← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K141403

# LITHO DK30 (K141403)

_Quanta System Spa · GEX · Aug 7, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141403

## Device Facts

- **Applicant:** Quanta System Spa
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Aug 7, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The device Litho DK30 including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.

## Device Story

Litho DK30 is a desktop-version pulsed Holmium:YAG laser surgical instrument (30 W, 2.1 µm). It delivers laser energy via a fiber optic delivery system to perform soft tissue surgery and lithotripsy. Operated by clinicians in surgical settings (OR, clinic), the device features a touch-screen interface and an RFID system for automatic fiber diameter recognition. Modifications from the predicate (Litho K091909) include a new mechanical structure, compact power supply, improved cooling (pump and heat-exchanger), updated focusing optics, and revised control hardware/software. The device assists surgeons in precise tissue management and stone fragmentation, providing a more compact form factor for clinical use.

## Clinical Evidence

Bench testing only. Verification and validation activities were conducted to demonstrate that the modified device meets the requirements of the third edition of IEC 60601-1 and its collateral standards. No clinical data was required or provided.

## Technological Characteristics

Pulsed Holmium:YAG laser; 30 W max output power; 2.1 µm wavelength. Desktop form factor. Features: RFID fiber recognition system, touch-screen GUI, water-to-air heat exchanger, and water pump cooling. Complies with IEC 60601-1 (3rd ed).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Litho ([K091909](/device/K091909.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 7. Special 510(K) SUMMARY - Device Modifications

## Introduction:

This document contains the 510(k) Summary for the device Litho DK30. The basis of this submission is Modifications to Device already cleared. The content of this summary is based on the requirements of 21 CFR 807.92(c).

| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058           |
|--------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Contact Person:                           | Maurizio Bianchi<br>Regulatory Affairs Manager<br>Quanta System SPA                        |
|                                                  | Email: maurizio.bianchi@quantasystem.com<br>Phone: +39 0331 376797<br>Fax: +39 0331 367815 |
| Date Prepared:                                   | May 24th, 2014                                                                             |
| Device Name:                                     | Litho DK30                                                                                 |
| Classification:                                  | Class II                                                                                   |
| Classification Name:                             | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology.    |
| Regulation Number:                               | 21 CFR 878.4810                                                                            |
| Product Code:                                    | GEX                                                                                        |
| Basis for Submission:                            | Device modifications                                                                       |
| Legally Marketed Device                          | Litho (K091909)                                                                            |

The modified device Litho DK30 is claimed to be derived from the legally marketed (unmodified) device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).

#### Performance Standards:

There are no mandatory performance standards for this device.

{1}------------------------------------------------

#### General Device Description:

The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).

This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909): Litho DK30 is the desktop version of the already cleared device Litho (K09109).

The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.

Both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 um.

The sum of the incremental changes from the original clearance K091009 has been taken into account and all the occurred modifications will be listed and described within this submission.

The device Litho including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.

The device Litho including a fiber optic delivery system is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to: Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.

The modified device Litho DK30 is claimed to be derived from the legally marketed (unmodified) device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).

#### Description of the modifications:

The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).

This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).

The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 µm, with no change in the fundamental scientific technology of the device.

The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.

{2}------------------------------------------------

The following modifications have been implemented on the cleared device Litho (K091909) in order to get the desktop version named Litho DK30:

- A new mechanical structure and related covers have been developed
- A more compact and lightweight power-supply with extended working range has been qualified.
- The water pump of the cooling systern has been changed to improve the cooling efficiency. -
- The water-to-air heat-exchanger has been changed to improve the cooling efficiency.
- i The technical specifications of lens focusing the laser beam into the optical fiber has been changed to improve the incoupling.
- An RFID recognition system has been added to read the fiber diameter directly from the n fiber equipped with the suitable transponder.
- Review of the combinations of the laser parameters (energy and frequency) achievable within 30 W maximum power.
- A larger touch-screen has been adopted
- The control hardware (boards) and related software have been reviewed
- The GUI (Graphic User Interface) has been reviewed to improve the readability i

All the modifications implemented have been evaluated following the requirements of third edition of IEC 60601-1 and to its collateral standards. The introduction of the third edition of the IEC 60601-1 and to its collateral standards has been taken into account: it leaded to a basic change in the (regulatory) design input requirements with the subsequent revision of risk analysis and specific verification and validation activities to demonstrate that modified device meet the new requirements.

Picture 7-1 shows the the external appearance of the modified device and the unmodified device.

Image /page/2/Picture/13 description: The image contains two medical devices. The device on the left is a larger, wheeled machine with a display screen and various controls. It has a light-colored body with a darker top section and a foot pedal. The device on the right is smaller and more compact, labeled "Litho DK". It also features a display screen and several buttons, indicating it is a medical device for lithotripsy.

Picture 7-1 - device Litho (on left) and Litho DK30 (on right) 7-3

{3}------------------------------------------------

## Intended Use/Indications for Use

The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).

This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).

The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 um, with no change in the fundamental scientific technology of the device.

The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.

The Intended Use/Indication for use statement of the unmodified device (Litho - K091009) has been analyzed. In the original statement there is not a clear definition (separation) of the Intended use and of the Indications for use and thus a revised statement is proposed.

The revised version of the statement has separated sections for the Intended Use and for the Indications for use.

Moreover, in order to have a more understandable definition of the Indications for use, it has been removed the (long) list of diseases/surgical operations given in K091909 because this list, as said by several physicians, does not add useful information.

#### Thus the Intended Use is:

The device Litho including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic, to perform incision, excision, excision, resection, ablation, vaporization, coaqulation and hemostasis of soft tissue and in Lithotripsy of stones.

#### And the Indications for use are:

The device Litho including a fiber optic delivery system is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to: Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology , Plastic Surgery and General Surgery.

#### The modified device Litho DK30 has the same intended use of the unmodified device.

Moreover the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.

{4}------------------------------------------------

## Substantial Equivalence:

This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because because Litho DK30 is the desktop version of the already cleared device Litho (K09109).

The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 µm, with no change in the fundamental scientific technology of the device.

The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.

Thus the modified device Litho DK30 is substantially equivalent to the previously legally marketed device Litho (K091909).

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined strands. The caduceus is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is presented in black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 7, 2014

Quanta System SPA Mr. Maurizio Bianchi Regulatory Affairs Manager Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058

Re: K141403

Trade/Device Name: Litho DK30 (and its accessories) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 7, 2014 Received: July 8, 2014

Dear Mr. Bianchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices . . Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K141403

Device Name Litho DK30

#### Indications for Use (Describe)

#### INTENDED USE

The device Litho DK30 including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.

#### INDICATIONS FOR USE

The device Litho DK30 including a fiber optic delivery system is insurgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to:

- Urology
- · Gastroenterology
- Arthroscopy
- Neurosurgery
- Pulmonary
- Gynecology
- ENT
- Dermatology
- Plastic Surgery
- General Surgery

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (2) CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

David Krause -S 2014.08.18 13:46:18 -04'00'

{8}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141403](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K141403)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
