← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K162837

# Smart Laser Indirect Ophthalmoscope (LIO); Laser delivery device for Lumenis Novus Spectra Laser System, Lumenis Vision One Laser System, and Lumenis Smart532 Laser System (K162837)

_Lumenis, Ltd. · GEX · Feb 28, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K162837

## Device Facts

- **Applicant:** Lumenis, Ltd.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Feb 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Lumenis Smart Laser Indirect Ophthalmoscope (LIO) is a battery-powered optical instrument intended for the viewing of the posterior segment of the eye and to deliver laser energy for photocoagulation procedures on the peripheral retina of the eye (in conjunction with the use of a hand-held condensing lens). The Smart LIO is indicated for use in the following photocoagulation procedures: - Panretinal photocoagulation; - Segmental peripheral photocoagulation; - Retinopexy; and, - Pediatric retinal repairs (under general anesthesia). The Smart LIO is intended to work in conjunction with the following Lumenis laser systems in ophthalmic photocoagulation procedures: - Lumenis Novus Spectra Laser System; - Lumenis Vision One Laser System; and - Lumenis Smart532TM Laser System.

## Device Story

Battery-powered wireless headset worn by ophthalmologists; used to view posterior eye segment and deliver laser energy for photocoagulation. Input: light for visualization; laser energy from compatible Lumenis laser systems (Novus Spectra, Vision One, Smart532). Operation: physician wears headset, uses hand-held condensing lens to focus fundus image; laser beams are precisely delivered through LIO optics to patient retina. Used in clinic or OR for patients in supine position or those unsuitable for slit-lamp delivery. Output: magnified fundus visualization; targeted laser photocoagulation. Benefits: enables treatment of patients who cannot be treated via standard slit-lamp systems; provides precise retinal repair.

## Clinical Evidence

Bench testing only. No clinical data. Verification and validation included: laser beam emission tests, spot size measurements, aiming beam transmission, eye safety filter transmission, and communication testing with compatible laser systems. Environmental testing performed for storage/transport. All tests met specifications.

## Technological Characteristics

Battery-powered wireless headset; LED illumination; optical delivery of laser energy (Nd:YAG, DPSS, Nd:GdVO4). Compatible with 532, 577, 659 nm treatment wavelengths. Fixed eye safety filter (OD>5). Standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-2, ISO 14971. Weight: 800g. Connectivity: wireless communication to Lumenis laser systems.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Laser Indirect Ophthalmoscope, Keeler model ([K022181](/device/K022181.md))

## Reference Devices

- Keeler Vantage Plus Binocular Indirect Ophthalmoscope ([K060822](/device/K060822.md))
- Lumenis Smart532™ Laser System ([K151109](/device/K151109.md))
- Lumenis Vision One Laser Systems ([K111213](/device/K111213.md))
- Lumenis Novus Spectra Laser System ([K022327](/device/K022327.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure, composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 28, 2017

Lumenis Ltd. Naama Jacoby Head of Regulatory Affairs Ophthalmic Platform 6 Hakidma Street PO Box 240 Yokneam Industrial Park Yokneam, IL 2069204

Re: K162837

Trade/Device Name: Smart Laser Indirect Ophthalmoscope (LIO); Laser delivery device for Lumenis Novus Spectra Laser System, Lumenis Vision One Laser System, and Lumenis Smart5327™ Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 22, 2017 Received: January 24, 2017

Dear Ms. Jacoby:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162837

#### Device Name

Smart Laser Indirect Ophthalmoscope (LIO); Laser delivery device for Lumenis Novus Spectra Laser System, Lumenis Vision One Laser System, and Lumenis Smart532TM Laser System

### Indications for Use (Describe)

The Lumenis Smart Laser Indirect Ophthalmoscope (LIO) is a battery-powered optical instrument intended for the viewing of the posterior segment of the eye and to deliver laser energy for photocoagulation procedures on the peripheral retina of the eye (in conjunction with the use of a hand-held condensing lens).

The Smart LIO is indicated for use in the following photocoagulation procedures:

- Panretinal photocoagulation;
- Segmental peripheral photocoagulation;
- Retinopexy; and,
- Pediatric retinal repairs (under general anesthesia).

The Smart LIO is intended to work in conjunction with the following Lumenis laser systems in ophthalmic photocoagulation procedures:

- Lumenis Novus Spectra Laser System;
- Lumenis Vision One Laser System; and
- Lumenis Smart532TM Laser System.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Lumenis, a company that provides energy to healthcare. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller blue font. The logo is simple and modern, and it conveys the company's focus on providing energy-based solutions for the healthcare industry.

