Valleylab ™FT10 Energy Platform Software Version 5.0 (VLFT10GEN VLFT10FXGEN VLFT10LSGEN)
K241085 · Covidien, LLC · GEI · Jul 12, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241085
Device Name
Valleylab ™FT10 Energy Platform Software Version 5.0 (VLFT10GEN VLFT10FXGEN VLFT10LSGEN)
Applicant
Covidien, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm thick tissue (tissue bundles), and lymphatics. When used with compatible ablation devices it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium. The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. The VLFT10FXGEN is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. The VLFT10LSGEN is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics.
Device Story
High-frequency electrosurgical generator; provides RF energy for monopolar/bipolar surgery, tissue fusion, vessel sealing, and cardiac ablation. Used in hospitals/healthcare facilities by surgeons. Inputs: monopolar/bipolar instruments, footswitches, return electrode pads. Device processes electrical signals to deliver controlled power for cutting, desiccating, and fulgurating tissue. Output: RF energy delivered to surgical site via compatible accessories. Healthcare providers use device interface to select modes; output affects tissue state (cutting/coagulation/sealing). Benefits: enables precise surgical tissue management and vessel hemostasis.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including system verification, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software verification/validation (IEC 62304), and mechanical/environmental testing.
Indicated for patients requiring electrosurgical cutting, coagulation, vessel sealing (up to 7mm tissue bundles/lymphatics), or cardiac tissue ablation. Contraindicated for patients with active endocarditis or for ablation in a pool of blood. Not indicated for tubal sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Valleylab FT10 Energy Platform with Software Version 4.0 (K191601)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. To the left of the FDA logo is the Department of Health & Human Services logo.
July 12, 2024
Covidien LLC Aliva Mahunta Sr. Regulatory Affairs Specialist 200 Medtronic Drive Lafayette, Colorado 80026
# Re: K241085
Trade/Device Name: Valleylab ™FT10 Energy Platform Software Version 5.0 (VLFT10GEN VLFT10FXGEN VLFT10LSGEN) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, OCL Dated: May 23, 2024 Received: May 24, 2024
# Dear Aliva Mahunta:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen -> Digitally signed by Long H.Chen -S
Date: 2024.07.12 16:00:47 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241085
Device Name
Valleylab ™FT10 Energy Platform Software Version 5.0 (VLFT10GEN VLFT10FXGEN VLFT10LSGEN)
Indications for Use (Describe)
VLFT10GEN
The Valleylab FT10 energy platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7mm thick tissue (tissue bundles), and lymphatics. When used with compatible ablation devices it is indicated for cardiac tissue ablation.
The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures.
Valleylab ™FT10 Electrosurgical Generator (VLFT10FXGEN)
The VLFT10FXGEN is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
Valleylab ™FT10 Vessel Sealing Generator (VLFT10LSGEN)
The VLFT10LSGEN is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K241085
# 510(k) Summary Complying with 21 CFR 807.92
#### l. SUBMITTER
Covidien IIc. 200 Medtronic Drive Lafayette, CO, 80026
Contact: Aliva Mahunta, Sr. Regulatory Affairs Specialist Phone Number : 800-255-6774 Email: aliva.mahunta@medtronic.com Date Summary Prepared: May 20, 2024
#### ll. PROPOSED DEVICE
| Trade Name: | Valleylab FT10 Energy Platform with Software Version 5.0 |
|----------------------|----------------------------------------------------------------------------------|
| Catalog Numbers: | VLFT10GEN; VLFT10FXGEN; VLFT10LSGEN |
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical cutting and coagulating device and accessories (21 CFR 878.4400) |
| Regulatory Class: | II |
| Product Code: | GEI, OCL |
### PREDICATE DEVICE
| Trade Name: | Valleylab FT10 Energy Platform with Software Version 4.0 |
|----------------------|----------------------------------------------------------------------------------|
| Catalog Numbers: | VLFT10GEN |
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical cutting and coagulating device and accessories (21 CFR 878.4400) |
| 510(k) Number: | K191601 |
| Regulatory Class: | II |
| Product Code: | GEI, OCL |
#### lll. DEVICE DESCRIPTION
Valleylab™ FT10 Energy Platform (VLFT10GEN): The Valleylab™ FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. The generator provides radio frequency (RF) energy for monopolar and bipolar surgical applications, tissue-fusion, and vessel-sealing applications (LigaSure/vessel sealing function) and cardiac applications. It is a combination of a full-featured general surgery electrosurgical unit and a bipolar vessel sealing system. The monopolar and bipolar sections, including the LigaSure/Bipolar section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery.
