NEW DOUBLO 2.0

K251334 · Hironic Co., Ltd. · GEI · Sep 29, 2025 · General, Plastic Surgery

Device Facts

Record IDK251334
Device NameNEW DOUBLO 2.0
ApplicantHironic Co., Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateSep 29, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NEW DOUBLO 2.0 with RM and RM(B) handpiece is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. The NEW DOUBLO 2.0 with PS handpiece is intended for the removal and destruction of skin lesions and coagulation of tissue.

Device Story

Electrosurgical system for dermatologic/general surgical procedures; utilizes high-frequency current to generate electric fields within tissue, causing heating via impedance. Consists of main unit, LCD touchscreen, foot switch, and three handpieces (RM, RM(B), PS). RM/RM(B) handpieces deliver bipolar RF energy via micro-needle tips for coagulation/hemostasis; PS handpiece delivers monopolar RF (plasma) for lesion removal/destruction. Operated by trained healthcare professionals in clinical settings. Energy output controlled by foot/hand switches; provider monitors via touchscreen. Benefits include controlled, non-invasive tissue coagulation and lesion destruction.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including electrical safety, electromagnetic compatibility, biocompatibility (ISO 10993), and ex vivo thermal/histological analysis of tissue coagulation.

Technological Characteristics

Electrosurgical system; RF energy source. RM/RM(B) handpieces: Bipolar, 0.5-3.5mm needle depth, 19.88W max power. PS handpiece: Monopolar, 41 kHz frequency. Connectivity: LCD touchscreen interface. Sterilization: EO Gas for needle tips. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 62304, IEC 62366-1, ISO 14971, ISO 10993.

