← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K973648

# MINI FORCEPS WITH HF, 3.5MM; MICRO FORCEPS WITH HF, 2MM (K973648)

_Richard Wolf Medical Instruments Corp. · GEI · Dec 17, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K973648

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corp.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Dec 17, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The forceps are used for grasping, manipulating and cutting as well as the dissection and biopsy of soft tissue/ organs under endoscopic control. The modular scissors can also be used for cutting of suture material. Minor hemorrhages can be coagulated using HF current, provided that products are marked accordingly. The modular forceps can be disassembled for easy cleaning and replacement.

## Device Story

MINI and MICRO forceps/scissors are endoscopic instruments for grasping, manipulating, cutting, dissecting, and biopsying soft tissue/organs. Devices feature 2mm or 3.5mm diameters for minimally invasive, pediatric, and outpatient laparoscopy. Instruments include HF connection for monopolar coagulation of minor hemorrhages. Modular design allows disassembly of jaws, shaft, and handle for cleaning/replacement; includes Luer connector for rinsing channels. Jaws are rotatable via handle knob. Operated by physicians in OR or clinic settings under endoscopic control. Output is mechanical tissue manipulation or electrosurgical coagulation. Benefits include reduced invasiveness and improved cosmetic outcomes compared to larger instruments.

## Clinical Evidence

No clinical tests performed. Evidence consists of bench testing, including mechanical load tests to verify jaw/part integrity and prevent deformation, and steam sterilization validation using the fractional method.

## Technological Characteristics

Electrosurgical endoscopic instruments (2mm/3.5mm diameter). Materials comparable to existing R. Wolf devices. Features: modular design (disassemblable jaws/shaft/handle), Luer connector for cleaning, rotatable jaws, monopolar HF connection. Steam sterilizable.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- 1 forceps/ scissors with H.F. 8383.03, .12, 930L, 930S (pre-enact.)
- Modular Forceps and Scissors System ([K935270](/device/K935270.md))
- Dissecting and Grasping Forceps 3.5/ 5 mm ([K935070](/device/K935070.md))
- Biopsy Instruments, Endoscopic Instrumentation 3.5/ 5 mm (Jarit)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

353 Corporate Woods Parkway - DEC 17 1997 Vernon Hills, Illinois 60064 Phone: 847-913.1113 Fax: 847.913.1488

K 97 3648

RICHARD W MEDICAL INSTRUMENTS COR

# 510(k) Summary of Safety and Effectiveness

| Submitter:                                                          |                                        |                 |          |                      | Date of Preparation:                             | September 22, 1997                     |                |
|---------------------------------------------------------------------|----------------------------------------|-----------------|----------|----------------------|--------------------------------------------------|----------------------------------------|----------------|
| Company / Institution name:                                         | RICHARD WOLF MEDICAL INSTRUMENTS CORP. |                 |          |                      |                                                  | FDA establishment registration number: | 14 184 79      |
| Division name (if applicable):                                      | N.A.                                   |                 |          |                      |                                                  | Phone number (include area code):      | (847) 913-1113 |
| Street address:                                                     | 353 Corporate Woods Parkway            |                 |          |                      |                                                  | FAX number (include area code):        | (847) 913-0924 |
| City:                                                               | Vernon Hills                           | State/Province: | Illinois | Country:             | USA                                              | ZIP / Postal Code:                     | 60061          |
| Contact name:                                                       | Mr. Robert L. Casarsa                  |                 |          |                      |                                                  |                                        |                |
| Contact title:                                                      | Quality Assurance Manager              |                 |          |                      |                                                  |                                        |                |
| Product Information:                                                |                                        |                 |          |                      |                                                  |                                        |                |
| Trade name:                                                         | MINI and MICRO Forceps with HF         |                 |          | Model number:        | 8390.xxx, 8391.xxx, 8393.96x, 8394.96x, 8756.xxx |                                        |                |
| Common name:                                                        | Forceps, biopsy, electric              |                 |          | Classification name: | Endoscopic electrosurgical unit and accessories  |                                        |                |
| Information on devices to which substantial equivalence is claimed: |                                        |                 |          |                      |                                                  |                                        |                |

|   | 510(k) Number | Trade or proprietary or model name                         |  | Manufacturer   |  |
|---|---------------|------------------------------------------------------------|--|----------------|--|
| 1 | pre-enact.    | 1 forceps/ scissors with H.F. 8383.03, .12, 930L, 930S     |  | 1 Richard Wolf |  |
| 2 | K935270       | 2 Modular Forceps and Scissors System                      |  | 2 Richard Wolf |  |
| 3 | K935070       | 3 Dissecting and Grasping Forceps 3.5/ 5 mm                |  | 3 Karl Storz   |  |
| 4 |               | 4 Biopsy Instruments, Endoscopic Instrumentation 3.5/ 5 mm |  | 4 Jarit        |  |

# 1.0 Description

.

