← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K173199

# PicoSure Workstation (K173199)

_Cynosure, Inc. · GEX · Feb 7, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K173199

## Device Facts

- **Applicant:** Cynosure, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Feb 7, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

755 nm: The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal. The PicoSure workstation with the 2mm and 6mm handpieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV. 532 nm: The PicoSure 532 nm delivery system is indicated for tattoo removal and benign pigmented lesion removal in Skin Types I-III. 1064 nm: The PicoSure 1064 nm Laser Delivery System is indicated for tattoo and benign pigmented lesions removal.

## Device Story

PicoSure Workstation is a high-powered laser system delivering energy at 755 nm (Alexandrite), 532 nm, and 1064 nm wavelengths. System uses articulated arm with interchangeable handpieces or delivery systems to convert 755 nm energy to 532 nm or 1064 nm. Laser activation controlled by footswitch. Used in clinical settings by trained professionals for dermatological procedures including tattoo removal, pigmented lesion removal, acne scar treatment, and wrinkle reduction. Output is pulsed laser energy; healthcare providers use visual assessment and clinical scales (e.g., GAIS) to monitor treatment progress and efficacy. Benefits include safe, effective removal of lesions and skin rejuvenation with transient, expected side effects.

## Clinical Evidence

Prospective clinical study of 15 subjects (30 treatment areas) aged 53-70, Fitzpatrick Skin Types I-III. Evaluated 532nm delivery system for benign pigmented lesion removal. Subjects received up to 4 treatments at 4-week intervals. Primary endpoint: GAIS score at 1-month follow-up. Results: 91.1% of subjects scored '1 – Very Much Improved'. Safety: All adverse effects (redness, blistering) were transient, resolving in 2-11 days.

## Technological Characteristics

Laser types: Alexandrite (755 nm), Nd:YVO4 (532 nm, 1064 nm). Pulse duration: 450-900 ps. Repetition rates: 1-10 Hz. Spot sizes: 1.4-10 mm. Energy delivery via articulated arm and interchangeable handpieces. Footswitch activation. Software-controlled system with incremental updates for reliability. EMC testing performed on added components.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- PicoSure Workstation ([K160480](/device/K160480.md))
- RevLite Q-Switched Nd:YAG Laser System ([K133254](/device/K133254.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

February 7, 2018

Cynosure, Inc Avinash Purohit Regulatory Affairs Specialist 5 Caelisle Road Westford, Massachusetts 01886

Re: K173199

Trade/Device Name: PicoSure Workstation Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 28, 2017 Received: October 2, 2017

Dear Avinash Purohit:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173199

Device Name PicoSure Workstation

Indications for Use (Describe) 755mm:

The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal. The PicoSure Workstation with the 2mm and 6mm hand pieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I - IV.

532mm:

The PicoSure 532mm Laser Delivery System is indicated for tattoo removal and benign pigmented lesion removal in Skin Types I - III.

1064nm:

The PicoSure 1064nm Laser Delivery System is indicated for tattoo and benign pigmented lesions removal.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary for the PicoSure Workstation                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 807.92(a)(1) Submitter Information                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Applicant                                                                | Cynosure, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Address                                                                  | 5 Carlisle Road<br>Westford, MA 01886                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Phone Number                                                             | (978) 367-2450                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Fax Number                                                               | (978) 256-6556                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Establishment Registration Number                                        | 1222993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Contact Person                                                           | Mr. Avinash Purohit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Preparation Date                                                         | 28 September 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 807.92(a)(2) Name of Device                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade or Proprietary Name                                                | PicoSure Workstation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common or Usual Name                                                     | Laser Workstation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name                                                      | Powered Laser Surgical Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Panel                                                     | General & Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulation                                                               | 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code(s)                                                          | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                          | PicoSure Workstation - K160480                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                          | RevLite Q-Switched Nd:YAG Laser System - K133254                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 807.92(a)(4) Device Description                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                          | The PicoSureTM Workstation is a high-powered, Alexandrite<br>system that delivers laser energy in the 755-nm nominal<br>wavelength. The system offers fast and efficient treatment<br>through a variety of spot sizes, fluences and repetition rates. Laser<br>activation is by footswitch. In addition to the 755nm handpiece,<br>optional 532nm Laser Delivery System and/or 1064nm Laser<br>Delivery System can replace the 755nm handpiece at the distal<br>end of the articulated arm. These Delivery Systems convert the<br>755nm laser energy into a 532nm wavelength or a 1064 nm<br>wavelength, and are available in multiple spot sizes. |
| 807.92(a)(5) Intended Use of the Device                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                          | 755 nm:<br>The PicoSure Workstation is indicated for tattoo and benign<br>pigmented lesions removal. The PicoSure workstation with the<br>2mm and 6mm handpieces and the Focus Array are indicated for<br>the treatment of acne scars and wrinkles in Skin Types I-IV.                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                          | 532 nm:<br>The PicoSure 532 nm delivery system is indicated for tattoo<br>removal and benign pigmented lesion removal in Skin Types I-<br>III.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                          | 1064 nm:<br>The PicoSure 1064 nm Laser Delivery System is indicated for<br>tattoo and benign pigmented lesions removal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{4}------------------------------------------------

