← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K202041

# Opera Evo (K202041)

_Quanta System Spa · GEX · Oct 15, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K202041

## Device Facts

- **Applicant:** Quanta System Spa
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Oct 15, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Opera Evo laser device and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical Traditionalties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, General Surgery and Arthroscopy.

## Device Story

Opera Evo is a surgical diode laser system emitting 1940 nm and 1470 nm wavelengths with up to 40W combined output. Laser energy is delivered via quartz optical fibers (up to 1000 µm diameter) to target soft tissue. Subsystems include diode lasers, power electronics, control electronics, and cooling system. Operated by a physician via footswitch in clinical settings (OR, endoscopic suites). Software manages device settings and laser emission. Output provides precise tissue incision, resection, ablation, vaporization, coagulation, and hemostasis. Benefits include minimally invasive surgical access and controlled tissue interaction across multiple specialties.

## Clinical Evidence

Bench testing only. Device evaluated against recognized consensus standards including IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), IEC 60825-1 (laser product safety), and IEC 60601-1-6 (usability). Software verification and validation conducted per FDA guidance.

## Technological Characteristics

Diode laser system; wavelengths 1940 nm and 1470 nm; max power 40W. Delivery via quartz optical fiber (up to 1000 µm). Aiming beam: 520 nm (< 5 mW). Emission: CW/pulsed (1 ms to CW). Frequency: up to 250 Hz. Cooling system included. Software-controlled settings. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-6.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- MultiPulse TM+1470 - Asclepion Laser Technologies Gmbh ([K133891](/device/K133891.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 15, 2020

Quanta System SpA Francesco Dell'Antonio Vice President Regulatory Affairs and QA Via Acquedotto 109 Samarate, Varese 21017 Italy

Re: K202041

Trade/Device Name: Opera Evo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 20, 2020 Received: July 23, 2020

Dear Francesco Dell'Antonio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202041

Device Name Opera Evo

Opera Evo laser device and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical Traditionalties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, General Surgery and Arthroscopy.

Urology:

Open and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

- · Urethral Strictures
- · Bladder Neck Incisions (BNI)
- · Ablation and Resection of Bladder Tumors, Urethral and Ureteral Tumors
- · Ablation of Benign Prostatic Hypertrophy (BPH)
- · Transurethral Incision of the Prostate (TUIP)
- · Laser Resection of the Prostate
- Laser Enucleation of the Prostate
- Laser Ablation of the Prostate
- · Condylomas
- Lesions of external genitalia

### ENT:

Endoscopic endonasal surgery (incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:

- · Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy
- Frontal sinusotomy
- · Ethmoidectomy
- Maxillary antrostomy
- · Functionale endoscopic sinus surgery
- · Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal area
- · Tonsillectomy
- Adenoidectomy

Gastroenterology:

Open and endoscopic gastroenterology surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) including:

- Polyps
- · Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- · Duodenal ulcers

### FORM FDA 3881 (1/14)

{3}------------------------------------------------

- · Non-bleeding ulcers
- · Pancreatitis
- Hemorrhoids
- · Cholecystectomy
- · Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- · Telangiectasias
- · Telangiectasias of the Osler-Weber-Rendu disease (OWRD)
- Vascular Malformation
- · Gastritis
- · Esophagitis
- · Esophageal ulcers
- · Varices
- · Colitis
- · Mallory-Weiss tear
- · Gastric erosion

### Thoracic and Pulmonary

Open and endoscopic thoracic and pulmonary surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue including:

- · Laryngeal lesions
### Gynecology:

Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) including:

- Intra-uterine treatment of submucous fibroids
- · Benign endometrial polyps and uterine septum by incision, ablation and vessel coagulation
- · Soft tissue excision procedures such as excisional conization of the cervix

### General Surgery:

Open laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

- · Cholecystectomy
- · Lysis of adhesion
- Appendectomy
- · Biopsv
- Skin incision
- · Tissue dissection
- · Excision of external tumors and lesions
- · Complete or partial resection of internal organs, tumors and lesions
- Mastectomy
- · Hepatectomy
- · Pancreatectomy
- · Splenectomy
- Thyroidectomy
- · Parathyroidectomy
- · Herniorrhaphy
- · Tonsillectomy
- Lymphadenectomy
- · Partial Nephrectomy
- Pilonidal Cystectomy
- Resection of lipoma
- Debridement of Decubitus Ulcer

{4}------------------------------------------------

### • Hemorrhoids

· Debridement of Statis Ulcer

Arthroscopy:

Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) Ablation of soft, cartilaginous and bony tissue in Minimal invasive Spinal Surgery including:

· Percutaneous Laser Disc Decompression/Discectomy

· Foraminoplasty

· Ablation and coagulation of soft vascular and nonvascular tissue in minimally invasi ve spinal surgery

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

K202041

### 5. 510(K) SUMMARY

| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy, 21017                       |  |  |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------|--|--|
| 510(k) Contact Person:                           | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA         |  |  |
|                                                  | Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |  |  |
| Date Prepared:                                   | June 30th 2020                                                                                  |  |  |
| Device Name:                                     | Opera Evo                                                                                       |  |  |
| Common name:                                     | Surgical laser                                                                                  |  |  |
| Classification:                                  | Class II                                                                                        |  |  |
| Classification Name:                             | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology.         |  |  |
| Regulation Number:                               | 21 CFR 878.4810                                                                                 |  |  |
| Product Code:                                    | GEX                                                                                             |  |  |
| Predicate Device                                 | MultiPulse TM+1470 - Asclepion Laser Technologies Gmbh<br>(K133891)                             |  |  |

