K061720 · Penntack Enterprises, Inc. · GEX · Jan 31, 2007 · General, Plastic Surgery
Device Facts
Record ID
K061720
Device Name
COSMELIGHT
Applicant
Penntack Enterprises, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 31, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cosmelight Intense Pulsed Light System (and its accessories) are indicated for use in surgical, aesthetic and cosmetic applications (requiring photothermolysis, photocoagulation and dermatology) in the treatment of acne, various benign pigmented lesions and hair removal as follow: 1. Intense Pulse Light Energy Wavelengths from 400 - 950 nm are indicated for the treatment of acne. 2. Intense Pulse Light Energy Wavelengths from 560 - 1200 nm are indicated for the treatment of benign pigmented (epidermal and coetaneous) lesions including warts, scars and striae. For the treatment of benign (cutaneous) vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider angiormas, poikiloderma of civatte, leg veins, facial veins and venous malformations. 3. Intense Pulse Light Energy Wavelengths from 700 - 1200 nm are indicated for the treatment of unwanted hair (i.e., hair removal). 4. Quick Cool Handle ( QCH ) The Quick Cool Handle is indicated for use in cooling the epidermis at the treatment Site prior to, during and after light or laser treatment in general surgery, plastic surgery And dermatology to : Reduce pain associated with light or laser treatments (via partial anesthesia from Cooling). Minimize thermal injury, including thermal necrosis, to non target skin and skin Structures during light or laser treatments, thus reducing the possibilities of scabbing, scarring, hypo and / or hyper pigmentation. Allows the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatme of vascular or pigmented lesions.
Device Story
Cosmelight Intense Pulsed Light (IPL) system; delivers light energy via removable handpieces; wavelengths 400-950nm (acne), 560-1200nm (pigmented/vascular lesions), 700-1200nm (hair removal). Includes Quick Cool Handle (QCH) for epidermal cooling; reduces pain/thermal injury; allows higher fluences. System comprises control console, display panel, power supply, handpieces with integrated trigger switch. Used in clinical settings by healthcare professionals. Output: pulsed light energy for photothermolysis/photocoagulation. Clinical benefit: treatment of dermatological conditions and hair removal with reduced risk of thermal necrosis/pigmentation changes.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison to legally marketed predicate devices.
Technological Characteristics
IPL system; wavelengths 400-1200nm; includes control console, power supply, and handpieces with integrated cooling (QCH). Energy source: Intense Pulsed Light. Class II device (21 CFR 878.4810).
Indications for Use
Indicated for patients requiring dermatological treatment for inflammatory acne, benign pigmented lesions (epidermal/cutaneous), benign vascular lesions (e.g., port wine stains, hemangiomas, telangiectasias, rosacea, melasma, angiomas, venous malformations), and unwanted hair removal. Includes use of Quick Cool Handle for epidermal cooling to reduce pain and thermal injury during light/laser procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## 6 1720
## Section 3 - 510(k) Summary Re: 061720
JAN 3 1 2007
This 510(k) summary of safety and effectiveness for the Cosmelight Intense Pulsed Light system by Penntack Enterprises Inc. is submitted in accordance with the requirements of the Safe Medical Devise Act (SMDA) of 1990 and follows the office of Devise Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.
