← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K141511

# SIDELAZE800 LASER BEAM DELIVERY ACCESSORY FOR CYNOSURE 1440MM WAVELENGTH LASERS (K141511)

_Cynosure, Inc. · GEX · Jan 23, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141511

## Device Facts

- **Applicant:** Cynosure, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jan 23, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440mm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers for surgical incision, vaporization, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, meniscus, mucous membrane, lymph vessels and nodes, organs and gland) laser assisted lipolysis and the treatment of primary axillary hyperhidrosis.

## Device Story

SideLaze800™ is a side-firing fiber optic accessory for Cynosure 1440nm Nd:YAG laser systems. It delivers focused laser energy to target and ablate soft tissue, including sweat glands in the underarm area for hyperhidrosis treatment. The device is used by physicians in clinical settings. The laser system is activated via footswitch. The accessory enables precise energy delivery for surgical procedures and sweat gland reduction. Clinical benefits include effective treatment of primary axillary hyperhidrosis with demonstrated sweat gland ablation and fragmentation.

## Clinical Evidence

Prospective clinical study with 59 subjects across 4 centers. Subjects received a single treatment and were followed for 12 months. Primary endpoints included Hyperhidrosis Disease Severity Scale (HDSS) scores, Physician Satisfaction Scale (PSS), and blinded evaluation of starch iodine test photographs. Results showed sweat gland reduction, ablation, and fragmentation. No deaths, serious adverse events (SAEs), or unanticipated adverse device effects (UADEs) were reported.

## Technological Characteristics

Side-firing fiber optic accessory; 800 µm diameter; 3-meter length; 632.8 nm aiming beam. Compatible with Cynosure 1440nm Nd:YAG laser systems. Maximum delivered energy limit 300 J/cm². Materials are identical to previously cleared K121127 device.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- SideLaze800™ Laser Beam Delivery Accessory for Cynosure 1440nm Wavelength Lasers ([K121127](/device/K121127.md))
- miraDry System ([K103014](/device/K103014.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 23, 2015

Cynosure Incorporated Ms. Aastha Kohli Senior Regulatory Affairs Specialist 5 Carlisle Road Westford, Massachusetts 01886

Re: K141511 Trade/Device Name: SideLaze800™ Laser Beam Delivery Accessory for Cynosure 1440nm Wavelength Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 18, 2014 Received: December 19, 2014

Dear Ms. Kohli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K141511

Device Name

SideLaze800™Laser BeamDelivery Accessory for Cynosure 1440nmWavelength Lasers

Indications for Use (Describe)

SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440mm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers for surgical incision, vaporization, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, meniscus, mucous membrane, lymph vessels and nodes, organs and gland) laser assisted lipolysis and the treatment of primary axillary hyperhidrosis.

| Type of Use (Select one or both, as applicable)                                            |                                             |
|--------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-family:Wingdings;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510K Summary

### 510(k) Summary for SideLaze800™ Laser Beam Delivery Accessory

#### A. Sponsor

Cynosure, Inc. 5 Carlisle Road Westford, MA 01886

### B. Contact

Ms. Aastha Kohli Senior Regulatory Affairs Specialist 781-993-2394 akohli@cynosure.com

#### C. Device Name

| Trade Name:          | SideLaze800  TM Laser Beam Delivery Accessory for Cynosure 1440nm<br>Wavelength Lasers        |
|----------------------|-----------------------------------------------------------------------------------------------|
| Common/usual Name:   | Powered Laser Surgical Instrument (Laser for Surgery and Dermatology)                         |
| Classification Name: | GEX-Powered laser surgical instrument, General & Plastic Surgery<br>21 CFR 878.4810, Class II |

#### D. Predicate Device

| Trade Name:             | SideLaze800TM Laser Beam Delivery Accessory for Cynosure 1440nm<br>Wavelength Lasers          |  |  |
|-------------------------|-----------------------------------------------------------------------------------------------|--|--|
| Common/usual Name:      | Powered Laser Surgical Instrument (Laser for Surgery and Dermatology)                         |  |  |
| Classification Name:    | GEX-Powered laser surgical instrument, General & Plastic Surgery<br>21 CFR 878.4810, Class II |  |  |
| Premarket Notification: | Cynosure Inc, K121127 (05/13/2012)                                                            |  |  |
| Trade Name:             | miraDry System                                                                                |  |  |
| Common/usual Name:      | Microwave And Accessories                                                                     |  |  |

#### Classification Name: NEY-System, Ablation, Microwave and Accessories, General & Plastic Surgery 21 CFR 878.4400, Class II Premarket Notification: Miramar Labs, Inc., K103014 (01/08/2011)

#### E. Device Description

The Cynosure 1440nm Wavelength laser is solid state Nd: YAG laser, having a neodymium rod as a lasing medium. Laser activation is by footswitch. Overall weight of the laser is 220lbs, and the size is 38.6"x16.5"x35.5" (HxWxD). Electrical requirement is 230 VAC, 16A, 50-60 Hz, single phase. SideLaze800™ Laser Beam Delivery Accessory is an optional side-firing fiber optic accessory used with the Cynosure Lasers with 1440nm wavelength. The SideLaze800™ option is intended to offer physicians a convenient accessory.

