← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K193367

# elysion-pro (K193367)

_High Technology Products S.L.U · GEX · Mar 12, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193367

## Device Facts

- **Applicant:** High Technology Products S.L.U
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Mar 12, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Indications for use for ELYSION diode laser hair removal system with 755nm and 810nm applicators include: Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defmed as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of Pseudofolliculitis barbae (PFB) Use on all skin types (Fitzpatrick I-VI).

## Device Story

ELYSION is a diode laser system for hair removal and PFB treatment; utilizes selective photothermolysis via near-infrared light (755nm/810nm). System comprises a central unit and three removable applicators; applicators contain laser diodes and sapphire contact-cooling tips. Operated by clinicians in a professional setting; energy delivery controlled by central unit; emission triggered by applicator button. Static mode (1-3 Hz) and dynamic mode (5-15 Hz) available. Sapphire tip maintains constant 5°C to cool skin, providing partial anesthesia and epidermal protection. Output is pulsed laser energy absorbed by hair follicle melanin; results in thermal destruction of follicle. Benefits include permanent hair regrowth reduction and PFB management.

## Clinical Evidence

No clinical data included in this submission. Substantial equivalence supported by non-clinical bench testing and adherence to recognized safety standards (IEC 60601 series, IEC 60825-1, ISO 10993).

## Technological Characteristics

Diode laser system (755nm/810nm); AlGaAs laser diode array; sapphire (Al2O3) contact cooling (5°C). Dimensions: 650x500x450 mm. Power: 100-240V, 50-60Hz. Static/Dynamic modes. Class IV laser. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-6, ISO 14971, ISO 10993-5, ISO 10993-10.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Primelase Excellence ([K191321](/device/K191321.md))
- Soprano XL / Soprano ICE ([K140009](/device/K140009.md))
- Lightsheer Desire / Lightsheer Desire Light / Lightsheer Duet / Lightsheer Infinity ([K170179](/device/K170179.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

High Technology Products SLU Mrs. Vardhini Kirthivas Freyr Global Regulatory Solutions and Services Level 4 Building No. H-08, Phoenix SEZ Phase 2 Gachibowli, Hyderabad, 500081 India

Re: K193367

Trade/Device Name: elysion-pro Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 22, 2019 Received: December 4, 2019

Dear Vardhini Kirthivas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193367

Device Name

EL YSION Diode Laser Hair Removal System

Indications for Use (Describe)

Indications for use for ELYSION diode laser hair removal system with 755nm and 810nm applicators include: · Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defmed as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

· Treatment of Pseudofolliculitis barbae (PFB)

· Use on all skin types (Fitzpatrick I-VI).

| Type of Use (Select one or both, as applicable)                                                                                                                           |                                                                                                                                                           |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)   </span> </span> | <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ELYSION

# K193367_510(k) Summary

{4}------------------------------------------------

## TRADITIONAL 510(K)

#### ELYSION

# K193367_510(K) Summary (21 CFR 868.5160)

## 5.1 Submitter Information:

| Application Correspondent: | Vardhini Kirthivas                             |
|----------------------------|------------------------------------------------|
|                            | Freyr Global Regulatory Solutions and Services |
|                            | Level 4, Building No. H-08, Phoenix SEZ, Phas  |
|                            | 2, Gachibowli, Hyderabad,                      |
|                            | Telangana 500081                               |
| Phone:                     | +91 - 9940445965                               |
| Email:                     | vardhini.kirthivas@FreyrSolutions.com          |
| Company:                   | High Technology Products, S.L.U.               |
|                            | Passatge Masoliver, 24 - 28                    |
|                            | 08005 Barcelona, Catalonia Spain               |
| Phone:                     | +34 93 458 85 66                               |
| Contact Person:            | Mr. Sergi Lozano                               |
| Email:                     | sergi@cocoonmedical.com                        |
| Date Prepared:             | March 7th, 2020                                |
|                            |                                                |

#### 5.2 Device Identification

| Trade Name:          | elysion-pro                             |
|----------------------|-----------------------------------------|
| Common Name:         | ELYSION Diode Laser Hair Removal System |
| Classification Name: | Powered Laser Surgical Instrument       |
| 510(K) Number        | K193367                                 |

