K962119 · Microline Pentax, Inc. · GEI · Aug 9, 1996 · General, Plastic Surgery
Device Facts
Record ID
K962119
Device Name
RE-NEW LAPAROSCOPIC INSTRUMENTS
Applicant
Microline Pentax, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 9, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The “Re-New” Laparoscopic Scissors are surgical instruments used to cut and dissect various abdominal tissue for use in endoscopic, including laparoscopic, surgical procedures.
Device Story
Manual detachable surgical instruments for endoscopic/laparoscopic procedures; system comprises reusable handles and interchangeable surgical tips (disposable or reusable). Tip attachment via proprietary double-thread mechanism. Tip styles include Metzenbaum, Mini, Hook, Endocut, and Micro. Used by surgeons in OR settings to cut/dissect abdominal tissue. Flushing port included for cleaning. Device provides mechanical cutting/dissecting capability; standard electrocautery compatibility.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Manual surgical instruments; 5mm diameter; 34cm/42cm length options. Reusable handle/shaft with proprietary double-thread tip attachment. Includes flushing port for cleaning. Compatible with standard electrocautery.
Indications for Use
Indicated for use in endoscopic, including laparoscopic, surgical procedures for cutting and dissecting abdominal tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K982287 — ASLAN SENSOR MODULAR INSTRUMENT · Aslan Medical Technologies, Ltd. · Aug 6, 1998
Submission Summary (Full Text)
{0}
AUG - 9 1996
K 96 2119
10. **SMDA Summary of Safety and Effectiveness - "510(k) Summary"**
A. **Submitter Information**
Microline, Inc.
181 Elliott Street
Beverly, MA 01915
Telephone: (508) 922-9810
Contact Person: Mr. Huges de Laforcade
President
Date Prepared: May 24, 1996
B. **Device Identification**
Common/Usual Name: Manual Detachable Surgical Instruments
Proprietary Name: “Re-New” Laparoscopic Instruments
C. **Identification of Predicate Device(s)**
The “Re-New” Laparoscopic Instruments are substantially equivalent to their predicate, “Quick Switch” (K932648, previously cleared and currently marketed.
D. **Device Description**
The “Re-New” Laparoscopic Scissors are surgical instruments used to cut and dissect various abdominal tissue for use in endoscopic, including laparoscopic, surgical procedures. The product line includes reusable common handles that can be easily attached and detached from a series of surgical tips available, which include:
- disposable tips, and
- reusable tips.
The tip assembly is designed with a proprietary double thread, which permits simple and reliable attachment and detachment of the desired tip to the shaft of the handle/shaft component. Tip configurations include:
- Metzenbaum,
- Mini,
- Hook,
- Endocut, and,
- Micro.
{1}
E. Substantial Equivalence
The technical characteristics of the "Re-New" Laparoscopic Instruments are almost identical to those of the Microline "Quick Switch". Refer to Table A for a comparison with their predicate. Differences that exist between these devices relating to technical specifications, materials, and physical appearance do not affect the relative safety or effectiveness of the "Re-New" relative to its predicate.
The "Quick Switch" and "Re-New" are intended for use in endoscopic, including laparoscopic surgical procedures where instruments are introduced into the body through a cannula.
000013
{2}
Table A
Comparison Table: Manual Detachable Surgical Instruments
| Manufacturer: | Microline, Inc. | Microline, Inc. |
| --- | --- | --- |
| Model | “Quick-Switch” | “Re-New” |
| K-Number | K932648 | |
| Class | II | Request II |
| Handle/Shaft: | | |
| Length (cm) | | |
| Disposable | 32 | 34,42 |
| Reusable | 32 | 34,42 |
| Diameter (mm) | 5 | 5 |
| Reusable Handles | Yes | Yes |
| Flushing Port for Cleaning | Yes | Yes |
| Tips: | | |
| Tip Length Range (cm) | | |
| Disposable | 3.5 | 4.0 - 4.5 |
| Reusable | 3.5 | 4.1 - 4.5 |
| Styles | Scissors | Scissors |
| Electrocautery | Standard | Standard |
| Disposable Assembly | Tips | Tips |
| Clinicals: | | |
| Clinical Claims | Endoscopic (inclusive of laparoscopic) Surgical Procedures | Endoscopic (inclusive of laparoscopic) Surgical Procedures |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.