K203054 · Gme German Medical Engineering GmbH · GEX · Mar 5, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203054
Device Name
FlexSys
Applicant
Gme German Medical Engineering GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device FlexSys is a multi-modality skin surface treatment . Depending on the chosen modules the intended use and the mode of operation varies. Module MultiLite: The intended use of the device FlexSys with MultiLite module is: · Blue 420 nm light: The GME FlexSys with Blue 420 MultiLite is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. · Yellow 585 nm light: The GME FlexSys with module Yellow 585 is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides. · Red 635 nm Light: The GME FlexSys with Red 635 MultiLite is indicated in general to treat dermatological conditions and specifically indicated to treat superficial, benign vascular, and pigmented lesions. Module Green 532: The GME FlexSys with module Green 532 (wavelength: 532nm) is intended for vaporization and photocoagulation of benign vascular and benign pigmented lesions in soft tissue. Module Yellow 577: The GME FlexSys with module Yellow 577 (wavelength: 577mm) is intended for treatment of benign vascular and benign pigmented lesions.
Device Story
FlexSys is a multi-modality skin treatment platform comprising a base unit with touchscreen controls, foot switch, and interchangeable laser/LED modules (Green 532, Yellow 577, MultiLite). Modules connect via optical fiber and electrical cables to handpieces/scanners. Principle of operation: Green/Yellow laser modules emit light absorbed by melanin/hemoglobin, causing thermal coagulation/vaporization of target lesions; MultiLite LED module uses non-thermal photomodulation to stimulate collagen or destroy acne-associated bacteria via oxygen radical generation. Device performs automated self-tests upon startup and continuously monitors laser power/fluence during operation. If parameters deviate >20% from nominal, emission stops and user is notified. Used in clinical settings by trained professionals. Output allows clinicians to perform targeted dermatological procedures, benefiting patients through lesion removal, wrinkle reduction, or acne management.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including laser safety (EN 60825-1, EN 60601-2-22), electrical safety (EN 60601-1), and software validation.
Technological Characteristics
Multi-modality platform; diode-pumped solid state lasers (532nm, 577nm) and LED modules (420nm, 585nm, 635nm). Stainless-steel body-contacting components. Connectivity via electrical signal cables and optical fibers. Power/fluence monitored continuously with +/- 20% tolerance. 10-year shelf life. Software-controlled parameter selection and safety monitoring.
Indications for Use
Indicated for patients requiring dermatological treatment, including moderate inflammatory acne vulgaris, periorbital wrinkles/rhytides, and superficial benign vascular or pigmented lesions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 5, 2021
GME German Medical Engineering GmbH Katja Kerl Regulatory Affairs Dreikoenigstrasse 6-8 Erlangen, Bavaria 91054 Germany
Re: K203054
Trade/Device Name: FlexSys Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 18, 2020 Received: October 7, 2020
Dear Katja Kerl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203054
Device Name FlexSys
#### Indications for Use (Describe)
The device FlexSys is a multi-modality skin surface treatment . Depending on the chosen modules the intended use and the mode of operation varies.
Module MultiLite:
The intended use of the device FlexSys with MultiLite module is:
· Blue 420 nm light: The GME FlexSys with Blue 420 MultiLite is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
· Yellow 585 nm light: The GME FlexSys with module Yellow 585 is generally indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.
· Red 635 nm Light: The GME FlexSys with Red 635 MultiLite is indicated in general to treat dermatological conditions and specifically indicated to treat superficial, benign vascular, and pigmented lesions.
Module Green 532:
The GME FlexSys with module Green 532 (wavelength: 532nm) is intended for vaporization and photocoagulation of benign vascular and benign pigmented lesions in soft tissue.
Module Yellow 577:
The GME FlexSys with module Yellow 577 (wavelength: 577mm) is intended for treatment of benign vascular and benign pigmented lesions.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;"><b>X</b></span> Registration for OFF-HIGHWAY Vehicle | <span style="padding-left: 20px;"></span> Use for Competition OFF-HIGHWAY Vehicle |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K203054
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date 01-19-2021 [21 CFR 807.92(a)(1)].
