UNCOATED ELECTROSURGICAL ELECTRODE

K973346 · Megadyne Medical Products, Inc. · GEI · Nov 7, 1997 · General, Plastic Surgery

Device Facts

Record IDK973346
Device NameUNCOATED ELECTROSURGICAL ELECTRODE
ApplicantMegadyne Medical Products, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateNov 7, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

This device is intended to be used in any application which requires electrosurgical cutting or coagulation.

Device Story

MegaDyne uncoated electrosurgical electrode; stainless steel rod with polyolefin insulation. Used in conjunction with hand- or foot-activated pencil connected to standard electrosurgical unit (ESU). Conducts high-frequency energy from ESU to target tissue for cutting or coagulation. Available in blade, needle, and ball configurations; 2.5" to 8" lengths. Used by surgeons in clinical/surgical settings. Single-use device.

Clinical Evidence

Bench testing only. Device conforms to ANSI/AAMI HF18-1993, clause 4.3.5.4 (Dielectric withstand of accessories).

Technological Characteristics

300 series stainless steel electrode; polyolefin insulation; standard 3/32" shaft compatibility; blade, needle, and ball configurations; single-use; conforms to ANSI/AAMI HF18-1993.

Indications for Use

Indicated for use in any surgical procedure requiring electrosurgical cutting and/or coagulation. For prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with the letters "MMP" in a bold, sans-serif font. The letters are arranged horizontally and enclosed within an oval shape. The logo appears to be a low-resolution image, with some pixelation and jagged edges visible. The overall impression is of a simple, recognizable brand mark. # K973346 NOV - 7 1997 ### 510(k) Summary September 4, 1997 #### Submitter's Name/Address: John W. Smith Director of Regulatory Affairs / Quality Assurance MegaDyne Medical Products, Inc. 11506 South State Street Draper, UT 84020 (801) 576-9669 (801) 576-9698 fax ### Contact Person: Same as above #### Device Name: Common Name: Trade Name: Classification (if known): Electrosurgical Electrode Not Yet Determined 21 CFR 878.4400, Electrosurgical cutting and coagulation device and accessories #### Predicate Device: Electrosurgical Blade, manufactured by National Wire and Stamping (K861495) ## Applicant Device Description: The MegaDyne uncoated electrosurgical electrode is very similar to other, commercially available, stainless steel electrodes. It is used in various electrosurgical procedures, in conjunction with a hand- or foot-activated "pencil", which is connected to a standard electrosurgical unit (ESU). The device consists of a formed stainless steel rod, insulated over most of its length by polyolefin insulation. The flat, or blade, end of the electrode conducts energy from a standard high-frequency electrosurgical generator to the target tissue. The round end fits into a standard electrosurgical handpiece. Various electrode configurations, such as blade, needle, and ball ends are available in assorted lengths from 2.5" to 8" overall. The blade, needle, and ball configurations are all identical to currently available uncoated stainless steel electrodes on the U.S. market. MegaDyne Uncoated Blade 510(k) Page 18 of 29 MEGADYNE MEDICAL PRODUCTS, INC. 11506 South State Street Draper, Utah 84020 Phone (801) 576-9669 FAX (801) 576-9698 Toll Free (800) 747-6110 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the letters "MMP" in a stylized, bold font. The letters are arranged horizontally within an oval shape. The logo has a distressed or textured appearance, giving it a vintage or worn look. The background is white, and the letters are black. # Applicant Device Intended Use: This device is intended to be used in any application which requires electrosurgical cutting or coagulation. ਸ ## Technological Characteristics: | Component/ Feature | MegaDyne's<br>Electrosurgical<br>Electrode | Predicate Device:<br>Electrosurgical Blade<br>(K861495) | |-------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------| | Intended use | to be used in any<br>application which<br>requires electrosurgical<br>cutting or coagulation | same | | Electrode Material | 300 series stainless steel | same | | Insulation Material | Polyolefin | unknown | | Configurations available | Blade, needle, and ball<br>end electrode | same | | Compatibility | Standard 3/32" shaft | same | | Single use | yes | same | | Conforms with ANSI / AAMI HF<br>18 - 1993 | yes | unknown | ## Performance Data: The Electrosurgical Electrode conforms with the relevant clause of ANSI / AAMI HF18 -1993: 4.3.5.4, Dielectric withstand of accessories. > MEGADYNE MEDICAL PRODUCTS, INC. 11506 South State Street Draper, Utah 84020 Phone (801) 576-9669 FAX (801) 576-9698 Toll Free (800) 747-6110 MegaDyne Uncoated Blade 510(k) Page 19 of 29 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans. The text "AN SERVICES USA" is also included in the logo. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. John W. Smith 'Director of Regulatory Affairs and Quality Assurance MegaDyne Medical Products, Inc. 11506 South State Street Draper, Utah 84020 NOV - 7 1997 Re: K973346 Trade Name: Electrosurgical Electrode Regulatory Class: II Product Code: GEI Dated: September 3, 1997 Received: September 5, 1997 Dear Mr. Smith: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. John W. Smith This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely vours. Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): KaJ3342 Device Name: Electrosurgical Electrode Indications for use: The MegaDyne uncoated electrode is indicated for use in any surgical procedure which requires electrosurgical cutting and/or coagulation. Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices | 510(k) Number | K973346 | |---------------|---------| |---------------|---------| | Prescription Use ✗<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use ______ | |--------------------------------------------|----|-----------------------------| |--------------------------------------------|----|-----------------------------| MegaDyne Uncoated Blade 510(k) Page 8 of 29
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...