K211335 · Zhengzhou Bestview ST Co., Ltd. · GEX · Jul 28, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211335
Device Name
Diode Laser Machine, Model Number: BM-100
Applicant
Zhengzhou Bestview ST Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Diode Laser Machine (Model: BM-100) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Device Story
Diode laser system (808 nm) for hair removal; consists of laser console, LCD touchscreen, and handpiece. Operates via emission of infrared laser pulses; parameters (pulse width, frequency, energy) controlled by user via touchscreen. Used in hospital settings by trained professionals. Water and air cooling systems manage thermal output. Device provides permanent hair reduction by targeting hair follicles; clinical benefit is long-term, stable reduction of hair regrowth. Healthcare provider uses device to perform treatments; output affects clinical decision-making regarding treatment frequency and energy settings.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-22), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Diode laser (808 nm); Class IV laser. Powered by AC (220/110V). Features LCD touchscreen, water/air cooling. Dimensions: 112x42x60cm; Weight: 63kg. Compliance: IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2, IEC 60825-1. Biocompatibility: ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K152898 — DIODE LASER THERAPY SYSTEMS · Beijing Sincoheren Science and Tech. Development Co., Ltd. · Feb 9, 2016
K191970 — Diode Laser Hair Removal System · Beijing Sano Laser S&T Development Co.,Ltd · Apr 6, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 28, 2021
Zhengzhou Bestview St Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, BeiJing 102401 China
Re: K211335
Trade/Device Name: Diode Laser Machine, Model: BM-100 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 28, 2021 Received: May 3, 2021
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
#### 510(k) Number (if known)
K211355
Device Name Diode Laser Machine Model: BM-100
Indications for Use (Describe)
The Diode Laser Machine (Model: BM-100) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
- Date of Preparation 1.
07/27/2021
- 2. Applicant Name and Address
## Zhengzhou Bestview St Co., Ltd.
Room 2004, 20F, Lande Center, Huayuan Road, Jinshui District, 450000 Zhengzhou, Henan Province, China
- 3. Contact Person Information YangChun Jia General Manager Tel: +86-15803801506 Email: info@bestviewmedical.com
- 4. Submission Correspondent
Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
- 5. Identification of Proposed Device Trade Name: Diode Laser Machine Common Name: Powered Laser Surgical Instrument Model(s): BM-100 Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General& Plastic Surgery;
- Identification of Primary Predicate 6. 510(k) Number: K162659 Product Name: Diode Laser Hair Removal System Manufacturer: Shandong Huamei Technology Co., Ltd.
- 7. Device Description
The proposed device incorporates a diode laser that emits invisible infrared laser radiation centered on a wavelength of 808 nm. Main components of the device include the laser console, liquid crystal display (LCD) touchscreen control panel, and laser handpiece. The device emits pulses of the 808 nm infrared laser emissions at pulse widths, repetition frequency, and energy levels, conductive to hair removal. The device incorporates water cooling, air cooling fan, and switches/buttons to control the laser output. The device is powered via an alternating current power source. Specification for the device are provided below.
- 8. Indication For Use
The Diode Laser Machine (Model: BM-100) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Substantially Equivalent (SE) Comparison 9.
### Table 3 General Comparison
| ITEM | Proposed Device | Predicate Device (K162659) | Remark |
|----------------|---------------------------------|----------------------------------------|--------|
| Product Code | GEX | GEX | SAME |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Class | 2 | 2 | SAME |
| Where used | Hospital | Hospital | SAME |
| Intended Use | The Diode Laser Machine (Model: | The Diode Laser Hair Removal System is | SAME |
{5}------------------------------------------------
## 510(k) Summary
| | BM-100) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Configuration | Main Unit | Main Unit | SAME |
| | Handpiece | Handpiece | SAME |
| Principle of Operation | Diode Laser | Diode Laser | SAME |
# Table 4 Performance Comparison
| ITEM | Proposed Device | Predicate Device (K162659) | Remark |
|----------------------|------------------------|----------------------------|---------|
| Laser Type | Diode Laser | Diode Laser | SAME |
| Laser Classification | Class IV | Class IV | SAME |
| Laser wavelength | 808 nm | 808 nm | SAME |
| Spot Size | 2.25 cm2 (15mm x 15mm) | 1.44 cm2 | |
| Fluence | 0-120 J/cm2 | 1-120J/cm2 | Similar |
| Frequency | 1-10 Hz | 05-15Hz | Similar |
| Pulse Duration | 10-400 ms | 5-400 ms | Similar |
| Power Supply | 220/110 VAC/50Hz-60Hz | AC 110V/60Hz | SAME |
| Dimension | 112 cm x 42 cm x 60cm | 450mm x 550mm x 350mm | |
| Weight | 63 kg | 52kg | |
# Table 5 Safety Comparison
| Item | Proposed Device | Predicate Device (K162659) | Remark |
|------------------------------------------------|-----------------------------------------|-----------------------------------------|--------|
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | SAME |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SAME |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | SAME |
| Patient Contact Materials and Biocompatibility | | | |
| Patient Contact<br>Materials | handpiece | handpiece | SAME |
| Cytotoxicity | No Cytotoxicity | Comply with ISO 10993-1 | SAME |
| Sensitization | No evidence of sensitization | | |
| Irritation | No evidence of irritation | | |
{6}------------------------------------------------
#### 10. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device conform to the following standards:
- > IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- > IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;
- > IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity
- ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation and A delay-type hypersensitivity
- > Software Validation & Verification Test
- 11. Clinical Testing
No clinical study is included in this submission.
- 12. Conclusion
The proposed device utilizes technological characteristics that are the same or are similar to the predicate device. The proposed device's technological characteristics do not raise new types of questions regarding safety and effectiveness, and performance testing conducted supports that the device can be used safety and effectively for the proposed indication for use above. Based on the comparison and analysis above, the proposed device is considered to be Substantially Equivalent (SE) to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.