← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K062871

# LUMACARE LC-122M NON-COHERENT LIGHT SOURCE (K062871)

_Lynton Lasers Limited · GEX · Nov 14, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K062871

## Device Facts

- **Applicant:** Lynton Lasers Limited
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Nov 14, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The LumaCare LC-122M Non-coherent Light Source is intended to provide therapeutic light to the body The LumaCare LC-122M Non-coherent Indications for Use : Light Source is generally indicated to treat dermatological conditions. The LumaCare LC-122M Non-coherent Light Source is specifically indicated to treat moderate inflammatory acne vulgaris.

## Device Story

LumaCare LC-122M is a non-coherent light therapy device for dermatological treatment. System consists of a desktop base unit with a 250W halogen bulb and a Fibre Optic Probe (FOP) delivery system. The base unit powers the lamp; a timing PCB controls illumination duration. An optical filter within the FOP selectively transmits therapeutic blue light (400-440nm) at fluences of 10-100mW/cm². An optional mechanical fixture holds the FOP at an adjustable distance/direction relative to the skin. The device is used by clinicians to expose the skin surface to precise wavelengths of light to treat moderate inflammatory acne vulgaris.

## Clinical Evidence

No clinical data provided. Bench testing only, including electrical safety and EMC testing per IEC 60601-1:1988/A1/A2 and IEC 60601-1-2:2002.

## Technological Characteristics

Non-coherent light source; 250W halogen bulb; 400-440nm blue light output; 10-100mW/cm² fluence. System includes desktop base unit, timing PCB, and Fibre Optic Probe (FOP) with optical filter. Mechanical fixture for probe positioning. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- BLU-U Blue Light Photodynamic Therapy Illuminator Model 4170 ([K031805](/device/K031805.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 062871

## Appendix A6 The 510(k) Summary

| Applicant & Submitter : | Lynton Lasers Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | NOV 14 2006 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address :               | Lynton House, Manor Lane,<br>Holmes Chapel, Cheshire, CW4 8AF, UK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |             |
| Telephone :             | +44 (0)1477 536977                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |
| Fax :                   | +44 (0)1477 536978                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |
| Contact Person :        | Dr. Andrew J Berry<br>AJBerry@lynton.co.uk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |             |
| Preparation Date :      | 3rd July 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |
| Device Submitted :      | LumaCare LC-122M<br>Non-coherent Light Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |
| Common Name :           | Light Therapy Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |             |
| Classification Name :   | Laser surgical instrument for use in general<br>and plastic surgery and in dermatology.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |             |
| Product Code :          | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |             |
| Predicate Device :      | BLU-U Blue Light Photodynamic Therapy<br>Illuminator Model 4170, manufactured by<br>DUSA Pharmaceuticals, Inc. (K031805)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |             |
| Device Description :    | The LumaCare LC-122M Non-coherent<br>Light Source is a high intensity lamp<br>intended for the therapy of dermatological<br>disorders such as moderate inflammatory<br>acne vulgaris by emitting visible light in the<br>wavelength range 400-440nm (blue) with<br>fluences of order 10-100mW/cm². The<br>principle parts of the system comprise of a<br>desktop base unit and a Fibre Optic Probe<br>(FOP) delivery system. The base unit<br>contains a mains supplied power supply unit<br>which powers a 250W halogen bulb, the<br>duration of the illumination being controlled<br>by a timing pcb with user-accessible<br>controls. The FOP delivery system<br>comprises of a ruggedised fibre bundle<br>assembly and (crucially) an optical filter<br>which selectively transmits only the<br>therapeutic blue light. A mechanical fixture |             |

{1}------------------------------------------------

is also optionally available for holding the output of the FOP delivery system at an adjustable distance and direction relative to the skin treatment area.

- Intended Use : The LumaCare LC-122M Non-coherent Light Source is intended to provide therapeutic light to the body
The LumaCare LC-122M Non-coherent Indications for Use : Light Source is generally indicated to treat dermatological conditions. The LumaCare LC-122M Non-coherent Light Source is specifically indicated to treat moderate inflammatory acne vulgaris.

- The pre-clinical testing includes Electrical Performance Data : Safety and EMC testing including the requirements of IEC 60601-1:1988/A1/A2 "Medical electrical equipment - General requirements for safety" and "IEC 60601-1-2:2002 "Medical electrical equipment -General requirements for safety. Collateral standard - Electromagnetic compatibility. Requirements and tests"
- Substantial Equivalence : The LumaCare LC-122M Non-coherent Light Source is substantially equivalent to the previously cleared BLU-U Blue Light Photodynamic Therapy Illuminator Model The LumaCare LC-122M has the 4170. same intended use and the same general and specific indications for use as the BLU-U. The spectral output, mode of operation and general operating principles for the LumaCare LC-122M are similar to or the same as the BLU-U. The LumaCare LC-122M and the BLU-U are both light device that are used to treat dermatological conditions by exposing the surface of the skin to light at precise wavelengths. Although there are some differences in method by which each device produces light and is delivered to the treatment area. these differences do not raise new questions of safety of efficacy. Thus, the LumaCare LC-122M Non-coherent Light Source is substantially equivalent to the BLU-U.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring an abstract design resembling an eagle or bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Lynton Lasers Limited % Dr. Andrew J. Berry Technical Director Lynton House, Manor Lane Holmes Chapel, Chesire CW 4 8 AF United Kingdom

Re: K062871

Trade/Device Name: LumaCare LC-122M Non-coherent Light Source Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 21, 2006 Received: September 25, 2006

NOV 1 4 2006

Dear Dr. Berry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Dr. Andrew J. Berry

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Penh

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K062871

LumaCare LC-122M Non-coherent Light Source Device Name:

Indications for Use: The LumaCare LC-122M Non-coherent Light Source is intended to provide therapeutic light to the body

> The LumaCare LC-122M Non-coherent Light Source is generally indicated to treat dermatological conditions. The LumaCare LC-122M Non-coherent Light Source is specifically indicated to treat moderate inflammatory acne vulgaris.

イ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Noh
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

**510(k) Number** *l(a6287)*

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K062871](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K062871)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