## 510(K) SUMMARY

## Smart Laser Indirect Ophthalmoscope (LIO);

## Laser delivery device for Lumenis Novus Spectra Laser System, Lumenis Vision One Laser System, and Lumenis Smart532™ Laser System

| Date Prepared:                 | February 26th, 2016                                                                                                                                                                                                                                              |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name<br>and Address: | Lumenis Ltd.<br>6 Hakidma Street PO Box 240<br>Yokneam Industrial Park,<br>Yokneam 2069204. Israel<br>Tel: +972-4-9599000<br>Fax: +972-4-9599046                                                                                                                 |
| Contact Person:                | Naama Jacoby - Head of Regulatory Affairs Ophthalmic Platform<br>Lumenis Ltd.<br>Tel: +972-4-9599085<br>Fax: +972-4-9599198<br>Mail: Naama.Jacoby@lumenis.com                                                                                                    |
| Trade Name:                    | Smart Laser Indirect Ophthalmoscope (LIO); Laser delivery device for<br>Lumenis Novus Spectra Laser System, Lumenis Vision One Laser<br>System, and Lumenis Smart532™ Laser System                                                                               |
| Common Name:                   | Laser Indirect Ophthalmoscope (LIO)                                                                                                                                                                                                                              |
| Classification Name:           | Powered laser surgical instrument                                                                                                                                                                                                                                |
| Product Code:                  | GEX                                                                                                                                                                                                                                                              |
| Device Class:                  | Class II                                                                                                                                                                                                                                                         |
| Regulation Number:             | 21 CFR 878.4810                                                                                                                                                                                                                                                  |
| Panel:                         | General & Plastic Surgery                                                                                                                                                                                                                                        |
| Predicate Device:              | Laser Indirect Ophthalmoscope, Keeler model, cleared under K022181.                                                                                                                                                                                              |
| Reference Devices:             | Keeler Vantage Plus Binocular Indirect Ophthalmoscope cleared under<br>K060822, Lumenis Smart532™ Laser System cleared under K151109,<br>Lumenis Vision One Laser Systems, cleared under K111213 and<br>Lumenis Novus Spectra Laser System cleared under K022327 |

{4}------------------------------------------------

### Intended Use/ Indications for Use:

The Lumenis Smart Laser Indirect Ophthalmoscope (LIO) is a battery-powered optical instrument intended for the viewing of the posterior segment of the eye and to deliver laser energy for photocoagulation procedures on the peripheral retina of the eye (in conjunction with the use of a hand-held condensing lens).

The Smart LIO is indicated for use in the following photocoagulation procedures:

- Panretinal photocoagulation:
- Segmental peripheral photocoagulation;
- Retinopexy; and,
- Pediatric retinal repairs (under general anesthesia).

The Smart LIO is intended to work in conjunction with the following Lumenis laser systems in ophthalmic photocoagulation procedures:

- Lumenis Novus Spectra Laser System;
- Lumenis Vision One Laser System; and
- Lumenis Smart532™ Laser System.

### Modified Device Description:

The Lumenis Smart Laser Indirect Ophthalmoscope (LIO) is a battery-powered optical instrument intended for the viewing of the posterior segment of the eye and to deliver laser energy for photocoagulation procedures on the peripheral retina of the eye (in conjunction with the use of a hand-held condensing lens). The Smart LIO is used in conjunction with a compatible Lumenis ophthalmic laser systems. The LIO illuminates and magnifies the fundus image for observation, and when connected to a Lumenis ophthalmic laser system, the laser aiming and treatment beams are precisely focused and delivered to the patient's eye through the LIO.

The Smart LIO is a wireless headset worn on the physician's head and is used to treat patients in a supine position or who could not otherwise be treated using a standard slit lamp delivery system.

### Technological Characteristics and Substantial Equivalence

The intended use and indications for use of the Smart LIO are the same as the selected predicate devices. In addition, the same technological characteristics and principles of operation apply for the Smart LIO and the predicate device. The modifications introduced to the subject Smart LIO as compared to the predicate device are designed and intended mainly for increased user convenience.

