← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K160413

# Sterling 5W Diode Laser (K160413)

_CAO Group, Inc. · GEX · Sep 21, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K160413

## Device Facts

- **Applicant:** CAO Group, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Sep 21, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Sterling Diode Laser is indicated for: Dental soft tissue indications: Dental, oral and soft tissue surgery including: Sulcular debridement of diseased or fibrous tisue, excision and biopsy, gingivectory and gingivoplasty, lesion (tumor) removal, fibroma removal, tissue retraction (troughing), aphthous ulcers, gingival hyperplasia (excision and recontour), crown lengthening, operculectomy, frenectomy, and photocoagulation. Laser periodontal procedures, including: Laser soft tissue curettage; laser removal of diseased, infected and necrosed soft tissue within the periodontal pocket; removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.

## Device Story

Sterling Diode Laser is a powered surgical instrument delivering laser energy to human soft tissue. Input: electrical power; Output: 810nm laser energy (max 5W) via quartz optical fiber. Principle: AlGaAs diodes generate laser energy absorbed by target tissue, converting to heat for hemostasis, ablation, or vaporization. Used in dental clinics by dentists/clinicians. Features user-definable settings, 630nm aiming beam, and continuous or pulse delivery. Activation via wireless foot pedal. Benefits: precise soft tissue surgery with reduced collateral damage.

## Clinical Evidence

Bench testing only. Performance verified through final assembly inspection (QAC-P02265) and main/safety specification verification testing.

## Technological Characteristics

AlGaAs diode laser; 810nm wavelength; 5W max power; quartz optical fiber delivery; heatsink/fan air cooling; wireless foot pedal activation; LED display/membrane keypad interface. Complies with ANSI/AAMI ES60601-1:2005, IEC 60601-1-2:2007, IEC 60601-1-6:2006, IEC 60601-2-22:2007, and IEC 60825-1:2007.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Odyssey 2.4G Diode Laser ([K050453](/device/K050453.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized eagle symbol. The eagle is depicted with three profiles facing right, representing the department's focus on health and human well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 21, 2016

CAO Group, Inc. Mr. Robert K. Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, UT 84084

Re: K160413

Trade/Device Name: Sterling 5W Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 22, 2016 Received: August 24, 2016

Dear Mr. Larsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Christopher J. Ronk -S

Fox Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K160413

Device Name Sterling Diode Laser

Indications for Use (Describe) The Sterling Diode Laser is indicated for: Dental soft tissue indications:

Dental, oral and soft tissue surgery including:

Sulcular debridement of diseased or fibrous tisue, excision and biopsy, gingivectory and gingivoplasty, lesion (tumor) removal, fibroma removal, tissue retraction (troughing), aphthous ulcers, gingival hyperplasia (excision and recontour), crown lengthening, operculectomy, frenectomy, and photocoagulation.

Laser periodontal procedures, including:

Laser soft tissue curettage; laser removal of diseased, infected and necrosed soft tissue within the periodontal pocket; removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CAO GROUP, INC. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name is the tagline "Easier · Faster · Better" in a smaller font.

4628 W. Skyhawk Drive
West Jordan, UT 84084
**www.caogroup.com**

301.256.9282 [phone]
877.877.9778 [toll free]
301.256.9287 [fax]

# 510(k) Summary

Prepared By: Robert K. Larsen Preparation Date: August 22, 2016

#### Device Name:

| Trade Name:             | Sterling Diode Laser                          |
|-------------------------|-----------------------------------------------|
| Common Name:            | Soft Tissue Diode Laser                       |
| Product Code:           | GEX                                           |
| Regulation:             | 878.4810                                      |
| Product Classification: | Powered Laser Surgical Instrument<br>Class II |

## Legally Marketed Predicate Devices for Substantial Equivalence:

Odyssey 2.4G Diode Laser, manufactured for Ivoclar Vivadent, Inc. (K050453)

#### Rationale for Substantial Equivalence:

The submitted device and identified predicate device share exactly identical indications for use: Dental soft tissue indications:

#### Dental, oral and soft tissue surgery including:

Sulcular debridement of diseased or fibrous tissue, excision and biopsy, gingivectorny and gingivoplasty, lesion (tumor) removal, fibroma removal, tissue retraction (troughing), aphthous ulcers, gingival hyperplasia (excision and recontour), crown lengthening, operculectomy, frenectomy, and photocoagulation.

