ENDOVASCULAR LAER VEIN SYSTEM KIT

K101712 · Biolitec, Inc. · GEX · Oct 27, 2010 · General, Plastic Surgery

Device Facts

Record IDK101712
Device NameENDOVASCULAR LAER VEIN SYSTEM KIT
ApplicantBiolitec, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateOct 27, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is intended for endovascular coagulation of blood vessels. The device is indicated for the endovascular coagulation of the greater saphenous vein in the thigh in patients with superficial vein reflux.

Device Story

Endo Laser Vein System Kit (ELVeS) with Radial Fiber facilitates endovascular coagulation of greater saphenous vein. Kit includes radial fiber, access needle, introducer sheath/dilator, and guidewire. Used in clinical settings by physicians to treat superficial vein reflux. Device functions as accessory to diode laser systems; laser energy delivered via radial fiber to induce thermal coagulation of vein wall. Procedure results in vein closure, providing therapeutic benefit for venous reflux. Output is thermal energy; clinical decision-making based on physician assessment of venous insufficiency.

Clinical Evidence

Bench testing only. Device complies with voluntary consensus standards including 21 C.F.R. §§ 1040.10 & 1040.11, ANSI/AAMI ES1, IEC 601-1, IEC 601-2-22, EN 60825-1, and ANSI/AAMI/ISO 10993-7.

Technological Characteristics

Kit components: radial fiber, access needle, introducer sheath/dilator, guidewire. Energy source: external diode laser system (1470nm, 980nm, 810nm). Standards compliance: 21 C.F.R. §§ 1040.10, 1040.11; ANSI/AAMI ES1; IEC 601-1; IEC 601-2-22; EN 60825-1; ANSI/AAMI/ISO 10993-7.

Indications for Use

Indicated for endovascular coagulation of the greater saphenous vein in the thigh in patients with superficial vein reflux.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K101712 ## 510(k) Summary Endo Laser Vein System Kit with Radial Fiber # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared OCT 2 7 2010 Biolitec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 (413) 525-0600 Phone: Facsimile: (413) 525-0611 Contact Person: Harry Hayes, Ph.D. - Regulatory Consultant Date prepared: June 7, 2010 #### Name of Device and Name/Address of Sponsor Endo Laser Vein System Kit (ELVeS®) with Radial Fiber Biolitec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 #### Classification Name Surgical laser accessories ## Predicate Devices Ceralas D Diode Laser System (1470nm, 980 nm and 810 nm) with ELVeS Kit ## Intended Use/Indication for Use The device is intended for endovascular coagulation of blood vessels. The device is indicated for the endovascular coagulation of the greater saphenous vein in the thigh in patients with superficial vein reflux. ## Technological Characteristics The ELVeS kit with Radial Fibers contain the following components: (1) radial fiber; (2) access needle; (3) introducer sheath/ dilator; and (4) a guidewire. ## Performance Data The device complies with the following voluntary consensus standards: 21 C.F.R. §§ 1040.10 & 1040.11; ANSI/AAMI ES1; IEC 601-1; IEC 601-2-22; EN 60825-1, and ANSI/AAMI/ISO 10993-7. {1}------------------------------------------------ #### Substantial Equivalence The ELVes with Radial Fiber uses previously cleared radial fiber technology (K924258) and has the same intended use and indications for use as the cleared Ceralas D 1470, 980 and 810nm ELVeS kits. The ELVeS kit is not a new technological characteristic for diode lasers for endovascular coagulation of the greater saphenous vein in the thigh in patients with superficial vein reflux. Thus, the ELVeS is substantially equivalent to its predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized image of an eagle or bird with outstretched wings, rendered in black lines. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is also in black and is arranged to follow the curve of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Biolitec, Inc. % Genmarhay BDA Mr. Harry Hayes 1349 Main Road Granville, Massachusetts 01034 OCT 2 7 2010 Re: K101712 Trade/Device Name: Endo Laser Vein System Kit with Radial Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 12, 2010 Received: October 13, 2010 Dear Mr. Hayes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Harry Hayes Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use Statement 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ OCT 2 7 2010 Device Name: Endo Laser Vein System Kit with Radial Fiber For endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux. # (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _ V (Per 21 C.F.R. 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Neil K. Arlen for nxn Division of Oil, Gas (Division S. Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K101712
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...