← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K192160

# EVO Platform (K192160)

_Quanta System Spa · GEX · Sep 6, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K192160

## Device Facts

- **Applicant:** Quanta System Spa
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Sep 6, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The EVO Platform is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows. 1064 & 532 nm (O-Switched, nanosecond mode) The EVO Platform is intended for treatment of vascular lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of: 532 nm (Q-Switched, nanosecond mode) Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos Treatment of vascular lesions including, but not limited to: - port wine birthmarks - telangiectasias - spider angiomaa - Cherry angioma - Spider nevi Treatment of benign pigmented lesions including, but not limited to: - cafe-au-Iait birthmarks - Ephalides, solar lentigines - senile lentigines - Becker's nevi - freckles - common nevi - nevus spilus - Ota Nevus Treatment of seborrheic keratosis Treatment of post inflammatory hyperpigmentation Skin resurfacing procedures for the treatment of acne scars and wrinkles. 1064 nm (O-Switched, nanosecond mode) Removal of dark ink (black, blue and brown) tattoos Removal of benign pigmented lesions including; - nevus of Ota - Café au lait spot - Ephalides, solar lentigo (lentigines) - Becker Nevus - Nevus spilus Treatment of common nevi Removal or lightening of unwanted hair Skin resurfacing procedures for the treatment of acne scars and wrinkles 1064 nm (non O-Switched - free running mode) Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Treatment of wrinkles. Treatment of mild to moderate inflammatory acne vulgaris. 1064 nm (pulsed) Dermatology/Plastic Surgery: Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lessons such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. The laser is also indicated for the treatment of facial wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The laser is also indicated for pigmented lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. It is indicated for use on all skin types (Fitzpatrick I-VI ) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types. 532 nm (pulsed) Intended for the coagulation and hemostasis of vascular lesions. Dermatology/Plastic Surgery: For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or ex-tremities telangiectasias, venous anomalies, leg veins]; benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size). 694 nm (Q-Switched) Indicated for: Tattoo removal: Suggested for blue, sky blue, black, green and violet ink Pigmented lesion removal (benign): - Cafe au lait spot - Ephalides, solar lentigo lentigines) - Becker Nevus - Ota and Ito Nevus - Nevus spilus - Mongolian spot 694 nm (non q-switch – free running mode) Intended to remove benign dermal and epidermal pigmented lesions, and, to effect hair removal of patients with skin types 1-4 through selective targeting of melanin in hair follicles in dermatology and plastic surgery. 755 nm (pulsed) Indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles. 1320 nm (pulsed) Indicated for use in general surgery and dermatology for the incision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars. IPL 590-1200nm; 625-1200nm; 650-1200nm Indicated for permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime IPL 550-1200nm; 570-1200nm Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions. IPL 400-1200nm Indicated for inflammatory acne (acne vulgaris). Integrated Skin Cooler The intended use of the integrated cooling system in the EVO Platform is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments.

## Device Story

EVO Platform is a multi-wavelength laser and IPL system for dermatological and surgical soft tissue applications; utilizes various wavelengths (532nm, 694nm, 755nm, 1064nm, 1320nm, and IPL ranges) to target tissue via photothermolysis, ablation, or coagulation; includes integrated skin cooling system to reduce pain and side effects; operated by physicians in clinical settings; provides therapeutic energy delivery for hair removal, tattoo removal, lesion treatment, and skin resurfacing; output controlled by clinician to achieve specific tissue effects; benefits include precise tissue management and improved aesthetic/clinical outcomes.

## Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.

## Technological Characteristics

Multi-wavelength laser and IPL system (532nm, 694nm, 755nm, 1064nm, 1320nm, IPL 400-1200nm); includes integrated skin cooling system; energy delivery via pulsed, Q-switched, and free-running modes; intended for soft tissue incision, excision, ablation, vaporization, and coagulation; Class II device (21 CFR 878.4810).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 6, 2019

Quanta System, S.p.A Francesco Dell'Antonio V.P. Regulatory Affairs & Quality Assurance Via Acquedotto 109 Samarate, IT 20826 VA

Re: K192160

Trade/Device Name: EVO Platform Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 30, 2019 Received: August 9, 2019

Dear Francesco Dell'Antonio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Acting Assistant Director Light Based Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name EVO Family

Indications for Use (Describe)

The EVO Platform is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows.

