K210142 · Quanta System Spa · GEX · Mar 11, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210142
Device Name
Fiber Dust
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Fiber Dust is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology - Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH] - Laser Resection of the Prostrate (LRP) - Laser Enucleation of the Prostate (LEP) - Laser Ablation of the Prostate (LAP) - Transurethral Incision of the Prostate (TUIP) - Condylomas - Urethral strictures - Lesions of external genitalia - Bladder neck incisions (BNI) - Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors - Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi - Treatment of distal impacted fragments remaining in the ureters following lithotripsy Lithotripsy and Percutaneous Urinary Lithotripsy Indications - Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones - Endoscopic fragmentation of renal calculi - Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed Gastroenterology Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: - Appendectomy - Angiodysplasia - Polyps - Colorectal cancer - Biopsy - Telangiectasias - Gall Bladder calculi - Telangiectasias of the Osler-Weber-Renu disease - Biliary/Bile duct calculi - Vascular Malformation - Ulcers - Gastritis - Gastric ulcers - Esophagitis - Duodenal ulcers - Esophageal ulcers - Non Bleeding Ulcers - Varices - Pancreatitis - Colitis - Haemorrhoids - Mallory-Weiss tear - Cholecystectomy - Gastric Erosions - Benign and Malignant Neoplasm Gynecology Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) of soft tissue.
Device Story
Fiber Dust is a Thulium surgical laser system (1940 nm) delivering up to 60W power via quartz optical fiber. Used in urology, gastroenterology, and gynecology for soft tissue surgery and lithotripsy. Operated by clinicians in clinical settings; laser emission triggered by footswitch. System comprises diode laser, power electronics, control electronics, and air cooling. Software manages device functions and user-selected settings. Aiming beam (532nm) assists targeting. Device enables precise tissue interaction (ablation, coagulation, etc.) and stone fragmentation, benefiting patients through minimally invasive endoscopic or open surgical procedures.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-6 standards. Software verification and validation performed per FDA guidance.
Technological Characteristics
Thulium laser source; 1.94 μm wavelength; 60W max average power; CW/pulsed emission. Pulse duration 0.1-15ms; pulse energy 0.02-6J; frequency 1-2500 Hz. Delivery via quartz optical fiber (up to 1000 μm). 532nm aiming beam (<5 mW). Air cooling system. Software-controlled functions.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, resection, ablation, coagulation, hemostasis, vaporization) or lithotripsy in urology, gastroenterology, and gynecology. Applicable to BPH, tumors, calculi, and various gastrointestinal/gynecological conditions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
SOLTIVE Laser System - Olympus Surgical Technologies America (K183647)
Related Devices
K220426 — Fiber Dust PRO · Quanta System Spa · Mar 16, 2022
K221306 — SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE SuperPulsed Laser Fibers, and Accessories · Gyrus Acmi, Inc. · Jan 27, 2023
K183647 — SOLTIVE Laser System(SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVELaser Fibers, and Accessories) · Olympus Surgical Technologies America · Aug 9, 2019
K242191 — SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) · Olympus Surgical Technologies America · Aug 23, 2024
K242987 — Thulium Fiber Laser (FiberLAZE+, FiberLAZE) · Allengers Global Healthcare Private Limited · Dec 17, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 11, 2021
Quanta System Spa Francesco Dell'Antonio Vice President Regulatory Affairs and QA via Acquedotto 109 Samarate (Va), 21017 Italy
Re: K210142
Trade/Device Name: Fiber Dust Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 29, 2020 Received: January 19, 2021
Dear Francesco Dell'Antonio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210142
Device Name Fiber Dust
Indications for Use (Describe)
Fiber Dust is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Urology
- · Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- · Laser Resection of the Prostrate (LRP)
- · Laser Enucleation of the Prostate (LEP)
- · Laser Ablation of the Prostate (LAP)
- · Transurethral Incision of the Prostate (TUIP)
- Condylomas
- · Urethral strictures
- Lesions of external genitalia
- · Bladder neck incisions (BNI)
- · Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- · Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- · Treatment of distal impacted fragments remaining in the ureters following lithotripsy
Lithotripsy and Percutaneous Urinary Lithotripsy Indications
- · Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate,
- monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
#### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- · Appendectomy
- · Angiodysplasia
- · Polyps
- · Colorectal cancer
- · Biopsy
- · Telangiectasias
- Gall Bladder calculi
- · Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
- Vascular Malformation
- Ulcers
- · Gastritis
- · Gastric ulcers
- · Esophagitis
- · Duodenal ulcers
- · Esophageal ulcers
{3}------------------------------------------------
- Non Bleeding Ulcers
- · Varices
- Pancreatitis
- Colitis
- Haemorrhoids
- Mallory-Weiss tear
- ·Cholecystectomy
- · Gastric Erosions
- ·Benign and Malignant Neoplasm
## Gynecology
Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) of soft tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K210142
## 5. 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy, 21017 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA |
| | Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | 4th March 2021 |
| Device Name: | Fiber Dust |
| Common name: | Surgical laser |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Predicate Device | SOLTIVE Laser System - Olympus Surgical Technologies America<br>(K183647) |
## Description of the device:
The Fiber Dust laser device includes a diode laser that emits a wavelength of 1940 nm with a laser output power up to 60W. Laser radiation is delivered to the patient via a quartz optical fiber having a diameter up to 1000 μm.
