← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K242044

# FiAPC plus probes (K242044)

_Erbe Elektromedizin GmbH · GEI · Aug 26, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K242044

## Device Facts

- **Applicant:** Erbe Elektromedizin GmbH
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Aug 26, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The flexible FiAPC probes are intended for argon plasma coagulation and ablation of tissue.

## Device Story

FiAPC plus probes are sterile, single-use flexible electrosurgical instruments for argon plasma coagulation (APC) and tissue ablation. Probes connect to compatible Erbe electrosurgical units (ESU) via a proximal plug containing an integrated filter and instrument recognition system. The device consists of a long tube for insertion through an endoscope working channel, terminating in a distal opening for argon plasma delivery. Available in axial (straight fire) and lateral (side fire) configurations. Operated by physicians in clinical settings; the device delivers monopolar high-frequency current via argon gas to achieve tissue coagulation or ablation. The integrated filter isolates the device from the patient side, and instrument recognition ensures compatibility with parent ESU systems. Clinical benefit includes precise thermal tissue treatment during endoscopic procedures.

## Clinical Evidence

Bench testing only. Performance validated via functional testing, tissue effect testing (comparing thermal effects and ignition distances to predicate), EMC/electrical safety (IEC 60601-1, 60601-2-2, 60601-2-18, 60601-1-2), biocompatibility (ISO 10993-1), sterilization (ISO 11135, SAL 10^-6), and shelf-life/packaging (ISO 11607-1, ASTM F 1980).

## Technological Characteristics

Materials: stainless steel, plastics, ceramics, silver. Energy: Monopolar high-frequency current (argon plasma). Max electrical load: 5000 Vp. Max gas flow: 2.4 l/min. Connectivity: Proprietary plug with instrument recognition for Erbe ESU compatibility. Sterilization: Ethylene Oxide. Form factor: Flexible tube for endoscopic use, axial or lateral distal openings.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ERBE APC INTEGRATED FILTER PROBES ([K060183](/device/K060183.md))
- ERBE APC INTEGRATED FILTER PROBE ADAPTER ([K060183](/device/K060183.md))

## Reference Devices

- APC 3 ([K191234](/device/K191234.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

August 26, 2024

Erbe Elektromedizin GmbH Matthias Kollek Regulatory Affairs Specialist Waldhoernlestrasse 17 Tuebingen, 72072 Germany

Re: K242044

Trade/Device Name: FiAPC plus probes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 12, 2024 Received: July 12, 2024

Dear Matthias Kollek:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long H. Chen -SDDate: 2024.08.26 10:37:08 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K242044

Device Name

FiAPC plus probes

Indications for Use (Describe)

The flexible FiAPC probes are intended for argon plasma coagulation and ablation of tissue.

Type of Use (Select one or both, as applicable)

| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

| Applicant                          | Erbe Elektromedizin GmbH                                       |
|------------------------------------|----------------------------------------------------------------|
|                                    | Waldhoernlestrasse 17                                          |
|                                    | 72072 Tuebingen                                                |
|                                    | Germany                                                        |
|                                    | Tel: 0049-7071-755-0                                           |
|                                    | Fax: 0049-7071-755-179                                         |
| Contact Person                     | Dr. Matthias Kollek                                            |
|                                    | Regulatory Affairs Specialist                                  |
|                                    | E-Mail: Matthias.Kollek@erbe-med.com                           |
| Date Prepared                      | July 12, 2024                                                  |
| Device Information                 |                                                                |
| Trade/Proprietary Name:            | FiAPC® plus probes                                             |
| Common Name:                       | Argon-enhanced endoscopic electrosurgical electrode            |
| Classification Name                | Electrosurgical cutting and coagulation device and accessories |
| Regulation Number:                 | 21 CFR 878.4400                                                |
| Class:                             | II                                                             |
| Product Code:                      | GEI                                                            |
| Legally Marketed Predicate Devices | ERBE APC INTEGRATED FILTER PROBES;                             |
|                                    | ERBE APC INTEGRATED FILTER PROBE                               |
|                                    | ADAPTER - K060183                                              |

{4}------------------------------------------------

### Device Description

The FiAPC® plus probes are sterile, single-use flexible probes intended for argon plasma coagulation (APC) and ablation of tissue. The probes have a plug on the proximal end that enables the connection to the compatible electrosurgical unit (ESU) and APC unit. The plug is ergonomically designed and contains a filter that isolates the device side from the patient side. The plug also contains an instrument recognition which ensures that the instrument can only be operated with the compatible parent devices. The connector is followed by a long tube so that the probe can be inserted into the working channel of an endoscope. At the distal end there is an opening through which the argon plasma is released. The flexible FiAPC® plus probes are available in different variants. All variants have the same outer diameter but differ in their length and distal opening, i.e. axial opening (straight fire) or lateral opening (side fire).

#### Indications for Use

The flexible FiAPC probes are intended for argon plasma coagulation and ablation of tissue.

