AESCULIGHT FAMILY OF FLEXIBLE FIBERS, HANDPIECES AND TIPS FOR CO2 SURGICAL LASERS
K081612 · Aesculight, LLC · GEX · Jun 25, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081612
Device Name
AESCULIGHT FAMILY OF FLEXIBLE FIBERS, HANDPIECES AND TIPS FOR CO2 SURGICAL LASERS
Applicant
Aesculight, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 25, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the product(s) is to communicate the laser beam of a CO, surgical laser to the target site for the incision, excision, ablation, or photocoagualtion of soft tissue. Representative examples of clinical applications include: Gynecology--excision and vaporization of cervical, vulvar, and perineal condyloma: ablation of vaginal and vulvar intraepithelial neoplasia; herpes vaporization of urethral caruncle; I&D Bartholin's and nubothian cysts Dermatology--port wine hemangioma removal; rhinophyma reduction; telangiectasia removal; wart removal; basal squamous cell carcinomoa removal; blepharoplasty, xanthalasma removal; removal of neurofibromas hemangiomas, nevi, and tircoeptiheliomas; dermabrasion for lentigos, keratoses, actinic keratosis and actinic chleilitis Dentistry/Oral Surgery-gingivectomy; frenum release; gingivoplasty; removal of soft tissue, cysts, and tumors General Surgery--hemorthoid removal; skin tag vaporization; pilodidal cyst removal and repair; debridement of decubitus ulcers and stasis ulcers; mastectomy; breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures Laparoscopic Surgery--vaporization, incision, excision, ablation, or photo-coagulation of soft tissue such as endometriosis ablation, excision of adhesions, salpingotomy Otorhinolaryngology--lymphangioma removal; turbinectomy, subglottic stenosis vaporization, tonsilectomy, removal of vocal cord papillomas, nodules, and polyps Podiatry-plantar wart vaporization; fungal nail treatment; partial and complete mattixectomy; porokeratoma ablation; Morton's neurome removal; ingrown toenail treatment Orthopedic -- menisectorny, chondromalacia abalation, partial synovectomy, lateral release, PMMA removal
Device Story
Detachable beam delivery system for CO2 surgical lasers; comprises flexible hollow lightpipe (fiber), handpieces, and rigid alumina tips. User selects handpiece/tip combination for specific clinical task. Fiber transmits laser energy to target site for soft tissue incision, excision, ablation, or photocoagulation. Used in clinical settings (e.g., OR, clinic) by physicians. Aesculight fiber uses internally-coated flexible seamless tubing for improved transmission efficiency compared to predicate molybdenum ribbon. Alumina tips provide conical lumen for beam concentration and higher power throughput. Some handpieces include integral/add-on smoke evacuation. Output is laser beam directed at tissue; enables precise tissue removal/vaporization. Benefits include clinical versatility across multiple surgical specialties and improved power handling/transmission efficiency.
Clinical Evidence
Bench testing only. Laboratory testing confirmed that Aesculight tips produce identical spot sizes to predicate tips, possess similar transmission efficiencies, and have power-handling capabilities that equal or exceed predicate devices.
Technological Characteristics
Flexible hollow lightpipe (fiber) with internally-coated seamless tubing; monolithic alumina tips with conical lumen; handpieces with optional smoke evacuation. Optical multi-mode transmission. Sterilization not specified. No software/algorithm.
