← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K222221

# DEKA HELIX (K222221)

_El.En Electronic Engineering Spa · GEX · Jan 4, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K222221

## Device Facts

- **Applicant:** El.En Electronic Engineering Spa
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jan 4, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Helix Laser System with its accessories is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

## Device Story

The DEKA HELIX is a laser surgical system for dermatology and plastic surgery. It utilizes two laser sources (10600 nm CO2 and 1570 nm fiber laser) to perform soft tissue ablation, vaporization, excision, and coagulation. The system includes a console with an optical bench, high-voltage power supply, cooling system, and an articulated arm for delivery. Operators interact with the device via an LCD touch-screen control panel to set parameters and monitor system status. The device uses a dual-axis scanner for fractional applications. It is intended for use by trained clinicians in a clinical setting. The output is laser energy delivered to the patient's skin to achieve therapeutic effects like resurfacing or lesion removal. The device is controlled by a footswitch and includes safety interlocks. It is designed to provide precise tissue interaction, benefiting patients by enabling controlled surgical and aesthetic procedures.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing per IEC 60601-2-22 and IEC 60825-1.

## Technological Characteristics

Class II laser surgical instrument. Dual-wavelength: 10600 nm (RF-excited CO2) and 1570 nm (fiber laser). Features: articulated arm, dual-axis scanner, cooled tips, LCD touch-screen interface. Electrical: 100-230V, 50/60 Hz. Cooling: closed-loop heat exchanger. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Alma Hybrid Laser System ([K203441](/device/K203441.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 4, 2023

El.En Electronic Engineering Spa Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, Firenze 50041 Italy

Re: K222221

Trade/Device Name: Deka Helix Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: December 2, 2022 Received: December 5, 2022

Dear Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222221

Device Name DEKA HELIX

#### Indications for Use (Describe)

The Helix Laser System with its accessories is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

#### CO2 Handpieces

The Helix CO2 handpieces with wavelength of 10600 nm are indicated for use for the particular indications as follows:

#### Dermatology & Plastic Surgery

The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

- * laser skin resurfacing
- * laser derm-abrasion
- * laser burn debridement.

Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

Scanning unit

The Helix Scanning unit, with wavelength of 10600 nm is indicated for:

·Laser skin resurfacing (ablation and/or vaporization) of soft tissue

The Helix Scanning unit, with wavelength of 1570 nm, is indicated for:

· Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue

The Helix Scanning unit, with wavelengths of 10600 nm & 1570mm is indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) Summary

#### K222221

## DEKA HELIX

#### Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

## Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

## Date Summary Prepared:

July 20, 2022

## Device Trade Name:

DEKA HELIX

#### Common Name:

Powered Laser Surgical Instrument

#### Classification Name:

Powered laser surgical Instrument

Powered laser surgical instrument with microbeam\fractional output

#### Product Code:

GEX

ONG

{5}------------------------------------------------

## Regulatory Class:

Class II

#### Classification Number:

21 CFR 878.4810

## Predicate Device:

The Alma Hybrid Laser System (K203441)

#### Device Description:

The Helix system consists of:

• Laser system console (containing the optical bench assembly and laser, the microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)

• LCD control panel with touch-screen technology: the LCD display provides information on the status and settings of the Helix system with touch-screen technology provided to input commands into the system.

- Two wavelengths CO2 and 1570 nm
- Articulated arm
- Footswitch
- Delivery devices (CO₂ non-fractional applicators, Scanner CO₂/1570nm fractional applicator)

Electrical specifications are:

100-230V ~ single phase, 50/60 Hz, Absorbed electric power 1500 VA (max)

#### Indications for Use:

The Helix Laser System with its accessories is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

#### CO₂ handpieces

{6}------------------------------------------------

The Helix CO handpieces, with wavelength of 10600nm, are indicated for use for the particular indications as follows:

Dermatology and Plastic Surgery

The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

laser skin resurfacing laser derm-abrasion laser burn debridement.

Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

#### Scanning unit

The Helix Scanning unit, with wavelength of 10600 nm is indicated for: Laser skin resurfacing (ablation and/or vaporization) of soft tissue

The Helix Scanning unit, with wavelength of 1570 nm, is indicated for: Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue

The Helix Scanning unit, with wavelengths of 10600nm & 1570nm is indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

## Comparison with The Predicate Device:

The DEKA HELIX is substantially equivalent to Alma Hybrid Laser System (K203441)

| Device Trade Name                    | Proposed Device<br>DEKA HELIX                                                                                                                                                                                                                                                                | Predicate Device<br>The Alma Hybrid Laser<br>System (K203441)                                                                                                                                                                                                                                                                                                                                          | Comment                                                   |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Indications for Use                  | The Helix Laser System<br>with its accessories is<br>intended for use in<br>surgical applications<br>requiring the ablation,<br>vaporization, excision,<br>incision, and coagulation<br>of soft tissue in medical<br>specialties including:<br>aesthetic surgery<br>(dermatology and plastic | The Alma Hybrid Laser<br>System, Delivery Devices,<br>Applicators and<br>Accessories are intended<br>for use in surgical<br>applications requiring the<br>ablation, vaporization,<br>excision, incision, and<br>coagulation of soft tissue<br>in medical specialties<br>including: aesthetic                                                                                                           | Identical                                                 |
| Device Trade Name                    | Proposed Device<br>DEKA HELIX                                                                                                                                                                                                                                                                | Predicate Device<br>The Alma Hybrid Laser<br>System (K203441)                                                                                                                                                                                                                                                                                                                                          | Comment                                                   |
|                                      | surgery).<br><br>CO2 handpieces<br><br>The Helix CO2 handpieces,<br>with wavelength of<br>10600nm, are indicated<br>for use for the particular                                                                                                                                               | surgery (dermatology and<br>plastic surgery).<br><br>Hy-Light CO2<br><br>The Alma Hybrid CO2 non-<br>fractional applicator, with<br>wavelength of 10600 nm is                                                                                                                                                                                                                                          |                                                           |
|                                      | indications as follows:<br><br>Dermatology and Plastic<br>Surgery<br><br>The ablation, vaporization,<br>excision, incision, and                                                                                                                                                              | cleared for use for the<br>particular indications as<br>follows:<br><br>Dermatology & Plastic                                                                                                                                                                                                                                                                                                          |                                                           |
|                                      | coagulation of soft tissue<br>in dermatology and plastic<br>surgery in the<br>performance of:<br><br>*laser skin resurfacing<br><br>*laser derm-abrasion                                                                                                                                     | Surgery<br><br>The ablation, vaporization,<br>excision, incision, and<br>coagulation of soft tissue<br>in dermatology and plastic<br>surgery in the<br>performance of:                                                                                                                                                                                                                                 |                                                           |
|                                      | *laser burn debridement.<br><br>Laser skin resurfacing<br>(ablation and/or<br>vaporization) of soft tissue                                                                                                                                                                                   | * laser skin resurfacing<br><br>* laser derm-abrasion<br><br>* laser burn debridement.                                                                                                                                                                                                                                                                                                                 |                                                           |
|                                      | for the reduction,<br>removal, and/or<br>treatment of: warts, acne<br>scars, nevi epidermal,<br>syringoma.<br><br>Vaporization/coagulation<br>of warts.                                                                                                                                      | Laser skin resurfacing<br>(ablation and/or<br>vaporization) of soft tissue<br>for the reduction,<br>removal, and/or<br>treatment of: warts, acne<br>scars, nevi epidermal,<br>syringoma.<br><br>Vaporization/coagulation                                                                                                                                                                               |                                                           |
| Device Trade Name                    | Proposed Device                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                       | Comment                                                   |
|                                      | DEKA HELIX                                                                                                                                                                                                                                                                                   | The Alma Hybrid Laser System (K203441)                                                                                                                                                                                                                                                                                                                                                                 |                                                           |
|                                      | Scanning unit                                                                                                                                                                                                                                                                                | of warts.                                                                                                                                                                                                                                                                                                                                                                                              |                                                           |
|                                      | The Helix Scanning unit, with wavelength of 10600 nm is indicated for:<br>Laser skin resurfacing (ablation and/or vaporization) of soft tissue                                                                                                                                               | Pixel<br>The Alma Hybrid Pixel CO2 fractional applicator, with wavelength of 10600 nm is indicated for: The ablation, vaporization and coagulation of soft tissues in dermatology and plastic surgery in the performance of skin resurfacing.                                                                                                                                                          |                                                           |
|                                      | The Helix Scanning unit, with wavelength of 1570 nm, is indicated for:<br>Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                           |
|                                      | The Helix Scanning unit, with wavelengths of 10600nm & 1570nm is indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.                                                                                                                                         | ProScan<br>The Alma Hybrid ProScan CO2 fractional applicator, with wavelength of 10600 nm is indicated for:<br>•Laser skin resurfacing (ablation and/or vaporization) of soft tissue<br>The Alma Hybrid ProScan 1570nm fractional applicator, with wavelength of 1570 nm, is indicated for:<br>• Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue |                                                           |
|                                      |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                           |
| Device Trade Name                    | Proposed Device<br>DEKA HELIX                                                                                                                                                                                                                                                                | Predicate Device<br>The Alma Hybrid Laser<br>System (K203441)                                                                                                                                                                                                                                                                                                                                          | Comment                                                   |
