K113390 · Meridian AG · GEX · Feb 14, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113390
Device Name
LASER INDIRECT OPHTHALMOSCOPE 500(LIO-500)
Applicant
Meridian AG
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Laser Indirect Ophthalmoscope 500 (LIO-500) is intended for use in photocoagulating ocular tissue in the treatment of diseases of the eye. The laser energy is delivered via either transpupillary delivery or intraccular endoprobe delivery.
Device Story
LIO-500 is an ophthalmic accessory for delivering 532 nm laser energy to ocular tissue. It functions as a delivery system for external ophthalmic treatment lasers (e.g., MERILAS 532a). Device attaches to standard indirect ophthalmoscopes (e.g., Heine Omega 500, Keeler Vantage Plus) for visualization of the treatment area. Operated by physicians in clinical settings. No internal contacts, displays, or user-accessible electronic functions. Laser energy is directed to the patient's eye via transpupillary or intraocular endoprobe delivery to perform photocoagulation for eye disease treatment.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Passive optical delivery system for 532 nm ophthalmic lasers. No internal electronics, displays, or software. Attaches to standard indirect ophthalmoscopes for visualization. Transpupillary or intraocular endoprobe delivery.
Indications for Use
Indicated for photocoagulation of ocular tissue in the treatment of eye diseases. Used via transpupillary or intraocular endoprobe delivery. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# LIO-500 510(k) Summary
#### Submitter:
113390
Regulatory Insight, Inc. 5401 S. Cottonwood Ct. Greenwood Village, Colorado 80121 Contact: Mr. Kevin Walls, RAC Telephone: 720-962-5412 Fax: 720-962-5413 Email: kevin@reginsight.com
for
MERIDIAN AG Bierigutstrasse 7 CH-3608 Thun Switzerland Phone: +41 33 334 11 11 Fax: + 41 33 334 11 19 Homepage: www.meridian.ch Email address: info@meridian.ch
Device Name:
Trade Name: Laser Indirect Ophthalmoscope 500 (LIO-500) Common Name: Ophthalmic Surgical Laser Classification Name: Powered Laser Surgical Instrument
### Predicate Device:
VISULAS 532S manufactured by CARL ZEISS, INC. and cleared under 510(k) # K013402
### Device Description:
The LIO-500 can be used on a 532 nm laser such as the MERILAS 532a or any other ophthalmic laser with 532 nm.
The LIO-500 does not incorporate any contacts, displays or other user accessible functions. Its main purpose is the delivery of the laser light, generated by an ophthalmic treatment laser unit to the eye of the patient. For the observation of the treatment area the attached indirect ophthalmoscope (such as the Heine Omega 500 or the Keeler Vantage Plus / Keeler Vantage Plus LED) gets used.
#### Intended Uses:
The Laser Indirect Ophthalmoscope 500 (LIO-500) is intended for use in photocoagulating ocular tissue in the treatment of diseases of the eye.
The laser energy is delivered via either transpupillary delivery or intraccular endoprobe delivery.
{1}------------------------------------------------
Comparison to Predicate Device:
The LIO-500 has the same technical characteristics as the VISULAS 532S. There are no clinical or non-clinical tests required to show substantial equivalence.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## FEB 1 4 2012
Meridian AG % Regulatory Insight, Inc. Mr. Kevin Walls, RAC 5401 South Cottonwood Court Greenwood Village, Colorado 80121
Re: K113390 ·
Trade/Device Name: Lase Indirect Ophthalmoscope 500 (LIO-500) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 15, 2011 Received: November 16, 2011
Dear Mr. Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Kevin Walls, RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part . 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Mellman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
Device Name: Laser Indirect Ophthalmoscope 500 (LIO-500)
Indications for Use:
The Laser Indirect Ophthalmoscope 500 (LiO-500) is intended for use in photocoagulating ocular tissue in the treatment of diseases of the eye.
The laser energy is delivered via either transpupillary delivery or intraocular endoprobe delivery.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Nel RPRgle for mxm
(D:i:n:i:tion
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113390
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.