Laser periodontal procedures, including:

Laser soft tissue curettage; laser removal of diseased, infected and necrosed soft tissue within the periodontal pocket; removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.

#### Description of Submitted Device:

The Sterling Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 10nm for a maximum of 5 watts of energy is delivered to the surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The target tissues absorb the laser energy converting it to heat. depending on the intensity or power output of the laser, the heat so generated can cell hemostasis, ablation, or vaporization. The device features

Page 1 of 4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for CAO GROUP, INC. The logo features a red square with a white Chinese character inside on the left. To the right of the square is the text "CAO GROUP, INC." in black, with a thin red line underneath. Below that, in a smaller font, is the text "E a s i e r . F a s t e r . B e t t e r" with a trademark symbol.

4628 W Skyhowk Drive West Jordan, UT 84084 www.caogroup.com

some user definable settings, including a selectable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options.

The optical fiber is stored within the device and dispensed to the desired length according to the clinician's need. A fiber is passed through a reusable, sterilizeable handpiece assembly and terminated with a single-use disposable tip. The operator uses the handpiece to position and direct the laser energy to the intended treatment site. The activation of the working beam diodes is accomplished by use of a foot-actuated switch.

## Description of the Predicate Device:

The Odyssey 2.4G Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The target tissues absorb the laser energy converting it to heat. depending on the intensity or power output of the laser, the heat so generated can cell hemostasis, ablation, or vaporization. The device features some user definable settings, including a selectable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options.

The optical fiber is stored within the device and dispensed to the desired length according to the clinician's need. A fiber is passed through a reusable, sterilizeable handpiece assembly and terminated with a single-use disposable tip. The operator uses the handpiece to position and direct the laser energy to the intended treatment site. The activation of the working beam diodes is accomplished by use of a foot-actuated switch.

#### Indications for Use of the Submitted Device:

The submitted device is indicated for use for -Dental soft tissue indications:

Dental, oral and soft tissue surgery including:

Sulcular debridement of diseased or fibrous tissue, excision and biopsy, gingivectomy and gingivoplasty, lesion (tumor) removal, fibroma removal, tissue retraction (troughing), aphthous ulcers, gingival hyperplasia (excision and recontour), crown lengthening, operculectomy, frenectomy, and photocoagulation.

Laser periodontal procedures, including:

Laser soft tissue curettage; laser removal of diseased, infected and necrosed soft tissue within the periodontal pocket; removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.