{4}------------------------------------------------
### K241085
The VLFT10GEN is used in hospitals and other health care facilities where surgical procedures are carried out.
The VLFT10GEN can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The VLFT10GEN connects to electrical mains and operates at an input line frequency of 47-63 Hz.
Valleylab™ FT10 Electrosurgical Generator (VLFT10FXGEN): The Valleylab™ FT10 Electrosurgical Generator is a high frequency electrosurgical only version of the VLFT10GEN Energy Platform. It is intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
Valleylab™ FT10 Vessel Sealing Generator (VLFT10LSGEN): The Valleylab™ FT10 Vessel Sealing Generator is the vessel sealing only version of the VLFT10GEN Energy Platform. It is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics.
#### IV. INDICATIONS FOR USE
### Valleylab™ FT10 Energy Platform (VLFT10GEN):
The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm thick tissue (tissue bundles), and lymphatics. When used with compatible ablation devices, it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.
### Valleylab™ FT10 Electrosurgical Generator (VLFT10FXGEN):
The VLFT10FXGEN is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
### Valleylab™ FT10 Vessel Sealing Generator (VLFT10LSGEN):
The VLFT10LSGEN is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, thick tissue (tissue bundles), and lymphatics.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.
## V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Valleylab FT10 Energy Platform with the proposed software modifications has the same technological and performance characteristics as the predicate Valleylab FT10 Energy Platform with software version 4.0 cleared in K191601. Both versions of the energy platform are a combination of a full-featured general surgery electrosurgical system, a vessel sealing system, and a cardiac ablation system. The proposed device VLFT10GEN with software version 5.0 (VLFT10GEN SW v5.0) provides
{5}------------------------------------------------
incremental improvements over the predicate while maintaining the same basic functionality and intended use.
| Characteristic | Proposed Device<br>Valleylab™ FT10 Energy<br>Platform with Software<br>version 5.0 | Predicate Device<br>Valleylab™ FT10 Energy<br>Platform with Software<br>version 4.0 (K191601) | Comparison |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 878.4400 | 878.4400 | Same |
| Regulation | | | |
| Class | II | II | Same |
| Product Code | GEI | GEI | Same |
| Indications for Use | The Valleylab FT10 Energy<br>Platform is a high frequency<br>electrosurgical generator<br>intended for use with<br>monopolar and bipolar<br>accessories for cutting and<br>coagulating tissue. When used<br>with compatible sealing devices,<br>it is indicated for sealing vessels<br>up to and including 7 mm thick<br>tissue (tissue bundles), and<br>lymphatics. When used with<br>compatible ablation devices it is<br>indicated for cardiac tissue<br>ablation. The generator can be<br>used with compatible<br>resectoscopes for<br>endoscopically controlled<br>removal or coagulation of tissue<br>using 0.9% NaCl solution as the<br>irrigation medium.<br><br>The tissue fusion function has<br>not been shown to be effective<br>for tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use this<br>function for these procedures.<br><br>VLFT10FXGEN Indications for<br>Use<br><br>The VLFT10FXGEN Energy<br>Platform is a high frequency<br>electrosurgical generator<br>intended for use with<br>monopolar and bipolar<br>accessories for cutting and<br>coagulating tissue. | The Valleylab FT10 Energy<br>Platform is a high frequency<br>electrosurgical<br>Generator intended for use<br>with monopolar and bipolar<br>accessories for cutting and<br>coagulating tissue. When<br>used with compatible<br>sealing devices, it is<br>indicated for sealing vessels<br>up to and including 7 mm,<br>tissue bundles, and<br>lymphatics. When used with<br>compatible ablation<br>devices, it is indicated for<br>cardiac tissue ablation. The<br>generator can also be used<br>with compatible<br>resectoscopes for<br>endoscopically controlled<br>removal or coagulation of<br>tissue using 0.9% NaCl<br>solution as the irrigation<br>medium.