(१);:

The forceps with HF are part of the MICRO and MINI instrument set for laparoscopic microsurgery, particularly suitable for diagnostic, smaller interventions, outpatient, and pediatric laparoscopy.

The instruments smaller diameter is less invasive and allows better cosmetic effects than bigger diameters.

{1}------------------------------------------------

#### 2.0 Intended Use

The forceps are used for grasping, manipulating and cutting as well as the dissection and biopsy of soft tissue/ organs under endoscopic control.

The modular scissors can also be used for cutting of suture material.

Minor hemorrhages can be coagulated using HF current, provided that products are marked accordingly.

#### Technological Characteristics 3.0

The forceps and scissors have comparable device materials as former R. Wolf devices. The dimensions are smaller for very minimal invasive laparoscopy. The design of the same, but miniaturized. The design of the handles is the same, but adapted to the smaller forceps.

The modular forceps and scissors with 3.5 mm diameter can be disassembled into jaws with shaft and handle for easy cleaning and replacement. They have a Luer connector to rinse cavities and channels of the forceps with a cleaning gun while reprocessing. The jaws are rotatable with an adjustment knob at the handle.

#### 4.0 Substantial Equivalence

These devices are substantially equivalent to existing pre-enactment devices and 510(k) devices sold by Richard Wolf and 510(k) devices sold by Karl Storz and Jarit.

#### 5.0 Performance Data

The Instruments have been tested to assure that there is no breakage of the jaws or other parts of the instrument.

Mechanical load tests show that there is no permanent deformation of the forceps if used normally.

The steam sterilization tests performed by Richard Wolf show that steam sterilization has no influence on the functional performance of the submitted devices when using the fractional method.

#### 6.0 Clinical Tests

No clinical tests performed.

#### 7.0 Conclusions Drawn

These devices are designed and tested to guarantee the safety and effectiveness when used according to the instructions manual.

By: Robert L. Casarcia

Robert L. Casarsa Quality Assurance Manager

Date: Sept 22, 97

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 17 1997

Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, Illinois 60061

Re: K973648

> Trade Name: MINI Forceps with HF, 3.5mm and MICRO Forceps with HF, 2mm Regulatory Class: II Product Code: GEI Dated: September 22, 1997 Received: September 25, 1997

Dear Mr. Casarsa:

We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for

Image /page/2/Picture/10 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with only the basic outline of a head and neck visible.

{3}------------------------------------------------

Page 2 - Mr. Robert L. Casarsa

devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

[signature]

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

k972648

510(k) Number (if known):

Device Name:

# MINI Forceps with HF, 3.5 mm and MICRO Forceps with HF, 2 mm

# Intended use:

The forceps are used for grasping, manipulating and cutting as well as the dissection and biopsy of soft tissue/ organs under endoscopic control.

The modular scissors can also be used for cutting of suture material.

Minor hemorrhages can be coagulated using HF current, provided that products are marked accordingly.

The modular forceps can be disassembled for easy cleaning and replacement.

# Indication and Applications:

For examination, diagnostics and/or therapy by qualified, suitably trained personnel in connection with endoscopically used accessories in different medical disciplines, such as surgery, urology and gynecology, particularly suitable for outpatient and pediatric laparoscopy.

# Contraindications:

There are no known contraindications directly relating to the product. The physician in charge must decide whether the intended application is possible taking into account the general condition of the patient. For further instructions please see the current technical literature.

### Combinations:

The forceps have a HF connection for the use with monopolar HF units.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANQTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (OD

Office of Device Evaluation (ODE)

neral Restorative De

Prescription Use
Per CRF 801.109

/ 3w

OR

Over-The Counter

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K973648](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K973648)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