### Description of the device:

The Opera Evo laser device includes a diode laser that emits a wavelength of 1940 nm and includes a diode laser that emits a wavelength of 1470 nm, with a combined cumulated laser output power up to 40W. Laser radiation is delivered to the patient via a quartz optical fiber having a diameter up to 1000 um. The main subsystems of the device are the two diode lasers, the power electronics, the optical delivery system, the control electronics, and the cooling system. Software controls the device functions and allows the user select device settings. Laser emission is triggered by a footswitch

{6}------------------------------------------------

|                      | Subject device      | Predicate device(s)                  | Comparison to predicate devices                                                                  |
|----------------------|---------------------|--------------------------------------|--------------------------------------------------------------------------------------------------|
| model name           | Opera Evo           | MultiPulse TM+1470                   | -                                                                                                |
| manufacturer         | Quanta System SpA   | Asclepion Laser<br>Technologies Gmbh | -                                                                                                |
| 510(k)               | -                   | K133891                              |                                                                                                  |
| Laser Source         | diode lasers        | Thulium:YAG laser and<br>diode laser | The differences<br>in technology<br>result in the<br>same laser<br>output<br>parameters<br>range |
| Wavelength<br>(nm)   | 1.94 µm and 1.47 µm | 1.94 µm and 1.47 µm                  | Same as                                                                                          |
| Emission             | CW/pulsed           | CW/pulsed                            | Same as                                                                                          |
| Pulse duration       | 1 ms up to CW       | 0.5 ms up to CW                      | Within the range<br>of the predicate                                                             |
| Frequency            | Up to 250 Hz        | Up to 1000 Hz                        | Within the range<br>of the predicate                                                             |
| Max average<br>power | 40 W                | 150 W                                | Within the range<br>of the predicate                                                             |
|                      | -                   | -                                    |                                                                                                  |
| Delivery<br>system   | Optical fibers      | Optical fibers                       | Same as                                                                                          |
| Aiming beam          | 520nm laser < 5 mW  | 635nm laser < 5 mW                   | Equivalent                                                                                       |

### Summary of the technological characteristics

#### Indications for Use

Opera Evo laser device and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical Traditionalties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, General Surgery and Arthroscopy.

### Urology:

Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

- Urethral Strictures
- Bladder Neck Incisions (BNI)
- . Ablation and Resection of Bladder Tumors, Urethral and Ureteral Tumors
- Ablation of Benign Prostatic Hypertrophy (BPH)
- Transurethral Incision of the Prostate (TUIP)
- Laser Resection of the Prostate
- Laser Enucleation of the Prostate
- . Laser Ablation of the Prostate
- Condylomas
- Lesions of external genitalia ●

{7}------------------------------------------------

### ENT:

Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:

- Endonasal/sinus Surgery
- Partial turbinectomy ●
- Polypectomy
- Dacryocystorhinostomy
- Frontal sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- Functionale endoscopic sinus surgery
- Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal area ●
- Tonsillectomy
- Adenoidectomy

### Gastroenterology:

Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

- Polyps
- Biopsy ●
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non-bleeding ulcers
- . Pancreatitis
- Hemorrhoids
- Cholecystectomy
- Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Rendu disease (OWRD) ●
- Vascular Malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric erosion

{8}------------------------------------------------

### Thoracic and Pulmonary

Open and endoscopic thoracic and pulmonary surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue including:

- . Laryngeal lesions
# Gynecology:

Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

- . Intra-uterine treatment of submucous fibroids
- . Benign endometrial polyps and uterine septum by incision, ablation and vessel coagulation
- . Soft tissue excision procedures such as excisional conization of the cervix

# General Surgery:

Open laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

- Cholecystectomy
- Lysis of adhesion
- Appendectomy
- Biopsy
- Skin incision
- Tissue dissection
- Excision of external tumors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- Mastectomy
- Hepatectomy
- Pancreatectomy
- Splenectomy
- Thyroidectomy
- Parathyroidectomy
- Herniorrhaphy
- Tonsillectomy
- Lymphadenectomy
- Partial Nephrectomy
- Pilonidal Cystectomy
- Resection of lipoma
- Debridement of Decubitus Ulcer
- Hemorrhoids
- Debridement of Statis Ulcer

# Arthroscopy:

Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) Ablation of soft, cartilaginous and bony tissue in Minimal invasive Spinal Surgery including:

{9}------------------------------------------------

- Percutaneous Laser Disc Decompression/Discectomy
- Foraminoplasty
- Ablation and coagulation of soft vascular and nonvascular tissue in minimally invasi ve spinal surgery

### Accessories

This device is intended to be used together with delivery optical fiber that separately received a FDA clearance for an intended use compatible with the one of this device.

### Performance testing

The subject device was subject to testing according to the following recognized consensus standards related to electromagnetic compatibility, electrical safety and performances.

- -IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance
- -IEC 60601-1-2:2014, ed 4, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements And Tests.
- । IEC 60601-2-22: 2012, ed 3.1, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment
- -IEC 60825-1:2014, ed 3.0, Safety of laser products - Part 1: Equipment classification and requirements
- -IEC60601-1-6:2013, ed 3.1, Medical Electrical Equipment Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- -Software Verification and Validation Testing: Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

The results of the non-clinical performance standards testing support that the device can be used safely and effectively.

### Comparison with predicate device:

The subject device's technological characteristics and indications for use are similar to the predicate device, and its laser output ranges are within the those of the predicate device. Other differences between the subject and predicate device to not raise new types of questions regarding the subject

{10}------------------------------------------------

device's safety and efficacy.

### Conclusions

The non-clinical performance testing conducted supports that the device can be used safely and effectively. The differences in the indications for use and technological characteristics between the subject and predicate device do not raise new types of questions regarding safety and effectiveness, and the subject device is considered substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K202041](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K202041)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
