← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K024335

# CANDELA GENTLELASE FAMILY OF LASER SYSTEMS (K024335)

_Candela Corp. · GEX · Mar 12, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K024335

## Device Facts

- **Applicant:** Candela Corp.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Mar 12, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The GentleLASE Family of lasers is indicated for use for the treatment of wrinkles.

## Device Story

The Candela GentleLASE Family of Laser Systems is a dermatology laser device used for the treatment of wrinkles. It utilizes an Alexandrite crystal to emit pulsed 755 nm near-infrared laser energy. The system includes a high-voltage power supply, optical laser head, circulator, optical delivery system, software control system, and a Dynamic Cooling Device (DCD). The DCD sprays cryogen onto the skin immediately before laser pulses to provide cooling. The device is operated by a clinician via an onboard control panel. Laser energy is delivered through an optical fiber to a handpiece, which produces a circular beam on the target skin area. Safety interlocks are integrated to protect patients and operators. The device is intended for clinical use.

## Clinical Evidence

The clinical evidence for the treatment of wrinkles is based on a controlled clinical study using a device previously cleared for the market. No new clinical data was generated for this specific 510(k) submission.

## Technological Characteristics

Alexandrite crystal laser source; 755 nm wavelength; pulsed energy delivery; optical fiber/handpiece delivery system; integrated Dynamic Cooling Device (cryogen spray); safety interlock systems; onboard control panel. Conforms to 21 CFR 1040 (Laser Performance Standard), UL 2601 (Electrical Safety), and EN 60601-1-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Candela GentleLASE GL ([K994260](/device/K994260.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

$$
\varkappa \circ \mathfrak{Z} \star \mathfrak{Z} \circ \mathfrak{Z} \qquad \forall \nu
$$

MAR 1 2 2003

# General Information:

This 510(k) is to provide notification of substantial equivalence for the Candela GentleLASE Family of Laser Systems, which is substantially equivalent to previously marketed devices . The GentleLASE Family of lasers is indicated for use for the treatment of wrinkles.

| Submitted by:      | Candela Corporation<br>530 Boston Post Road<br>Wayland, MA 01778-1886                                               |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person:    | William H. McGrail                                                                                                  |
| Date prepared:     | December 26, 2002                                                                                                   |
| Classification:    | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use in<br>General and Plastic Surgery and in Dermatology) |
| Common Name:       | Dermatology Laser, GentleLASE Family of Laser Systems                                                               |
| Predicate Devices: | Candela GentleLASE GL (K994260)                                                                                     |

### Description:

The Candela GentleLASE Family of Lasers utilizes an Alexandrite rod (crystal) which emits pulsed energy at 755 nanometers in the near infrared region. Energy from the laser is directed to the target area via optical fiber/handpiece delivery system. The Dynamic Cooling Device, which provides a short burst of cryogen spray prior to firing the laser pulse. The laser output energy is delivered via an optical fiber to a handpiece, which produces a circular beam on the skin. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece. The GentleLASE Family of Laser Systems are designed with six major components:

- High voltage power supply and modulator system 1.
- 2. Optical laser head
- Circulator system 3.
- 4. Optical delivery system
- Software control system న్.
- 6. Dynamic cooling device

The Candela GentleLASE Family of Laser Systems are equipped with safety interlock systems to protect patients and operators. Users of the device make selections from an onboard control panel to regulate operation during treatment.

{1}------------------------------------------------

#### GentleLASE Family of Laser Systems 510k Summary

KO24335 24

# Testing:

As laser products, the GentleLASE Family of laser Systems are required to conform and does conform to the Laser Performance Standard (21 CFR 1040). In addition the GentleLASE Family of Laser Systems conforms to the UL 2601 Electrical Safety Standard and with the Harmonized Standard EN 60601-1-2, Part 2 established by and required by the European Community.

# Safety and Effectiveness Information:

The indications for use for the treatment of wrinkles is based on a controlled clinical study using a device that has been cleared for use in the market. We therefore believe that there are no questions of safety or effectiveness raised by the introduction of the Candela GentleLASE Family of Laser Systems.

### Summary of Substantial Equivalence:

The Candela GentleLASE Family of Laser Systems have the same intended use, utilizes similar operating principles and matches key design aspects, including similar spot size, the same wavelength and the same maximum delivered fluence as the predicate device.

On the basis of similarities in methods of assembly, method of operation, and intended uses, Candela Corporation believes that the Candela GentleLASE Family of Laser Systems is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUM." is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 2 2003

Mr. William H. McGrail Vice President Research & Development and Operations Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778

Re: K024335

Trade/Device Name: GentleLASE Family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 26, 2002 Received: December 27, 2002

Dear Mr. McGrail:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general, controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Miriam C. Provost

ful Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Candela. The logo consists of a stylized flame on the left and the word "CANDELA" on the right. The flame is black on the left side and has horizontal lines on the right side. The word "CANDELA" is in a simple, sans-serif font.

# INDICATION FOR USE STATEMENT

K024331 510(k) Number (if known):

Candela GentleLASE Family of Laser Systems Device Names

Indications For Use:

- 1. Treatment of Wrinkles
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) OR Over-The-Counter Use Prescription Use (Per 21 CFR 801.109) Division of General, Restorative and Neurologica: Devices 1842-1 4024335 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K024335](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K024335)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