#### A. Applicant Name and Address [21 CFR 807.92(a)(1)]
GME German Medical Engineering GmbH. Dreikoenigstrasse 6-8 91054 Erlangen Bavaria, Germany Tel: +49 9131 934159 10 Fax: +49 9131 934159 99
#### B. Contact Information
GME German Medical Engineering GmbH. Dreikoenigstrasse 6-8 91054 Erlangen Bavaria, Germany Tel: +49 9131 934159 00 Fax: +49 9131 934159 99 Contact person: 1.)
Dr. Dietmar Fischer (technical director), Phone: +49 9131 93415911 e-mail: dietmar.fischer@gmeonline.de
- 2.) Katja Kerl (regulatory affairs) Phone: +49 9131 93415918 e-mail: katja.kerl@gmeonline.de
#### C. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: FlexSys
Device Common Name: Laser Instrument for Dermatology Classification Name: Laser Instrument, Surgical Powered 21 CFR 878.4810 Product Code: GEX Device Classification: Class II
#### D. Predicate Devices [21 CFR 807.92(a)(3)]
The FlexSys uses similar technology and physical output characteristics as the following predicate devices: K031425 LightBio Science L.L.C., Gentle Waves, and K120460 Medmix Co. Ltd., SMARTLUX K133297 Asclepion Laser Technologies GmbH, QuadroStar PRO
{5}------------------------------------------------
### E. Device Description [21 CFR 807.92(a)(4)]
The device FlexSys consists of a base unit with up to two installed internal laser modules, handpieces and/or scanners for these modules and attachable applicators. All applicators, handpieces, and scanners are connected by an electrical cable.
The base unit contains most of the controls, such as the touch-screen, the on / off switch or the emergency stop button. In addition the power cord, foot switch, and the door plug are connected to the base unit. The touch-screen serves as the main control panel.
The applicators, handpieces, and scanners direct the light onto the skin. Some applicators contain a hand switch on the front of the handle by which light emission can be triggered.
The scanners and handpieces of the internal modules are connected to the base unit by an electric signal cable and by an optical fiber which guides the aiming beam and working beam laser light to the handpiece/scanner in its core. The light is then guided onto the skin via several mirrors and lenses. Due to the round shape of the fiber core the resulting laser spot on the skin is circular. The size of the working beam spot corresponds to the nominal value when the distance tip touches the skin. For larger distances the spot size will increase, i.e. the laser beam is divergent.
During development and manufacturing applicable directives, legal requirements and standards for medical devices were considered. Our product complies to them.
If you turn on the device, then the unit performs several tests automatically. Thereby the proper functioning of the device - according to the specifications - is ensured. As soon as these tests have been completed successfully, the device is ready for parameter selection and patient treatment. In addition to the initial tests a number of device functions and parameters are monitored continuously.
If the unit detects abnormalities or defects, then it reacts depending on the nature and importance by a notice, a warning or error message. This message needs to be confirmed by the user. Depending on the message one of the following cases is possible:
- the user needs to perform an action to eliminate the cause of the message (Example: ● The user has to take his foot off the foot switch or order a device service)
- the user must wait for a change of parameters (Example: the temperature of the device is too low, the user must wait until the unit has warmed up sufficiently.)
- . the device attempts to resolve the cause of the error and performs a re-test (Example: a too high / low laser power was measured. The device performs an adjustment of power. )
- A reboot of the device is required. The device asks the user to initiate a restart. .
Essential performance of the device:
{6}------------------------------------------------
- -The radiation wavelength of the single modules has to lie in the range specified in the tables of section 7.1 of this user manual. This device property is tested by the manufacturer during production. No repeat measurement of this necessary.
- -The light power of all laser modules has to stay within +-20% of its nominal value. This device property is check by the manufacturer during production. The light power is continuously monitored during laser emission. Laser emission will be stopped and error messages will be shown if values outside the +-20% range are detected. The calibration of the laser modules is checked by the service technician during the maintenance specified in section 6.3. The device will notify the user about necessary maintenance works. Appropriate laser power meters are used to calibrate the device in service mode according to the service instructions provided by GME.
