K200525 · Weifang Mingliang Electronics Co., Ltd. · GEX · May 27, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200525
Device Name
Medical Diode Laser Hair Removal System
Applicant
Weifang Mingliang Electronics Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medical Diode Laser Hair Removal System is used for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types ( Fitzpatrick I-VI), including tanned skin.
Device Story
Medical Diode Laser Hair Removal System (Model V16) is a diode laser device for permanent hair reduction. System consists of a console with an LCD touch-screen control panel and a handpiece. Operator (clinician) selects laser parameters via the interface, which communicates with an internal micro-controller. Device emits 808nm laser light from the handpiece to target hair follicles. Used in clinical settings. Output parameters (fluence, pulse duration, repetition rate) are adjusted by the operator to manage treatment. Device provides non-invasive hair reduction, benefiting patients through long-term, stable reduction in hair regrowth.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Diode laser source; 808nm wavelength; 1.44cm2 spot size; 48 J/cm2 max fluence; 1-300ms pulse duration; 1-10Hz repetition rate. System controlled by micro-controller with LCD touch-screen interface. Non-sterile, cleanable. Conforms to IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-2 standards.
Indications for Use
Indicated for permanent hair reduction in patients of all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Weifang Mingliang Electronics CO., LTD % Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building#15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China
Re: K200525
Trade/Device Name: Medical Diode Laser Hair Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 28, 2020 Received: March 2, 2020
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K200525
Device Name
Medical Diode Laser Hair Removal System
Indications for Use (Describe)
The Medical Diode Laser Hair Removal System is used for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types ( Fitzpatrick I-VI), including tanned skin.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <div style="display:inline-block;"> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block;"> <svg height="15px" width="15px"> <rect fill="none" height="15px" stroke="black" stroke-width="1" width="15px" x="0" y="0"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black" stroke-width="2"></path> </svg> </div> </span> </div> </div> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <div style="display:inline-block;"> <div style="display:inline-block; vertical-align:middle;"> <span> <div style="display:inline-block;"> <svg height="15px" width="15px"> <rect fill="none" height="15px" stroke="black" stroke-width="1" width="15px" x="0" y="0"></rect> </svg> </div> </span> </div> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
| Department of Health and Human Services |
|-----------------------------------------|
| Food and Drug Administration |
| Office of Chief Information Officer |
| Paperwork Reduction Act (PRA) Staff |
| PRAStaff@fda.hhs.gov |
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
# Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K200525
- Date of Preparation 1.
04/23/2020
- Sponsor 2.
### Weifang Mingliang Electronics CO., LTD.
No.13426, Yuqing Street, High-tech district, Weifang, Shandong, China
Contact Person: Li MingLiang Position: General Manager Tel: +86-15854424546 Email: 290659723@qq.com
- 3. Submission Correspondent
Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
- 4. Identification of Proposed Device
Trade Name: Medical Diode Laser Hair Removal System
Common Name: Powered Laser Surgical Instrument
Model(s): V16
Regulatory Information:
Classification Name: Powered Laser Surgical Instrument
Classification: II;
Product Code: GEX;
Regulation Number: 21 CFR 878.4810;
Review Panel: General& Plastic Surgery;
Intended Use:
The Medical Diode Laser Hair Removal System is used for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI), including tanned skin.
#### Device Description 5.
Laser parameters and other system features are controlled from the control panel on the console, which provides an interface to the system's micro-controller through an LCD touch-screen. In the handpiece of the machine, the light emitted by each discharge at 808nm.
- Identification of Predicate Device 6.
Primary Predicate Device 510(k) Number: K181019 Product Name: Diode Laser System Manufacturer: Guangzhou Huafei Tongda Technology Co., Ltd. Secondary Predicate Device 510(k) Number: K153718 Product Name: Spirit Hair Removal laser Family Manufacturer: ACTIVE OPTICAL SYSTEMS , LTD.
