K072682 · Biolitec, Inc. · GEX · Apr 18, 2008 · General, Plastic Surgery
Device Facts
Record ID
K072682
Device Name
50W CERALAS D 1950NM DIODE LASER
Applicant
Biolitec, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Urology Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Urethral strictures . Bladder neck incisions . Ablation and resection of bladder tumors, urethral tumors and . ureteral tumors Ablation of Benign Prostatic hypertrophy (BPH) Transurethral incision of the prostate (TUIP) Laser Resection of the Prostate (HoLRP) Laser Enucleation of the Prostate (HoLEP) Laser Ablation of the Prostate (HoLAP) Condylomas Lesions of the external genitalia Gastroenterology Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Appendectomy . Polyps Biopsv . Gall Bladder calculi . Biliary/bile duct calculi t Ulcers . Gastric ulcers . Duodenal ulcers . . Non-bleeding ulcers . Pancreatitis Hemorrhoids . Cholecystectomy . Benign and malignant neoplasm . . Angiodysplasia Colorectal cancer . Telangiectasias . Telangiectasias of the Osler-Weber-Renu disease . Vascular malformation . Gastritis Esophagitis Esophageal ulcers Varices Colitis Mallory-Weiss tear Gastric erosions ● Thoracic/Pulmonary Surgery Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including Laryngeal lesions . Airway obstruction including carcinoma . Polyps and granulomas . Palliation of obstructing carcinomas of the tracheobronchial tree . Gynecology Open and endoscopic surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) including Intra uterine treatment of submucous fibroids, benign . endometrial polyps and uterine septum by incision, excision, ablation and or vessel coagulation Soft tissue excision procedures such as excisional conization of the cervix Ear, Nose and Throat (Otolaryngology) Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Endonasal/sinus surgery . Partial turbinectomy Polypectomy Dacryocystorhinostomy ● Frontal sinusotomy Ethmoidectomy Maxillary antrostomy Functional endoscopic sinus surgery Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal Tonsillectomy Adenoidectomy Dermatology/Plastic Surgery Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue in therapeutic plastic, dermatologic and aesthetic procedures including: Basal Cell carcinomas . Lesions of the skin and subcutaneous tissue ● Skin tags . Plantar warts Arthroscopy Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) Ablation of soft and cartilaginous tissue in minimally invasive . spinal surgery including: Percutaneous laser disc decompression/discectomy ♥ Foraminoplasty . Ablation and coagulation of soft vascular and non . vascular tissue General Surgery Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Cholecystecomy Lysis of adhesions Appendectomy Biopsv Skin incision Tissue dissection Excision of external tumors and lesions Complete or partial resection of internal organs, tumors and lesions Mastectomy Hepatectomy Pancreatectomy Splenectomy Thyroidectomy Parathyroidectomy Herniorrhaphy Tonsillectomy Lymphadenectomy Partial nephrectomy Pilonidal Cystectomy Resection of lipoma Debridement of decubitus ulcers Hemorrhoids Debridement of statis ulcers Biopsy
Device Story
50W Ceralas D 1950nm Diode Laser (Model D1950) is a surgical laser system; delivers laser energy to target soft, mucosal, fatty, or cartilaginous tissue. Used by physicians in OR or clinic settings for incision, excision, resection, ablation, vaporization, coagulation, and hemostasis. Device transforms electrical power into 1950nm laser output; energy absorption by tissue facilitates surgical effects. Healthcare providers operate device via console controls; output allows precise tissue management during endoscopic or open procedures. Benefits include minimally invasive access and controlled tissue interaction.
Clinical Evidence
Bench testing only. Comparative study of vaporization, cutting, and coagulation rates between the Ceralas D1950 and a 2010nm laser using potato slices and porcine muscle, liver, and skin tissue. Results demonstrated substantially equivalent performance and tissue effects.
Technological Characteristics
Diode laser; 1950nm wavelength; 50W maximum power output. Used for surgical incision, excision, resection, ablation, vaporization, coagulation, and hemostasis. Class II device.