Indications for Use
Indicated for soft tissue incision, excision, ablation, or photocoagulation in gynecology, dermatology, dentistry/oral surgery, general surgery, laparoscopic surgery, otorhinolaryngology, podiatry, and orthopedic procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Modified LX-20 Laser System fibers, handpieces, and tips (K960475)
Submission Summary (Full Text)
{0}------------------------------------------------
SURGICAL LASERS
1-866-697-7548 425-483-8905 Fax 425-483-6015
JUN 2 5 2008 KO81612
## 510(k) Summary
SURICAL
Date prepared: June 2, 2008
### Submitter's name, address, telephone, fax, and contact person;
Aesculight LLC 16928 Woodinville-Redmond Rd N.E., Suite B104 Woodinville, WA 98072 (425)483-8905 telephone (425)483-6015 fax Paul Diaz, Vice President of Engineering
Product trade name: Aesculight family of fibers, handpieces, and tips for CQ surgical lasers
Product common name: Laser flexible fibers, handpieces, and tips
Classification names: General and Plastic Surgery Devices, 21 CFR, 878.4810, product code GEX; ENT Microsurgical CO2 Laser, 21 CFR874.4500, product code EWG; Gynecologic Surgical Laser, 21 CFR 884.4550, product code HHR; Soft Tissue Dental Laser, 21 CFR 878.4810, product code NVK
Legally marketed predicate device: Modified LX-20 Laser System fibers, handpieces, and tips, as described in 510(k) no. K960475
## Product description
This is a family of detachable devices that comprise the distal portion of a beam delivery system to be used with a CO, surgical laser such as that described in K96075. Specifically, the devices consist of a flexible hollow lightpipe (fiber), handpieces that connect to the fiber, and various tips that connect to the handpieces. The tips are short, rigid hollow alumina lightpipes producing a variety of spot sizes. The user may "mix and match" handpieces and tips to suit the clinical task at hand.
### Intended vse
The intended use of the product(s) is to communicate the laser beam of a CO, surgical laser to the target site for the incision, excision, ablation, or photocoagualtion of soft tissue. Representative examples of clinical applications include:
Gynecology--excision and vaporization of cervical, vulvar, and perineal condyloma: ablation of vaginal and vulvar intraepithelial neoplasia; herpes vaporization of urethral caruncle; I&D Bartholin's and nubothian cysts
{1}------------------------------------------------
Dermatology--port wine hemangioma removal; rhinophyma reduction; telangiectasia removal; wart removal; basal squamous cell carcinomoa removal; blepharoplasty, xanthalasma removal; removal of neurofibromas hemangiomas, nevi, and tircoeptiheliomas; dermabrasion for lentigos, keratoses, actinic keratosis and actinic chleilitis
Dentistry/Oral Surgery-gingivectomy; frenum release; gingivoplasty; removal of soft tissue, cysts, and tumors
General Surgery--hemorthoid removal; skin tag vaporization; pilodidal cyst removal and repair; debridement of decubitus ulcers and stasis ulcers; mastectomy; breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures
Laparoscopic Surgery--vaporization, incision, excision, ablation, or photo-coagulation of soft tissue such as endometriosis ablation, excision of adhesions, salpingotomy
Otorhinolaryngology--lymphangioma removal; turbinectomy, subglottic stenosis vaporization, tonsilectomy, removal of vocal cord papillomas, nodules, and polyps
Podiatry-plantar wart vaporization; fungal nail treatment; partial and complete mattixectomy; porokeratoma ablation; Morton's neurome removal; ingrown toenail treatment
Orthopedic -- menisectorny, chondromalacia abalation, partial synovectomy, lateral release, PMMA removal
No new indications for use are sought beyond those associated with the predicate devices.
#### Technological characterization
When used with a CO, laser, the effect on tissue of the laser beam, as produced by the Aesculight family of products, is identical to that of the predicate products. The Aesculight tips produce the same range of spot sizes as the predicate tips. Cosmetically, dimensionally, and ergonomically, the Aesculight products are very similar to the predicates, and use the same optical principles to conduct the laser light from the base unit to the target site. There are, in some cases however, some differences in construction. The predicate fiber is based upon an optically-coated (for enhanced reflectivity) molybdenum ribbon formed into a long skinny tube (with the optical coating on the inside surface), whereas the Aesculight fiber is based upon internally-coated flexible seamless tubing, This tubing, with no seams or gaps in the optical coating, exhibits somewhat better transmission efficiency than the predicate fiber. While some of the Aesculight tips are identical in construction to the predicate tips, some are not. To produce the smaller spot sizes, the predicate tips use a cone of coated molybdenum ribbon to concentrate the beam, whereas the Aesculight tips are monolithic alumina with a conical lumen. The alumina tips are capable of sustaining higher power throughputs than the predicate foil tips. These differences in construction do not introduce any new safety or performance concerns.