{4}------------------------------------------------

### F. Intended Use/Indications for Use

SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440mm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers' for surgical incision, excision, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and gland), laser assisted lipolysis and the treatment of primary axillary hyperhidrosis.

### G. Technological Characteristics

The SideLaze800™ option contains equivalent components and patient-contacting materials as previously cleared SideLaze800 Laser Beam Delivery Accessory for Cynosure 1440nm Wavelength Lasers (K121127). There is no change in device design or performance characteristics compared to previously cleared device. This proposed SideLaze800™ Fiber will have an additional indication for use for the treatment of primary axillary hyperhidrosis.

{5}------------------------------------------------

#### 510K Summary

|                                | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device                                                                                  |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k)#                        | K141511                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K121127                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K103014                                                                                           |
| Manufacturer                   | Cynosure, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Cynosure, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Miramar Labs, Inc.                                                                                |
| Device Name                    | SideLaze800™™, Laser Beam Accessory<br>for Cynosure 1440nm Wavelength Lasers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | SideLaze800™, Laser Beam Accessory<br>for Cynosure 1440nm Wavelength Lasers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | miraDry System                                                                                    |
| Clearance Date                 | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | May 13, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | January 8, 2011                                                                                   |
| Classification/ Regulation     | Class II/21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Class II/21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Class II/21 CFR 878.4400                                                                          |
| Indications for Use            | SideLaze800™™ is an accessory to delivery<br>optical energy for Cynosure 1440nm<br>wavelength lasers and intended to be used<br>with Cynosure 1440nm wavelength<br>lasers' cleared indications for use, such as<br>the surgical incision, excision,<br>vaporization, ablation, and coagulation of<br>soft tissue (including skin, cutaneous<br>tissue, subcutaneous tissue, striated and<br>smooth tissue, muscle, cartilage,<br>meniscus, mucous membrane, lymph<br>vessels and nodes, organs and gland) laser<br>assisted lipolysis and the treatment of<br>primary axillary hyperhidrosis. | SideLaze™ is an accessory to deliver<br>optical energy for Cynosure 1440nm<br>wavelength lasers and intended to be used<br>with Cynosure 1440nm wavelength<br>lasers' cleared indications for use, such as<br>the surgical incision, excision,<br>vaporization, ablation, and coagulation of<br>soft tissue (including skin, cutaneous<br>tissue, subcutaneous tissue, striated and<br>smooth tissue, muscle, cartilage<br>meniscus, mucous membrane, lymph<br>vessels and nodes, organs and glands) and<br>laser assisted lipolysis. SideLaze800™<br>may be used in combination with a<br>Cynosure SMA-compatible laser system<br>for the same indications | The miraDry System is indicated for use<br>in the treatment of primary axillary<br>hyperhidrosis. |
| Use Principle                  | Delivers focused laser energy for surgical<br>incision, excision, vaporization, ablation,<br>and coagulation of soft tissue, which<br>includes ablation of sweat glands.                                                                                                                                                                                                                                                                                                                                                                                                                      | The SideLaze800™ Laser Beam Delivery<br>Accessory targets the sweat glands in the<br>underarm area and uses focused laser<br>energy to ablate their function.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | MiraDry utilizes microwave energy, for<br>the reduction or removal of sweat glands.               |
| Wavelength                     | 1440 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 1440 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                   |
| Pulse Characteristics:         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                   |
| Maximum Delivered Energy Limit | 300 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 300 J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                   |
| Fiber Optic Characteristics:   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                   |
| Diameter                       | 800 µm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 800 µm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                   |
| Length                         | 3 meters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 3 meters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                   |
| Aimino Ream                    | 632.8 nm (wavelength)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 632.8 nm (wavelength)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                   |

{6}------------------------------------------------

#### 510K Summary

#### H. Clinical Information

Clinical testing showed that the device provides a safe and effective means to treat axillary hyperhidrosis using the SideLaze800™ Laser Beam Delivery Accessory. There were 59 subjects enrolled at 4 study centers, all subjects underwent single treatment and were followed up till 12 months. Treatment effectiveness was assessed through subject rated Hyperhidrosis Disease Severity Scale (HDSS) scores at each follow up visit, Physician Satisfaction Scale (PSS), blinded evaluation of pre and post photographs from starch iodine test results obtained at each follow up visit and recording of adverse events. Ninety Eight percent subjects returned for the three month follow up visit, Ninety six percent returned for the six month follow-up visit, eighty six percent returned for the nine month and twelve month follow up. All subjects demonstrated a reduction, ablation and fragmentation of the sweat glands. There were no deaths, serious adverse events (SAEs) or unanticipated adverse device effects (UADEs) reported in these studies.

#### I. Substantial Equivalence

There is no change in the SideLaze800™ accessory's design, principle of operation, material of construction and intended use compared to the previously cleared accessories for the Cynosure Lasers with 1440nm wavelength. The proposed indication for use "Primary axillary hyperhidrosis" is supported with clinical evidence demonstrating SideLaze800™ safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141511](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