#### Table 4: Device Identification

| Regulation No.  | Product Code | Device Class |
|-----------------|--------------|--------------|
| 21 CFR 878.4810 | GEX          | Class II     |

#### 5.3 Legally Marketed Equivalent Device

- 1. Primelase Excellence K191321 by High Technology Products, S.L.U (Same company).
- 2. The Modified Alma Lasers XL™ Family of Multi-Application and Multi-Technology Platforms [Soprano XL Soprano XL and Soprano ICE] – K140009 by Alma Lasers Ltd.
- 3. Lightsheer Desire; Lightsheer Desire Light; Lightsheer Duet; Lightsheer Infinity K170179 by Lumenis Ltd.

{5}------------------------------------------------

## TRADITIONAL 510(K)

#### ELYSION

#### 5.4 Device Description

ELYSION diode laser hair removal system is a medical electrical equipment intended for hair removal treatment, the basic principle of which is selective photothermolysis, which consist in the specific destruction of a follicle due to an increase of the temperature induced by a high-powered beam of light which is selectively absorbed by the melanin.

The ELYSION equipment consists of a central unit and a set of 3 removable applicators. The ELYSION equipment emits laser radiation (near-infrared light with a wavelength range of 755 nm & 810 nm), pulsed through the laser aperture situated at the tip of the applicator.

The device applicator contains the diode which emits the laser energy whereas the power delivered, and the working frequency being controlled by the machine's central unit. The applicator emits the energy through a sapphire window, in contact with the skin throughout the treatment, aimed at damaging the hair follicle.

The device has three types of applicators differentiated by the emitted wavelength and the area of emission. Two of the applicators emitting radiation at 810 nm wavelength for two areas 10×10 mm² and 18×10 mm² and another one emitting radiation at 755 nm for an area of 10×10 mm². Two different operation modes are available: static mode and dynamic mode, which are differentiated basically by the frequency range defined for each mode (1-2-3 Hz for static and 5-10-15 Hz for dynamic).

The emission of energy is activated in the form of continuous pulses when pressing the applicator button. The applicator sapphire tip is cooled to a constant temperature to cool the skin, so that it partially anaesthetizes the tissue reducing the risk of damage to the epidermis during treatment.

## 5.5 Indications for Use

Indications for use for ELYSION diode laser hair removal system with 755nm and 810nm applicators include:

- Hair Removal with Static and Dynamic modes intended for permanent reduction in hair . regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB). .
- Use on all skin types (Fitzpatrick I-VI) .

{6}------------------------------------------------

## TRADITIONAL 510(K)

#### ELYSION

#### 5.6 Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- 1. IEC 60601-1:2005/ (R)2012 and A1:2012 Ed 3.1, Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
- 2. EN IEC 60601-1-2:2014 Ed 4.0, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- 3. IEC 62304:2006 Ed.1.0 Medical device software software life cycle processes
- 4. IEC 60601-2-22:2012 Ed 3.1 Medical electrical equipment Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- 5. IEC 60825 1:2007 Ed 2.0 Safety of laser products Part 1: Equipment classification, and requirements.
- 6. IEC 60601-1-6:2013 Ed 3.1 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- 7. I SO 14971:2007 Medical devices Application of risk management to medical devices.
- UNE-EN-ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for 8. Vitro cytotoxicity
- 9. UNE-EN-ISO 10993-10:2013, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity

#### 5.7 Clinical Test Conclusion

No clinical study is included in this submission

{7}------------------------------------------------

# TRADITIONAL 510(K)

## 5.8 Substantially Equivalent (SE) Comparison

Table 5: Comparison table

| S. No. | Parameters for<br>Substantial Equivalence                    | Soprano ICE<br>(K140009)                                                            | Lightsheer Infinity<br>(K170179)             | Primelase Excellence<br>(K191321)                 | ELYSION                                           |
|--------|--------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------------------------|---------------------------------------------------|
| 1      | Company                                                      | Alma Lasers, Ltd,                                                                   | Lumenis Ltd.                                 | High Technology Products<br>SLU                   | High Technology Products<br>SLU                   |
| 2      | Product Code                                                 | GEX                                                                                 | GEX                                          | GEX                                               | GEX                                               |
| 3      | Regulation Number                                            | 21 CFR 878.4810                                                                     | 21 CFR 878.4810                              | 21 CFR 878.4810                                   | 21 CFR 878.4810                                   |
| 4      | Principle of Operation                                       | AlGaAs Laser diode<br>array                                                         | AlGaAs Laser diode array                     | AlGaAs Laser diode array                          | AlGaAs Laser diode array                          |
| 5      | Laser Wavelength                                             | 755nm, 810 nm                                                                       | 805nm, 1060 nm                               | 755 nm,<br>810 nm,<br>810 - 1060 nm               | 755nm, 810 nm                                     |
| 6      | Laser Contact                                                | Sapphire window                                                                     | Pure sapphire, AL2O3                         | Pure sapphire, AL2O3                              | Pure sapphire, AL2O3                              |
| 7      | Spot Sizes (mm x mm)                                         | 12x10, 15x10, 20x10                                                                 | 9x9, 27x9, 22x35                             | 20x9, 30x9, 30x17                                 | 10x10, 18x10                                      |
| 8      | Fluence (Energy Density)                                     | 120J/cm2                                                                            | 100 J/cm2                                    | 80 J/cm2                                          | 70 J/cm2                                          |
| 9      | Maximum Fluence<br>actually used (as per<br>treatment table) | 40 J/cm2                                                                            | 48 J/cm2                                     | 43 J/cm2                                          | 40 J/cm2                                          |
| 10     | Frequency                                                    | UP TO 3 Hz (HR)<br>5 - 10 Hz (SHR)                                                  | Up to 3 Hz                                   | UP TO 3 Hz (Static)<br>5 - 10 Hz (Dynamic)        | UP TO 3 Hz (Static)<br>5 - 15 Hz (Dynamic)        |
| 11     | Pulse Duration                                               | 3.3 - 200 ms                                                                        | 5 - 400 ms                                   | 3 - 400 ms                                        | 3 - 400ms                                         |
| 12     | Treatment Mode                                               | SHR, HR & LB                                                                        | Repetitive impulses                          | Static & Dynamic                                  | Static & Dynamic                                  |
| 13     | Tissue Cooling                                               | Contact continuous,<br>thermo-electrical                                            | Chilltip contact cooling                     | Contact continuous,<br>thermo-electrical          | Contact continuous,<br>thermo-electrical          |
| 14     | Cooling temperature                                          | 4°C (39°F)                                                                          | 2°C - 12°C                                   | 5°C                                               | 5°C                                               |
| 15     | User Interface                                               | LCD touchscreen                                                                     | LCD touchscreen                              | LCD touchscreen                                   | LCD touchscreen                                   |
| 16     | Pulsing Control                                              | Finger switch                                                                       | Finger switch                                | Finger switch                                     | Finger switch                                     |
| 17     | Configuration                                                | Main unit, Handpiece<br>& Foot Control                                              | Main unit and Handpiece                      | Main unit, Handpiece &<br>Foot Control (Optional) | Main unit, Handpiece &<br>Foot Control (Optional) |
| 18     | Laser Classification                                         | Class IV                                                                            | Class IV                                     | Class IV                                          | Class IV                                          |
| 19     | Power Supply                                                 | 120VAC, 11A, 50/60<br>Hz, single phase<br>220/230VAC, 6A,<br>50/60 Hz, single phase | 100-240 VAC +/-10%, 15<br>A max.   50/60 Hz. | Single phase, 100-240V<br>50-60HZ                 | Single phase, 100-240V<br>50-60HZ                 |
| 20     | Dimension                                                    | 53 cm wide x 57 cm<br>deep x 120 cm high                                            | 44 x 50 x 123 cm                             | 1140 x 480 x 550 mm                               | 650 x 500 x 450 mm                                |

{8}------------------------------------------------

# TRADITIONAL 510(K)

# 5.9 Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193367](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K193367)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
