ARTHROCARE ARTHROSCOPIC ELETROSURGERY SYSTEM (MODEL 980)

K963123 · Arthrocare Corp. · GEI · Oct 8, 1996 · General, Plastic Surgery

Device Facts

Record IDK963123
Device NameARTHROCARE ARTHROSCOPIC ELETROSURGERY SYSTEM (MODEL 980)
ApplicantArthrocare Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateOct 8, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ArthroCare Arthroscopic Electrosurgery System 980 is intended for use in arthroscopic surgical procedures of the knee, shoulder, ankle, elbow, hip and wrist for soft tissue resection and hemostasis.

Device Story

ArthroCare Arthroscopic Electrosurgery System 980; electrosurgical generator system for arthroscopic soft tissue resection and hemostasis. Components: Controller (generator), Probe, Probe Cable. Bipolar design with integral return electrode on probe shaft; eliminates need for patient return pad; energy localized to distal tip. Probe diameters 1.5-4.5 mm; straight or angled tips (up to 90 degrees); multi-electrode or loop configurations. Used in OR by orthopedic surgeons. Controller provides high-frequency energy; no software involved. System benefits: precise tissue resection and hemostasis without deep tissue energy penetration characteristic of monopolar devices.

Clinical Evidence

No clinical data provided; substantial equivalence established via design and technological comparison to predicate.

Technological Characteristics

Bipolar electrosurgical system; integral return electrode on probe shaft. Probe diameters 1.5-4.5 mm; straight or angled (up to 90 degrees) tips; multi-electrode or loop configurations. High-frequency electronic generator. Probe is sterile, single-use. Probe cable reusable via ethylene oxide or steam autoclaving. No software.

Indications for Use

Indicated for patients undergoing arthroscopic surgical procedures of the knee, shoulder, ankle, elbow, hip, or wrist requiring soft tissue resection and hemostasis.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0} OCT - 8 1996 ArthroCare Corporation Sunnyvale, California Premarket Notification August 9, 1996 # X. 510(k) Summary k963123 ## A. Name of Device Trade name: ArthroCare Arthroscopic Electrosurgery System 980 Common name: Electrosurgical Generator Classification name: Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) Accessory to an Arthroscope (21 CFR 888.1100 ## B. Predicate devices | Device | Premarket Notification | | --- | --- | | ArthroCare Arthroscopic Electrosurgery System 970 | K943450 | | ArthroCare Arthroscopic Electrosurgery System 970 (Addition of Hip Indication) | K961747 | | ArthroCare Arthroscopic Electrosurgery System 970 (Addition of Wrist Indication) | K962321 | ## C. Device description The ArthroCare Arthroscopic Electrosurgery System 980 is comprised of three components: the Probe, the Probe Cable, and an electrosurgical generator called the Controller. The Probe Cable connects the Controller to the Probe. The Probe is provided in a variety of models, ranging in diameters from 1.5 mm to 4.5 mm with the distal tip of the probe as straight or bent at angles up to 90 degrees. The distal tip may be configured with multi-electrodes or loop electrodes. The Probe is supplied sterile and intended for single patient use. The Probe Cable is designed for repeat sterilization by either ethylene oxide gas or steam autoclaving methods, as selected by the user. The Controller is a high frequency electronic instrument. There is no software utilized in the operation of the Controller. 023 {1} ArthroCare Corporation Sunnyvale, California Premarket Notification August 9, 1996 The ArthroCare Arthroscopic Electrosurgery System 980 is bipolar, incorporating a return electrode on the shaft of the device. This means that a return pad is not required for operation. The return energy in a bipolar device with an integral return electrode does not penetrate the tissue as in a monopolar device. In a monopolar device, the energy passes through the patient's body to reach the return pad. ## D. Intended use The ArthroCare Arthroscopic Electrosurgery System 980 is intended for use in arthroscopic surgical procedures of the knee, shoulder, ankle, elbow, hip and wrist for soft tissue resection and hemostasis. ## E. Technological characteristics The technological characteristics of the ArthroCare Arthroscopic Electrosurgery System 980 are the same as those of the ArthroCare Arthroscopic Electrosurgery System 970. The devices are substantially equivalent in terms of design, materials, principle of operation, product specifications and sterilization. ## F. Summary By virtue of design, materials, function and intended use, the ArthroCare Arthroscopic Electrosurgery System 980 is substantially equivalent to the ArthroCare Arthroscopic Electrosurgery System 970 which is currently marketed in the United States.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...