← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K983236

# BIPOLAR HF CABLE (K983236)

_Richard Wolf Medical Instruments Corp. · GEI · Dec 4, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K983236

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corp.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Dec 4, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Bipolar HF Cables are used to connect adequately equipped R. Wolf HF instruments to the corresponding HF devices.

## Device Story

Bipolar HF Cables serve as electrical conduits between high-frequency generators and R. Wolf HF surgical instruments. The device transmits high-frequency electrical current to enable electrosurgical procedures. It is intended for use in clinical settings by trained healthcare professionals. The cable design supports a maximum recurrent peak voltage of 4000 Vp for specific configurations and 1000 Vp for others. The device is reusable and compatible with steam sterilization at 134°C.

## Clinical Evidence

No clinical tests performed. Bench testing conducted to verify compliance with ANSI/AAMI HF18 and IEC601-2-18 standards.

## Technological Characteristics

Bipolar HF cable; maximum recurrent peak voltage of 4000 Vp or 1000 Vp depending on configuration; steam sterilizable at 134°C; compliant with ANSI/AAMI HF18 and IEC601-2-18 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Bipolar Cables 8108 ([K943776](/device/K943776.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

배

353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488

K983236

ਿੱਚ

RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION

## 510(k) Summary of Safety and Effectiveness

| Submitter:                             | Date of Preparation:                 |
|----------------------------------------|--------------------------------------|
| Company / Institution name:            | September 15, 1998                   |
| Richard Wolf Medical Instruments Corp. | FDA establishment regulation number: |
| Division name (if applicable):         | 14 184 79                            |
| N.A.                                   | Phone number (include area code):    |
| Street address:                        | (847) 913-1113                       |
| 353 Corporate Woods Parkway            | FAX number (include area code):      |
| City:                                  | (847) 913-0924                       |
| Vernon Hills                           |                                      |
| State/Province:                        | Country:                             |
| Illinois                               | USA                                  |
| Contact name:                          | ZIP/Postal Code:                     |
| Mr. Robert L. Casarsa                  | 60061                                |
| Contact title:                         |                                      |
| Quality Assurance Manager              |                                      |

Product Information:

| Trade name:      | Model number:        |
|------------------|----------------------|
| Bipolar HF Cable | 8108.xxx             |
| Common name:     | Classification Name: |
| Bipolar HF Cable | Bipolar HF Cable     |

Information on devices to which substantial equivalence is claimed:

| 510(k) Number   | Trade or proprietary or model name | Manufacturer        |
|-----------------|------------------------------------|---------------------|
| 1 K943776       | 1 Bipolar Cables 8108              | 1 Richard Wolf GmbH |
| 2 pre-enactment | 2 Bipolar Cables 8108              | 2 Richard Wolf GmbH |
| 3               | 3                                  | 3                   |
| 4               | 4                                  | 4                   |

#### 1.0 Description

The Bipolar HF Cables are used to direct a high frequency electrical current from an electrical generator made by unspecified manufacturers to a R. Wolf HF instrument.

{1}------------------------------------------------

#### 2.0 Intended Use

The Bipolar HF Cables are used to connect adequately equipped R.Wolf HF instruments to the corresponding HF devices.

#### Technological Characteristics 3.0

The bipolar HF Cables may be operated at a maximum recurrent peak voltage of 4000 Vp and the bipolar cables at a maximum 1000 Vp.

The cables can be sterilized by steam at 134℃.

#### 4.0 Substantial Equivalence

Bipolar HF Cables 8108 are substantially equivalent to Bipolar HF Cables cleared on K943776. The material changes have not reduced safety or effectiveness.

#### 5.0 Performance Data

No known FDA performance standard exists.

The Bipolar HF Cables were tested to comply with the appropriate sections of ANSI/AAMI standard on high frequency devices, HF18 and IEC601-2-18.

#### 6.0 Clinical Tests

No clinical tests performed.

#### 7.0 Conclusions Drawn

The devices are designed and tested to guarantee the safety and effectiveness when used according to the instruction manuals.

By: Robert J. Casassa

Robert L. Casarsa Quality Assurance Manager

Date: Sept 9, 98

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings or feathers. The eagle is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the border of the circle. The text is in all capital letters and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 4 1998

Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, Illinois 60061

Re: K983236 Trade Name: Bipolar HF Cable Regulatory Class: II Product Code: GEI Dated: September 15, 1998 Received: September 15, 1998

Dear Mr. Casarsa:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2 - Mr. Robert L. Casarsa

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

ia M. Witten, Ph.D., M.D. Cel Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Us

K983236

510(k) Number (if known):

Device Name:

Bipolar HF Cable

## Intended Use:

The Bipolar HF Cables are used to connect adequately equipped R. Wolf HF instruments to the corresponding HF devices.

## Contraindications:

Contraindications directly related to the product are currently unknown. For further information refer to the latest medical literature.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General Res 510(k) Number

Image /page/4/Picture/13 description: The image shows the phrase "Prescription Use Per 21 CFR 801.109". The text is in a sans-serif font and is black. The text is surrounded by a hand-drawn oval shape.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K983236](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K983236)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
