K173700 · Lutronic Corporation · GEX · Jun 19, 2018 · General, Plastic Surgery
Device Facts
Record ID
K173700
Device Name
PICOPLUS Laser System
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The PICOPLUS Laser System is indicated for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery as follows: The 1064 nm wavelength and 450 pico-second pulse width of the PICOPLUS system is indicated for: · Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple. · Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. The 532 nm wavelength and 450 pico-second pulse width of the PICOPLUS system is indicated for: - · Removal of tattoos for Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - · Treatment of benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The 1064 nm wavelength and nanosecond pulse width of the PICOPLUS system is indicated for: - · Treatment of benign pigmented lesions on Fitzpatrick skin type I-VI - Removal of dark and multi-colored tattoos containing dark colored tattoo inks on Fitzpatrick skin types 1-VI The 532 nm wavelength and nanosecond pulse width of the PICOPLUS system is indicated for: - · Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV - Removal of lighter colored tattoo inks, including red and yellow inks, on Fitzpatrick skin types I-III
Device Story
Surgical laser system; delivers 1064 nm and 532 nm wavelengths; pulse durations of 450 picoseconds (ps) and 2 nanoseconds (ns). Components: console, articulated arm, handpieces. Used in dermatology, general, and plastic surgery clinics by physicians. Laser energy targets tattoo inks and benign pigmented lesions; thermal/photoacoustic effect breaks down pigments. Output: laser energy delivery to skin. Clinical decision-making: physician selects wavelength/pulse duration based on lesion type, skin type, and ink color. Benefits: removal of tattoos and pigmented lesions.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing and comparison to predicate devices.
Technological Characteristics
Laser surgical instrument; 1064 nm and 532 nm wavelengths; 450 ps and 2 ns pulse durations. Complies with ISO 13485:2003, IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for tattoo removal and treatment of benign pigmented lesions in dermatology and plastic surgery. Patient population: Fitzpatrick skin types I-VI (depending on wavelength/pulse duration). Contraindications: None explicitly stated beyond specific skin type/ink color limitations per wavelength.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 19, 2018
Lutronic Corporation Dr. James Childs Clinical Science Director, Lutronic Global 19 Fortune Drive Billerica, MA 01821
Re: K173700
Trade/Device Name: PICOPLUS Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 18, 2018 Received: May 18, 2018
Dear Dr. James Childs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173700
Device Name PICOPLUS Laser System
#### Indications for Use (Describe)
The PICOPLUS Laser System is indicated for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery as follows:
The 1064 nm wavelength and 450 pico-second pulse width of the PICOPLUS system is indicated for:
· Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
· Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
The 532 nm wavelength and 450 pico-second pulse width of the PICOPLUS system is indicated for:
- · Removal of tattoos for Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange.
- · Treatment of benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.
The 1064 nm wavelength and nanosecond pulse width of the PICOPLUS system is indicated for:
- · Treatment of benign pigmented lesions on Fitzpatrick skin type I-VI
- Removal of dark and multi-colored tattoos containing dark colored tattoo inks on Fitzpatrick skin types 1-VI
The 532 nm wavelength and nanosecond pulse width of the PICOPLUS system is indicated for:
- · Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV
- Removal of lighter colored tattoo inks, including red and yellow inks, on Fitzpatrick skin types I-III
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|------------------------------------------------------------|
|------------------------------------------------------------------------------------|------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 6. 510(k) SUMMARY OR STATEMENT
The Company's 510(k) Summary for the PICOPLUS Laser System is as follows:
{4}------------------------------------------------
# 510(k) Summary
# Lutronic Corporation PICOPLUS Laser System
This 510(k) Summary is being submitted in accordance with 21 CFR § 807.92.
## 1. General Information
| Applicant: | Lutronic Corporation<br>Lutronic Center, 219 Sowon-ro<br>Haengsin-dong, Deogyang-gu<br>Goyang-si, Gyeonggi-do 410-220<br>Republic of Korea<br>Tel: (82) 31-908-3440<br>FAX: (82) 31-907-3440 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | James J Childs, PhD<br>Clinical Science Director |
| | Lutronic Corporation<br>19 Fortune Drive<br>Billerica, MA 01821<br>Tel: 508-446-1126<br>FAX: 781-262-0647<br>Email: jchilds@lutronic.com |
| Summary Preparation Date: | June 18, 2018 |
# 2. Names
Trade Name: PICOPLUS Laser System
Common Name: Laser Surgical Instrument
{5}------------------------------------------------
Classification Name:
Powered Laser Surgical Instrument Product Code: GEX 21 CFR § 878.4810 Panel: General & Plastic Surgery
# 3. Predicate Devices
Cutera Enlighten III Laser System (K172077) Quanta System S.P.A. Discovery Pico Family (K163222)
# 4. Device Description
The PICOPLUS Laser System is a surgical laser instrument. The PICOPLUS Laser System is designed and manufactured in accordance with 21 CFR 820 (with emphasis on 820.30) and 21 CFR 1040 for medical devices in order to insure user safety and durability of use.
