Sterile single use tip cleaner (WK800-S10 , WK800-S20)
K241022 · Wickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd · GEI · Jul 11, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241022
Device Name
Sterile single use tip cleaner (WK800-S10 , WK800-S20)
Applicant
Wickimed (Huizhou) Medical Equipment Manufacturing Co.,Ltd
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Indications for Use
The sterile single use tip cleaner is a single use sterile device intended to be used as an electrosurgical accessory to remove eschar buildup from the tips of electrosurgical cauterization blades during surgical procedures.
Device Story
Sterile single-use foam pad with abrasive surface; used as electrosurgical accessory. Input: electrosurgical cauterization blade tip with eschar buildup. Operation: manual scratching of blade tip against abrasive surface to remove eschar. Output: cleaned cauterization blade. Used in surgical settings; operated by surgical staff. Adhesive back allows secure placement on surgical drapes or equipment. Benefits: maintains blade performance by removing eschar buildup during procedures. Contains barium monofilament for X-ray detection.
Clinical Evidence
Bench testing only. Testing included appearance, adhesion, peel strength (≥2N/CM), abrasive surface integrity (no sand shedding), and residual testing (ETO ≤4mg/pcs, ECH ≤9mg/pcs).
Technological Characteristics
Materials: Aluminum oxide abrasive, polyurethane foam base, pressure-sensitive adhesive, barium monofilament (X-ray detectable). Dimensions: 50mm x 50mm x 5mm or 50mm x 25mm x 5mm. Sterile, single-use, disposable. No energy source. Mechanical cleaning principle.
Indications for Use
Indicated for use as an electrosurgical accessory to remove eschar buildup from electrosurgical cauterization blades during surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo features the Department of Health & Human Services logo on the left. To the right of that is the FDA logo in blue, with the words "U.S. Food & Drug Administration" written in blue text.
July 11, 2024
WickiMed (Huizhou) Medical Equipment Manufacturing Co.,Ltd Haobin Li General Manager TangJiao XingWang Street, LiLin Town, ZhongKai Hi-Tech Zone HuiZhou, GuangDong 516003 China
Re: K241022
Trade/Device Name: Sterile single use tip cleaner (WK800-S10 , WK800-S20) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 20, 2024 Received: May 20, 2024
Dear Haobin Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Long H. Chen
Long H. Chen -S -s
Date: 2024.07.11 15:06:45 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K241022
Device Name
Sterile single use tip cleaner (WK800-S10 , WK800-S20)
Indications for Use (Describe)
The sterile single use tip cleaner is a single use sterile device intended to be used as an electrosurgical accessory to remove eschar buildup from the tips of electrosurgical cauterization blades during surgical procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K241022
# WickiMed (Huizhou) Medical Equipment Manufacturing Co., Ltd. 510(K) SUMMARY
This summary of 510(K) is being submitted in accordance with the requirements of 21 CFR 8807.92.
Type of submission :Traditional The assigned 510(K) number is:K241022 The date the summary was prepared: April 15, 2024
#### 1. Submitter information:
Manufacturer Name: WickiMed(Huizhou) Medical Equipment Manufacturing Co.,Ltd.
Address: TangJiao Street, LiLin Town, ZhongKai XingWang -Hi-Tech Zone, HuiZhou,GuangDong, China.
Establishment Registration Number:3010601992
Contact person:Haobin Li
Position: General Manager
Tel : 0086-0752-3860807
Fax : 0086-0752-3863017
E-mail: mac lai@wickimed.com
#### 2. Device Information:
Trade/Common Name: Sterile single use tip cleaner Models: WK800-S10 ,WK800-S20 Regulatory Class: Class II Regulatory Number: 21 CFR 878.4400 Review Panel: General & Plastic Surgery Product Code: GEI (Electrosurgical, Cutting & Coagulation & Accessories)
#### 3. Predicative Device
Manufacturer: Key Surgical Incorporated Trade/Device Name: Key Surgical® Cautery Tip Cleaner Product Code: GEI 510(k) Number: K151222
#### 4. Indication for Use
The sterile single use tip cleaner is a single use sterile device intended to be used as an electrosurgical accessory to remove eschar buildup from the tips of electrosurgical
{5}------------------------------------------------
# K241022
WickiMed (Huizhou) Medical Equipment Manufacturing Co., Ltd. cauterization blades during surgical procedures.