Indications for Use
Indicated for patients requiring open, laparoscopic, or endoscopic surgical procedures (incision, excision, resection, ablation, vaporization, coagulation, hemostasis) across urology, gastroenterology, thoracic/pulmonary, gynecology, ENT, dermatology/plastic surgery, orthopedics, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K112324 — CERALAS 980NM DIODE LASER FAMILY · Biolitec Medical Devices, Inc. · Sep 6, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K072682
## 510(k) SUMMARY
## Biolitec Inc.'s 50W Ceralas Diode 1950nm Laser System (Model D1950)
APR 1 8 2008
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Hogan & Hartson 555 13th Street NW Washington DC 20004
Phone: 202 637-5794 Facsimile: (202) 637-5910
Jonathan S. Kahan Contact Person:
September 20, 2007 Date Prepared:
### Name of Device and Name/Address of Sponsor
50W Ceralas D 1950nm Diode Laser (Model D1950)
Biolitec, Inc. 515 Shaker Road East Longmeadow, MA 01028
### Common or Usual Name
Diode Laser
#### Classification Name
Laser, Surgical Diode Laser System
#### Product Code and Regulation Number
GEX, 21 C.F.R. 878.4810
#### Predicate Devices
AllMed Revolix 120 Laser Systems (K033423, K051167, K070476)
{1}------------------------------------------------
K072682
pg 2 of 5
Intended Use / Indications for Use
# Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral strictures .
- Bladder neck incisions .
- Ablation and resection of bladder tumors, urethral tumors and . ureteral tumors
- Ablation of Benign Prostatic hypertrophy (BPH)
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostate (HoLRP)
- Laser Enucleation of the Prostate (HoLEP)
- Laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of the external genitalia
# Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy .
- Polyps
- Biopsv ●
- Gall Bladder calculi �
- Biliary/bile duct calculi t
- Ulcers .
- Gastric ulcers .
- Duodenal ulcers .
- . Non-bleeding ulcers
- . Pancreatitis
- Hemorrhoids .
- Cholecystectomy .
- Benign and malignant neoplasm .
- . Angiodysplasia
- Colorectal cancer .
- Telangiectasias .
- Telangiectasias of the Osler-Weber-Renu disease .
- Vascular malformation .
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
{2}------------------------------------------------
K072682
Pg 3-4-5
- Mallory-Weiss tear
- Gastric erosions ●
## Thoracic/Pulmonary Surgery
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including
- Laryngeal lesions .
- Airway obstruction including carcinoma .
- Polyps and granulomas .
- Palliation of obstructing carcinomas of the tracheobronchial tree .
# Gynecology
Open and endoscopic surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) including
- Intra uterine treatment of submucous fibroids, benign . endometrial polyps and uterine septum by incision, excision, ablation and or vessel coagulation
- Soft tissue excision procedures such as excisional conization of . the cervix
# Ear, Nose and Throat (Otolaryngology)
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Endonasal/sinus surgery .
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy ●
- Frontal sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- Functional endoscopic sinus surgery
- Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
- Tonsillectomy
- Adenoidectomy
{3}------------------------------------------------
072682 pg 4-5
# Dermatology/Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue in therapeutic plastic, dermatologic and aesthetic procedures including:
- Basal Cell carcinomas .
- Lesions of the skin and subcutaneous tissue ●
- Skin tags .
- Plantar warts
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in minimally invasive . spinal surgery including:
- Percutaneous laser disc decompression/discectomy ♥
- Foraminoplasty .
- Ablation and coagulation of soft vascular and non . vascular tissue
## General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystecomy �
- Lysis of adhesions
- Appendectomy
- Biopsv
- Skin incision
- Tissue dissection
- Excision of external tumors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- Mastectomy
- Hepatectomy
- Pancreatectomy
- Splenectomy
- Thyroidectomy
- Parathyroidectomy
- Herniorrhaphy
- Tonsillectomy
- Lymphadenectomy
- Partial nephrectomy
- Pilonidal Cystectomy
- Resection of lipoma
{4}------------------------------------------------
K072682
pg 5 - 1-5
- Debridement of decubitus ulcers
- Hemorrhoids
- Debridement of statis ulcers
- Biopsy
# Technological Characteristics
The Ceralas D1950 is a diode laser operating at 1950nm with a maximum power output of 50W.
# Performance Data
Bench testing was performed to compare the vaporization, cutting, and coagulation rates of the Ceralas D1950 and a 2010nm laser. The lasers were directed at potato slices and porcine muscle tissue to evaluate vaporization, and cutting and coagulation were evaluated with porcine liver, skin, and muscle tissue. Similar parameters were employed with each laser to test the effects of the Ceralas D1950 and the 2010nm lasers on the target tissue for each of these tests. The testing demonstrated substantially equivalent performance and tissue effects between the Ceralas D1950 and the 2010mm laser device.
