K170073 · Lightmed Dentaltechnology Corp. · GEX · Jul 25, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170073
Device Name
LightMed Dental Laser System
Applicant
Lightmed Dentaltechnology Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LightMed Dental Laser System is intended for use in intraoral general dental surgery procedures, and in max!Ilofacial and cosmetic dentistry. It is intended for the incision, excision, vaporization, and coagulation of soft tissue and a variety of hard tissues (tooth and bone).
Device Story
LightMed Dental Laser System is an Er:YAG solid-state laser (2940 nm) for intraoral dental/maxillofacial surgery. Device comprises control panel, system controller, laser applicator, power supplies, cooling module, water spray module, and footswitch. Laser energy delivered via handpiece with sapphire tips to treatment site. Operated by dental professionals in clinical settings. System transforms electrical energy into pulsed laser output via flashlamp pumping. Output used for incision, excision, ablation, vaporization, and coagulation of hard (tooth/bone) and soft tissues. Water spray module provides cooling during operation. Clinical benefit includes precise tissue management and improved periodontal indices. Healthcare providers use device to perform surgical procedures; output directly facilitates tissue removal or modification.
Clinical Evidence
Bench testing only. Device performance validated against federal and international standards including IEC 60825-1 (laser safety), AAMI/ANSI 60601-1 (electrical safety), IEC 60601-2-22 (laser surgical equipment), and ISO 10993 series (biocompatibility). No clinical data presented.
Technological Characteristics
Er:YAG solid-state laser; 2940 nm wavelength; flashlamp pumped; pulse system; max pulse energy 700 mJ; pulse duration < 800 µs; pulse rate up to 50 Hz; average power 0.5-8.4 W. Handpiece: aluminum alloy. Tips: sapphire. Class I ME equipment, Type BF. Microprocessor-controlled. Water spray module for cooling. Complies with IEC 60601-1, IEC 60601-2-22, IEC 62304, and ISO 10993.
Indications for Use
Indicated for oral and maxillofacial surgery and dentistry including caries removal, cavity preparation, enamel etching, pit/fissure excavation, bone shaving/contouring/resection, apicoectomy, osteotomy, osseous crown lengthening, soft tissue biopsies, gingival incision/excision, gingivoplasty, gingivectomy, implant recovery, frenectomy, fibroma removal, lesion removal, operculectomy, papillectomy, gingival hypertrophy reduction, soft tissue crown lengthening, preprosthetic surgery, sulcular debridement, and endodontal procedures (root canal access, pulpotomy, pulp extirpation, root canal debridement/cleaning/preparation).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2017
LightMed Dental Technology Corp. Ms. April Cheng Regulatory Affairs Engineer 5F., No. 96, Luke 5th Rd., Luzhu Dist. Kaohsiung, Taiwan (R.O.C.) 821 Taiwan
Re: K170073
Trade/Device Name: LightMed Dental Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 23, 2017 Received: June 26, 2017
Dear Ms. Cheng
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
Lightmed Dental Laser System
#### Indications for Use (Describe)
The LightMed Dental Laser System is intended for use in intraoral general dental surgery procedures, and in max!Ilofacial and cosmetic dentistry. It is intended for the incision, excision, vaporization, and coagulation of soft tissue and a variety of hard tissues (tooth and bone).
The LightMed Dental Laser System is indicated for the following oral and maxillofacial surgery and dentistry:
- · Hard Tissue Indications of Erbium Laser Energy:
- * Caries removal
- * Cavity preparation
- * Enamel etching
- * Excavation of pits and fissures for placement of sealant
- · Bone Indications of Erbium Laser Energy:
- * Shaving, contouring, and resection of oral osseous tissue (bone)
- * Apicoectomy amputation of the root end
- * Cutting bone to prepare a window access to the apex (apices) of the root(s)
- * Osteotomy
- * Osseous crown lengthening
- · Soft Tissue and Periodontal Indications of Erbium Laser Energy:
- * Excisional and incisional biopsies
- * Exposure of unerupted teeth
- * Incision and drainage of abscesses
- * Gingival incision and excision
- * Gingivonlasties
- * Gingivectomies
- * Gingivectomy in case of hyperplasias of the gingival or excision of hyperplasias
- * Implant recovery
- * Frenectomies and frenotomies
- * Fibromatosis (fibroma removal)
- * Benign and malignant lesion removal
- * Operculectomy
- * Oral papillectomies
- * Reduction of gingival hypertrophy
- * Soft tissue crown lengthening
* Preprosthetic surgery: flabby alveolar ridge, vestibuloplasty, exposure of implants, papillomas, fibromatoses benign growths, vestibuloplasty
* Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- · Endodontal Indications of Erbium Laser Energy:
- * Tooth preparation to obtain access to root canal
- * Pulpotomy
- * Pulpotomy as an adjunct to root canal therapy
- * Pulp extirpation
- * Root canal debridement and cleaning
{3}------------------------------------------------
| * Root canal preparation including enlargement |
|------------------------------------------------|
|------------------------------------------------|
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Lightmed Dental. The logo consists of a blue abstract symbol on the left, followed by the word "LIGHTMED" in black, sans-serif font. Below "LIGHTMED" is the word "DENTAL" in white, set against a blue rounded rectangle.
