← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K960255

# CERAMIC BLADE ELECTRODE (K960255)

_Utah Medical Products, Inc. · GEI · Mar 27, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K960255

## Device Facts

- **Applicant:** Utah Medical Products, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 27, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Utah Medical Products, Inc. (UMP) Ceramic Blade Electrodes (CBE) are intended for use in virtually every surgical discipline where flat, paddle-type electrosurgical blades are used for making straight cuts through tissue. The CBE-2XX blades are intended for use in procedures that require precise cuts where only light or no hemostasis is desired. The CBE-1XX blades are intended for use in procedures where a greater level of hemostasis is desired. When an electrosurgical generator blend mode is used for cutting with the CBE-1XX blades, cauterization of the wound will occur while the cut is being made.

## Device Story

Ceramic blade electrode; paddle-type electrosurgical accessory. Input: electrical energy from standard electrosurgical generator via electrosurgical pen. Operation: blade conducts energy to tissue to perform cutting and cauterization. Output: physical tissue incision and wound cauterization. Used in surgical settings by clinicians. Benefits: precise cutting; controlled hemostasis; minimizes bleeding.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Ceramic blade electrode; paddle-type form factor. Compatible with standard electrosurgical pens and generators. Sterilization via gamma radiation. Materials selected for biocompatibility, electrical/thermal properties, and strength.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- UMP DLP-L11 Loop Electrode ([K905263](/device/K905263.md))
- Mega Dyne E-Z Clean Blade ([K913473](/device/K913473.md))
- Aspen Labs Blade Electrode #60-0180-001 ([K780977](/device/K780977.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

UTAH MEDICAL PRODUCTS INC.
K 960255
SMDA 510(k) Summary
MAR 27 1996

Trade Name: Ceramic Blade Electrode
Common Name: Electrosurgical Blade
Classification: Electrosurgical Cutting and Coagulation Device and Accessories, 21 CFR 878.4400.

The Utah Medical Products, Inc. (UMP) Ceramic Blade Electrodes (CBE) are intended for use in virtually every surgical discipline where flat, paddle-type electrosurgical blades are used for making straight cuts through tissue. The CBE-2XX blades are intended for use in procedures that require precise cuts where only light or no hemostasis is desired. The CBE-1XX blades are intended for use in procedures where a greater level of hemostasis is desired. When an electrosurgical generator blend mode is used for cutting with the CBE-1XX blades, cauterization of the wound will occur while the cut is being made.

The CBE device is substantially equivalent to the UMP DLP-L11 Loop Electrode (K905263), the Mega Dyne E-Z Clean Blade (K913473), and the Aspen Labs Blade Electrode #60-0180-001 (K780977). The posterior end of the CBE device fits snugly into a standard electrosurgical pen which is connected to an FDA-approved electrosurgical generator. The CBE is used in exactly the same way as the Aspen stainless steel blade and Mega Dyne Teflon®-coated blade electrodes are used for cutting tissues and cauterizing the wound to minimize bleeding. The cutting and cauterization takes place in exactly the same way as with the UMP DLP-L11 Loop Electrode.

Materials that comprise the CBE were selected based on properties of strength, biocompatibility, compatibility with gamma sterilization, conformance with commonly used medical grade plastics and other materials, electrical and thermal properties, ease of processing and low cost. Biocompatibility and sterilization information is available from UMP upon request.

Kevin L. Cornwell
President &amp; CEO

-37-
UTAH MEDICAL PRODUCTS, INC • 7043 South 300 West • Midvale, Utah 84047-1048 • 801/566-1200 • FAX#801/566-2062

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K960255](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K960255)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