|                                      |                                                                                                                                                                                                                                                                                              | CO2 &1570nm fractional<br>applicator, with<br>wavelengths of 10600 nm<br>&<br>1570nm is indicated for<br>laser skin resurfacing<br>(ablation and/or<br>vaporization) of soft<br>tissue.                                                                                                                                                                                                                |                                                           |
| Regulation number                    | 21 CFR 878.4810 Laser<br>surgical instrument for<br>use in general and<br>plastic surgery and in<br>dermatology                                                                                                                                                                              | 21 CFR 878.4810 Laser<br>surgical instrument for<br>use in general and<br>plastic surgery and in<br>dermatology                                                                                                                                                                                                                                                                                        | Identical                                                 |
| Product Code                         | GEX<br>ONG                                                                                                                                                                                                                                                                                   | GEX<br>ONG                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                 |
| System Control                       | Fully computerized                                                                                                                                                                                                                                                                           | Fully computerized                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                 |
| Emission Control                     | Footswitch<br>Interlock                                                                                                                                                                                                                                                                      | Footswitch<br>Interlock                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                 |
| User Interface                       | Touch-screen display<br>Emergency button<br>Key switch                                                                                                                                                                                                                                       | Touch-screen display<br>Emergency button<br>Key switch                                                                                                                                                                                                                                                                                                                                                 | Identical                                                 |
| Cooling System                       | Closed loop, heat<br>exchanger                                                                                                                                                                                                                                                               | Closed loop, heat<br>exchanger                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                 |
| Emission indicator                   | Visual<br>Audible                                                                                                                                                                                                                                                                            | Visual<br>Audible                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                 |
| Electrical<br>Requirements           | 100-230 V, 16A 50/60 Hz,<br>single phase                                                                                                                                                                                                                                                     | 120 VAC, 11 A, 50/60 Hz,<br>single phase<br>220/230 VAC, 6 A, 50/60<br>Hz, single phase                                                                                                                                                                                                                                                                                                                | Differences do not<br>affect safety and<br>effectiveness. |
|                                      |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                           |
| Device Trade Name                    | Proposed Device<br>DEKA HELIX                                                                                                                                                                                                                                                                | Predicate Device<br>The Alma Hybrid Laser<br>System (K203441)                                                                                                                                                                                                                                                                                                                                          | Comment                                                   |
| Physical<br>Dimensions and<br>Weight | 62 cm x 63 cm x 138/202<br>cm (erected)<br>~ 70 Kg                                                                                                                                                                                                                                           | 465mm x 533 mm x<br>1345/1973 mm (erected)<br>~ 86 Kg                                                                                                                                                                                                                                                                                                                                                  | Differences do not<br>affect safety and<br>effectiveness. |
| Laser Type                           | RF-excited CO2 laser, Class<br>4<br>Fiber Laser, Class 4                                                                                                                                                                                                                                     | RF-excited CO2 laser, Class<br>4<br>Fiber Laser, Class 4                                                                                                                                                                                                                                                                                                                                               | Identical                                                 |
| CO2 Handpieces                       |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                           |
| Wavelength                           | 10600 nm                                                                                                                                                                                                                                                                                     | 10600 nm                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                 |
| Pilot beam                           | 635nm, 4mW                                                                                                                                                                                                                                                                                   | 650 nm, 3mW                                                                                                                                                                                                                                                                                                                                                                                            | Differences do not<br>affect safety and<br>effectiveness. |
| Spot size (mm)                       | 0.125 – 2.0 mm                                                                                                                                                                                                                                                                               | 0.15 – 3.0 at 50 mm<br>working distance<br>0.25- 3.1 (Dia.) at 200 mm<br>working distance                                                                                                                                                                                                                                                                                                              | Differences do not<br>affect safety and<br>effectiveness  |
| Pulse Duration                       | 10 - 900ms                                                                                                                                                                                                                                                                                   | 1-1000 ms                                                                                                                                                                                                                                                                                                                                                                                              | Differences do not<br>affect safety and<br>effectiveness. |
