← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K032704

# INNOVATECH ILLUMINATING LASER PROBE, MODEL 420-30 (K032704)

_Innovatech Surgical, Inc. · GEX · Oct 24, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K032704

## Device Facts

- **Applicant:** Innovatech Surgical, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Oct 24, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

For photocoagulation and illumination during ophthalmic surgery. This Indications for Use: r or photoougaliation, as well as laser energy, to target tissue causing coagulation. Spot size can be varied by altering the distance between the tissue and the probe tip.

## Device Story

Innovatech Illuminating Laser Probe is an ophthalmic laser delivery device. It does not generate or intensify energy; it acts as a conduit. Input: laser energy from an external source. Components: connector for laser source, glass silica core fiber for laser delivery, acrylic fiber for illumination, PVC jacket, Delrin handpiece, and 304 stainless steel 20-gauge needle. Operation: clinician connects probe to a compatible laser source; laser energy is delivered to target tissue for photocoagulation; illumination is provided via acrylic fiber. Spot size is adjusted by varying the distance between the probe tip and the tissue. Used in clinical ophthalmic surgical settings. Benefits: provides simultaneous illumination and laser delivery to facilitate precise tissue coagulation.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Materials: 304 stainless steel needle, Delrin handpiece, glass-silica core optical fiber (.008"/200 microns), acrylic illumination fiber, PVC jacket. Energy: passive laser delivery (max 1 watt). Connectivity: connects to external laser source. Sterilization: ETO (AAMI Overkill Method).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Peregrine Illuminated Laser Probe ([K031023](/device/K031023.md))
- HGM Illuminating Laser Probe ([K931784](/device/K931784.md))
- Peregrine Straight Laser Probe ([K024061](/device/K024061.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Innovatech Surgical, Inc. The logo features a stylized sunburst graphic to the left of the word "Innovatech" in large, bold font. To the right of "Innovatech" is the text "SURGICAL, INC." in a smaller, but still prominent font. The text "KO32704 1 of 2" is at the top right of the image.

# OCT 2 4 2003

August 26, 2003

### Premarket Notification [510(k)] Summary

Submitter: Innovatech Surgical Inc. 1000 Atlantic Avenue, Suite 514 Camden, NJ 08104 Phone: 800-240-3123 856-225-1203 Fax:

Official Correspondent: Michael J. McGowan, Sr.

Trade Name: Innovatech Illuminating Laser Probe

Common Name: Ophthalmic Laser Probe

Registration Number: 3003988504

Classification: Class II

- Class Name: We were unable to find the device listed in the Disposable classification regulations, 21 CFR Parts 862-892 [807.87 (c)]
Panel: Ophthalmic

Product Code: HQF

- Device Description: The Innovatech Illuminating Laser Probe is an ophthalmic laser delivery device. By its design, it does not generate, intensify or significantly reduce energy. It consists of a connector that is plugged into an existing laser source, a glass fiber for laser delivery and acrylic fiber for illumination with PVC jacket, a Delrin handpiece and 304 stainless needle. The specific laser source to which the probe is connected will be specified in the "Indications for Use."
Statement of Indications for use: For photocoagulation during ophthalmic surgery. To be used with (specified equipment).

Page 3-1

1000 Atlantic Avenue, Suite 514, Camden, New Jersey 08104 800-240-3123 innovatechsurgical.com

{1}------------------------------------------------

K032704 2 of 2

#### Substantial Equivalence Comparison

| Application for 510(k) | Substantial Equivalence to: |                                    |                        |
|------------------------|-----------------------------|------------------------------------|------------------------|
| Innovatech             | Peregrine Illuminated       | HGM Illuminating                   | Peregrine Straight     |
| Illuminating Laser     | Laser Probe                 | Laser Probe (Gamp<br>& Associates) | Laser Probe            |
| Probe                  | Product: PD600.10           | 510(k): K931784                    | Product: PD600.00      |
| Product: 420-30        | Manufactured by             |                                    | Manufactured by        |
|                        | Peregrine Surgical Ltd      |                                    | Peregrine Surgical Ltd |
|                        | 510(k): K031023             |                                    | 510(k): K024061        |
| Illumination and Light | Illumination and Light      | Illumination and Light             | Light transmission for |
| transmission for       | transmission for            | transmission for                   | photocoagulation       |
| photocoagulation       | photocoagulation            | photocoagulation                   |                        |
| Aluminum connector     | Aluminum connector          | Aluminum connector                 | Aluminum connector     |
| Delrin Handpiece       | Delrin Handpiece            | Delrin Handpiece                   | Delrin Handpiece       |
| Optical Fiber          | Optical Fiber               | Optical Fiber                      | Optical Fiber          |
| Glass - Silica Core    | Glass - Silica Core         | Glass - Silica Core                | Glass - Silica Core    |
| .008" (200 microns)    | .008" (200 microns)         | .008" (200 microns)                | .008" (200 microns)    |
| PVC Jacket             | PVC Jacket                  | Teflon Jacket                      | PVC Jacket             |
| Length 101 inches      | Length 101 inches           | Length 96 inches                   | Length 101 inches      |
| 304 Stainless Needle   | 304 Stainless Needle        | 304 Stainless Needle               | 304 Stainless Needle   |
| 20 Gauge               | 20 Gauge                    | 20 Gauge                           | 20 Gauge               |
| Max power output       | Max power output            | Max power output                   | Max power output       |
| 1 watt                 | 1 watt                      | 1 watt                             | 1 watt                 |
| Illumination           | Illumination                | Illumination                       | No illumination        |
| Acrylic Optical fiber  | Acrylic Optical fiber       | Acrylic Optical fiber              | N/A                    |

## Sterility

The Device will be ETO Sterilized.

The method used to validate the sterilization cycle is AAMI Overkill Method.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular fashion around the caduceus symbol.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 4 2003

Mr. Michael J. McGowan, Sr. Innovatech Surgical, Inc. 1000 Atlantic Avenue, Suite 514 Camden, New Jersey 08104

Re: K032704

Trade/Device Name: Innovatech Illuminating Laser Probe Regulation Number: 21 CFR 878.4810, 886.4390 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology, Ophthalmic laser Regulatory Class: II Product Code: GEX, HQF Dated: August 26, 2003 Received: September 2, 2003

Dear Mr. McGowan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Michael J. McGowan, Sr.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

K032704

510(k) Number (if known):

Device Name:

Innovatech Illuminating Laser Probe

For photocoagulation and illumination during ophthalmic surgery. This Indications for Use: r or photoougaliation, as well as laser energy, to target tissue causing coagulation. Spot size can be varied by altering the distance between the tissue and the probe tip.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED:

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ______xx_____________________________________________________________________________________________________________________________________________________

OR

Over-The-Counter Use

Musiam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________

Page 3-3

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K032704](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K032704)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