- -The applied light dose (=fluence) of all modules has to stay within +-20% of its nominal value. This device property is checked by the manufacturer during production. The applied dose is calculated from the actual light power and the emission time during light emission. In case of too high or too low dose values light emission will be stopped and the user will be notified. The calibration of the modules is checked by the service technician during the maintenance specified in section 6.3. The device will notify the user about necessary maintenance works. Appropriate laser or light power meters are used to calibrate the device in service mode according to the service instructions provided by GME.
#### Principle of Operation:
- For the Green 532 and Yellow 577 modules: Visible light penetrates tissue up to a ● depth of about 1 mm. When the laser radiation hits the skin, the light is mainly absorbed by melanin and hemoglobin in the skin. Melanin and hemoglobin-containing structures such as pigment spots or blood vessels heat up more strongly than the surrounding skin. This causes irreversible damage or coagulation of the structures.
- . For the MultiLite module: Visible light penetrates tissue up to a depth of about 10 mm. Light at this wavelength is beneficial in treating bengin dermatological lesions. LED photo modulation with visible light reverses signs of aging using a nonthermal mechanism.
Blue, LED therapy can destroy the bacteria associated with acne lesions because the light generates oxygen radicals in very active cells (like bacteria or cancer cells) which destroy the cell walls and thus lead to necrosis or apostosis of the cells.
#### F. Device Specifications and Comparison to Predicates [21 CFR 807.92(a)(6)]
The FlexSys Platform is a multi-modality skin treatment device that can be equipped with up to three different modules. Thus the predicate devices will vary from module to module.
The predicate devices used to demonstrate substantial equivalence are the the Asclepion QuadroStar Pro (K133297), the Medmix Smartlux (K120460), and the Light Bioscience GentleWaves (K031425).
The equivalence comparisons for the visible 532 nm and 577 nm laser wavelengths use the predicate Asclepion QuadroStar Pro (K133297).
The equivalence comparisons for the visible LED module Multilite use the predicates Medmix Smartlux (K120460) and Light Bioscience GentleWaves (K003142).
{7}------------------------------------------------
| Characteristic | GME FlexSys | GME FlexSys | Asclepion QuadroStarPRO |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Module Green 532<br>"FlexSys Green 532" | Module Yellow 577<br>“FlexSys Yellow 577" | "QuadroStar" |
| Applicable<br>510(k)s | NA | NA | K133297 |
| Panel/ | General and Plastic<br>Surgery | General and Plastic Surgery | General and Plastic Surgery |
| Product Code/ | GEX | GEX | GEX |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Indications for<br>Use Statement | NA | The GME FlexSys with<br>module Yellow 577<br>(wavelength: 577nm) is<br>intended for treatment of<br>benign vascular and benign<br>pigmented lesions. | The QuadroStarPRO<br>YELLOW (wavelength:<br>577nm) is<br>intended for treatment of<br>benign vascular and benign<br>pigmented lesions. |
| | The GME FlexSys with<br>module Green 532<br>(wavelength: 532nm) is<br>intended for vaporization<br>and photocoagulation of<br>benign vascular and<br>benign pigmented lesions<br>in soft tissue. | NA | The QuadroStarPRO GREEN<br>(wavelength: 532nm) is<br>intended for vaporisation and<br>photocoagulation of benign<br>vascular and benign<br>pigmented lesions in soft<br>tissue. |
| Classification | Class II | Class II | Class II |
| Common Name | Diode-pumped solid state<br>laser | Diode-pumped solid state laser | Diode-pumped solid state laser |
| Mechanism of<br>Action | Heats chromophores<br>(vessels or pigments) and<br>water in the skin through<br>the absorption of light.<br>The heating denatures the<br>target structures. | Heats chromophores (vessels<br>or pigments) and water in the<br>skin through the absorption of<br>light. The heating denatures<br>the target structures. | Heats chromophores (vessels<br>or pigments) and water in the<br>skin through the absorption of<br>light. The heating denatures<br>the target structures. |
# 1. Green 532 and Yellow 577 module
# 2. MultiLite module
| Characteristic | GME FlexSys | Medmix Smartlux | Light Bioscience<br>GentleWaves LED<br>Photomodulation Device |
