ISURGICAL LASER SYSTEM & ACCESSORIES · Laserscope · Nov 17, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
June 6, 2023
Cutera, Inc. Amogh Kothare VP, Clinical and Regulatory Affairs 3240 Bayshore Blvd. Brisbane, California 94005
Re: K230660
Trade/Device Name: AviClear Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 10, 2023 Received: March 10, 2023
Dear Amogh Kothare:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230660
Device Name
AviClear Laser System
#### Indications for Use (Describe)
The AviClear Laser System is indicated for the long-term treatment of mild to severe inflammatory acne vulgaris.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K230660 510(k) Summary
This 510(k) Summary of safety and effectiveness for the AviClear Laser System is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) Summary.
| Applicant: | Cutera, Inc. |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3240 Bayshore Blvd., Brisbane, CA 94005 |
| Contact Person: | Amogh Kothare<br>akothare@cutera.com<br>m. (415) 690-6796<br>o. (415) 657-5730<br>f. (415) 715-3730 |
| Preparation Date: | March 10, 2022 |
| Device Trade Name: | AviClear Laser System |
| Common Name: | Dermatology Laser |
| Classification Name: | Powered Laser Surgical Instrument, GEX, 21 CFR 878.4810 |
| Legally Marketed<br>Predicate Device: | AviClear Laser System (K213461) |
| Indications for Use: | The AviClear Laser System is indicated for the long-term treatment of mild<br>to severe inflammatory acne vulgaris. |
| Device Description: | The AviClear Laser System is an infrared diode laser device with a nominal<br>wavelength of 1726 nm. Similar to other well-established infrared laser<br>devices, the device uses a combination of the treatment spot size, beam<br>characteristics, and tissue absorption and scattering coefficients at the |
e output wavelength to deliver energy to tissues at depth. The wavelength of the laser energy, when combined with the pre, parallel (during energy delivery), and post cooling of epidermal and superficial dermal structures, causes selective heating of dermal tissue at different depths. The 1726 nm laser enerqy heats the chromophore, sebum within sebaceous glands, which results in controlled thermal injury of the sebaceous glands, the ultimate treatment target, thus reducing or eliminating sebum production and causing an improvement in acne vulgaris.
> The diode laser and associated beam delivery optics, laser and electronics power supplies, control electronics, and cooling system are housed inside a console equipped with a touchscreen user interface. The laser treatment parameters are selected using the touchscreen.
> The treatment handpiece has an integrated scanner for delivering treatment spot(s) in an operator-selected pattern; a temperature-controlled skincontact cooling window to provide thermal protection for the epidermis and superficial dermis and through which energy is delivered to the patient; and skin-contact pressure sensors that enable laser energy delivery when the system is in Ready mode and the footswitch is depressed.
{4}------------------------------------------------
## Section 7 510(k) Summary
- Performance Data: IEC 60601-1/A1:2012 Medical Electrical Equipment - Part 1: ● General Requirements for Safety
- IEC 60601-1-6:2010/A1:2013 Medical electrical equipment -O Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- IEC 60601-2-22:2007/A1:2012 Medical Electrical Equipment - O Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories
- IEC 60825-1:2014 Safety of laser products Part 1: Equipment o classification and requirements
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Disturbances
- Software Verification and Validation Testing ●
- Biocompatibility testing of patient-contact materials according to ISO-● 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Data for the clinical study was previously submitted under K213461, Results of Clinical Study: resulting in FDA 510k clearance for the treatment of mild to severe inflammatory acne vulqaris. This is a prospective, multi-center, clinical study to assess the safety and effectiveness of the Cutera AviClear Laser System for the treatment of mild to severe acne vulgaris. The study's primary effectiveness objective was to show more that than 50% of subjects enrolled were Responders who achieved Treatment Success, where Treatment Success was defined as a subject with ≥50% fewer inflammatory acne lesions with respect to baseline. Safety was assessed by a review of the totality of all reported adverse events and their resolutions, including assessments of treatment discomfort.
104 subjects, 57% female and 43% male, 16 to 40 years (avg. 22.2 ± 5.5), diagnosed with mild, moderate, and severe acne vulgaris were enrolled. Subjects received up to 3 treatments and were followed up through 52weeks post final treatment, with 12-weeks being primary end-point.
Details surrounding the Primary, Secondary Effectiveness as well as Safety Analysis have been outlined in the previous 510(K) summary submitted under K213461.