{2}------------------------------------------------
The Aesculight handpieces, for the most part, exhibit only minor cosmetic differences to the predicate handpieces. In one case (the backstop handpiece), the Aesculight handpiece is a shorter version of the predicate, and in other cases, Aesculight produces more handpiece models with integral or add-on smoke evacuation capabilities, but these differences merely affect the convenience of use of the handpieces, and have no effect on the laser beam-tissue interaction.
#### Conclusion
The Aesculight fiber, although differing somewhat in construction, has very similar physical and optical characteristics as the predicate fiber. Both are flexible, optically multi-mode, hollow lightpipes, and both have the same lumen size and length, with the Aesculight version exhibiting superior transmission efficiency. In some cases, the Aesculight handpieces are almost identical to the predicate handpieces; in other cases, the Aesculight handpieces are simple mechanical modifications of the predicate handpieces with the object of providing more convenience for the user. The Aesculight tips are, in some cases, exactly identical to the predicate tips; in other cases, the construction (but not the internal geometry) differs from the predicates, and laboratory bench testing confirms that the spot sizes produced by the Aesculight tips are identical to the predicate tips, that the Aesculight tips have similar transmission efficiencies as the predicate tips and have power-handling capabilities that equal or exceed those of the predicate tips. From a clinical efficacy standpoint, a user could not distinguish a difference between the predicate products and the Aesculight products, nor do the Aesculight products raise any new safety issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUN 2 5 2008
Aesculight, LLC % Mr. Paul Diaz Vice President of Engineering 16928 Woodinville-Redmond Road Northeast, Suite B104 Woodinville, Washington 98072
K081612 Re:
Trade/Device Name: The Aesculight family of flexible fibers, handpieces,
and tips for CO2 surgical lasers
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: June 5, 2008 Received: June 9, 2008
Dear Mr. Diaz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Paul Diaz
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
Device name: The Aesculight family of flexible fibers, handpieces, and tips for CO, surgical lasers
#### Indications for use: Representative examples are listed below.
Gynecology-excision and vaporization of cervical, vulvar, and perineal condyloma; ablation of vaginal and vulvar intraepithelial neoplasia; herpes vaporization; vaporization of urethral caruncle; I&D Bartholin's and nubothian cysts.
Dermatology--port wine hemangioma removal; rhinophyma reduction; telangiectasia removal; wart removal; basal squamous cell carcinomoa removal; blepharoplasty, xanthalasma removal; removal of neurofibromas hemangiomas, nevi, and tircoeptiheliomas; dermabrasion for lentigos, keratoses, actinic keratosis and actinic chleilitis
Dentistry/Oral Surgery-gingivectomy; frenum release; gingivoplasty; removal of soft tissue, cysts, and tumors
General Surgery--hemorthoid removal; skin tag vaporization; pilodidal cyst removal and repair; debridement of decubitus ulcers and stasis ulcers; mastectomy; breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease; many dermatological procedures
Laparoscopic Surgery-vaporization, incision, excision, ablation, or photo-coagulation of soft tissue such as endometriosis ablation, excision of adhesions, salpingotomy
Qtorhinolaryngology--lymphangioma removal; turbinectomy, subglottic stenosis vaporization, tonsilectomy, removal of vocal cord papillomas, nodules, and polyps
Podiatry-plantar wart vaporization; fungal nail treatment; partial and complete matrixectorny; porokeratoma ablation; Morton's neurome removal; ingrown toenail treatment
Orthopedic-menisectomy, chondromalacia abalation, partial synovectomy, lateral release, PMMA removal
| Prescription Use | Yes | AND/OR | Over-The-Counter Use | No |
|-----------------------------|-----|--------|------------------------|----|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) | |
Concurrence of CDRH, Office of Division Sign-Off (ODE)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | 11081612 |
|---------------|----------|
|---------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.