{5}------------------------------------------------

Comparison table of technological characteristics of the Smart LIO compared to those of the predicate device is provided below:

| Parameter                                    | Lumenis<br>LIO, Keeler model (K022181)<br>(Predicate Device)                                                                                                              | Lumenis<br>Smart LIO<br>(Subject device)                                                        |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| LIO Headset                                  | Keeler, Fison model                                                                                                                                                       | Keeler, Vantage Plus model<br>(K060822)                                                         |
| Lumenis Laser System<br>Compatibility        | • Smart532 (K151109)<br>• Vision One (K111213)<br>• Novus Spectra (K022327)<br>• Novus Varia,<br>• Novus Verdi,<br>• Novus Omni,<br>• Novus 2000,<br>• Ultima & Ulima SE, | • Smart532 (K151109)<br>• Vision One (K111213)<br>• Novus Spectra (K022327)                     |
| Laser Source &<br>Wavelengths Compatibility: |                                                                                                                                                                           |                                                                                                 |
| • Laser Source:                              | Nd:YAG, Argon, Krypton                                                                                                                                                    | Nd:YAG, DPSS, Nd:GdVO4                                                                          |
| • Treatment beam:                            | 532, 561, 659 nm                                                                                                                                                          | 532, 577, 659 nm                                                                                |
| • Aiming beam:                               | 635nm                                                                                                                                                                     | 635nm                                                                                           |
| Spot Sizes,<br>at LIO focal plane            | $910 \mu m \pm 10%$                                                                                                                                                       | $1100 \mu m \pm 20%$                                                                            |
| Working distance:                            | 366 mm (14.41")                                                                                                                                                           | 280mm (11.02")                                                                                  |
| Eye Safety Filter                            | Fixed, blocking treatment<br>wavelengths, OD>5                                                                                                                            | Fixed, blocking treatment<br>wavelengths, OD>5                                                  |
| Transmission:                                | >70%                                                                                                                                                                      | >70%                                                                                            |
| Optical Fiber Length                         | 4.6 meters                                                                                                                                                                | 5 meters                                                                                        |
| Illumination source                          | Xenon Bulb                                                                                                                                                                | LED                                                                                             |
| Dimensions (HxWxL):                          |                                                                                                                                                                           |                                                                                                 |
| • Headset                                    | 17.7cm x 20.3cm x 40.6cm                                                                                                                                                  | 18.0-22.0cm× 25.0cm × 30.0-38.0cm                                                               |
| • Power Supply / Charging<br>Station         | 26.1cm x 23.1cm x 36cm                                                                                                                                                    | 12.0cm x 18.4cm x 20.7cm                                                                        |
| Weight:                                      |                                                                                                                                                                           |                                                                                                 |
| • Headset                                    | 545 grams (1.2lbs)                                                                                                                                                        | 800 grams (1.87 lbs)                                                                            |
| • Power Supply / Charging<br>Station         | 5.44 Kg (12 lbs)                                                                                                                                                          | 0.4 Kg                                                                                          |
| Electrical Requirements                      | Power Supply;<br>• Voltage: 120/230V AC±10%;<br>• Frequency:50/60Hz<br>• Current: 0.8 Amp max                                                                             | Charging Station;<br>• Voltage: 100-240V AC,<br>• Frequency: 50/60Hz,<br>• Current: 0.7 Amp max |

Lumenis Smart Laser Indirect Ophthalmoscope (LIO); Laser delivery device for Lumenis Novus Spectra Laser 510(k) Summary - Page 3 System, Lumenis Vision One Laser System, and Lumenis Smart532TM Laser System

{6}------------------------------------------------

Performance testing was conducted in order to demonstrate the performance of the Smart LIO and its substantial equivalence, with respect to the safety and effectiveness of the cleared predicate device. The following activities were performed:

- Risk analysis activities in compliance with the requirements of ISO 14971.
- Electrical safety and electromagnetic compatibility testing as required to conform with the following performance standards:
	- IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic । safety and essential performance.
	- । IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - requirements and tests.
- . Light hazard protection compatibility testing as required to conform with the following performance standard:
	- ISO 15004-2 Ophthalmic Instruments - Fundamental requirements and test methods Part 2: Light hazard protection
- Verification and validation testing: ●
	- The binocular viewing optics and illumination controls of the Binocular Indirect — Ophthalmoscope (by Keeler Instruments, Inc within their cleared device).
	- । The modification of the laser delivery optics were verified and confirmed the Smart LIO:
		- Laser beam emission test for all laser wavelengths,
		- Laser spot size test at working distance of 280mm for all wavelengths
		- " Aiming beam transmission testing through the Smart LIO.
	- Eye safety filters transmission test; performed at laser treatment wavelengths of l 532nm, 577nm, 659 nm and aiming beam wavelength of 635nm
	- -Smart LIO communication to laser systems test; test preformed for all compatible laser systems for all connection modes.
- Environmental testing demonstrating the ability of the subject device to withstand variant ● operation, storage and transportation conditions.

Test results indicated that the subject Smart LIO performs in accordance with its requirements and specifications, in similarity to its predicate device. Consequently, the Smart LIO was found to perform as well as its predicate, to be as safe and effective for its intended use as its predicate, and is substantially equivalent to its predicate device without raising any new safety and/or effectiveness issues.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K162837](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K162837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