The main subsystems of the device are the diode laser, the power electronics, the optical delivery system, the control electronics, and the cooling system. Software controls the device functions and allows the user select device settings. Laser emission is triggered by a footswitch.
{5}------------------------------------------------
## Summary of the technological characteristics
| | Subject device | Predicate device(s) | Comparison to predicate devices |
|-------------------|--------------------|------------------------------------------|-----------------------------------------|
| model name | Fiber Dust | SOLTIVE Laser System | - |
| manufacturer | Quanta System SpA | Olympus Surgical Technologies<br>America | - |
| 510(k) | - | K183647 | - |
| Laser Source | Thulium laser | Thulium laser | Same<br>technology |
| Wavelength (nm) | 1.94 μm | 1.94 μm | Same |
| Emission | CW/pulsed | CW/pulsed | Same |
| Pulse duration | 0.1 to 15ms | 0.2 to 50ms | Within the<br>range of the<br>predicate |
| Pulse energy | 0.02 to 6J | 0.025 to 6J | Within the<br>range of the<br>predicate |
| Frequency | 1 to 2500 Hz | 1 to 2400 Hz | Similar |
| Max average power | 60 W | 60 W | Same |
| Delivery system | Optical fibers | Optical fibers | Same as |
| Aiming beam | 532nm laser < 5 mW | 500-550nm laser < 5 mW | Equivalent |
| Cooling system | Air cooling system | Air cooling system | Equivalent |
## Indications for Use
Fiber Dust is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
#### Urology
- Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- Laser Resection of the Prostrate (LRP)
- Laser Enucleation of the Prostate (LEP)
- Laser Ablation of the Prostate (LAP)
- Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral strictures
- Lesions of external genitalia
- Bladder neck incisions (BNI)
- Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy
{6}------------------------------------------------
## Lithotripsy and Percutaneous Urinary Lithotripsy Indications
• Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
#### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including:
- · Appendectomy
- Angiodysplasia
- Polyps
- Colorectal cancer
- Biopsy
- Telangiectasias
- Gall Bladder calculi
- Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
- Vascular Malformation
- Ulcers
- Gastritis
- Gastric ulcers
- Esophagitis
- Duodenal ulcers
- Esophageal ulcers
- Non Bleeding Ulcers
- Varices
- Pancreatitis
- Colitis
- Haemorrhoids
- Mallory-Weiss tear
- •Cholecystectomy
- Gastric Erosions
- •Benign and Malignant Neoplasm
#### Gynecology
Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
#### Accessories
This device is intended to be used together with delivery optical fiber that separately received an FDA clearance for an intended use compatible with the one of this device.
{7}------------------------------------------------
#### Performance testing
The subject device was subject to testing according to the following recognized consensus standards related to electromagnetic compatibility, electrical safety and performances.
- -IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance
- -IEC 60601-1-2:2014, ed 4, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements And Tests.
- -IEC 60601-2-22: 2012, ed 3.1, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment
- -IEC 60825-1:2014, ed 3.0, Safety of laser products – Part 1: Equipment classification and requirements
- -IEC60601-1-6:2013, ed 3.1, Medical Electrical Equipment Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- -Software Verification and Validation Testing: Software verification testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The results of the non-clinical performance standards testing support that the device can be used safely and effectively.
#### Comparison with predicate device:
The subject device's technological characteristics and indications for use are similar to the predicate device, and its laser output ranges are equivalent to those of the predicate device. Other differences between the subject and predicate device to not raise new types of questions regarding the subject device's safety and efficacy.
#### Conclusions
The non-clinical performance testing conducted supports that the device can be used safely and effectively. The differences in the indications for use and technological characteristics between the subject and predicate device do not raise new types of questions regarding safety and effectiveness, and the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.