The intended use of the subject device and legally marketed predicate device is the same. Compared to the predicate device, the indications for use statement of the subject device additionally includes the term ablation of tissue for clarification in accordance with the parent device APC 3 (K191234) which was cleared for "argon plasma coagulation and ablation of tissue" ... "when used in conjunction with a compatible Erbe VIO Electrosurgical Generator (ESU) and applicators or probes"

|                        | Subject Device                                                                                       | Predicate Device                                                                                                                                                                                       | Comments                |
|------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Characteristics        | FiAPC® plus probes                                                                                   | Argon Plasma Coagulation (APC)<br>Integrated Filter Probes<br>and Adapter<br>K060183                                                                                                                   |                         |
| Manufacturer           | Erbe Elektromedizin GmbH<br>(Germany)                                                                | Erbe Elektromedizin GmbH<br>(Germany)                                                                                                                                                                  | Same                    |
| Regulation<br>number   | 878.4400                                                                                             | 878.4400                                                                                                                                                                                               | Same                    |
| Regulatory<br>class    | II                                                                                                   | II                                                                                                                                                                                                     | Same                    |
| Product code           | GEI                                                                                                  | GEI                                                                                                                                                                                                    | Same                    |
| Indications for<br>use | The flexible FiAPC probes are<br>intended for argon plasma<br>coagulation and ablation of<br>tissue. | The APC Integrated Filter Probes<br>are intended for use in Argon<br>Plasma Coagulation<br>(APC). The devices are used to<br>treat many conditions in<br>endoscopy for various surgical<br>procedures. | Same<br>Intended<br>Use |
| Prescription or<br>OTC | Prescription                                                                                         | Prescription                                                                                                                                                                                           | Same                    |
| Compatibility          | Erbe APC 2 with VIO D or S<br>series                                                                 | Erbe APC 300 with ICC series<br>Erbe APC 2 with VIO series                                                                                                                                             | Similar                 |

## Comparison of Technological Characteristics

{5}------------------------------------------------

### Erbe Elektromedizin GmbH Traditional 510(k) for FiAPC® plus probes

|                                   | Erbe APC 3 with VIO 3                                                                 | Erbe APC 3 with VIO 3                              |                                |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
|-----------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------|--------------------------------|--|--|--|--|--|--|--|--|--|--|----------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-----------------------------------------------------------|
| Materials                         | Stainless steel, plastics, ceramics, silver                                           | Stainless steel, plastics, tungsten, ceramics      | Similar                        |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Beam Type                         | Axial and lateral (sidefire)                                                          | Axial, lateral (sidefire) and circumferential      | Similar                        |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Probe Dimensions                  | Beam form Diameter [mm] Length [m]   Axial 2.3 2.2   Axial 2.3 3.0   Sidefire 2.3 2.2 |                                                    |                                |  |  |  |  |  |  |  |  |  |  | Beam form Diameter [mm] Length [m]   Axial 1.5 1.5   Axial 1.5 3.0   Axial 2.3 2.2   Axial 2.3 3.0   Axial 3.2 2.2   Sidefire 2.3 2.2   Circumf. 2.3 2.2 |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | Similar; optimized design of the distal end (rounded tip) |
| Plug                              | Filter integrated plug with instrument recognition                                    | Filter integrated plug with instrument recognition | Similar; optimized plug design |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Max. electrical load capacity     | 5000 Vp                                                                               | 5000 Vp                                            | Same                           |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Max. gas flow rate                | 2.4 l/min                                                                             | 2.4 l/min                                          | Same                           |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Energy delivery                   | Argon plasma (Monopolar high frequency current)                                       | Argon plasma (Monopolar high frequency current)    | Same                           |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Condition Provided/ Use Condition | Sterile, single-use                                                                   | Sterile, single-use                                | Same                           |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |
| Sterilization Method              | Ethylene Oxide                                                                        | Ethylene Oxide                                     | Same                           |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                          |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                           |

The specifications of the subject device are largely identical compared to the predicate device. Both devices are sterile, single-use devices which apply the same kind of energy, offer the same beam types, have the same probe dimensions, a filter integrated plug and the same max, electrical load capacity,

Minor differences are related to the pluq design of the distal tip and the used materials.

To exclude a negative impact on safety and effectiveness and to show substantial equivalence, side-by-side tissue testing was conducted in compliance with FDA Guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued on March 9, 2020. Compared to the predicate device, the subject device showed comparable tissue effects and comparable ignition distances.

As shown by bench testing, all design and performance requirements were successfully verified/validated, and all acceptance criteria are met. Biocompatibility testing in compliance with FDA recognized consensus standard ISO 10993-1 shows that the new materials are biocompatible. The provided testing demonstrates that the subject device is as safe and effective as the legally marketed predicate devices and does not raise new or different questions of safety and effectiveness.

{6}------------------------------------------------

#### Summary of non-clinical bench performance testing

Verification/validation activities from non-clinical testing as described below demonstrate that the differences do not raise any new issues of safety or effectiveness of the subject device compared to the predicate device.

Functional testing and design controls to verify both safety and performance of the subject device was performed in compliance with 21 CFR 820.30 to ensure that the subject device performs as intended and meets design specifications.

Tissue testing was performed in compliance with FDA Guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued on March 9, 2020, to validate the thermal effect on tissue compared to the predicate device.

EMC and Electrical safety of the subject device was tested in compliance with FDA recognized versions of IEC 60601-1; IEC 60601-2-2; IEC 60601-2-18: and IEC 60601-1-2, as applicable. Biocompatibility testing was performed in compliance with ISO 10993-1 and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" to demonstrate biocompatibility of the used materials.

Sterilization validation was performed in compliance with ISO 11135 and documentation was provided according to FDA Guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" showing an SAL of 10-6. EO residual testing and limits are in compliance with ISO 10993-7.

Packaging and shelf-life validation was performed in compliance with ISO 11607-1 and accelerated aged devices (ASTM F 1980).

#### Conclusion

The subject device has the same intended use, the same fundamental design, substantially equivalent performance characteristics, and the same energy source as the predicate device. The subject device was tested as described above and the minor differences in technological characteristics were assessed with regards to safety and effectiveness. Taken together, the subject device does not raise new or different questions of safety and effectiveness, and the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K242044](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K242044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