The PICOPLUS Laser System is a laser system, delivering energy at wavelengths of 1064 nm and 532 nm, both at pulse durations of 450 picoseconds (ps) and 2 nanoseconds (ns). The laser system console, an articulated arm and attached handpieces.
# 5. Indications for Use
The PICOPLUS Laser System is indicated for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery as follows:
The 532 nm wavelength and 450 pico-second pulse duration of the PICOPLUS system is indicated for
{6}------------------------------------------------
- Removal of tattoos for Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. .
- Treatment of benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. .
The 1064 nm wavelength and 450 pico-second pulse duration of the PICOPLUS system is indicated for:
- Removal of tattoos on all skin types (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, . blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. .
The 1064 nm wavelength and nanosecond pulse duration of the PICOPLUS system is indicated for:
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-VI .
- Removal of dark and multi-colored tattoos containing dark colored tattoo inks on Fitzpatrick skin types I-VI .
The 532 nm wavelength and nanosecond pulse duration of the PICOPLUS system is indicated for:
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV .
- . Removal of lighter colored tattoo inks, including red and yellow inks, on Fitzpatrick skin types I-III
{7}------------------------------------------------
# 6. Substantial Equivalence
Tables 1a & 1b below compare the device specifications for each Indication for Use of the Subject and Predicate devices.
## Table 1a. Comparison of IFUs and device specifications at Nanosecond Pulse Durations
(Note the Predicate device IFUs and device specifications table or side-by-side comparison to the Subject device's Hand 2.)
| | Comparison of device specifications at Nanosecond Pulse Duration | | | | | | | | | | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------------------|-----------------------------------|-----------------------|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------------------|-----------------------------------------------------|------------------------------------------|
| Lutronic Corporation PICOPLUS Laser System | | | | | | Cutera Enlighten III | | | | | |
| Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence | Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence |
| Handpiece<br>(2 ns) | Indicated for:<br>• Treatment of benign<br>pigmented lesions on<br>Fitzpatrick skin type I-<br>VI<br>• Removal of dark and<br>multi-colored tattoos<br>containing dark<br>colored tattoo inks on<br>Fitzpatrick skin types<br>1-VI | 1064<br>nm | 800 mJ | 2, 3, 4,<br>5, 6<br>mm | 0.18~<br>10<br>J/cm² | Handpiece<br>(2 ns) | Indicated for treatment<br>of benign pigmented<br>lesions on patients with<br>all skin types<br>(Fitzpatrick I-VI); tattoo<br>removal for dark<br>colored tattoo inks and<br>for multicolored tattoos<br>containing inks on<br>patients with all skin<br>types (Fitzpatrick I-VI) | 1064<br>nm | 800 mJ | 2, 2.5, 3,<br>3.5, 4, 5,<br>6, 7, 8,<br>9, 10<br>mm | 10 J/cm²<br>(range<br>not<br>available) |
| | Indicated for:<br>• Treatment of benign<br>pigmented lesions on<br>Fitzpatrick skin types I-<br>IV<br>• Removal of lighter<br>colored tattoo inks,<br>including red and yellow<br>inks, on Fitzpatrick skin<br>types I-III | 532<br>nm | 300 mJ | 2.3,<br>3.3,<br>4.3,<br>5.3<br>mm | 0.05~<br>2.5<br>J/cm² | | Indicated for treatment<br>of benign pigmented<br>lesions for Fitzpatrick<br>skin types I-IV; tattoo<br>removal with lighter<br>colored inks including<br>red and yellow inks for<br>Fitzpatrick skin types I-<br>III | 532<br>nm | 400 mJ | 2, 2.5, 3,<br>3.5, 4, 5,<br>6, 7, 8,<br>9, 10<br>mm | 2.5 J/cm²<br>(range<br>not<br>available) |
{8}------------------------------------------------
