VNUS CLOSURE SYSTEM

K974521 · Vnus Medical Technologies, Inc. · GEI · Feb 20, 1998 · General, Plastic Surgery

Device Facts

Record IDK974521
Device NameVNUS CLOSURE SYSTEM
ApplicantVnus Medical Technologies, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 20, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VNUS Closure™ Catheter System is intended for use in the coagulation of blood vessels during general surgery.

Device Story

The VNUS Closure™ System is an electrosurgical device used for blood vessel coagulation during general surgery. It comprises a sterile, single-use Closure Probe, a non-sterile RF Generator, and an autoclave-sterilized instrument cable; an optional footswitch is available. The system operates by delivering high-frequency bipolar RF energy to a target site. The RF Generator is software-controlled and utilizes temperature feedback from a thermocouple located in the Closure Probe to regulate power delivery in a temperature-controlled, power-limited manner. The operator sets power, temperature, and time parameters on the generator, which displays real-time measurements of these values along with impedance. The device is intended for use by clinicians in a surgical setting to achieve hemostasis or vessel coagulation.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

High-frequency bipolar electrosurgical system. Components include a disposable probe with an integrated thermocouple for temperature feedback, an RF generator, and an instrument cable. The system is software-controlled, providing temperature-regulated and power-limited energy delivery based on user-defined setpoints for power, temperature, and time. Connectivity is limited to the physical connection between the probe, cable, and generator.

Indications for Use

Indicated for the coagulation of blood vessels during general surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness - Determination of Substantial Equivalence A. VNUS Closure™ System FEB 2 0 1998 в. Common Name Electrosurgery System/Electrosurgical Coagulator ## C. Predicate Device(s) ArthroCare Electrosurgery System (K971532) Boston Surgical Products, Inc. Model 50 Bipolar Coagulator (K972350) Somnus Model 215 Electrosurgical Generator and Accessories (K971711) Valley Forge Scientific Mini-Symm Bipolar Coagulator (K964143) Bergen Model 500 Electrosurgery Generator (K972299) Bergen 610 Coagulator (K964736) Kirwan 28 1000 Coagulator (K971341) K974521 #### D. Device Description The VNUS Closure™ System consists of three main components: The VNUS Closure Probe, the VNUS RF Generator and the VNUS Instrument Cable. The Closure Probe is provided Sterile, and is a single-use, disposable device. The RF Generator is non-sterile. The Instrument Cable is autoclave sterilized by the user. An optional Footswitch for RF ON/RF OFF is provided for use at the physician's discretion. The RF Generator is a high frequency electronic, bipolar, software controlled instrument. It allows the user to set Power, Temperature and Time values, and provides user displays of Power. Temperature and Time (setpoints and measured values) as well as measured impedance and other messages. The RF Generator works in a temperature controlled, power limited manner, based on operator settings and temperature feedback provided by a thermocouple in the Closure Probe. The Closure Probe is used to carry RF energy to the desired treatment site and provide temperature feedback to the RF Generator. It is designed to deliver the RF energy in a bipolar manner. The Instrument Cable is used to connect the Closure Probe to the RF Generator. #### ய் Intended Use The VNUS Closure™ Catheter System is intended for use in the coagulation of blood vessels during general surgery. This premarket notification is being submitted as the VNUS Closure System is a new device. Electrosurgical Coagulation Devices and Accessories are Class 2 per 21 CFR 878.4400. {1}------------------------------------------------ # F. Intended Use of Predicate Devices The specified predicate devices are indicated for "soft tissue resection", "soft body tissue coagulation", "the coagulation of tissue", and/or for "ablation and coagulation of blood vessels during general surgical procedures." # Technological Comparison G. The technological characteristics and principals of operation of the VNUS Closure System are substantially equivalent to the noted predicate devices. All devices rely on the delivery of RF energy to achieve their intended use. ## H. Discussion of Non-Clinical Tests and Conclusions Non-clinical tests performed by VNUS have demonstrated the substantially equivalent performance of the Closure System with predicate electrosurgery systems used for substantially equivalent indications. ## .. Summary of Safety and Effectiveness Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-ciinical testing performed by VNUS, it is concluded that the Closure System is substantially equivalent to the noted predicate devices in safety and effectiveness. Tom M. Holdych Vice President, Regulatory and Clinical Affairs VNUS Medical Technologies, Inc. December 1, 1997 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are black against a white background. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 0 1998 Mr. Tom Holdych Vice President, Regulatory and Clinical Affairs VNUS Medical Technologies, Inc. 238 East Caribbean Drive Sunnyvale, California 94089 Re: K974521 VNUS™ Closure™ System Trade Name: Regulatory Class: II Product Code: GEI Dated: December 1, 1997 December 2, 1997 Received: Dear Mr. Holdych: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act · include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with most with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. Holdych This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your . . device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Ph.D., M.D. a M. Witten, Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Device Name: Closure™ System 510(k) Number (if known): 974521 Indications for Use: The VNUS Closure™ System is indicated for use in the coagulation of blood vessels during general surgery. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) # CONCURRENCE OF CORH, OFFICE OF DEVICE EVALUATION (ODE) Over the Counter Use: (Per 21 CFR 801.109) 1 or Prescription Use: coello
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...