TATTOOSTAR Y

K071451 · Asclepion Laser Technologies GmbH · GEX · Aug 1, 2007 · General, Plastic Surgery

Device Facts

Record IDK071451
Device NameTATTOOSTAR Y
ApplicantAsclepion Laser Technologies GmbH
Product CodeGEX · General, Plastic Surgery
Decision DateAug 1, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TattooStar Y is indicated for incision, excision, ablation and vaporization of soft tissue in general dermatology and the removal of tattoos, pigmented lesions, vascular lesions and unwanted hair.

Device Story

TattooStar Y is a Q-switched solid-state laser system emitting wavelengths of 1064 nm, 532 nm, and 585 nm. The device consists of a laser enclosure and an articulated mirror arm for optical delivery. It is intended for use by clinicians in a dermatology setting for soft tissue procedures and the removal of tattoos, pigmented lesions, vascular lesions, and hair. The laser energy is applied to the target tissue to achieve the desired clinical effect (ablation, vaporization, or pigment destruction). The device is operated by a physician or qualified healthcare professional.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Q-switched solid-state laser; wavelengths: 1064 nm, 532 nm, 585 nm; delivery system: articulated mirror arm; classification: Class II, 21 CFR 878.4810, Product Code GEX.

Indications for Use

Indicated for incision, excision, ablation, and vaporization of soft tissue in general dermatology, and for the removal of tattoos, benign pigmented lesions, vascular lesions, and unwanted hair in patients requiring dermatological laser treatment.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071451 ## 510(k) SUMMARY ASCLEPION LASER TECHNOLOGIES GmbH TattooStar Y AUG - 1 2007 This 510(k) summary of safety and effectiveness for the Asclepion Laser Technologies GmbH TattooStar Y is submitted in accordance with the requirements of 21 CFR 907.92 and follows Office of Device Evaluation Guidance concerning the organization and content of a 510(k) summary. | Applicant: | ASCLEPION LASER TECHNOLOGIES GmbH<br>Am Semmicht 1A<br>07751 Jena, Germany | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Mrs Antje Katzer<br>Product Management and<br>International Regulatory Affairs | | Phone:<br>Fax:<br>e-mail: | +49 3641 77 00 309<br>+49 3641 77 00 302<br>antje.katzer@asclepion.com | | Preparation Date: | May 21th, 2007 | | Device Name: | TattooStar Y | | Common Name: | TattooStar Y | | Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.481 | | Equivalent Device: | Medlite C6 | | Device Description: | The TattooStar Y is a q-switched solid state lasers emitting<br>wavelengths of 1064, 532 and 585 nm. It consists of a laser<br>enclosure and optic delivery system (articulated mirror arm). | | Intended Use: | The TattooStar Y is indicated for incision, excision, ablation<br>and vaporization of soft tissue in general dermatology and<br>the removal of tattoos, pigmented lesions, vascular lesions<br>and unwanted hair. | | Comparison to: | The TattooStar Y is substantially equivalent to the Medlite<br>C6, with the same principles of operation, and the same<br>indication for use. | | Nonclinical Performance Data: | None | | Clinical Performance Data: | None | | Conclusion: | The TattooStar Y is another safe and effective device<br>for incision, excision, ablation, vaporization of soft<br>tissue for general dermatology and the removal of<br>tattoos, pigmented lesions, vascular lesions and hair. | | Additional Information : | None | {1}------------------------------------------------ Qaita 9 unn 2 : - ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {2}------------------------------------------------ ## EXHIBIT F ## INDICATION FOR USE STATEMENT : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with a double helix design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Asclepion Laser Technologies % Antje Katzer 1A; IM SEMMICHT JENA, GERMANY 07751 AUG = 1 2007 Re: K071451 Trade/Device Name: TattooStar Y Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and and in dermatology Regulatory Class: II Product Code: GEX Dated: June 29, 2007 Received: July 2, 2007 Dear Antje Katzer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 – Antje Katzer This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ハモ Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: TattooStar Y Indications for Use: The TattooStar Y is intended for use for incision, excision, ablation, vaporization The Tattoostal T is Intended for use for the belows of tattoos, the removal of hair of soft tissue for general dematology and the romovial of hair. benign pigmented lesion, the removal of vascular lesion and the removal of hair. Division Signa vision Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number k071481 k11 Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...