|                                       | Technological Characteristics and Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                       | CAO Group, Inc. Sterling Diode Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ivoclar Vivadent, Inc. Odyssey 2.4G                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Working<br>Beam Output<br>Power       | $810\pm10$ nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | $810\pm20$ nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                       | CAO Group, Inc. Sterling Diode Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ivoclar Vivadent, Inc. Odyssey 2.4G                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Working<br>Beam Output<br>Wavelength  | 0.5 - 5.0 watts                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0.1 - 5.0 watts                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Aiming Beam<br>Output Power           | < 3mW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | < 3mW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Aiming Beam<br>Output<br>Wavelength   | 630-650 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 630-650 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Laser Source                          | AlGaAs Diode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | AlGaAs Diode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Laser<br>Activation                   | Wireless foot pedal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Wireless foot pedal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Pulse Control                         | Digital emission control<br>Fixed pulse duration of 0.05 seconds, 10Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Digital emission control<br>Fixed pulse duration of 0.05 seconds, 10Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Cooling<br>Method                     | Heatsink / Fan air cooled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Heatsink / Fan air cooled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Electrical<br>Power Input<br>(System) | 100-240 VAC @ 47-63 Hz, 1.5A (switchable)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 100-240 VAC @ 47-63 Hz, 1.6A (switchable)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| User Interface                        | Illuminated LED Display;<br>Membrane keypad                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Illuminated LED Display;<br>Membrane keypad                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Laser Delivery<br>System              | Quartz optical fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Quartz optical fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Dimensions                            | 9" x 6" x 5"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 10" x 8" x 4"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Weight                                | 5.0 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 6.5 lbs (Control unit only)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 510(k)<br>Number                      | Pending this application.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K050453                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications<br>for Use                | Dental soft tissue indications:<br><br>Dental, oral and soft tissue surgery including:<br>Sulcular debridement of diseased or fibrous<br>tissue, excision and biopsy, gingivectomy and<br>gingivoplasty, lesion (tumor) removal, fibroma<br>removal, tissue retraction (troughing),<br>aphthous ulcers, gingival hyperplasia (excision<br>and recontour), crown lengthening,<br>operculectomy, frenectomy, and<br>photocoagulation.<br><br>Laser periodontal procedures, including:<br>Laser soft tissue curettage; laser removal of<br>diseased, infected and necrosed soft tissue<br>within the periodontal pocket; removal of<br>highly inflamed edematous tissue affected by | Dental soft tissue indications:<br><br>Dental, oral and soft tissue surgery including:<br>Sulcular debridement of diseased or fibrous<br>tissue, excision and biopsy, gingivectomy and<br>gingivoplasty, lesion (tumor) removal, fibroma<br>removal, tissue retraction (troughing),<br>aphthous ulcers, gingival hyperplasia (excision<br>and recontour), crown lengthening,<br>operculectomy, frenectomy, and<br>photocoagulation.<br><br>Laser periodontal procedures, including:<br>Laser soft tissue curettage; laser removal of<br>diseased, infected and necrosed soft tissue<br>within the periodontal pocket; removal of<br>highly inflamed edematous tissue affected by |

Page 2 of 4

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CAO GROUP, INC. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the text is the tagline "E a si e r · F a s t e r · B e t t e r™" in a smaller font.

4628 W. Skyhawk Drive 801.256.9282 [phone] West Jordan, UT 84084 877.877.9778 [toll free] www.caogroup.com 801.256.9287 [fax]

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for CAO Group, Inc. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name, there is a tagline that reads "Easier · Faster · Better™".

| 4628 W. Skyhawk Drive | 801.256.9282 [phone]     |
|-----------------------|--------------------------|
| West Jordan, UT 84084 | 877.877.9778 [toll free] |
| www.caogroup.com      | 801.256.9287 [fax]       |

| CAO Group, Inc. Sterling Diode Laser                                    | Ivoclar Vivadent, Inc. Odyssey 2.4G                                     |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| bacteria penetration of the pocket lining and<br>junctional epithelium. | bacteria penetration of the pocket lining and<br>junctional epithelium. |

#### Conformity to Standards:

The Sterling Diode Laser is designed to comply with the performance requirements of ANSJ/AAMI ES60601-1:2005, IEC 60601-1-2:2007, IEC 60601-1-6:2006, IEC 60601-2-22:2007, and IEC 60825-1:2007.

#### Performance Data:

Bench testing on an evaluation sample of the submitted device was performed consistent with internal requirements:

- QAC-P02265 - Final assembly inspection of Sterling Diode Laser
- . Sterling 5W Main and Safety Specification Verification Test

#### Conclusion:

The Sterling 5W Diode Laser is substantially equivalent to the listed predicate. This device shares identical intended uses, identical operating principles, similar design features, and identical functional and performance characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K160413](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K160413)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