#### Section 5 510(k) Summary
### I. SUBMITTER
LightMed Dental Technology Corp. 5F., No. 96, Luke 5th Rd., Luzhu Dist., Kaohsiung City, Taiwan, R.O.C. TEL: +886 7 695 5111 FAX: +886 7 695 5211 Contact Person: April Cheng / Regulatory Affairs Engineer Date Prepared: Jun 03, 2017
## II. DEVICE
Trade name: LightMed Dental Laser System Model name: Sapphire Classification name: Laser Instrument, Surgical, Powered Classification Panel: General & Plastic Surgery Regulation number: 21 CFR 878.4810, Class II Product code: GEX
#### III. PREDICATE DEVICE
Substantial equivalence to the following predicate device is as follows:
LITETOUCH Dental Laser System Decision Date:9/27/2006 K061966
# IV. DESCRIPTION OF THE DEVICE SUBJECT TO PREMERKET NOTIFI-CATION
LightMed Dental Laser System is an erbium: yttrium aluminium garnet (Er: YAG) solid-state laser that provides optical energy at 2940 nm. LightMed Dental Laser System is a delicate integrate device of optical, mechanic, electronic, water and gas, which is consists of a control panel, system controller, laser applicator, power supplies and discharger, laser cooling module, spray water module, footswitch and external remote interlock connector. The laser console is embedded in the laser handpiece, delivers laser energy to the treatment site through a laser tip attached to a handpiece. The laser
1
EL-FDA-Summary-A2 lof7
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Lightmed Dental. The logo consists of a blue geometric shape resembling a stylized flower or starburst on the left. To the right of the shape, the word "LIGHTMED" is written in a bold, sans-serif font. Below "LIGHTMED", the word "DENTAL" is written in white letters inside a blue rounded rectangle.
is activated by means of a footswitch. Various laser tips are available for different clinical applications. The LightMed Dental Laser System is intended for use in intraoral general dental surgery procedures, and in maxillofacial and cosmetic dentistry. It is intended for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue and a variety of hard tissues (tooth and bone).
# V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Trade name | LightMed Dental Laser System | LITETOUCH Dental Laser System |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model name | Sapphire | LITETOUCH™ |
| 510(k)<br>number | K170073 | K061966 |
| Unit | 11.8 x 17.3 x 35.8 in<br>(30 x 44 x 91 cm) | 10.25 x 15.75 x 15.75 in<br>(26 x 40 x 40 cm) |
| Weight | 77 lbs (35 kg) | 44 lbs (20 kg) |
| Operating<br>Voltage | 100-230 VAC | 100-230 VAC |
| Electrical<br>Frequency | 50 / 60 Hz | 50 / 60 Hz |
| Electrical<br>Requirement | Single Phase | Single Phase |
| Current<br>Rating | 12 A | 13 A |
| Optical<br>Frequency | Up to 50 Hz | Up to 50 Hz |
| Spot Size | 0.2-1.0 mm (Tip dependent) | 0.2-1.0 mm (Tip dependent) |
| Tip category | 1. 1.3×19 mm<br>2. 1.3×14 mm<br>3. 1.0×17 mm<br>4. 0.8×14 mm<br>5. 0.6×17 mm<br>6. Side Firing 1.3×19 mm<br>7. Chisel | 1. 1.3 x 19 mm<br>2. 1.3 x 17 mm<br>3. 1.3 x 14 mm<br>4. 1.0 x 17 mm<br>5. 1.0 x 14 mm<br>6. 0.8 x 17 mm<br>7. 0.8 x 14 mm<br>8. 0.6 x 17 mm<br>9. 0.4 x 17 mm<br>10. Implants tips, 1.3 x 19 mm<br>11. Chisel tip |
| Laser Tip<br>Material | sapphire | sapphire |
| Handpiece<br>Material | Aluminum alloy | Aluminum alloy |
| Type of pro-<br>tection against<br>electrical<br>shock | Class I ME Equipment | Class I ME Equipment |
| Degree of pro-<br>tection against<br>electrical<br>shock | Type BF Equipment | Type BF Equipment |
| Laser<br>Classification | 4 | 4 |
| Laser Energy<br>Source | Er: YAG (Erbium: Yttrium-<br>Aluminium-Garnet) | Er: YAG (Erbium: Yttrium-<br>Aluminium-Garnet) |
| Wavelength | 2.94 µm (2940 nm) | 2.94 µm (2940 nm) |
| Pumping<br>Source | Flashlamp | Flashlamp |
| Delivery<br>Component | Applicator with reflector and optics | Applicator with reflector and optics |
| Pulsing<br>System | Pulse | Pulse |
| Pulse Rate | Up to 50Hz | Up to 50Hz |
