CANDELA DYNAMIC COOLING DEVICE

K001589 · Candela Corp. · GEX · Aug 18, 2000 · General, Plastic Surgery

Device Facts

Record IDK001589
Device NameCANDELA DYNAMIC COOLING DEVICE
ApplicantCandela Corp.
Product CodeGEX · General, Plastic Surgery
Decision DateAug 18, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Candela Dynamic Cooling Device is: (1) cooling of the skin prior to laser treatment (2) reduction of pain during laser treatment (3) allows for use of higher laser fluences for laser treatments, such as for hair removal and vascular lesions (4) reduces potential side effects of laser treatments, such as for hair removal and vascular lesions

Device Story

Device consists of HFC 134a refrigerant source, electronically controlled delivery valve, and electronic timing circuitry. Integrated with laser trigger-switch system; delivers pulsed spray of refrigerant immediately preceding laser pulse. Evaporative cooling minimizes thermal injury to non-vascular skin structures and reduces treatment-associated pain. Used in clinical settings by laser operators. Enables use of higher laser fluences; reduces side effects like scabbing, scarring, and hyperpigmentation.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate devices.

Technological Characteristics

System utilizes HFC 134a refrigerant. Components include fluid source, electronically controlled delivery valve, and electronic timing circuitry. Integrated with laser trigger-switch. Class II device (21 CFR 878.4810).

Indications for Use

Indicated for patients undergoing laser treatment for hair removal or vascular lesions to provide skin cooling, pain reduction, and protection against thermal injury to non-vascular skin structures.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Koois P9 AUG 1 8 2000 # 510(k) Summary ### General Information: This 510(k) is to provide notification of substantial equivalence for the Candela Dynamic Cooling Device, which is substantially equivalent to previously marketed devices intended for cooling of the skin and roduction of pain during laser treatment (K951033) and for protecting the dermis from themal injury, which reduces possible side effects such as scabbing, scarring, and hyperpigmentation (K972347) and allows for the utilization of higher laser fluences for laser treatment, | Submitted by: | Candela Corporation | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 530 Boston Post Road | | | Wayland, MA 01778-1886 | | Contact Person: | Joan M. Clifford | | Date prepared: | May 19, 2000 | | Classification: | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use in<br>General and Plastic Surgery and in Dermatology) | | Trade Name: | Candela Dynamic Cooling Device | | Common Name: | Skin Refrigerant | | Predicate Devices: | Candela Dynamic Cooling Device cleared under 510(k) K951033 and<br>the Epidermal Chiller tip accessory to Coherent, Inc.'s Versanulse<br>Aesthetic Surgical Laser cleared under 510(k) K972347 | ### Description: Candela's Dynamic Cooling Device consists of 1) a source of skin refrigerant fluid (HFC 1348), 2) an electronically controlled soleasid delivery valve, and 3) electronic timing circuity. The Dynamic Cooling Device is connected in line with a laser's triggerswitch system so that activation of the triggerswitch controls the delivery of a pulsed spray of HFC 134a just prior to the delivery of a laser pulse. The pulsed spray of skin refrigerant cools the skin as it evaporates. Thermal injury to non-vascular structures is minimized and pain associated with the laser treatment is reduced. The intended use of the Candela Dynamic Cooling Device is to minimize injury to the skin structures during laser therapy of vascular lesions and hair removal, to reduce pain associated with laser treatment, to allow for use of higher laser fivences and to reduce potential side offects of laser treatments, such as hair removal and vascular lesions. #### Summary of Substantial Equivalence: The Candela Dynamic Cooling Device for the use of higher laser fluences and reduction of potential side effects of laser treatments is substantially equivalent to the Candela Dynamic Cooling Device which has been previously cleared for cooling of the skin prior to and reduction of pain during laser treatment. (K951033). Operating principles materials, design, construction, methods of assembly and other intended uses are the same as those for the predicate device. With respect to indications for use, Candela's Dynamic Cooling device is also substantially equivalent to Coherent Inc.'s Epidermal Chiller tip accessory (K972347). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure with three wing-like lines above it. The bird-like figure appears to be in flight. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 1 8 2000 K001589 Ms. Joan M. Clifford Clinical Research Manager Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778 Re: Trade Name: Candela Dynamic Cooling Device Regulatory Class: II Product Code: GEX Dated: May 19, 2000 Received: May 23, 2000 Dear Ms. Clifford: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class iff (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Ms. Joan M. Clifford This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, e R. lochner. Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): Candela Dynamic Cooling Device Device Name: Indications for Use: The intended use of the Candela Dynamic Cooling Device is: - (1) cooling of the skin prior to laser treatment - (2) reduction of pain during laser treatment - (3) allows for use of higher laser flucaccs for laser treatments, such as for hair removal and vascular lesions - (4) reduces potential side effects of laser treatments, such as for hair removal and vascular lesions (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) . OR Over-The Counter Use _ (Optional format 1-2-96) Dumas R. Jochner. (Division Sign-Off) (Division of General Restorative Devices 510(k) Number Koole Koole & T
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...