- 7. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device conforms with the following standards:
{5}------------------------------------------------
>
- IEC 60601-1:2005/A1:2012 Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance;
- > IEC 60601-2-22:2012, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment;
- > IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- > Software Validation & Verification Test
- Clinical Test Conclusion 8.
No clinical study is included in this submission.
- Substantially Equivalent (SE) Comparison 9.
| | Table 1 General Comparison | |
|--|----------------------------|--|
| | | |
| ITEM | Proposed Device | Predicate Device K181019 | Predicate Device K153718 | Remark |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | GEX | GEX/ILY | GEX/ILY | SAME |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME |
| Class | 2 | 2 | 2 | SAME |
| Intended Use | The Medical Diode Laser Hair<br>Removal System is used for<br>permanent reduction in hair<br>regrowth defined as a long term,<br>stable reduction in the number of<br>hairs re-growing when measured at<br>6, 9 and 12 months after the<br>completion of a treatment regimen.<br>Use on all skin types ( Fitzpatrick<br>I-VI), including tanned skin. | The Diode Laser System is<br>intended for hair removal,<br>permanent hair reduction on all<br>skin types (Fitzpatrick skin type<br>I-VI), including tanned skin.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9,<br>and 12 months after the<br>completion of a treatment regime. | The Spirit Hair Removal<br>Laser Family is generally<br>intended for dermatological<br>use. The devices are<br>specifically indicated for hair<br>removal, permanent hair<br>reduction by using selective<br>laser energy.<br>The Spirit Hair Removal<br>Laser Family is intended for<br>use on all skin types<br>(Fitzpatrick skin types I-VI),<br>including tanned<br>skin.Permanent reduction in<br>hair regrowth is defined as<br>long term, stable reduction in<br>the number of hairs | SAME |
{6}------------------------------------------------
| | | | after the completion of the | |
|--|--|--|-----------------------------|--|
| | | | treatment regimen. | |
{7}------------------------------------------------
| ITEM | Proposed Device | Predicate Device | Predicate Device | Remark |
|-------------------------------------|------------------------|------------------------|--------------------------------------|---------|
| | | K181019 | K153718 | |
| Wavelength | 808 nm | 808nm | 810nm | SAME |
| Light/Laser source | Diode | Diode | Diode | SAME |
| How supplied | Non-sterile, cleanable | Non-sterile, cleanable | Non-sterile, cleanable | SAME |
| Fluence (Energy<br>Density) [J/cm2] | 48 | 40 | 6-90 (Model 918) | SIMILAR |
| Rep Rate [Hz] | 1-10 | 1-5 | ≤ 10 (Model 918) | SAME |
| Pulse Duration<br>[ms] | 1-300 | 30-200 | Up to 310 ms (Model<br>918) | SIMILAR |
| Spot Size | 1.44cm2 (1.2 x 1.2 cm) | 1.20cm2 | 1.92 cm2 (12 x 16<br>mm) (Model 918) | SIMILAR |
Table 2 Performance Comparison
{8}------------------------------------------------
| Item | Proposed Device | Predicate Device | Predicate Device | Remark | |
|----------------------------------|-----------------------------------------------|--------------------------------------------|--------------------------------------------|------------------------------------------|------------------------------------------|
| | | K181019 | K153718 | | |
| EMC, Electrical and Laser Safety | | | | | |
| Electrical Safety | Comply with IEC<br>60601-1, IEC<br>60601-2-22 | Comply with IEC 60601-1,<br>IEC 60601-2-22 | Comply with IEC 60601-1,<br>IEC 60601-2-22 | SAME | |
| | EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SAME |
| | | Laser Safety | Comply with IEC<br>60601-2-22, IEC 60825 | Comply with IEC<br>60601-2-22, IEC 60825 | Comply with IEC<br>60601-2-22, IEC 60825 |
| Table 3 Safety Comparison | | |
|---------------------------|--|--|
| | | |
Analysis
The proposed device has similar indication for use with predicate devices, the main differences are output parameters, such as spot size, fluence, frequency range, pulse duration.
#### 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.