| Average<br>Power | 0.5-8.4 W | 0.5-8.4 W |
| Maximum<br>Pulse Energy | 700 mJ | 700 mJ |
| Pulse<br>Duration | < 800 µs | < 800 µs |
| Device System | Sapphire is an advanced micropro-<br>cessor-controlled laser system, com-<br>posed of the following sub-systems:<br>• Control Panel<br>• System Controller<br>• Laser Applicator<br>• Water Spray Module | LiteTouch is an advanced micro-<br>processor-controlled laser system,<br>composed of the following sub-<br>systems:<br>• Control Panel<br>• System Controller<br>• Laser Applicator |
| | Power Supply and Discharger Laser Cooling Module Footswitch Remote Interlock Connector | Water Spray Module Power Supply and Discharger Laser Cooling Module Footswitch Remote Interlock Connector |
| Intended Use | <b>Hard Tissue Indications of Erbium Laser Energy:</b> Caries removal Cavity preparation Enamel etching Excavation of pits and fissures for placement of sealant <b>Bone Indications of Erbium Laser Energy:</b> Shaving, contouring, and resection of oral osseous tissue (bone) Apicoectomy - amputation of the root end Cutting bone to prepare a window access to the apex (apices) of the root(s) Osteotomy Osseous crown lengthening <b>Soft Tissue and Periodontal Indications of Erbium Laser Energy:</b> Excisional and incisional biopsies Exposure of unerupted teeth Incision and drainage of abscess- | <b>Hard Tissue Indications of Erbium Laser Energy:</b> Caries removal Cavity preparation Enamel etching Enameloplasty Excavation of pits and fissures for placement of sealant <b>Bone Indications of Erbium Laser Energy:</b> Contact and non-contact cutting Shaving, contouring, and resection of oral osseous tissue (bone) Apicoectomy - amputation of the root end Cutting bone to prepare a window access to the apex (apices) of the root(s) Osteoplasty Osteotomy Osseous crown lengthening <b>Soft Tissue and Periodontal Indications of Erbium Laser Energy:</b> Excisional and incisional biopsies Exposure of unerupted teeth Incision and drainage of abscess- |
| * Gingival incision and excision | * Gingival incision and excision | |
| * Gingivoplasties | * Gingivoplasties | |
| * Gingivectomies | * Gingivectomies | |
| * Gingivectomy in case of hyper-<br>plasias of the gingival or excision<br>of hyperplasias | * Gingivectomy in case of hyper-<br>plasias of the gingival or excision<br>of hyperplasias | |
| * Implant recovery | * Gingival troughing for crown im-<br>pressions | |
| • Frenectomies and frenotomies | * Hemostasis | |
| * Fibromatosis (fibroma removal) | * Implant recovery | |
| * Benign and malignant lesion re-<br>moval | • Frenectomies and frenotomies | |
| * Operculectomy | * Fibromatosis (fibroma removal) | |
| * Oral papillectomies | * Benign and malignant lesion re-<br>moval | |
| * Reduction of gingival hypertro-<br>phy | * Operculectomy | |
| * Soft tissue crown lengthening | * Oral papillectomies | |
| * Preprosthetic surgery: flabby al-<br>veolar ridge, vestibuloplasty, expo-<br>sure of implants, hyperplasia, epu-<br>lides, papillomas, fibromatoses be-<br>nign growths, vestibuloplasty | * Reduction of gingival hypertro-<br>phy | |
| * Sulcular debridement (removal of<br>diseased or inflamed soft tissue in<br>the periodontal pocket to improve<br>clinical indices including gingival<br>index, gingival bleeding index,<br>probe depth, attachment loss and<br>tooth mobility) | * Soft tissue crown lengthening | |
| | * Preprosthetic surgery: flabby al-<br>veolar ridge, vestibuloplasty, expo-<br>sure of implants, hyperplasia, epu-<br>lides, papillomas, fibromatoses be-<br>nign growths, vestibuloplasty | |
| | * Sulcular debridement (removal of<br>diseased or inflamed soft tissue in<br>the periodontal pocket to improve<br>clinical indices including gingival<br>index, gingival bleeding index,<br>probe depth, attachment loss and<br>tooth mobility) | |
| Endodontal Indications of Erbi- | Endodontal Indications of Erbi-…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.