← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K060680

# ATRILAZE SURGICAL ABLATION SYSTEM (K060680)

_Medical Cv, Inc. · GEX · Apr 11, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K060680

## Device Facts

- **Applicant:** Medical Cv, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Apr 11, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Medical CV Atrilaze™ Surgical Ablation System is indicated for the delivery of 810mm or 1064nm laser light to soft tissue to include cardiac tissue, during surgical procedures. Indications include the incision, excision, dissection, vaporization, ablation or coagulation of soft tissue.

## Device Story

Atrilaze Surgical Ablation System delivers laser energy to soft/cardiac tissue; consists of laser generator and handheld fiber optic probe. System provides choice of 810nm (GaAlAs semiconductor diode) or 1064nm (Nd:YAG) laser light. Probe connects to generator via standard SMA 905 connector. Used by clinicians in surgical settings for hemostasis, incision, coagulation, and vaporization. Output is laser radiation; clinician directs energy to target tissue. Device benefits include precise tissue interaction for surgical procedures. Complications include potential damage to adjacent organs, perforation, rupture, tearing, or burns if energy is misapplied.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included power output verification and visual/histological evaluation of lesions created on cardiac tissue using the 1064nm wavelength, confirming equivalence to the 810nm predicate. Biocompatibility testing (non-toxic, non-hemolytic, non-pyrogenic) performed under GLP per 21 CFR Part 58.

## Technological Characteristics

Laser surgical instrument. Energy source: 810nm GaAlAs semiconductor diode laser or 1064nm Nd:YAG laser. Delivery: Disposable fiber optic probe with SMA 905 connector. Materials: Biocompatible, non-toxic, non-hemolytic, non-pyrogenic. Operation: Standalone generator. Sterilization: Not specified (no change from predicate).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Atrilaze™ Surgical Ablation System ([K040744](/device/K040744.md))
- Atrilaze™ Surgical Ablation System ([K052495](/device/K052495.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

APR 1 1 2006

Page I of 2

## Section II

K060680

## Summary of Safety and Effectiveness (as required by 21 CFR 807.92)

## Atrilaze™ Surgical Ablation System

| Submitter:       | MedicalCV, Inc.<br>9725 South Robert Trail<br>Inver Grove Heights, MN 55077<br>USA | Contact:             | Denny Steger<br>V.P. RA/QA<br>Phone: 651 452 3000<br>Fax: 651 452 4948 |
|------------------|------------------------------------------------------------------------------------|----------------------|------------------------------------------------------------------------|
| Date of Summary: | March 14, 2006                                                                     | Classification Name: | Laser Instrument,<br>Surgical Powered                                  |
| Common Name:     | Surgical Laser Instrument                                                          | Proprietary          | AtrilazeTM Surgical<br>Ablation System                                 |

The Altrilaze Surgical Ablation System consists of a generator Description of Device: designed for the delivery of 810nm or 1064nm laser light and a hand held fiber optic light delivery device (probe) fitted with a standard SMA 905 connector at the proximal end. The system may be used in conjunction with surgical treatment for hemostasis, incision, coagulation and vaporization of tissue as required by the clinician.

The MedicalCV Atrilaze Surgical Ablation System is indicated Statement of Intended Use: for the delivery of 810mm or 1064nm laser light to soft tissue to include cardiac tissue, during surgical procedures. Indications include the incision, excision, dissection, vaporization, ablation or coagulation of soft tissue.

The Atrilaze™ Surgical Ablation System is not indicated for the treatment of Warning: cardiac arrhythmias.

The risk of actual damage to adjacent organs from the instrument exists and perforation, rupture or tearing of tissue, may occur as a complication of laser use. Burns can occur if the faser energy is not correctly applied. These complications may be serious.

{1}------------------------------------------------

Technological Comparison: The current Atrilaze Surgical Ablation System (K040744 & K052495) consists of a standalone surgical laser emitting 810mm light and a disposable fiber optic delivery system (probe). The fiber optic delivery system is coupled to the laser via an SMA 905 connector to deliver laser radiation to the targeted tissue. The proposed modification to the currently marketed Atrilaze™ Surgical Ablation System is to provide the end-user with an alternate laser generator (wavelength) for ablating soft tissue by providing the end user the option of using either a gallium aluminum arsenide (GaAlAs) semiconductor diode laser at a wavelength of 810nm or a Neodymium-yttrium-aluminum-garnet (Nd:YAG) laser at a wavelength of 1064nm. There is no change to the currently available fiber optic delivery systems (probes). which are coupled to either laser via an SMA 905 connector to deliver laser radiation to the target tissue. For purposes of this submission, the Nd: YAG laser (1064nm) used in conjunction with the Atrilaze Surgical Ablation System was compared to the following predicate device:

Medical CV Atrilaze™ Surgical Ablation System (K040744 & K052495) .

The results of biocompatibility testing conducted on the disposable fiber optic Testing: delivery systems (probes) contained in the original submissions (K040744 & K052495) support that the materials used in the manufacture of the disposable probe are non-toxic, non-hemolytic, and non-pyrogenic. All biocompatibility testing was conducted under Good Laboratory Practices per 21 CFR Part 58. Since there has been no change to the design, packaging and/or sterilization process for the disposable probe additional biocompatibility or sterilization testing was determined not to be required.

Performance testing for the Atrilaze Surgical Ablation System utilizing the Nd: Y AG 1064nm laser included compliance to manufacturing specifications for Power Output, with visual and histology evaluation of the lesions obtained using 1064nm wavelength on cardiac tissue. Testing demonstrated that lesions obtained using the Nd: Y AG laser at 1064nm wavelength are substantially equivalent to those obtained with the currently cleared laser at 810nm wavelength.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.

FEB 2 1 2008

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MedicalCV, Inc. c/o Mr. Denny Steger VP. Regulatory Affairs/Ouality Assurance 9725 South Robert Trail Inver Grove Heights, MN 55077

Re: K060680 Trade/Device Name: Atrilaze™ Surgical Ablation System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II (two) Product Code: OCL, GEX Dated: March 14, 2006 Received: March 15, 2006

Dear Mr. Steger:

This letter corrects our substantially equivalent letter of April 11, 2006.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Denny Steger

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

elzmell

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

510(K) Number:

Device Name: Atrilaze™ Surgical Ablation System

Indications for Use: The Medical CV Atrilaze™ Surgical Ablation System is indicated for delivery of 810nm or 1064nm laser light to soft tissue to include cardiac tissue during surgical procedures. Indications include the incision, dissection, vaporization, ablation, or coagulation of soft tissue.

The Atrilaze™ Surgical Ablation System is not indicated for the treatment Warning: of cardiac arrhythmias.

The risk of actual damage to adjacent organs from the instrument exists and perforation, rupture or tearing of tissue, may occur as a complication of laser use. Burns can occur if the laser energy is not correctly applied. These complications may be serious.

Prescription Use X (Per 21 CFR 801.109) OR

Over-the-Counter Use_

(Please do not write below this line -- Continue on another page if necessary)

| Concurrence of CDRH, Office of Device Evaluation (ODE)     |         |
|------------------------------------------------------------|---------|
| (Division Sign-Off)                                        |         |
| Division of General, Restorative, and Neurological Devices |         |
| 510(k) Number                                              | K060680 |

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K060680](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K060680)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