Indications for Use

Indicated for dermatologic and general surgical procedures requiring electro-coagulation, hemostasis, and the removal/destruction of skin lesions.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 29, 2025 Hironic Co., Ltd. Lee Gwijin Official Correspondent 19F, U-TOWER, 767, Sinsu-ro, Suji-gu Yongin-si, Gyeonggi-do 16827 Korea, South Re: K251334 Trade/Device Name: New Doublo 2.0 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: April 30, 2025 Received: April 30, 2025 Dear Lee Gwijin: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251334 - Lee Gwijin Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251334 - Lee Gwijin Page 3 Sincerely, Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2025.09.29 20:37:53 -04'00' Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} NEW DOUBLO 2.0 Page 8 of 49 | Indications for Use | | | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251334 | | Please provide the device trade name(s). | | | NEW DOUBLO 2.0 | | | Please provide your Indications for Use below. | | | The NEW DOUBLO 2.0 with RM and RM(B) handpiece is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. The NEW DOUBLO 2.0 with PS handpiece is intended for the removal and destruction of skin lesions and coagulation of tissue. | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | {4} K251334 HIRONIC # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92 | 807.92(a)(1) – Submitter Information | | | --- | --- | | Applicant | Hironic Co., Ltd. | | Address | 19F, U-TOWER, 767, Sinsu-ro, Suji-gu, Yongin-si, Gyeonggi-do, 16827, Republic of Korea | | Phone Number | +82-31-525-7000 | | Contact Person | Gwijin Lee | | Contact Information | 19F, U-TOWER, 767, Sinsu-ro, Suji-gu, Yongin-si, Gyeonggi-do, 16827, Republic of Korea t. +82-31-525-7000, e. ra@hironic.com | | Date Prepared | 24 September 2025 | | 807.92(a)(2) – Device Name and Code | | | --- | --- | | Trade/Device Name | NEW DOUBLO 2.0 | | Common Name | Electrosurgical System | | Classification Name | Electrosurgical cutting and coagulation device and accessories | | Product Code | GEI | | Regulation Number | 21 CFR 878.4400 | | Regulatory Class | II | | Review Panel | General & Plastic Surgery (ODE) | | 807.92(a)(3) – Legally marketed device(s) to which equivalence is claimed | | | --- | --- | | Primary Predicate Device | K233123 SILKRO | | Predicate/Reference Device | K241099 PLASONIC | | 807.92(a)(4) – Device Description | | --- | | This device is an instrument used for tissue coagulation by means of high-frequency current. It consists of the main unit, three handpieces, foot switch, LCD touchscreen and power cable. The NEW DOUBLO 2.0 has three handpieces: The device is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis with RM and RM(B) handpieces and the removal and destruction of skin lesions and coagulation of tissue with PS handpiece. The device is intended for use by trained healthcare professionals in a clinical setting to provide controlled radiofrequency therapy through non-invasive procedure. RF uses electromagnetic wave frequencies to generate an electric field within tissue to cause heating. The amount of energy delivered to tissue is proportional to the current, time, and resistance within the tissue, also known as impedance. Tissue impedance is the main variable in the amount of heat delivered by the generated electric current. | {5} K251334 HIRONIC High frequency amplification is performed and it is transmitted to the handpiece. The corresponding output is controlled through the Foot Switch and Hand Switches, and when the Foot Switch and Hand Switches are pressed, each handpiece irradiates the transmitted high-frequency energy to the skin. # 807.92(a)(5) - Indication for Use The NEW DOUBLO 2.0 with RM and RM(B) handpiece is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. The NEW DOUBLO 2.0 with PS handpiece is intended for the removal and destruction of skin lesions and coagulation of tissue. # 807.92(a)(6) - Summary of the Technological Characteristics of the Device Compared to the Predicate 1. Primary Predicate Device | | Proposed Device | Primary predicate Device | Comparison Comment | | --- | --- | --- | --- | | 510(k) Number | K251334 | K233123 | | | Manufacturer | Hironic Co., Ltd. | Hironic Co., Ltd. | | | Trade/Device Name | NEW DOUBLO 2.0 | SILKRO | | | Classification Name | Electrosurgical cutting and coagulation device and accessories | Electrosurgical cutting and coagulation device and accessories | | | Indication for use | Intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. | Intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. | Identical | | Output energy type | Radio Frequency | Radio Frequency | Identical | | User interface | Color Touch Panel | Color Touch Panel | Identical | | Operating Frequency | 0.5, 1, 2 Wz | 2 Wz | Similar | | Rated Input | 100 - 240 VAC, 50/60 Hz | 100 - 240 VAC, 50/60 Hz | Identical | | Max. Power | RM, RM(B) Handpiece: 19.88 W ± 20% | RM Handpiece: 20.520 W ± 20% | Similar | | RF Duration | RM, RM(B) Handpiece: 50 - 950 ms | RM Handpiece: 50 - 950 ms | Identical | | Needle Insert Depth | RM, RM(B) Handpiece: 0.5 - 3.5 mm | RM Handpiece: 0.5 - 3.5 mm | Identical | | Repetition | RM, RM(B) Handpiece: 0.2 - 0.8, 1.0, 2.0 sec, Single | RM Handpiece: 0.2, 0.5, 0.8, 1.0, 2.0 sec, Single | Similar | | Mode of Operation | RM/RM(B) Handpiece: Bipolar Type | RM Handpiece: Bipolar Type | Identical | {6} K251334 HIRONIC | Electrode(Needle) Type | RF Micro Needle Tip for RM, RM(B) Handpiece: 25 pin, 49 pin | RF Micro Needle Tip for RM Handpiece: 25 pin, 49 pin | Identical | | --- | --- | --- | --- | | Single Use | RF Micro Needle Tip for RM, RM(B) Handpiece | RF Micro Needle Tip for RM Handpiece | Identical | | Sterilization | EO Gas | EO Gas | Identical | 2. Predicate/Reference Device | | Proposed Device | Predicate/Reference Device | Comparison Comment | | --- | --- | --- | --- | | 510(k) Number | K251334 | K241099 | | | Manufacturer | Hironic Co., Ltd. | Hironic Co., Ltd. | | | Trade/Device Name | NEW DOUBLO 2.0 | PLASONIC | | | Classification Name | Electrosurgical cutting and coagulation device and accessories | Electrosurgical cutting and coagulation device and accessories | | | Indication for use | Used in the removal and destruction of skin lesions and coagulation of tissue. | Used in the removal and destruction of skin lesions and coagulation of tissue. | Identical | | Mode of Operation | Radio Frequency (Plasma) | Radio Frequency (Plasma) | Identical | | Output | Monopolar | Monopolar | Identical | | Power Supply | 100 – 240 VAC, 50/60 Hz | 100 – 240 VAC, 50/60 Hz | Identical | | Frequency | 41 kHz | 41 kHz | Identical | 807.92(b)(1) – Non-clinical tests submitted - IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance - IEC 60601-2-2:2017, IEC 60601-2-2:2017/AMD1:2023 Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories - IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests - IEC 60601-1-6:2010, AMD1:2013, AMD2:2020 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability - IEC 62304:2006 Medical device software – Software life cycle processes is evaluated according to the FDA-recognized consensus standard - IEC 62366-1:2015 Medical device – Part 1: Application of usability engineering to medical devices is evaluated in accordance with the FDA-recognized consensus standard - ISO 14971:2019 Medical devices – Application of risk management to medical devices {7} K251334 HIRONIC - ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity - ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization - ISO 10993-11:2017 Biological evaluation of medical devices – Part 11: Tests for systemic toxicity - In the ex vivo model, the handpieces were applied in the same way as applied to the actual human body and the temperature change of the application site was analyzed through a thermal imaging camera. In addition, the length, width and depth of tissue coagulation at the application site were analyzed through photography and histological evaluation. | 807.92(b)(2) – Clinical tests submitted | | --- | | No clinical test was performed. | ## Conclusions drawn from clinical and non-clinical tests submitted The NEW DOUBLO 2.0 is substantially equivalent to its primary predicate device and predicate/reference device with same indication for use and technological characteristics. The non-clinical data for proposed device, including biocompatibility, bench testing, hardware and software documentation shows that the device should perform as intended in the specified use. In addition, the Electromagnetic Compatibility and Electrical Safety test shows that the device is safe to use and meets required standards. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that we conclude that substantially equivalent with predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...