| | | Lutronic Corporation PICOPLUS Laser System | | | | Cutera Enlighten III | | | | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------|------------------------------|-----------------------|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------------------|-----------------------------------------------------|------------------------------------------|
| Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence | Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence |
| Handpiece<br>2<br>(2 ns) | Indicated for:<br>• Treatment of benign<br>pigmented lesions on<br>Fitzpatrick skin type I-<br>VI<br>• Removal of dark and<br>multi-colored tattoos<br>containing dark<br>colored tattoo inks on<br>Fitzpatrick skin types<br>1-VI | 1064<br>nm | 800 mJ | 6, 7, 8,<br>9, 10<br>mm | 0.07~<br>2.8<br>J/cm² | Handpiece<br>(2 ns) | Indicated for treatment<br>of benign pigmented<br>lesions on patients with<br>all skin types<br>(Fitzpatrick I-VI); tattoo<br>removal for dark<br>colored tattoo inks and<br>for multicolored tattoos<br>containing inks on<br>patients with all skin<br>types (Fitzpatrick I-VI) | 1064<br>nm | 800 mJ | 2, 2.5, 3,<br>3.5, 4, 5,<br>6, 7, 8,<br>9, 10<br>mm | 10 J/cm²<br>(range<br>not<br>available) |
| | Indicated for:<br>• Treatment of benign<br>pigmented lesions on<br>Fitzpatrick skin types I-<br>IV<br>• Removal of lighter<br>colored tattoo inks,<br>including red and yellow<br>inks, on Fitzpatrick skin<br>types I-III | 532<br>nm | 300 mJ | 4.3,<br>5.3,<br>6.5, 8<br>mm | 0.02~<br>2.0<br>J/cm² | | Indicated for treatment<br>of benign pigmented<br>lesions on patients with<br>Fitzpatrick skin types I-<br>IV; tattoo removal for<br>lighter colored tattoo<br>inks, including red and<br>yellow inks, on patients<br>with Fitzpatrick skin<br>types I-III | 532<br>nm | 400 mJ | 2, 2.5, 3,<br>3.5, 4, 5,<br>6, 7, 8,<br>9, 10<br>mm | 2.5 J/cm²<br>(range<br>not<br>available) |
Table 1a. Comparison of IFUs and device specifications at Nanosecond Pulse Durations (Cont.)
{9}------------------------------------------------
{10}------------------------------------------------
## Table 1b. Comparison of IFUs and device specifications at Picosecond Pulse Durations (Cont.)
(Note the Predicate device IFUs and device specifications table-by-side comparison to the Subject device's Hand 2.
| Comparison of device specifications at Picosecond Pulse Duration | | | | | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Lutronic Corporation PICOPLUS Laser System | | | Discovery Pico Family | | |
| Handpiece | Indications for Use | Wave-<br>length | Handpiece | Indications for Use | Wave-<br>length |
| Handpiece<br>1 (450 ps) | Indicated for:<br>• Removal of tattoos<br>on all skin type<br>(Fitzpatrick skin<br>types I-VI) with the<br>following tattoo<br>colors: black,<br>brown, green, blue<br>and purple.<br>Indicated for:<br>• Removal of tattoos for<br>Fitzpatrick skin types I-<br>III with the following<br>tattoo colors: red,<br>yellow and orange.<br>• Treatment of benign<br>pigmented lesions<br>removal for Fitzpatrick<br>Skin Types I-IV | 1064 nm<br><br>532 nm | Handpiece<br>(450 ps) | Indicated for removal of<br>tattoos on all Fitzpatrick<br>skin types I- VI with<br>following colors: black,<br>brown, green, blue, and<br>purple.<br>Indicated for treatment of<br>benign pigmented lesions<br>on Fitzpatrick types I-IV<br>Indicated for removal of<br>tattoos on Fitzpatrick<br>skin types I-III with the<br>following colors: red,<br>yellow and orange.<br>Indicated for treatment of<br>benign pigmented lesions<br>on Fitzpatrick Skin Types<br>I-IV | 1064 nm<br><br>532 nm |
| Max<br>Pulse<br>Energy | Spot<br>Size | Fluence<br>(range) | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence |
| 800 mJ<br><br>300 mJ | 3, 4, 5,<br>6 mm<br><br>3.3, 4.3,<br>5.3 mm | 0.18 -<br>5.6 J/cm²<br><br>0.05 -<br>2.8 J/cm² | 800 mJ<br><br>300 mJ | 3, 4.5,<br>6, 7.5<br>9, 10.5 mm<br><br>3, 4.5,<br>6, 7.5<br>9, 10.5 mm | 5.6 J/cm²<br>(range not<br>available)<br><br>2.8 J/cm²<br>(range not<br>available) |
{11}------------------------------------------------
