← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K193477

# Nd: YAG Laser Therapy Systems (K193477)

_Beijing Kes Biology Technology Co., Ltd. · GEX · Jun 24, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193477

## Device Facts

- **Applicant:** Beijing Kes Biology Technology Co., Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jun 24, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The device is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm wavelength: -Tattoo removal: light ink(red, tan, purple, orange, skyblue, green) -Removal of Epidermal Benign Pigmented Lesions -Removal of Minor Benign Vascular Lesions including but not limited to telangiectasias -Treatment of Lentigines -Treatment of Cafe-Au-Lait -Treatment of Seborrheic Keratoses -Treatment of Post Inflammatory Hyper-Pigmentation -Treatment of Becker's Nevi, Freckles and Nevi Spilus 1064nm Wavelength: -Tattoo removal: dark ink (black, blue and brown) -Removal of Nevus of Ota -Removal or lightening of unwanted hair with or without adjuvant preparation. -Treatment of Common Nevi -Skin resurfacing procedures for the treatment of acne scars and wrinkle

## Device Story

Multi-wavelength, pulsed Nd:YAG laser system (Model MED-810A) for dermatological and surgical soft tissue treatment. System components: power supply, microprocessor control, laser handpiece, display, and cooling system. Physician selects wavelength (1064nm or 532nm) and energy via control panel. Device uses Q-switched pulses to target specific skin pigments or tissues. Used in clinical settings by physicians. Output energy affects tissue via ablation, vaporization, or coagulation, facilitating lesion removal, tattoo removal, and skin resurfacing. Benefits include precise tissue management and cosmetic improvement.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Nd:YAG laser system; 1064nm and 532nm wavelengths; Q-switched pulse mode; Xenon lamp energy source; 5-10ns pulse duration; 2-10mm spot size; 1-6Hz repetition rate. Materials: Aluminum alloy, ABS. Compliance: IEC 60601-1 (safety), IEC 60601-2-22 (laser safety), IEC 60601-1-2 (EMC). Biocompatibility: ISO 10993-1, -5, -10.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- CuRAS Nd: YAG Laser ([K173038](/device/K173038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

Beijing Kes Biology Technology Co., Ltd. % Daina Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box. 120-119 Shanghai, 200120 Cn

June 24, 2020

## Re: K193477

Trade/Device Name: Nd:YAG Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 4, 2019 Received: December 16, 2019

Dear Daina Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193477

Device Name Nd:YAG Laser Therapy Systems

The device is indicated for the incision, ablation, vaporization of soft tissues for general dermatology,

dermatologic and general surgical procedures for coagulation and hemostasis.

532nm wavelength:

-Tattoo removal: light ink(red, tan, purple, orange, skyblue, green)

-Removal of Epidermal Benign Pigmented Lesions

-Removal of Minor Benign Vascular Lesions including but not limited to telangiectasias

-Treatment of Lentigines

-Treatment of Cafe-Au-Lait

- -Treatment of Seborrheic Keratoses
-Treatment of Post Inflammatory Hyper-Pigmentation

-Treatment of Becker's Nevi, Freckles and Nevi Spilus

1064nm Wavelength:

-Tattoo removal: dark ink (black, blue and brown)

-Removal of Nevus of Ota

-Removal or lightening of unwanted hair with or without adjuvant preparation.

- -Treatment of Common Nevi
-Skin resurfacing procedures for the treatment of acne scars and wrinkle

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.

The assigned 510(k) Number: K193477

- Date of Preparation: 06/24/2020 1.
- Establishment Registration Number: 3009444646 2.

| 3. Applicant    |                                                                                                                                    |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------|
| Name:           | BEIJING KES BIOLOGY TECHNOLOGY CO., LTD.                                                                                           |
| Address:        | No.15, Zone 2, Mafang Industrial Park, PingguyuanZhongguancun Science And<br>Technology Park Pinggu District, Beijing CHINA 101204 |
| Contact Person: | Wei Meng Meng                                                                                                                      |
| Telephone:      | +86-10-60998698                                                                                                                    |
| Fax:            | +86-10-61227575                                                                                                                    |
| Email:          | weimengmeng@keslaser.com                                                                                                           |

#### Submission Correspondent 4.

Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850 Fax: 360-925-3199 Email: info@mid-link.net

- 5. Identification of the Proposed Device

| Trade Name:          | ND:YAG Laser Therapy Systems                                                        |
|----------------------|-------------------------------------------------------------------------------------|
| Common Name:         | Powered Laser Surgical Instrument                                                   |
| Model(s):            | MED-810A                                                                            |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification:      | II                                                                                  |
| Product Code:        | GEX                                                                                 |
| Regulation Number:   | 21 CFR 878.4810                                                                     |
| Review Panel:        | General& Plastic Surgery                                                            |

- 6. Identification of Predicate Device 510(k) Number: K173038 Product Name: CuRAS Nd: YAG Laser
- 7. Device Description

The proposed device is a multi-wavelength, pulsed laser system designed for the treatment of benign pigmented lesions. The device can produce 1064nm and 532nm two different laser wavelengths to treat the different color skin. It consists of power supply module, microprocessor control module, laser treatment handpiece module, operation display module and cooling system module. The physician is able to select the desired wavelength and the related output energy via control panel.