Indications for use

1064 & 532 nm (O-Switched, nanosecond mode) The EVO Platform is intended for treatment of vascular lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of:

532 nm (Q-Switched, nanosecond mode)

Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos

Treatment of vascular lesions including, but not limited to:

- port wine birthmarks
- telangiectasias
- spider angiomaa
- Cherry angioma
- Spider nevi

Treatment of benign pigmented lesions including, but not limited to:

- cafe-au-Iait birthmarks
- Ephalides, solar lentigines
- senile lentigines
- Becker's nevi
- freckles
- common nevi
- nevus spilus
- Ota Nevus

Treatment of seborrheic keratosis

Treatment of post inflammatory hyperpigmentation

Skin resurfacing procedures for the treatment of acne scars and wrinkles.

1064 nm (O-Switched, nanosecond mode)

Removal of dark ink (black, blue and brown) tattoos

Removal of benign pigmented lesions including;

- nevus of Ota
- Café au lait spot
- Ephalides, solar lentigo (lentigines)
- Becker Nevus
- Nevus spilus

Treatment of common nevi

Removal or lightening of unwanted hair

Skin resurfacing procedures for the treatment of acne scars and wrinkles

FORM FDA 3881 (1/14)

{3}------------------------------------------------

1064 nm (non O-Switched - free running mode)

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin.

Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to port wine stains,

hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins.

Coagulation and hemostasis of soft tissue.

Treatment of wrinkles.

Treatment of mild to moderate inflammatory acne vulgaris.

1064 nm (pulsed)

Dermatology/Plastic Surgery:

Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis.

The laser is also intended for the treatment of benign pigmented lessons such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, verrucae, skin tags, keratosis and plaques.

The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. The laser is also indicated for the treatment of facial wrinkles.

Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

The laser is also indicated for pigmented lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

It is indicated for use on all skin types (Fitzpatrick I-VI ) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types.

532 nm (pulsed)

Intended for the coagulation and hemostasis of vascular lesions.

## Dermatology/Plastic Surgery:

For photocoagulation and hemostasis of vascular and cutaneous lesions in dermatology including but not limited to the following general categories: vascular lesions [angiomas (port wine), telangiectasia (facial or ex-tremities telangiectasias, venous anomalies, leg veins]; benign pigmented lesions (nevi, lentigines, chloasma, cafe au- lait, tattoos (red and green ink), verrucae, skin tags, keratoses, plaques, cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).

694 nm (Q-Switched)

Indicated for:

Tattoo removal: Suggested for blue, sky blue, black, green and violet ink

Pigmented lesion removal (benign):

- Cafe au lait spot
- Ephalides, solar lentigo lentigines)
- Becker Nevus
- Ota and Ito Nevus
- Nevus spilus
- Mongolian spot

{4}------------------------------------------------

694 nm (non q-switch – free running mode)

Intended to remove benign dermal and epidermal pigmented lesions, and, to effect hair removal of patients with skin types 1-4 through selective targeting of melanin in hair follicles in dermatology and plastic surgery.

755 nm (pulsed)

Indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.

1320 nm (pulsed)

Indicated for use in general surgery and dermatology for the incision, ablation, vaporization, coagulation and haemostasis of soft tissue. It is also indicated for the treatment of perioral wrinkles, fine lines and wrinkles, and the treatment of back acne and atrophic acne scars.

IPL 590-1200nm; 625-1200nm; 650-1200nm

Indicated for permanent hair removal.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime

IPL 550-1200nm; 570-1200nm

Indicated for photocoagulation of dermatological vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions.

IPL 400-1200nm

Indicated for inflammatory acne (acne vulgaris).

Integrated Skin Cooler

The intended use of the integrated cooling system in the EVO Platform is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesion, and to reduce the potential side effects of laser treatments.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

{5}------------------------------------------------

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K192160](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K192160)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