|-----------------------|----------------------------------------|-----------------------------|---------------------------------------------------------------|
| | Module MultiLite<br>"FlexSys MutiLite" | "Smartlux" | "GentleWaves" |
| Applicable<br>510(k)s | NA | K120460 | K031425 |
| Panel/ | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery |
{8}------------------------------------------------
| Product Code/ | GEX | GEX | GEX |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Indications for<br>Use Statement | Blue: Dermatological<br>conditions and specifically<br>treatment of moderate<br>inflammatory acne vulgaris | 415nm wavelength:<br>Dermatological conditions<br>and specifically indicated to<br>treat moderate inflammatory<br>acne vulgaris | NA |
| | Yellow: Treatment of<br>periorbital wrinkles and<br>rhytides | | 588nm wavelength: The<br>GentleWaves LED<br>Photomodulation Device is<br>indicated/intended for use in<br>the treatment of periorbital<br>wrinkles and rhytides |
| | Red: Treatment of superficial,<br>benign vascular, and<br>pigmented lesions | 633nm wavelength:<br>Dermatology for treatment of<br>superficial, benign vascular,<br>and pigmented lesions | NA |
| Classification | Class II | Class II | Class II |
| Common Name | LED light source | LED light source | LED light source |
| Mechanism of<br>Action | Reduces inflammation and<br>stimulates collagen<br>production via<br>photomodulation effect of the<br>light penetrating the skin. | Reduces inflammation and<br>stimulates collagen<br>production via<br>photomodulation effect of the<br>light penetrating the skin. | Stimulates collagen<br>production via<br>photomodulation effect of the<br>light penetrating the skin. |
## G. Indications for Use [21 CFR 807.92(a)(5)]
The device FlexSys is a multi-modality skin surface treatment . Depending on the chosen modules the intended use and the mode of operation varies.
#### Module MultiLite:
The intended use of the device FlexSys with MultiLite module is:
- . Blue 420 nm light: The GME FlexSys with Blue 420 MultiLite is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
- Yellow 585 nm light: The GME FlexSys with module Yellow 585 is generally ● indicated to treat dermatological conditions and specifically indicated for treatment of periorbital wrinkles and rhytides.
- Red 635 nm Light: The GME FlexSys with Red 635 MultiLite is indicated in general ● to treat dermatological conditions and specifically indicated to treat superficial, benign vascular, and pigmented lesions.
{9}------------------------------------------------
### Module Green 532:
The GME FlexSys with module Green 532 (wavelength: 532nm) is intended for vaporization and photocoagulation of benign vascular and benign pigmented lesions in soft tissue.
### Module Yellow 577:
The GME FlexSys with module Yellow 577 (wavelength: 577mm) is intended for treatment of benign vascular and benign pigmented lesions.
#### H. Discussion of nonclinical tests [21 CFR 807.92(b)(1)]
The Guidance Document, Laser Products – Conformance with EN 60825-1:2015-07 and EN 60601-2-22: 2015-08 (Laser Notice 56: 2018-05-08) was used. Testing reports for EN 60825-1:2015-07 and EN 60601-2-22:2015-08 were submitted. Testing reports for EN 60601-1:2013-12 and EN 60601-1-2:2016-05 were also submitted. The European version of the standards (EN) have been used for the tests. but they are identical in all test requirements to the IEC versions of the standards.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Sofware documentation provided is consistent with a moderate level of concern.
Biocompatibility:The body-contacting component for the FlexSys uses identical stainless-steel material as in FDA cleared reference devices Wavelight IDAS (K053604) and Wavelight ARION (K042474). Therefore, biocompatibility testing is not required.
Usability was established based on standards EN 62366-1:2015 + AC:2015 and EN 60601-1-6:2010+A1:2015
Shelf life was established to be 10 years.
Clinical Performance Data: None
#### I. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of technology, function, indications for use, design, materials and performance, The GME FlexSys is to be concluded substantially equivalent to the predicate devices. There are no new questions of safety or efficacy raised by the introduction of the FlexSys.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.