- 8. Indications for Use
The device is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm wavelength:

-Tattoo removal: light ink (red, tan, purple, orange, skyblue, green)

{4}------------------------------------------------

- -Removal of Epidermal Benign Pigmented Lesions
- -Removal of Minor Benign Vascular Lesions including but not limited to telangiectasias
- -Treatment of Lentigines
- -Treatment of Cafe-Au-Lait
- -Treatment of Seborrheic Keratoses
- -Treatment of Post Inflammatory Hyper-Pigmentation
- -Treatment of Becker's Nevi, Freckles and Nevi Spilus

1064nm Wavelength:

- -Tattoo removal: dark ink (black, blue and brown)
-Removal of Nevus of Ota

- -Removal or lightening of unwanted hair with or without adjuvant preparation.
-Treatment of Common Nevi

- -Skin resurfacing procedures for the treatment of acne scars and wrinkle
- 9. Substantially Equivalent (SE) Comparison

| ITEM                | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device K173038                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code        | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Class               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for use | The device is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.<br><br>532nm wavelength:<br>-Tattoo removal: light ink(red, tan, purple, orange, skyblue, green)<br>-Removal of Epidermal Benign Pigmented Lesions<br>-Removal of Minor Benign Vascular Lesions including but not limited to telangiectasias<br>-Treatment of Lentigines<br>-Treatment of Cafe-Au-Lait<br>-Treatment of Seborrheic Keratoses<br>-Treatment of Post Inflammatory Hyper-Pigmentation<br>-Treatment of Becker's Nevi, Freckles and Nevi Spilus<br>1064nm Wavelength:<br>-Tattoo removal: dark ink (black, blue and brown)<br>-Removal of Nevus of Ota<br>-Removal or lightening of unwanted hair with or without adjuvant preparation. | The CuRAS Nd; YAG laser systems indicated for: the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.<br><br>532nm wavelength:<br>-Tattoo removal: light ink(red, tan, purple, orange, skyblue, green)<br>-Removal of Epidermal Pigmented Lesions<br>-Removal of Minor Vascular Lesions including but not limited to telangiectasias<br>-Treatment of Lentigines<br>-Treatment of Cafe-Au-Lait<br>-Treatment of Seborrheic Keratoses<br>-Treatment of Post Inflammatory Hyper-Pigmentation<br>-Treatment of Becker's Nevi, Freckles and Nevi Spilus<br>1064nm Wavelength:<br>-Tattoo removal: dark ink (black, blue and brown)<br>-Removal of Nevus of Ota<br>-Removal or lightening of unwanted hair with or without adjuvant preparation. |

{5}------------------------------------------------

|                           | -Treatment of Common Nevi                            | -Treatment of Common Nevi                            |
|---------------------------|------------------------------------------------------|------------------------------------------------------|
|                           | -Skin resurfacing procedures for the                 | -Skin resurfacing procedures for the                 |
|                           | treatment of acne scars and wrinkle                  | treatment of acne scars and wrinkle                  |
| Energy Source             | Xenon Lamp                                           | Xenon Lamp                                           |
| Wavelength (nm)           | 1064nm and 532nm                                     | 1064nm and 532nm                                     |
| Aiming beam wavelength    | 635nm                                                | 635nm                                                |
| Aiming laser output power | 5mW                                                  | 5mW                                                  |
| Laser output mode         | Q-switched pulse                                     | Q-switched pulse                                     |
| Maximum pulse energy      | @1064nm wavelength:1600mJ<br>@532nm wavelength:400mJ | @1064nm wavelength:1600mJ<br>@532nm wavelength:400mJ |
| Pulse Duration            | 5-10ns                                               | 5~20ns                                               |
| Spot Size                 | 2mm-10mm                                             | 2mm-10mm                                             |
| Fluence range             | 1064nm<br>0.4-51.0 J/cm2<br>532nm<br>0.4-12.7 J/cm2  | Unknown                                              |
| Repetition rate           | 1-6Hz                                                | 1-15Hz                                               |
| Patient contact material  | Aluminum alloy, ABS                                  | Unknown                                              |
|                           | Biocompatibility                                     |                                                      |
| Cytotoxicity              | No Cytotoxicity                                      | No Cytotoxicity                                      |
| Sensitization             | No evidence of Sensitization                         | No evidence of sensitization                         |
| Irritation                | No evidence of Irritation                            | No evidence of irritation                            |
| Electrical Safety and EMC |                                                      |                                                      |
| Electrical Safety         | Comply with                                          | Comply with                                          |
|                           | IEC 60601-1                                          | IEC 60601-1                                          |
|                           | IEC 60601-2-22,                                      | IEC 60601-2-22,                                      |
| EMC                       | Comply with                                          | Comply with                                          |
|                           | IEC 60601-1-2                                        | IEC 60601-1-2                                        |

#### 10. Non-Clinical Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following documents:

- ▲ IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance:
- A IEC 60601-2-22:2012, Medical Electrical Equipment - Part 2-22: Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment;
- ▲ IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- > ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)]
- A ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- A Software Validation & Verification Test, documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

{6}------------------------------------------------

Additionally, Bench Testing was conducted to verify the performance.

- 11. Clinical Testing
No clinical study is included in this submission.

- 12. Conclusion
Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193477](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193477)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
