← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K981724

# O.R.PRO 300 ELECTROSURGICAL GENERATOR MODEL 40001-100 (K981724)

_Medtrex, Inc. · GEI · Jun 22, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K981724

## Device Facts

- **Applicant:** Medtrex, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jun 22, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The O. R. Pro 300™ Electrosurgical Generator is a general purpose solid state generator used to supply the RF signals to electrosurgical handpieces used on soft tissues where a wide range of tissue types, patient conditions and load impedances are encountered.

## Device Story

The O.R. Pro 300™ is a solid-state electrosurgical generator providing RF energy for soft tissue cutting and coagulation. It features separate monopolar and bipolar output sections. The monopolar section is floating and isolated, requiring a patient return electrode; the bipolar section is also floating and isolated for use with bipolar instruments. The device converts AC mains power to DC, then uses power MOSFETs in an RF amplifier to generate the output signal. Activation is controlled via handswitching or footswitching. Front panel controls allow selection of three cutting modes (Pure, Blend 1, Blend 2) and two coagulation modes (Pinpoint, Spray), plus bipolar coagulation. Used in clinical settings by healthcare professionals, the device provides the energy necessary for electrosurgical procedures, allowing surgeons to perform tissue cutting and hemostasis.

## Clinical Evidence

Bench testing only. The device was tested to all applicable requirements of AAMI HF18-R-10/93 (American National Standard for Electrosurgical Devices) and met all applicable requirements.

## Technological Characteristics

Solid-state electrosurgical generator; RF energy output via power MOSFETs; monopolar and bipolar isolated outputs; front panel push-button controls for modes (Pure, Blend 1, Blend 2, Pinpoint, Spray, Bipolar Coag); AC mains power input; tested to AAMI HF18-R-10/93 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Valley Laboratories Inc. Force FX™ Electrosurgical Generator ([K944602](/device/K944602.md))
- Aspen Surgical Systems ExcaliburPlus Electrosurgical Generator ([K953007](/device/K953007.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 22 1998

# SAFETY AND EFFECTIVENESS SUMMARY Medtrex O.R.Pro 300™ Electrosurgical Generator

## Name and address of Device Manufacturer submitting 510(k) Notification:

Medtrex Incorporated 14 Inverness Drive East, Building B, Suites 116-120 Englewood, Colorado 80112

Regulatory Correspondent of Device Manufacturer:

William J. Bowers Medtrex Incorporated 14 Inverness Drive East Building B Suites 116-120 Englewood. Colorado 80112 Phone: (303)790-2032 FAX: (303)790-0302

Date Summary was prepared:

May 11, 1998

Name of the device:

Classification:

Indications for Use:

Description of the device:

O.R.Pro 300™ Electrosurgical Generator

Electrosurgical cutting and coagulation device and accessories. Class II per 21CFR 878.4400

The O. R. Pro 300™ Electrosurgical Generator is a general purpose solid state generator used to supply the RF signals to electrosurgical handpieces used on soft tissues where a wide range of tissue types, patient conditions and load impedances are encountered.

The OR Pro 300™ is a full powered electrosurgical generator with separate monopolar and bipolar sections. The monopolar section is designed for cutting and coagulation outputs, is intended for use with a patient return electrode, and is a floating, isolated output. The bipolar section incorporates a separate output from the monopolar output and has independently controlled power. The Bipolar section

## 180 OF 181

{1}------------------------------------------------

is a floating isolated output and is intended to be used with bipolar instruments.

AC power mains is applied to the generator from a detachable line cord. This power mains voltage is converted to a DC voltage and applied to a single RF amplifier. An RF amplifier utilizes power MOSFETs to convert the DC voltage into a RF energy that is suitable for electrosurgical procedures. The generator is activated by either handswitching or footswitching active accessories. Three different types of CUTTING, Pure, Blend 1 and Blend 2, can be selected by the push buttons on the front panel. Two different types of COAG, Pinpoint and Spray, can also be selected on the front panel. In addition, BIPOLAR COAG is provided through the bipolar output jacks.

The Medtrex O.R. Pro 300 Electrosurgical Generator is substantially equivalent to the following two legally marketed electrosurgical generators:

Valley Laboratories Inc. Boulder, Colorado 80301 Force FX™ Electrosurgical Generator 510(k) K944602, cleared 6/5/95

Aspen Surgical Systems Div of ConMed /Corporation 311 Broad Street Utica, NY 13501 ExcaliburPlus Electrosurgical Generator 510(k) K953007 Cleared 8/21/95

The O.R.Pro has been tested to all applicable requirements of AAMI HF18-R-10/93 - American National Standard for Electrosurgical Devices. The O.R.Pro300 met all the applicable requirements of the Standard. The requirements of the Standard that were not applied to the O.R. Pro were those associated with the electrosurgical pencil and return pad. These devices were cleared for marketing by separate submissions listed under Accessories.

### Substantial Equivalence:

Safety and Efficacy:

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains an emblem of an eagle with three stylized lines representing its wings. The text "DEPARTMENT OF HEALTH & HUM. SERVICES - USA" is arranged around the emblem in a circular fashion. The seal is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 22 1998

Medtrex Incorporated c/o Mr. William E. McKay President RCMDI 9712 South Altamont Drive Sandy, Utah 84092

Re: K981724 O.R. Pro 300™ Electrosurgical Generator Trade Name: Regulatory Class: II Product Code: GEI Dated: May 13, 1998 Received: May 15, 1998

Dear Mr. McKay:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{3}------------------------------------------------

#### Page 2 - Mr. William E. McKay

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

- -

510(k) Number (if known): Not-yet-assigned

Device Name: _ O. R. Pro 300™ Electrosurgical Generator

Indications For Use:

The O. R. Pro 300™ Electrosurgical Generator is a general purpose solid state generator used to supply the RF signals to electrosurgical handpieces used on soft tissues where a wide range of tissue types, patient conditions and load impedances are encountered.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office if Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

(Division Sign-Off) Division of General Restorative Devic 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K981724](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K981724)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