<br><br>The tissue fusion function<br>has not been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use this<br>function for these<br>procedures | Same<br>There is no change to<br>the intended use or<br>indications for use of the<br>device. The indications<br>for use for the separate<br>VLFT10FXGEN and<br>VLFT10LSGEN functions<br>are simply a subset of<br>the fully functional<br>VLFT10GEN indications<br>for use, which are not<br>changing.<br><br>The intended use was<br>revised to add thick<br>tissue (tissue bundles)<br>while keeping tissue<br>bundles.<br>This revision is for<br>clarification purposes<br>only for the benefit of<br>the surgeon.<br>There is no change to<br>the intended use of the<br>generator, but rather a<br>clarification of terms for<br>the benefit of the<br>surgeon. |
| | VLFT10LSGEN Indications for<br>Use<br>The VLFT10LSGEN Energy<br>Platform is a high frequency<br>electrosurgical generator. When<br>used with compatible sealing<br>devices, it is indicated for<br>sealing vessels up to and<br>including 7 mm tissue bundles,<br>and lymphatics. | | |
| Contraindications | The cardiac tissue ablation<br>feature is contraindicated for<br>patients that have active<br>endocarditis at the time of<br>surgery.<br>Ablation in a pool of blood is<br>contraindicated (for example<br>through a purse string suture). | The cardiac tissue ablation<br>feature is contraindicated<br>for patients that have active<br>endocarditis at the time of<br>surgery.<br>Ablation in a pool of blood<br>is contraindicated (for<br>example through a purse<br>string suture). | Same |
| Physical Characteristics<br>(approximate | Height: 7 in. (17.78 cm)<br>Width: 14.5 in. (35.8 cm)<br>Length: 18.2 in. (46.2 m)<br>Weight: 22.3 lbs. (10.1 kg) | Height: 7 in. (17.78 cm)<br>Width: 14.5 in. (35.8 cm)<br>Length: 18.2 in. (46.2 m)<br>Weight: 22.3 lbs. (10.1 kg) | Same |
| Touchscreen | LCD; Resistive Technology | LCD; Resistive Technology | Same |
| Connector Receptacles | LED Illuminated connector<br>readers | LED Illuminated connector<br>readers | Same |
| Monopolar Ports | 2 | 2 | Same |
| Bipolar/LigaSure Ports | 2 | 2 | Same |
| Hand switch Activation | Yes | Yes | Same |
| Footswitch Activation | Yes | Yes | Same |
| REM Quality Contact<br>Monitoring | Yes | Yes | Same |
| Wireless Connectivity | No | No | Same |
| Instrument<br>ldentification | Barcode/RFID | Barcode/RFID | Same |
{6}------------------------------------------------
{7}------------------------------------------------
### VI. PERFORMANCE DATA
The following performance data were conducted in support of the substantial equivalence determination. Verification and Validation testing confirmed that the modifications to the Valleylab™ FT10 Energy Platform do not raise any different issues of safety and effectiveness. The testing confirmed the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:
- System Verification showed that VLFT10GEN with SW v.5.0 possesses all required functionality and meets system specifications.
- . Confirmatory Electrical safety and EMC testing were conducted on the Proposed Device and showed that the system complies with IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
- . Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Device Software Functions" [2023] and IEC 62304 Medical Device Software- Software life cycle processes".
- Mechanical Testing in the form of Environmental and Packaging Test and Chemical Soak and Abrasion Test was conducted and met all the acceptance criteria.
### VII. CLINICAL STUDIES
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
### VIII. CONCLUSIONS
Verification and Validation activities demonstrate that the Valleylab™FT10 Energy Platform with software version 5.0 (VLFT10GEN SW v5.0) is substantially equivalent to the predicate Valleylab™FT10 Energy Platform with software version (VLFT10GEN SW v4.0) cleared in K191601. The intended use of the Valleylab™FT10 Energy Platform has not changed. The non-clinical data support the safety of the device and the hardware and software verification, and validation demonstrate that the Proposed Device Valleylab™FT10 Energy Platform with software version 5.0 (VLFT10GEN SW v5.0) should perform as intended in the specified use conditions.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.