| Applicant : | Penntack Enterprises Inc. |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address : | 3321 NW 79 ST. Miami, FL 33147 |
| Contact Person : | Paul Pena ( Director ) |
| Telephone : | 305-835-0411 |
| Preparation Date : | May 5, 2006 |
| Device Trade Name: | Cosmelight |
| Common Name : | Intense Pulsed Light System |
| Classification Name | Laser surgical instrument for use in general and plastic and<br>dermatology.<br>21 CFR 878.4810<br>Product Code: GEX<br>Panel: 79 |
| Legally-marketed | The Cosmelight is substantially equivalent to the following currently |
| Predicate Devices: | marketed devices: |
| | Lumenis Inc IPL Quantum, K020839 |
| | Palomar Inc. EsteLux, K020453 |
| | McCue Plc. McCue Energist ULTRA, K040659 |
| | DDD A/S , Ellipse I2 PL, K060516 |
| System Description: | The Cosmelight Intense Pulsed Light System is comprised of the<br>following main components: |
| | • A control console unit. |
| | • A control and color display panel. |
| | • Two removable handpieces with an integrated trigger switch<br>and a Cooling System (a cooling device integrated into the<br>handpieces. Quick Cool Handle ( QCH ) |
| | • Power supply. |
| Indicated Use : | The Cosmelight Intense Pulsed Light System (and its accessories)<br>are indicated for use in surgical, aesthetic and cosmetic<br>applications (requiring photothermolysis, photocoagulation and<br>dermatology) in the treatment of acne, various benign pigmented<br>lesions and hair removal as follow: |
| | 1. Intense Pulse Light Energy Wavelengths from 400 - 950 nm<br>are indicated for the treatment of acne. |
| | 2. Intense Pulse Light Energy Wavelengths from 560 - 120 0 nm<br>are indicated for the treatment of benign pigmented (epidermal |
| | and coetaneous) lesions including warts, scars and striae. For the<br>treatment of benign (cutaneous) vascular lesions including port<br>wine stains, hemangiomas, facial, truncal and leg telangiectasias,<br>rosacea, melasma, angiomas and spider angiormas, poikiloderma<br>of civatte, leg veins, facial veins and venous malformations. |
| | 3. Intense Pulse Light Energy Wavelengths from 700 - 1200 nm<br>are indicated for the treatment of unwanted hair (i.e., hair<br>removal). |
| | 4. Quick Cool Handle ( QCH ) |
| | The Quick Cool Handle is indicated for use in cooling the<br>epidermis at the treatment Site prior to, during and after<br>light or laser treatment in general surgery, plastic surgery<br>And dermatology to :<br>Reduce pain associated with light or laser treatments<br>(via partial anesthesia from Cooling). |
| | Minimize thermal injury, including thermal necrosis, to<br>non target skin and skin Structures during light or laser<br>treatments, thus reducing the possibilities of scabbing,<br>scarring, hypo and / or hyper pigmentation. |
| | Allows the use of higher light or laser fluences for light<br>or laser treatments (such as for hair removal and the treatme<br>of vascular or pigmented lesions. |
| Performance Data : | The differences in specifications of the Cosmelight system and the<br>Predicate devices do not result in different performance or raise<br>new questions of safety and efficacy. |
| Conclusion : | Based on the foregoing, the Cosmelight system is substantially<br>Equivalent to the legally-marketed predicate devices mentioned<br>above. |
{1}------------------------------------------------
## |2061720
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Penntack Enterprises, Inc. % Mr. Paul Pena Director 3321 NW 79th Street Miami, Florida 33147
JAN 3 1 2007
Re: K061720
Trade/Device Name: Cosmelight Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 24, 2006 Received: November 28, 2006
Dear Mr. Pena:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Mr. Paul Pena
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
son Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(K) Number (if known): K061720
Device Name: Cosmelight
Indications for Use:
The Cosmelight Intense Pulsed Light System (and its accessories) are indicated for use in surgical, aesthetic and cosmetic applications (requiring photothermolysis, photocoagulation and dermatology) in the treatment of acne, varlous benign pigmente lestons and hair removal as follow:
Intense Pulse Light Energy Wavelengths from 400 - 950 nm are indicated for the reatment of: inflammatory acne.
2. Intense Pulse Light Energy Wavelengths from 560 - 1200 nm are indicated for the treatment of: benign pigmented (epidermal and coetaneous) lesions including warts, cars and striae. For the treatment of benign (cutaneous) vascular lessons including ort wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider anglormas, poikiloderma of civatte, leg veins, facial veins and venous malformations.
3. Intense Pulse Light Energy Wavelengths from 700 - 1200 nm are indicated for the treatment of: unwanted hair (i.e., hair removal).
Prescription Use Part 21 CFR 801 Subpart D ) OR
Over-The-Counter Use ( 21CFR 807 Subpart C )
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE
L
510(k) Number
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.