- . Long Term Effectiveness Analysis: At the 52-week post-finaltreatment follow-up visit, 91.5% of subjects achieved treatment success (p<0.001); thereby showing durable and efficacious results from the 12week post-final-treatment follow-up visit.
#### Secondary Effectiveness Analyses: ●
Improvement in Investigator's Global Assessment
- By the 52-week follow-up visit, the number of subjects had O increased to 66.2% for the Clear or Almost Clear skin categories.
{5}------------------------------------------------
## Section 7 510(k) Summary
Improvement from Baseline in Non-Inflammatory Lesion Counts
- The non-inflammatory lesion counts reduced from 25% (p=0.001) o at 12-week mark to 57.6% reduction from Baseline at the 52week follow-up assessment (p<0.001).
Subject Satisfaction
- Approximately 70% of subjects reported they were either "Very O likely" or "Likely" to have the laser treatment aqain at both the 12and 52-week follow-up visits.
- Safety Variable Analysis: All treatment sessions were well tolerated ● with an average discomfort score of 5.23 ± 1.15. No treatment sessions ended prematurely due to excessive treatment discomfort. No reports of serious adverse events (SAEs) and unanticipated adverse device effects of any severity (SUADEs/UADEs).
There were no residual adverse events at the 52-week mark. Adverse events through 12-week follow-up can be found outlined in previous 510(K) summary submitted under K213461. Most commonly reported adverse events from the 12-week mark are summarized below:
- As is expected with laser procedures, all subjects experienced o mild, transient erythema, while 98.1% of subjects experienced edema. Erythema self-resolved within minutes to hours, while edema self-resolved within hours to days.
- 45.2% of subjects experienced mild acne purging / flare-up; O typically resolving with no medical intervention.
- 18.3% of subjects experienced mild dryness; typically resolving O with application of topical skin moisturizers.
- There were no reports of blistering, hyper- or hypopigmentation, O or scarring adverse events in any subjects.
As demonstrated by the results of the clinical study, the AviClear Laser System continues to demonstrate safe and effective performance at the oneyear follow-up, thus supporting the claim for long-term treatment of mild to severe inflammatory acne vulgaris.
The technical specifications of the previously cleared AviClear Laser System and the subject device are identical, as tabulated below. However, the following minor design modifications were made to the subject device:
- . Slimmer handpiece body (no change to patient contact end of handpiece)
- Updated touchscreen user interface ●
Comparison of the Proposed Device and Predicate Device:
{6}------------------------------------------------
## Section 7 510(k) Summary
## Technical Specification Comparison
| | AviClear Laser System<br>(current submission) | AviClear Laser System<br>K213461 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Wavelength | 1726 nm | 1726 nm |
| Max Fluence | Single pulse mode: 30 J/cm²<br>Double pulse mode: 20 J/cm² | Single pulse mode: 30 J/cm²<br>Double pulse mode: 20 J/cm² |
| Max Pulse Energy | 5 J | 5 J |
| Pulse Duration | Up to 50 ms | Up to 50 ms |
| Spot Size | 3 mm or 10 mm 7-spot<br>hexagonal array | 3 mm or 10 mm 7-spot<br>hexagonal array |
| Laser Type | Diode | Diode |
| Output Mode | Quasi-CW | Quasi-CW |
| User Interface | Touchscreen | Touchscreen |
| Treatment Beam Activation | Footswitch and handpiece contact<br>sensors verifying firm and even<br>contact between cooling window<br>and skin is made and maintained | Footswitch and handpiece contact<br>sensors verifying firm and even<br>contact between cooling window and<br>skin is made and maintained |
| Delivery System | Optical fiber handpiece | Optical fiber handpiece |
| Skin Cooling | 0°C to 5°C, Sapphire window | 0°C to 5°C, Sapphire window |
| Aiming Beam | 650 nm | 650 nm |
| Handpiece | Non-sterile, reusable, cleanable | Non-sterile, reusable, cleanable |
Conclusion: As demonstrated by the results of the clinical study, the AviClear Laser System is safe and effective for the requested indication for use. The clinical study also confirmed that there are no new concerns for safety or efficacy for the updated indication for use. The minor cosmetic design changes to the handpiece and touchscreen do not impact the technical specifications nor the safety and effectiveness of the device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.