| 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to the Predicate |                                    |                                 |                                      |                                   |                                 |                                      |                                                        |         |
|---------------------------------------------------------------------------------------------------|------------------------------------|---------------------------------|--------------------------------------|-----------------------------------|---------------------------------|--------------------------------------|--------------------------------------------------------|---------|
| 510(k) Number                                                                                     | PicoSure Workstation<br>(KPending) |                                 |                                      | PicoSure Workstation<br>(K160480) |                                 |                                      | RevLite Q-switched Nd:YAG<br>Laser System<br>(K133254) |         |
| Laser Type                                                                                        | Nd:YVO4                            | Frequency<br>doubled<br>Nd:YVO4 | Alexandrite                          | Nd:YVO4                           | Frequency<br>doubled<br>Nd:YVO4 | Alexandrite                          | Flashlamp Excited Q-Switched<br>Nd:YAG                 |         |
| Wavelength<br>(nm)                                                                                | 1064 nm                            | 532 nm                          | 755 nm                               | 1064 nm                           | 532 nm                          | 755 nm                               | 1064 nm                                                | 532 nm  |
| Maximum<br>Average<br>Fluence (J/cm²)                                                             | 3.6 J/cm²                          | 1.5 J/cm²                       | 6.37 J/cm²                           | 3.6 J/cm²                         | 1.5 J/cm²                       | 6.37 J/cm²                           | 12 J/cm²                                               | 5 J/cm² |
| Repetition<br>Rate (Hz)                                                                           | 1, 2.5, 5, 10 Hz                   | 1, 2.5, 5, 10 Hz                | Single, 1, 2.5, 5,<br>10 Hz          | 1, 2.5, 5,<br>10 Hz               | 1, 2.5, 5,<br>10 Hz             | Single, 1, 2.5, 5,<br>10 Hz          | Single Shot, 1, 2, 5, 10 Hz                            |         |
| Pulse Duration                                                                                    | 450-900 ps                         |                                 |                                      | 450-900 ps                        |                                 |                                      | 7 – 20 ns                                              |         |
| Spot Sizes<br>(mm)                                                                                | Fixed 1.4 –<br>4.0 mm              | Fixed 1.5 –<br>3.5 mm           | Zoom 2-6 mm,<br>Fixed 6, 8, 10<br>mm | Fixed 1.4 –<br>4.0 mm             | Fixed 1.5 –<br>3.5 mm           | Zoom 2-6 mm,<br>Fixed 6, 8, 10<br>mm | 2 – 8.5 mm with 0.1 mm<br>increments                   |         |

# 807.92(b)(1) Non-clinical tests submitted

For this submission, software and electrical safety testing were not necessary for the expanded indication. However, incremental software updates required testing and was completed successfully. With regard to electrical safety, the device experienced only constructional changes as part of normal maintenance, so additional testing was not required. EMC testing was still performed on two components added to improve product reliability. All changes have been analyzed and successary. There necessary. There software updates/constructional changes do not affect the safety and efficacy of the PicoSure Workstation.

## 807.92(b)(2) Clinical tests submitted

A prospective clinical study was conducted using 30 treatment areas (2 areas for each of the 15 subjects), both male and female, in which the removal of benign pigmented lesions was performed using the PicoSure Workstation's 532nm Delivery System. Subjects in the study were 93% female, 7% male, and were between the ages (53 – 70) years old. They were a range of Fitzpatrick Skin Types I-III. All potential subjects were evaluated against the study's inclusion criteria to determine enrollment eligibility. During the course of the study, subjects received up to 4 treatments that were 4 weeks (+/- 2) apart using the PicoSure Workstation's 532mm wavelength haser energy. These subjects were then followed up on a (1 month) timeframe to assess for any adverse patient impacts at the treatment site. The clinical efficacy endpoint was evaluated through comparison of the pre-treatment images by three blinded, board certified Dermatologists as

{5}------------------------------------------------

independent reviewers. The clinical investigator evaluated progress via the GAIS [Global Aesthetic Improvement Scale], where 91.1% of the subjects at [1 month follow-up] had a score of "1 – Very Much Improved". The safety endpoint sought to course of the study. All adverse effects from the clinical study were transient (resolving in 2-11 days) in form of reching, and blistering. These are all expected side effects from laser treatments. The objectives of the study were met and concluded that the use of PicoSure's 532nm Delivery System is a safe and effective method for the removal of benign pigmented lesions in Skin Types I-III.

## 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted

The non-clinical data demonstrates support for safety and effectiveness of the device in regards to herded for treatment conditions. The clinical study demonstration using the 532mm Delivery System operated as expected and delivered safe and effective results with minimal side effects in subjects being pigmented lesion removal in Skin Types I-III.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K173199](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K173199)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
