← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K150587

# Cortex Laser System (K150587)

_Cynosure, Inc. Dba Ellman · GEX · Jul 21, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K150587

## Device Facts

- **Applicant:** Cynosure, Inc. Dba Ellman
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jul 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The CO2 laser is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing. The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery. Dermatology, Plastic Surgery and General Surgery procedures: - Laser skin resurfacing. - . Treatment of furrows and wrinkles. - Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation. - Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications. - Blepharoplasty. ● - . Site preparation for hair transplants. - The fractional scanner is for skin resurfacing. The Er: Y AG handpiece is designed specifically for superficial skin ablation resulting in skin dermabrasion, and the treatment of wrinkles. In addition this system is intended for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in dermatology, plastic surgery (including aesthetic surgery).

## Device Story

Cortex Laser System is a dual-modality surgical laser console (CO2 and Er:YAG). CO2 laser energy is delivered via a 7-joint articulated arm; Er:YAG laser energy is delivered via a dedicated handpiece containing the laser cavity. System operated by physicians in clinical settings; user interface is an LCD touch screen; laser emission activated by footswitch. Device performs incision, ablation, vaporization, and coagulation of soft tissues. Fractional scanner accessory enables ablative skin resurfacing. Clinical benefits include precise tissue removal and resurfacing for aesthetic and surgical applications. System requires 110-120 VAC power; utilizes closed-loop liquid cooling.

## Clinical Evidence

No clinical data submitted; bench testing only.

## Technological Characteristics

CO2 laser source (10.6 micron wavelength); Er:YAG handpiece; 0.5-40W power output; 200W peak power; 1-8 Hz / 240 Hz Super Pulse repetition rate; 0.1ms-1s pulse length; Continuous Wave, Blend (PW), and Super Pulse modes; 3mW 650nm diode aiming beam; 7-joint articulated arm; closed-loop liquid cooling; LCD touch screen interface; 110-120 VAC power input; 150 x 46 x 31 cm dimensions.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Cortex Laser System ([K110897](/device/K110897.md))
- DEKA SmartXide2 Laser System ([K133895](/device/K133895.md))
- C02-2B3 Galaxy CO2 Laser System ([K133915](/device/K133915.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked slightly offset from each other. The profiles are simple and abstract, with a flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

July 21, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Cynosure. Incorporated dba Ellman Ms. Alison Sathe Director of Regulatory and Clinical Affairs 400 Karin Lane Hicksville, New York 11801

Re: K150587 Trade/Device Name: Cortex Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 15, 2015 Received: June 17, 2015

Dear Ms. Sathe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Joshua

For

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150587

Device Name Cortex Laser System

#### Indications for Use (Describe)

The CO2 laser is indicated for incision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology, neurosurgery, orthopedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, and coagulation of soft tissue in dermatology and plastic surgery, general surgery.

Dermatology, Plastic Surgery and General Surgery procedures:

- · Laser skin resurfacing.
- · Treatment of furrows and wrinkles.

· Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation.

- · Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
- · Blepharoplasty.
- · Site preparation for hair transplants.
- · The fractional scanner is for skin resurfacing.

The Er: YAG handpiece is designed specifically for superficial skin ablation resulting in skin dermabrasion, and the treatment of wrinkles. In addition this system is intended for coagulation, ablation, or cutting of soft tissue (skin) in dermatology, plastic surgery (including aesthetic surgery).

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <div> <span> <span style="font-size:20px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> |
|------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:20px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div>  |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

This 510(K) Summary of safety and effectiveness for the Cortex Laser System is submitted in accordance with the requirements of the SMDA 1990 and FDA guidance concerning the organization and content of a 510(K) summary.

| Applicant:                               | Cynosure, Inc. dba Ellman                                                                                      |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Address:                                 | Cynosure, Inc. dba Ellman<br>400 Karin Lane<br>Hicksville, NY 11801                                            |
| Contact Person:                          | Alison Sathe                                                                                                   |
| Telephone/Email/Fax:                     | 513-658-8960<br>asathe@ellman.com<br>516-267-6750                                                              |
| Preparation Date:                        | February 20, 2015                                                                                              |
| Device Trade Name:                       | Cortex Laser System                                                                                            |
| Common Name:                             | CO2 Laser, Er:YAG Laser                                                                                        |
| Classification Name:                     | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878.4810                                                |
| Legally Marketed Predicate<br>Device(s): | Cortex Laser System K110897<br>DEKA SmartXide2 Laser System K133895<br>C02-2B3 Galaxy CO2 Laser System K133915 |

Device Description:

The Cortex system and controls are contained in a single console. Electrical power is supplied to the console by the facility's power source. The CO2 laser is delivered via an articulated arm that is permanently attached to the console. The Er: Y AG handpiece attaches to a port on the console and contains the laser cavity in the head of the handpiece. The user interface is a touch screen located on the console. The user activates the laser emission by means of a footswitch.

Intended Use:

The CO2 laser is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

{4}------------------------------------------------

The CO2 System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery.

Dermatology, Plastic Surgery and General Surgery procedures:

- Laser skin resurfacing.
- . Treatment of furrows and wrinkles.
- Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basal cell carcinoma, warts and uneven pigmentation.
- Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
- Blepharoplasty. ●
- . Site preparation for hair transplants.
- The fractional scanner is for skin resurfacing. ●

The Er: Y AG handpiece is designed specifically for superficial skin ablation resulting in skin dermabrasion, and the treatment of wrinkles. In addition this system is intended for coagulation, vaporization, ablation, or cutting of soft tissue (skin) in dermatology, plastic surgery (including aesthetic surgery).

# Technological Characteristics:

|                          | Cortex Laser System                      |
|--------------------------|------------------------------------------|
| Source                   | CO2                                      |
| Wavelength               | 10.6 micron (10,600nm)                   |
| Power to tissue          | 0.5 - 40 W                               |
| Peak power               | 200 W                                    |
| Repetition Rate          | 1 - 8 Hz / 240 Hz Super Pulse            |
| Pulse Length             | 0.1ms to 1 s                             |
| Laser operation modes    | Continuous Wave, Blend (PW), Super Pulse |
| Aiming beam              | 3mW (650nm diode) adjustable             |
| Articulated arm          | 7-joint articulated arm                  |
| Cooling type             | Closed loop (liquid)                     |
| User Interface           | LCD Touch Screen                         |
| Power input requirements | 110 - 120 VAC, 10 A, 50-60 Hz            |
| Dimension (H x W x D)    | 150 x 46 x 31 cm                         |

Performance Data: None submitted.

Substantial Equivalence:

<sup>4</sup> The Er:YAG technical specifications of the Cortex Laser System are omitted as they remain unchanged and are not being compared to the predicate device.

{5}------------------------------------------------

The Cortex Laser System has the same intended uses and indications, technological characteristics, principles of operation as its predicate device. There are no new issues of safety or effectiveness. Thus, the Cortex Laser System is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K150587](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K150587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