| Pulse Repetition<br>Frequency        | 0.2 — 200 Hz                                                                                                                                                                                                                                                                                 | 1 — 100 Hz                                                                                                                                                                                                                                                                                                                                                                                             | Differences do not<br>affect safety and<br>effectiveness. |
| Min output energy                    | 5 mJ                                                                                                                                                                                                                                                                                         | 5 mJ                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                 |
| Output Power                         | 70W                                                                                                                                                                                                                                                                                          | 30W, 70W                                                                                                                                                                                                                                                                                                                                                                                               | Differences do not<br>affect safety and<br>effectiveness. |
| Scanning unit                        |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                        |                                                           |
| Wavelenght                           | 10600 nm and/or 1570<br>nm                                                                                                                                                                                                                                                                   | 10600 nm and/or 1570 nm                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                 |
| Pilot beam                           | 635nm, 4mW                                                                                                                                                                                                                                                                                   | 650 nm, 3mW                                                                                                                                                                                                                                                                                                                                                                                            | Differences do not<br>affect safety and<br>effectiveness. |
| Device Trade Name                    | Proposed Device<br>DEKA HELIX                                                                                                                                                                                                                                                                | Predicate Device<br>The Alma Hybrid Laser<br>System (K203441)                                                                                                                                                                                                                                                                                                                                          | Comment                                                   |
| Scanner                              | Dual axis scanner                                                                                                                                                                                                                                                                            | Dual axis scanner                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                 |
| Tip                                  | Cooled                                                                                                                                                                                                                                                                                       | Cooled                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                 |
| Spot size @ 10600<br>nm              | 0.374mm                                                                                                                                                                                                                                                                                      | 0.35 mm@ 100 mm<br>distance                                                                                                                                                                                                                                                                                                                                                                            | Differences do not<br>affect safety and<br>effectiveness. |
| Scan size @ 1570<br>nm               | 20x20 mm                                                                                                                                                                                                                                                                                     | Up to 30 mm diameter                                                                                                                                                                                                                                                                                                                                                                                   | Differences do not<br>affect safety and<br>effectiveness. |
| Spot size @ 1570<br>nm               | 0.75mm                                                                                                                                                                                                                                                                                       | 0.75 mm                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                 |
| Output energy @<br>10600 nm          | Max 251mJ                                                                                                                                                                                                                                                                                    | 120 (30W model)<br>240 mJ (70W model)                                                                                                                                                                                                                                                                                                                                                                  | Differences do not<br>affect safety and<br>effectiveness. |
| Output energy @<br>1570 nm           | Max 144mJ                                                                                                                                                                                                                                                                                    | 24mJ-144 mJ/pixel                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                 |
| Beam density @<br>10600 nm           | 61% - 98% untreated<br>tissue between spots                                                                                                                                                                                                                                                  | ~63% - 97% untreated<br>tissue between spots                                                                                                                                                                                                                                                                                                                                                           | Differences do not<br>affect safety and<br>effectiveness. |
| Beam density @<br>1570 nm            | Up to 400 pixels/cm2                                                                                                                                                                                                                                                                         | Up to 390 pixels/cm2                                                                                                                                                                                                                                                                                                                                                                                   | Differences do not<br>affect safety and<br>effectiveness. |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

## Clinical Performance Data:

None

# Non-Clinical Performance Data:

# Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Helix device, according to the following standards:

. ANSI AAMI ES60601-1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

{12}------------------------------------------------

. IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.

# Software Validation and Verification Testing

Software verification and validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

# Additional non-clinical testing conducted

Additional tests were conducted on the Helix device, according to the following standards:

. IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

- . IEC 60825-1: Safety of laser products - Part 1: Equipment classification and requirements.
# Conclusion:

On the basis of the comparison with the predicate device and on the non-clinical performance data, we can conclude that DEKA HELIX is as safe, as effective, and performs as well as the legally marketed predicate device (K203441).

# Additional Information:

None

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K222221](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K222221)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
