LASER ENDOSCOPIC DECOMPRESSION KIT, MODEL 1100, AND PERCUTANEOUS DISCECTOMY KIT, MODEL 1120
K040424 · Clarus Medical, LLC · GEX · May 11, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040424
Device Name
LASER ENDOSCOPIC DECOMPRESSION KIT, MODEL 1100, AND PERCUTANEOUS DISCECTOMY KIT, MODEL 1120
Applicant
Clarus Medical, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
May 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Clarus Model 1100 Laser Endoscopic Decompression Kit is intended to be used on patients with contained lumbar or cervical disc herniations or bulges. This is a identical, with the addition of cervical, to the predicate device, Model 1100 Laser Endoscopic Decompression Kit (K922881)
Device Story
Kit provides instruments for endoscopic laser disc decompression; includes deflectable endoscope, fixed laser fiber, flexible trocar, cannulas, dilators, trephine, obturator, measuring scale, marking pen, and scalpel. Used by physicians in clinical settings for visualization and laser surgical removal of inner disc material. Device enables minimally invasive access to spine; shortened working length compared to predicate allows cervical application. Output is visual field for surgeon and laser energy for tissue ablation; aids in decompression of herniated or bulging discs.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and intended use similarity to previously cleared predicate devices.
Technological Characteristics
Stainless steel cannulas, dilators, and trephine with molded plastic proximal ends. Deflectable endoscope with fixed laser fiber. System includes manual surgical instruments. Sterilization is identical to predicate. No software or electronic sensing/actuation described.
Indications for Use
Indicated for laser disc decompression in lumbar and cervical spine regions for patients with contained disc herniations or bulges; used to remove inner disc material.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Model 1100 Laser Endoscopic Decompression Kit (K922881)
{0}------------------------------------------------
#### 1/2 ૨૦૫૦૫૨૫
MAY 11 2004
Section 4
# 510(k) Summary
## General Information:
| Submitted by: | | Clarus Medical, LLC<br>1000 Boone Avenue North<br>Minneapolis, MN 55427 | | |
|----------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------|--|--|
| Contact: | Tom Barthel, President<br>Telephone 763-525-8401<br>Facsimile 763-525-8656 | | | |
| Summary Date | February 4, 2004 | | | |
| Device Name: | Model 1100 Laser Endoscopic Decompression Kit | | | |
| Common Name: | Spinal endoscope w/ laser fiber. | | | |
| Classification Name: | Laser Instrument, Surgical, Powered; 878.4810 | | | |
### Predicate Devices:
| 510(k) | Description | Manufacturer |
|---------|-----------------------------------------------|---------------------|
| K922881 | Model 1100 Laser Endoscopic Decompression Kit | Clarus Medical, LLC |
| K011454 | Model 2180 Spinescope Endoscope | Clarus Medical, LLC |
| K022610 | Model 1150 Laser Fiber | Clarus Medical, LLC |
### Intended Use:
The Clarus Model 1100 Laser Endoscopic Decompression Kit is intended to be used on patients with contained lumbar or cervical disc herniations or bulges. This is a identical, with the addition of cervical, to the predicate device, Model 1100 Laser Endoscopic Decompression Kit (K922881)
Page 4.1
{1}------------------------------------------------
# K040424 2/2
#### Device Description:
#### General
This 510(k) submission is a modification of the existing Clarus Model 1100 Laser Endoscopic Decompression Kit, previously filed as K922881, and found to be substantially equivalent by the FDA on November 16, 1992. The K922881 510(k) device is intended for lumbar disc decompression in the spine. The modifications, represented by this submission, is the addition of cervical indications.
The Clarus Model 1100 Laser Endoscopic Decompression Kit intended use is to be endoscopic laser decompression of discs in the spine (lumbar and cervical). The kit consists of components necessary for endoscopic laser surgery where visualization and laser surgical techniques are required. The Model 1100 consists of a deflectable endoscope with a fixed laser fiber, a flexible trocar, straight and curved cannulas with dilators, a trephine and obturator, a measuring scale, a skin marking pen, and a scalpel. The cervical LASE is identical to the Clarus Model 2180 Spinescope (previously cleared for cervical visualization, K011454) with a shorter working length and the addition of a fixed laser fiber. The fixed laser fiber is identical to Clarus Model 1150 Laser Fiber (previously cleared for cervical soft tissue, K022610).
Construction
The Clarus Model 1100 Laser Endoscopic Decompression Kit, contain the same items, and are manufactured, packaged, and sterilized identically, with one exception, to the device which have been previously filed with FDA under 510(k) application K922881 and found to be equivalent. This exception is that the working length of the device is being shortened for cervical applications. The cannulas, trocars, dilators, and trephine will likewise be changed to accommodate the working length of the device
As with the previous kits, the main components, (the endoscope, cannulas, and dilators) are manufactured by Clarus. The other individual components have been selected by Clarus to offer the user a comprehensive set of instruments for endoscopic laser disc decompression.
The cannulas and dilators are manufactured of stainless steel with a molded plastic proximal end. The trephine (coring needle) is of similar construction as well. These materials are standard to the industry for surgical instruments.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 1 2004
Mr. Tom Barthel President Clarus Medical, LLC 1000 Boone Avenue North Minneapolis, Minnesota 55427
Re: K040424
Trade/Device Name: Model 1100 Laser Endoscopic Decompression Kit Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 17, 2004 Received: February 18, 2004
Dear Mr. Barthel:
We have reviewed your Section 510(k) premarket notification of intent to market the device wfo have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Tom Barthel
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and will and on a finding of substantial equivalence of your device to a legally prematics notification: "The stassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific arrivation at (301) 594-4659. Also, please note the regulation entitled, Contact the Ories of Court Court Cation" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 10(k) Number (if known): | A 11 011 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | |
Model 1100 Laser Endoscopic Decompression Kit
Indications For Use:
Device Name:
The Clarus Model 1100 Laser Endoscopic Decompression Kit indicated for laser disc decompression in the lumbar and cervical regions of the spine, where the laser is used to remove inner disc material.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Prescription Use (Per 21 CFR 801.109)
iriam C. Provost
OR
Over-The-Counter Use_
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
(Optional Format 1-2-96)
1040424 510(k) Number_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.