Litho Evo

K192990 · Quanta System Spa · GEX · Dec 20, 2019 · General, Plastic Surgery

Device Facts

Record IDK192990
Device NameLitho Evo
ApplicantQuanta System Spa
Product CodeGEX · General, Plastic Surgery
Decision DateDec 20, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Litho Evo laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery.

Device Story

Litho Evo is a pulsed Holmium:YAG laser system (2100nm) for soft tissue surgery. Input: electrical power; Output: laser energy delivered via fiber optics to target tissue. Operated by physicians in OR/clinic settings using a footswitch. System includes laser cavity, power/control electronics, cooling system, and touchscreen interface. Software manages device functions and user selections. Laser energy enables incision, ablation, and coagulation, facilitating minimally invasive procedures across multiple specialties. Benefits include precise tissue removal and hemostasis, reducing surgical trauma compared to traditional methods.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through electrical safety, electromagnetic compatibility, and laser safety testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Software verification and validation conducted per FDA guidance.

Technological Characteristics

Pulsed Holmium:YAG laser source (2100nm); 35W max average power; 3-30Hz frequency; up to 1100µs pulse duration; up to 5.0J energy per pulse. Delivery via optical fibers. Aiming beam: 532nm green diode (<5mW). System includes touchscreen interface, footswitch control, and cooling system. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.