#### 5. Device Description
The product has 2 sizes: 50mm(L)×50mm(W)×5mm(T) and 50mm(L)×25mm(W)×
5mm(T). The product is composed by a small foam pad with an abrasive surface, and adhesive back layer, which are used to aid in the removal of eschar on electrosurgical blades during surgical procedures. The abrasive surface is made from aluminum oxide 'gravel' with a polyurethane foam base and pressure sensitive adhesive layer containing a barium monofilament for x-ray detection. The adhesive back allows for universal placement and keeps the device in place while the tip of the electrosurgical cauterization device is scratched on the abrasive surface to remove eschar buildup. The product does not come in contact with the patient, is provided sterile and is single-use.
#### 6. Comparison of Intended Use and Technological Characteristics
The Sterile single use tip cleaner predicate device is the Key Surgical, Inc., K151222, cleared on May 22, 2015. The following illustrates the similarities in the product design.
| Characteristics | Predicate Device | Proposed Device |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Key Surgical®<br>Cautery Tip Cleaner<br>K151222 | Sterile single use tip cleaner |
| Indications for Use | The disposable cautery tip cleaner is a single use sterile device intended to be used as an electrosurgical accessory to remove eschar buildup from the tips of electrosurgical cauterization blades during surgical procedures. | The sterile single use tip cleaner is a single use sterile device intended to be used as an electrosurgical accessory to remove eschar buildup from the tips of electrosurgical cauterization blades during surgical procedures. |
| Conditions of Use | Single Use, disposable | Single Use, disposable |
| Materials | Aluminum oxide 'gravel', with a polyurethane foam pad and polyurethane adhesive layer containing a barium monofilament for x-ray detection. | Aluminum oxide 'gravel' with a polyurethane foam base and pressure sensitive adhesive layer containing a barium monofilament for x-ray detection. |
| Adhesive back | Yes, for universal placement and aid in keeping the device in place. | Yes, for universal placement and aid in keeping the device in place. |
| Sterility | Provided Sterile | Provided Sterile |
| Principle of Operation | The cautery tip cleaner is used | The cautery tip cleaner is used |
Table 1 : Comparison to Predicate device
{6}------------------------------------------------
# K241022
| | to aid in the removal of eschar<br>on electrosurgical blades<br>during surgical procedures. | to aid in the removal of eschar<br>on electrosurgical blades<br>during surgical procedures. |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Interface with Electrosurgical<br>Cauterization Blade | The Cautery Tip Cleaner does<br>not impact the function of the<br>electrosurgical cauterization<br>blades. It is not required for<br>use with the blades, but an<br>accessory that aids in the<br>removal of eschar buildup. | The sterile single use tip<br>cleaner does not impact the<br>function of the electrosurgical<br>cauterization blades. It is not<br>required for use with the<br>blades, but an accessory that<br>aids in the removal of eschar<br>buildup. |
# WickiMed (Huizhou) Medical Equipment Manufacturing Co., Ltd.
### 7. Summary of Non-clinical Performance Testing
| Item | Requirement | Result |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1. Appearance | 1.1The surface of the product should be clean, free from damage, cracking, or other undesirable phenomena. | PASS |
| | 1.2 The product sandpaper surface, sponge base layer, pressure-sensitive adhesive, X-ray ray, and release paper should be tightly bonded together, and there should be no curling or falling off. | PASS |
| 2.Product performance | 2.1 The double-sided adhesive has good adhesion, smoothness and no warping. | PASS |
| | 2.2 There is no sand falling off on the polishing abrasive (sandpaper) surface. | PASS |
| | 2.3 The peel strength of the double-sided adhesive should be≥2N/CM. | PASS |
| 3. Residual Testing | 3. FTO Residual ≤4mg/pcs ECH≤9mg/pcs | PASS |
| | | Table 2 : Non-clinical Performance Testing | |
|--|--|--------------------------------------------|--|
|--|--|--------------------------------------------|--|
## 8. Conclusion
The sterile single use tip cleaner is as safe and effective as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.