← Product Code [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX) · K011677

# MEDLITE Q-SWITCHED ND:YAG LASER REPLACED BY THE MEDLITE C3 Q-SWTICHED ND:YAG LASER (K011677)

_Continuum Electro-Optics, Inc. · GEX · Jun 22, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K011677

## Device Facts

- **Applicant:** Continuum Electro-Optics, Inc.
- **Product Code:** [GEX](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX.md)
- **Decision Date:** Jun 22, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

Tattoo Removal Treatment of Vascular Lesions Treatment of Pigmented Lesions Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology

## Device Story

Medlite C3 Q-Switched Nd:YAG laser system; utilizes laser energy for dermatological procedures. Device integrates subsystems from predicate Continuum laser products. Operated by physicians in clinical settings for tattoo removal, vascular/pigmented lesion treatment, and soft tissue surgery. Output is controlled laser energy; clinical benefit derived from targeted tissue interaction (ablation/vaporization/pigmentation removal).

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Q-Switched Nd:YAG laser system. Design and construction based on integration of subsystems from predicate Continuum laser devices. Operates within established parameters for medical laser systems (Product Code GEX).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- MedliteTM Q-Switched Nd:YAG Laser
- MedliteTM IV Q-Switched Nd:YAG Laser

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 2 2 2001

510(k) Summary

| Submitter:                    | Continuum Electro-Optics, Inc.<br>3150 Central expressway<br>Santa Clara, CA 95051                                                                                                                                                                                                           | K011677 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Contact:                      | Ronald Kohlhardt                                                                                                                                                                                                                                                                             |         |
| Date Summary Prepared:        | December 15, 2000                                                                                                                                                                                                                                                                            |         |
| Device Trade Name:            | MedliteTM C3 Q-Switched Nd:YAG Laser                                                                                                                                                                                                                                                         |         |
| Common Name:                  | Medical Laser System                                                                                                                                                                                                                                                                         |         |
| Classification Name:          | Instrument, surgical, powered, laser<br>79-GEX                                                                                                                                                                                                                                               |         |
| Equivalent Device(s):         | MedliteTM Q-Switched Nd:YAG Laser<br>MedliteTM IV Q-Switched Nd:YAG Laser                                                                                                                                                                                                                    |         |
| Intended Use:                 | • Tattoo Removal<br>• Treatment of Vascular Lesions<br>• Treatment of Pigmented Lesions<br>• Incision, Excision, Ablation, Vaporization of Soft Tissue for<br>General Dermatology                                                                                                            |         |
| Comparison:                   | The MedliteTM C3 Q-Switched Nd:YAG Laser bases its<br>design/construction from the integration of primary subsystems<br>extracted from legally marketed predicate Continuum products. The<br>new device performs and is specified within all product parameters<br>of the predicate devices. |         |
| Nonclinical Performance Data: | None                                                                                                                                                                                                                                                                                         |         |
| Clinical Performance Data:    | None                                                                                                                                                                                                                                                                                         |         |
| Additional Information:       | None                                                                                                                                                                                                                                                                                         | 000123  |

:

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines.

Public Health Service

JUN 2 2 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Ronald Kohlhardt Director, Regulatory Compliance and Quality Assurance Continuum Electro-Optics, Inc. 3150 Central Expressway Santa Clara, California 95051

Re: K011677

Trade/Device Name: Medlite™ C3 Q-Switched Nd: YAG Laser Regulation Number: 878.4810 Regulatory Class: II Product Code: GEX Dated: May 25, 2001 Received: May 30, 2001

Dear Mr. Kohlhardt:

We have reviewed your Section 510(k) notification of intent to market the device referenced we nave leviewed your becalled is ro(x) nearceantially equivalent (for the indications for use above and we nave doloriminou the arteted predicate devices marketed in interstate commerce stated in the encrosure) to legally markets of the Medical Device Amendments, or to devices that provision in the masterial in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (Act. The general controls provisions of the Act include requirements for annual provisions of the rec. "The general onanufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (soo aso very anto additional controls. Existing major regulations (1 chiaries rippt val), it they or casyon in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good A substantially equirements, as set forth in the Quality System Regulation (QS) for Manufacturing Praction regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspections, the Food and Drug result in regulatory action. In addition, FDA may publish comply with the GMT regarersing your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might response to your promation in 11.642 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. Ronald Kohlhardt

This letter will allow you to begin marketing your device as described in your 510(k) premarket I ntification. The FDA finding of substantial equivalence of your device to a legally marketed preadicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Marlin N Milliken

Celia M. Witten, Ph.D., M.D. Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Device Name:

Indications for Use:

Medlite™ C3 Q-Switched Nd: YAG Laser.

## KO11677

Tattoo Removal Treatment of Vascular Lesions Treatment of Pigmented Lesions Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use

L. Mark N. Milliken

Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K011677

(Optional Format 1-2-96)

00:33

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K011677](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEX/K011677)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