Indications for Use

Indicated for surgical procedures (incision, excision, resection, ablation, vaporization, coagulation, haemostasis) in Urology (e.g., BPH, tumors, calculi), Gastroenterology (e.g., polyps, ulcers, neoplasms), Arthroscopy (e.g., ligament release, debridement), Gynaecology, ENT (e.g., sinus surgery), and General Surgery. Applicable to open, laparoscopic, and endoscopic approaches.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an abstract design of human figures. To the right of the seal, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG ADMINISTRATION' are written in a bold, sans-serif font. The words 'FOOD & DRUG' are on the top line, and the word 'ADMINISTRATION' is on the bottom line. December 20, 2019 Quanta System SPA Francesco Dell'Antonio Vice President Regulatory Affairs and QA via Acquedotto 109 Samarate (Va), 21017 IT Re: K192990 Trade/Device Name: Litho Evo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 27, 2019 Received: October 25, 2019 Dear Francesco Dell'antonio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192990 Device Name Litho EVO #### Indications for Use (Describe) The Litho EVO laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectory, Gynaecology, ENT and General Surgery. Urology Open and endoscopic surgery (incision, excision, resection, ablation, - vaporization, coagulation and haemostasis) including: - · Urethral Strictures - · Bladder Neck Incisions (BNI) - · Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors, - · Ablation of Benign Prostatic Hypertrophy (BPH). - · Transurethral incision of the prostate (TUIP) - · Holmium Laser Resection of the Prostrate (HoLRP) - · Holmium Laser Enucleation of the Prostate (HoLEP) - · Holmium laser Ablation of the Prostate (HoLAP) - · Condylomas - · Lesions of external genitalia Lithotripsy and Percutaneous Urinary Lithotripsy · Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate, - monohydrate and calcium oxalate dehydrate stones. - · Endoscopic fragmentation of kidney calculi - · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. # Gastroenterology Open and endoscopic Gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: - · Appendectomy - Polyps - · Biopsy - · Gall Bladder calculi - Biliary/Bile duct calculi - · Ulcers - · Gastric ulcers - · Duodenal ulcers - · Non Bleeding Ulcers - Pancreatitas - Haemorrhoids #### FORM FDA 3881 (1/14) {3}------------------------------------------------ - · Cholecystectomy - Benign and Malignant Neoplasm - Angiodysplasia - Colorectal cancer - · Telangiectasias - · Telangiectasias of the Osler-Weber-Renu disease - Vascular Malformation - · Gastritis - · Esophagitis - Esophageal ulcers - Varices - · Colitis - · Mallory-Weiss tear - · Gastric Erosions # Arthroscopy Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including: - · Ligament and tendon Release - · Contouring and sculpting of articular surfaces - · Capsulectomy in the Knee - · Chondreplasty in the Knee - Debridement of inflamed synovial tissue - Chondromalacia Ablation - Chondromalacia and tears - · Plica Removal - · Meniscectomy - Loose Body Debridement - · Lateral retinecular release Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including - · Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy - Percutaneous Thoracic Disc Decompression/Discectomy # Gynaecology Open and laparoscopic gynaecological surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) of soft tissue # ENT Endoscopic endonasal surgery (incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and cartilage) including: - · Endonasal/sinus Surgery - · Partial turbinectomy - · Polypectomy - Dacryocystorhinostomy - Frontal Sinusotomy - Ethmoidectomy - · Maxillary antrostomy {4}------------------------------------------------ #### · Functional endoscopic sinus surgery General Surgery Open, laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including: - · Appendectomy - Skin incision - · Excision of external and internal lesions - · Complete of partial resection of internal organs, tumors and lesions - Biopsv Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # K192990 510(K) SUMMARY | Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy, 21017 | |--------------------------------------------------|-------------------------------------------------------------------------------------------------| | 510(k) Contact Person: | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA | | | Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 | | Date Prepared: | 27th September 2019 | | Common name: | Holmium laser device | | Brand name: | Litho Evo | | Classification: | Class II | | Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. | | Regulation Number: | 21 CFR 878.4810 | | Product Code: | GEX | | Basis for Submission: | Device modifications | | Legally Marketed Device: | Olympus Empower H35 (K180426) – Quanta System SPA | | Reference predicate device: | Litho 100 (K180922) – Quanta System SPA | The modified device Litho Evo is derived from the legally marketed (unmodified) device Olympus Empower H35 (K180426). # Performance Standards: There are no mandatory performance standards for this device. {6}------------------------------------------------ #### Description of the device: Litho Evo is based on Holmium laser cavity emitting at 2100nm. Laser emission is triggered by mean of a footswitch and is delivered to the tissue through fiber optics. The main parts (subsystems) of the subject device are the Holmium laser source, the optical delivery system, the power electronics, the control electronics and the cooling system. Moreover, a specific software controls the device functions and allows the user selections through the touchscreen display. #### Description of the modifications: This Special 510(k) of Litho Evo is submitted due to Device Modifications of the already cleared device Olympus Empower H35 (K180426) due to some technical modifications. The subject device has the same technological characteristics (design, material, components, energy source) as the predicate device except for the differences shown in the table below. | | Subject device Litho Evo | Main predicate device<br>Empower H35 | Reference predicate<br>device