PHOTOSILK PLUS PULSED LIGHT SYSTEM AND LASER ATTACHMENT

K051442 · Cynosure, Inc. · GEX · May 24, 2006 · General, Plastic Surgery

Device Facts

Record IDK051442
Device NamePHOTOSILK PLUS PULSED LIGHT SYSTEM AND LASER ATTACHMENT
ApplicantCynosure, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateMay 24, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PhotoSilk Plus Pulsed Light System is intended for permanent hair reduction, and the treatment of vascular and pigmented lesions, facial and leg veins, and inflammatory acne. The PhotoSilk Plus Pulsed Light System Laser Attachment is intended for: 2,940 nm - for skin resurfacing and for the incision, excision, ablation or vaporization of soft bodily tissues. 1,064 nm - for treatment of benign cutaneous vascular lesions, pigmented lesions, and hair removal. QS 1,064 nm - for tattoo removal, and the treatment of pigmented lesions.

Device Story

PhotoSilk Plus Pulsed Light System and Laser Attachment provides multi-wavelength light/laser energy for dermatological procedures; includes 400-1200nm pulsed light, 2,940nm Er:YAG, 1,064nm Nd:YAG, and 1,064nm Q-Switched Nd:YAG lasers. Device operated by physician in clinical setting; laser emission activated via foot switch. Energy delivery targets specific chromophores in skin/tissue for therapeutic effects (ablation, coagulation, hair follicle destruction). System requires 220 VAC power. Clinical benefit includes aesthetic improvement, lesion removal, and hair reduction.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Multi-wavelength system: 400-1200nm pulsed light, 2,940nm Er:YAG, 1,064nm Nd:YAG, 1,064nm Q-Switched Nd:YAG. Powered by 220 VAC, 20A, 50-60 Hz. Foot switch activation. Class II device.

Indications for Use

Indicated for permanent hair reduction, treatment of dermatological vascular lesions, facial/leg veins, benign pigmented lesions, inflammatory acne, skin resurfacing, incision/ablation/vaporization of soft tissue, wrinkle treatment, and dark tattoo removal.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ . | Submitter: | Cynosure, Inc.<br>10 Elizabeth Drive<br>Chelmsford, ΜΑ 01824 | 510(K) Summary | |-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------| | Contact: | George Cho<br>Senior Vice President of Medical Technology | K051442 | | Date Summary Prepared: | June 1, 2005 | | | Device Trade Name: | PhotoSilk Plus Pulsed Light System and Laser Attachment | | | Common Name: | Pulsed Light System.<br>Laser: Er:YAG laser, Nd:YAG laser, Q-Switched Nd:YAG laser. | | | Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.4810 | | | Equivalent Device: | StarLux Pulsed Light System, CO3 Er:YAG laser, ESC Nd:YAG<br>Accessory, and Medlite C6 Q-Switched Nd:YAG laser | | | Device Description: | The PhotoSilk Plus Pulsed Light System and Laser Attachment<br>provides 400-1200nm pulsed light, 2,940nm Er:YAG, 1064nm<br>Nd:YAG and 1064nm Q-Switched Nd:YAG wavelengths. Laser<br>emission activation is by foot switch. Electrical requirement is 220<br>VAC, 20A, 50-60 Hz, single phase. | | | Intended Use: | The PhotoSilk Plus Pulsed Light System is intended for permanent<br>hair reduction, and the treatment of vascular and pigmented lesions,<br>facial and leg veins, and inflammatory acne.<br>The PhotoSilk Plus Pulsed Light System Laser Attachment is intended<br>for:<br>2,940 nm - for skin resurfacing and for the incision, excision, ablation<br>or vaporization of soft bodily tissues.<br>1,064 nm - for treatment of benign cutaneous vascular lesions,<br>pigmented lesions, and hair removal.<br>QS 1,064 nm - for tattoo removal, and the treatment of pigmented<br>lesions. | | | Comparison: | The PhotoSilk Plus Pulsed Light System and Laser Attachment have<br>very similar indications for use, the same principle of operation, and<br>similar performance specifications as the predicate devices. | | | Nonclinical Performance Data: | none | | | Clinical Performance Data: | none | | | Conclusion: | The PhotoSilk Plus Pulsed Light System and Laser Attachment is a<br>safe and effective device for the indications specified. | | | Additional Information: | none | | : {1}------------------------------------------------ Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 4 2006 Cynosure, Inc. % Mr. George Cho Senior Vice President 5 Carlisle Road Westford, Massachusetts 01886 Re: K051442 Trade/Device Name: Cynosure PhotoSilk Plus Pulsed Light System Laser Attachment Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 2, 2006 Received: February 3, 2006 Dear Mr. Cho: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {2}------------------------------------------------ ## Page 2 - Mr. George Cho forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): _ K ( ( 5 | 4 2 Device Name: Cynosure PhotoSilk Plus Pulsed Light System Laser Attachment Indications For Use: The PhotoSilk Plus Pulsed Light System is intended for permanent hair reduction and the treatment of dermatological vascular lesions, facial and leg veins, benign pigmented lesions, and inflammatory acne. The PhotoSilk Plus Pulsed Light System Laser Attachment is intended for: Er: Y AG 2,940 mm - for skin resurfacing and for the incision, ablation or vaporization of soft bodily tissues. Long Pulse Nd: Y AG 1,064 nm - for the treatment of vascular lesions, pigmented lesions, wrinkles and for permanent hair reduction .. Q-Switched Nd: Y AG 1,064 nm - for dark tattoo removal, and the treatment of pigmented lesions. Prescriptive Use x OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) .m.de (Division Sign-Off Division of General, Restorative, and Neurological Devices Kos1442 510(k) Number.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...