# Substantial Equivalence
The Ceralas D1950 is as safe and effective as AllMed Revolix Laser Systems (K033423, K051167, K070476). The Ceralas D1950 has the same intended uses and indications, and similar technological characteristics, and principles of operation as its predicate devices. Thus, the Ceralas D1950 is substantially equivalent.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biolitec, Inc. % Hogan & Hartson, LLP Mr. Johnathan S. Kahan 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K072682
Trade/Device Name: 50W Ceralas D 1950nm Diode Laser (Model D1950) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 10, 2008 Received: March 10, 2008
APR 1 8 2008
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 – Mr. Johnathan S. Kahan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Pg 1 of 4
Indications for Use Statement
510(k) Number (if known): K072682
50W Ceralas D 1950nm Diode Laser (Model D1950) Device Name:
Indications for Use:
#### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral strictures �
- Bladder neck incisions �
- Ablation and resection of bladder tumors, urethral tumors and ureteral tumors
- Ablation of Benign Prostatic hypertrophy (BPH)
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostate (HoLRP)
- Laser Enucleation of the Prostate (HoLEP)
- Laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of the external genitalia
#### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
B' : "rice
- Appendectomy ●
- Polyps
- Biopsv
- Gall Bladder calculi
- Biliary/bile duct calculi
- Ulcers
- ਾ ਪ Gastric ulcers
- Duodenal ulcers
- Non-bleeding ulcers
- Pancreatitis
- Hemorrhoids
- Cholecystectomy
- Benign and malignant neoplasm
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
Nilkl Gel for mxn
Division Sign-Off
Division of General, Restorative. and Neurological Devices
510(k) Number K072682
{8}------------------------------------------------
107268
-
- Telangiectasias of the Osler-Weber-Renu disease
- Vascular malformation
- Gastritis Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric erosions
#### Thoracic/Pulmonary Surgery
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including
- Laryngeal lesions .
- Airway obstruction including carcinoma
- Polyps and granulomas
- Palliation of obstructing carcinomas of the tracheobronchial tree
#### Gynecology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including
- Intra-uterine treatment of submucous fibroids, benign . endometrial polyps and uterine septum by incision, excision, ablation and or vessel coagulation
- Soft tissue excision procedures such as excisional conization of the cervix
### Ear. Nose and Throat (Otolaryngology)
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: .
- Endonasal/sinus surgery .
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy
- Frontal sinusotomy
- Ethmoidectomy ●
- Maxillary antrostomy .
- � Functional endoscopic sinus surgery
- Lesions or tumors of the oral, nasal, glossal, pharyngeal and . laryngeal
(Division Sign=Of Division of General, Restorative, and Neurological Devices
- Tonsillectomy Adenoidectomy
510(k) Number_Ko72682
{9}------------------------------------------------
K072682 Pg 3 of 4
## Dermatology/Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue in therapeutic plastic, dermatologic and aesthetic procedures including:
- Basal Cell carcinomas .
- Lesions of the skin and subcutaneous tissue
1 ···
- Skin tags
- Plantar warts
## Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in minimally invasive . spinal surgery including:
r. 1 . " !!#
- Percutaneous laser disc decompression/discectomy
- Foraminoplasty
- Ablation and coagulation of soft vascular and non vascular tissue
## General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystecomy ●
- Lysis of adhesions
- Appendectomy
- Biopsy
- Skin incision
- Tissue dissection
- Excision of external tumors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- Mastectomy
- Hepatectomy .
- Pancreatectomy .
- Splenectomy
- Thyroidectomy
- Parathyroidectomy
- Herniorrhaphy
- Tonsillectomy
- Lymphadenectomy
- Partial nephrectomy
- Pilonidal Cystectomy
- Resection of lipoma
Neil Re Opln for mkm
Division Sign Off
Division of General, Restorative. and Neurological Devices
510(k) Number K072682
{10}------------------------------------------------
K072682
Pg 4 of 4
- 1 1 1 -
- Debridement of decubitus ulcers
- Hemorrhoids :
- Debridement of statis ulcers
- Biopsy
Prescription Use __ X. (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Ogden-Erskine
(Division Sign-Off Division of General, Restorative, and Neurological Devices
Page 4 of 4
510(k) Number_clock) 22682
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.