Litho 100 | |---------------------------------------------|-------------------------------------------------------|--------------------------------------|-----------------------------------------| | External<br>appearance | Cosmetic change from the<br>main predicate | - | - | | Graphic user<br>interface and<br>software | Same as the reference<br>predicate | - | - | | Footswitch | Double footswitch, Same as<br>the reference predicate | Single footswitch | Double footswitch | | Hardware,<br>components and<br>performances | Same as main predicate | - | - | The modified device has the same intended use of the unmodified device. Moreover, the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications. Based on the nature of the changes implemented, the device underwent and successfully passed performance testing and software verifications and validation according to the relevant standards. A comparison of the technical characteristics of the subject device and its predicates follows: | | Subject device | Main predicate device | Reference predicate device | |--------------|-----------------------------------|-----------------------------------|-----------------------------------| | model name | Litho Evo | Empower H35 | Litho 100 | | manufacturer | Quanta System SpA | Quanta System SpA | Quanta System SpA | | 510(k) | - | K180426 | K180922 | | Laser Source | Pulsed Holmium laser<br>(CHT:YAG) | Pulsed Holmium laser<br>(CHT:YAG) | Pulsed Holmium laser<br>(CHT:YAG) | {7}------------------------------------------------ | | Subject device | Main predicate device | Reference predicate<br>device | |----------------------|-------------------|--------------------------|-------------------------------| | Wavelength<br>(nm) | 2.1 μm | 2.1 μm | 2.1 μm | | Emission | pulsed | pulsed | pulsed | | Pulse<br>duration | Up to 1100µs | Up to 1500 µs | Up to 1100µs | | Energy per<br>pulse | up to 5.0 Joule | up to 5.0 Joule | up to 5.0 Joule | | Frequency | from 3 to 30 Hz | from 3 to 30 Hz | from 3 to 80 Hz | | Max average<br>power | 35 W | 35 W | 105W | | Delivery<br>system | Optical fibers | Optical fibers | Optical fibers | | Aiming beam | green 532nm < 5mW | Green diode laser < 5 mW | Green diode laser < 5<br>mW | # Intended Use/Indications for Use Litho Evo has the same intended use of the unmodified device, as follows: The Litho Evo laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery. # Urology Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: - Urethral Strictures - Bladder Neck Incisions (BNI) - Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors, ● - Ablation of Benign Prostatic Hypertrophy (BPH), - Transurethral incision of the prostate (TUIP) - . Holmium Laser Resection of the Prostrate (HoLRP) - Holmium Laser Enucleation of the Prostate (HoLEP) - Holmium laser Ablation of the Prostate (HoLAP) - Condylomas - Lesions of external genitalia # Lithotripsy and Percutaneous Urinary Lithotripsy {8}------------------------------------------------ - Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, ● calcium oxalate, monohydrate and calcium oxalate dehydrate stones. - Endoscopic fragmentation of kidney calculi - . Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. #### Gastroenterology Open and endoscopic Gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: - Appendectomy - Polyps - Biopsy - . Gall Bladder calculi - Biliary/Bile duct calculi - Ulcers - Gastric ulcers ● - Duodenal ulcers - Non Bleeding Ulcers - Pancreatitas - Haemorrhoids - Cholecystectomy - Benign and Malignant Neoplasm - Angiodysplasia - Colorectal cancer - Telangiectasias - Telangiectasias of the Osler-Weber-Renu disease - Vascular Malformation - Gastritis - Esophagitis - Esophageal ulcers - Varices - . Colitis - Mallory-Weiss tear ● - Gastric Erosions # Arthroscopy Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including: - . Ligament and tendon Release {9}------------------------------------------------ - Contouring and sculpting of articular surfaces - Capsulectomy in the Knee - Chondreplasty in the Knee - Debridement of inflamed synovial tissue - Chondromalacia Ablation - Chondromalacia and tears - Plica Removal - Meniscectomy - Loose Body Debridement - Lateral retinecular release Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including - Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, . including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy #### Gynaecology Open and laparoscopic gynaecological surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) of soft tissue # ENT Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and cartilage) including: - Endonasal/sinus Surgery - Partial turbinectomy - Polypectomy - Dacryocystorhinostomy - Frontal Sinusotomy - Ethmoidectomy - Maxillary antrostomy - Functional endoscopic sinus surgery # General Surgery Open, laparoscopic and endoscopic surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) including: - Appendectomy - Skin incision - Excision of external and internal lesions - Complete of partial resection of internal organs, tumors and lesions ● {10}------------------------------------------------ - . Biopsy #### Performace testing Litho Evo was subjected to performance testing in accordance with the following recognized consensus standards related to electromagnetic compatibility, electrical safety and performances: - -IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance - । IEC 60601-1-2 Edition 4: 2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests. - -IEC 60601-2-22: 2012-10 ed 3.1, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment - -IEC 60825-1 Ed. 3.0 (2014) Safety of laser products - Part 1: Equipment classification and requirements - -Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Litho Evo passed all the required testing and is in compliance will all applicable sections of the abovementioned performance standards. #### Comparison with predicate device: The subject and unmodified devices have intended use and the same fundamental scientific technology, based on Holmium laser sources. #### Summary The modified device Litho Evo is substantially equivalent to its identified predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...