| Comparison of device specifications at Picosecond Pulse Duration | | | | | | | | | | | |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------------------|---------------------------|------------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------------------|------------------------------------|------------------------------------------|
| Lutronic Corporation PICOPLUS Laser System | | | | | | Discovery Pico Family | | | | | |
| Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence<br>(range) | Handpiece | Indications for Use | Wave-<br>length | Max<br>Pulse<br>Energy | Spot<br>Size | Fluence |
| Handpiece<br>2 (450<br>ps) | Indicated for:<br>• Removal of tattoos<br>on all skin type<br>(Fitzpatrick skin<br>types I-VI) with the<br>following tattoo<br>colors: black,<br>brown, green, blue<br>and purple.<br>• Treatment of benign<br>pigmented lesions on<br>Fitzpatrick skin types I-<br>IV | 1064<br>nm | 800 mJ | 6, 7, 8,<br>9, 10<br>mm | 0.07 -<br>2.8<br>J/cm² | Handpiece<br>(450 ps) | Indicated for removal of<br>tattoos on all Fitzpatrick<br>skin types I-VI with<br>following colors: black,<br>brown, green, blue, and<br>purple.<br>Indicated for treatment of<br>benign pigmented lesions<br>on Fitzpatrick types I-IV | 1064<br>nm | 800 mJ | 3, 4.5,<br>6, 7.5<br>9, 10.5<br>mm | 5.6 J/cm²<br>(range<br>not<br>available) |
| | Indicated for:<br>• Removal of tattoos for<br>Fitzpatrick skin types I-<br>III with the following<br>tattoo colors: red,<br>yellow and orange.<br>• Treatment of benign<br>pigmented lesions<br>removal for Fitzpatrick<br>Skin Types I-IV | 532<br>nm | 300 mJ | 4.3, 5.3,<br>6.5, 8<br>mm | 0.02 -<br>2.0<br>J/cm² | | Indicated for removal of<br>tattoos on Fitzpatrick<br>skin types I-III with the<br>following colors: red,<br>yellow and orange.<br>Indicated for treatment of<br>benign pigmented lesions<br>on Fitzpatrick Skin Types<br>I-IV | 532<br>nm | 300 mJ | 3, 4.5,<br>6, 7.5<br>9, 10.5<br>mm | 2.8 J/cm²<br>(range<br>not<br>available) |
Table 1b. Comparison of IFUs and device specifications at Picosecond Pulse Durations (Cont.)
{12}------------------------------------------------
The PICOPLUS Laser System shares the same indications for use, device operation, technical and functional capabilities. The devices' fluences for the 1064 nm and 532 nm wavelengths at both 450 ps and 2 ns pulse durations are all the same.
Therefore the handpieces of the Subject device are substantially equivalent to the corresponding handpieces of the Predicate devices for the Indications for Use.
# 7. Performance Data
The Company's Performance Data for the PICOPLUS Laser System is as follows:
### Bench Testing
The PICOPLUS Laser System complies with all applicable standards, including ISO 13485:2003, ISO 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility.
# Clinical Testing
No performance data has been provided since the PICOPLUS Laser System is equivalent to the previously cleared predicate devices with no new issues regarding safety and effectiveness.
# 8. Conclusion
{13}------------------------------------------------
The PICOPLUS Laser System and the legally marketed Cutera Enlighten III Laser System cleared under 510(k) number K172077 and the legally marketed Quanta System S.P.A. Discovery Pico Family cleared under 510(k) number K163222, have the same intended use and Indications for Use statement. While the technological characteristics differ between the two systems, the differences have been established to be minor. Performance testing data established that the PICOPLUS Laser System is safe and effective as the legally marked predicate devices and that the PICOPLUS Laser System does not raise any different questions of safety and effectiveness than the predicate. On this basis and in accordance with 21 CFR§ 807.100(b), the PICOPLUS Laser System is substantially equivalent to the Cutera Enlighten III Laser System and the Quanta System S.P.A. Discovery Pico Family and can be legally marketed in the U.S.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.