← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K240999

# Legend X Applicator VO (K240999)

_Pollogen, Ltd. · GEI · May 8, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240999

## Device Facts

- **Applicant:** Pollogen, Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** May 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen100, gen 36, gen36L, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62 mJ/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.

## Device Story

Legend X Applicator VO is a handheld electrosurgical handpiece connecting to the Legend X console; delivers fractional radiofrequency (RF) energy to skin via micro-electrode pin arrays on disposable tips. Used for ablation, resurfacing, electrocoagulation, and hemostasis. Operated by clinicians in clinical settings; requires continuous direct user control and visualization. Console software manages energy delivery based on user-selected parameters; provides real-time status display. Benefits include non-invasive treatment of face and body skin conditions. Device is an extension of the cleared Legend X platform, adding the genXL tip for deeper tissue effects (up to 4mm).

## Clinical Evidence

Bench testing only. Performance verified through biocompatibility testing (ISO 10993-1, -5, -10, -11, -23), electrical safety/EMC compliance (IEC 60601-1), and software verification and validation per FDA guidance. No clinical data presented.

## Technological Characteristics

Handheld electrosurgical applicator; stainless steel and gold-plated micro-electrode pin arrays. Energy source: RF generator (Legend X console). Connectivity: Wired to console. Sterilization: EO (disposable tips). Software: Embedded control system. Standards: IEC 60601-1 (electrical/EMC), ISO 10993 (biocompatibility).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Legend X Platform ([K232903](/device/K232903.md))
- InMode System with the Morpheus8 Applicators ([K200947](/device/K200947.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

May 8, 2024

Pollogen Ltd. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

RE: K240999

Trade/Device Name: Legend X Applicator VO Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 11, 2024 Received: April 11, 2024

Dear Prithul Bom:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Long H. Chen -S Lona H. Chen-Date: 2024.05.08 13:39:12 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k)Number (if known) K240999

DeviceName Legend X Applicator VO

#### Indications for Use (Describe)

The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen361, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostass. At higher energy levels greater than 62 m.J/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|×| Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary K240999

## 1. General information

| 510(k) Submitter        | Pollogen Ltd<br>6 Kaufman Street<br>Tel Aviv, IL 6801298                                               |
|-------------------------|--------------------------------------------------------------------------------------------------------|
| FDA Registration Number | 3008753275                                                                                             |
| Primary Correspondent   | Karen Smith<br>Vice President, Regulatory & Quality<br>Lumenis Be, Inc., a Pollogen Ltd Family Company |
| Contact Information     | Email: karen.smith@lumenis.com                                                                         |
| Secondary Correspondent | Lola Fajinmi<br>Regulatory Affairs Consultant<br>MedTeCheck                                            |
| Contact Information     | Email: lola@MedTeCheck.com                                                                             |
| Date Prepared           | April 11, 2024                                                                                         |

#### 2. Device Identification

The Legend X Applicator VO consists of an applicator accessories (disposable tips) needed to perform some of Legend X dermatological procedures. The Legend X Applicator VO can be defined as follows:

| Proprietary Name            | Legend X Applicator VO                                            |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories              |
| Regulation Description:     | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Regulation Number:          | 878.4400                                                          |
| Device Class:               | Class II                                                          |
| Product Code:               | GEI                                                               |

#### 3. Predicate Devices

Primary Predicate:

| Proprietary Name            | Legend X Platform                                                 |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories              |
| Regulation Description:     | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| 510(k) Number:              | K232903                                                           |
| Regulation Number:          | 878.4400                                                          |

{5}------------------------------------------------

| Device Class: | Class II |
|---------------|----------|
| Product Code: | GEI, NGX |

Secondary Predicate:

| Proprietary Name            | InMode System with the Morpheus8 Applicators                      |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories              |
| Regulation Description:     | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| 510(k) Number:              | K200947                                                           |
| Regulation Number:          | 878.4400                                                          |
| Device Class:               | Class II                                                          |
| Product Code:               | GEI                                                               |

#### 4. Device Description

The Legend X Applicator VO ("Proposed Device") is connected to the Legend X softwarecontrolled capital equipment platform to enable application of radiofrequency onto the skin for ablation and resurfacing or electrocoagulation and hemostasis. The proposed device applicator and applicator accessories are identical to the recently cleared Legend X Platform (K232903), with the exception of the new optional Applicator VO genXL disposable tip. The application of radiofrequency with the Applicator VO Disposable Tip only occurs under the continuous and direct control of a user with direct visualization, via the user inputs into the Legend X Console and Applicator indicators.

The proposed device consists of applicators and applicator accessories (disposable tips) needed to perform some of Legend X dermatological procedures. The summary descriptions of the proposed device are outlined below.

Legend X Applicator VO: The Legend X Applicator VO is a handheld handpiece which delivers energy to the treatment area. Depending on the selected user inputs, the user can utilize the applicator to deliver fractional RF energy for ablation and resurfacing, or electrocoagulation and hemostasis, as a means for noninvasive skin treatment throughout the face and body. The Applicator VO is connected to the cleared Legend X Console via an Applicator Connector cord.

Legend X Software: The Legend X Software provides the user and patient with the ability to safely commence, drive, and stop the operation of the Applicator on the skin area of interest throughout the face and body. It receives user input from the Legend X Console and computes the appropriate output to the Applicator VO connected to the Console. It provides a graphical user interface where the treatment parameters are adjusted as well as status of operation is shown in real time and displays important system status information.

{6}------------------------------------------------

Legend X Applicator VO Disposable Tips: The Legend X Applicator VO Disposable Tips set is part of the Legend X Applicator VO kit. The Legend X Applicator VO Disposable Tips Set includes:

- 1. gen12 disposable tip
- 2. gen36 disposable tip
- 3. gen36L disposable tip
- 4. gen100 disposable tip
- 5. H7X7 disposable tip
- 6. genXL disposable tip

The patient-contacting portions of the disposable tips are constructed of stainless steel and gold plating. The disposable tips are connected to handheld Applicator VO for delivery of radiofrequency electrical current via an array of micro-electrode pins onto the skin surface.

## 5. Intended Use/Indications for Use

The Legend X Applicator VO and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (tips gen12, gen100, gen 36, gen36L, H7x7). The Legend X Applicator VO with genXL tip is intended for use in dermatological procedures for electrocoagulation and hemostasis. At higher energy levels greater than 62 mJ/pin, use of the Applicator VO with genXL tip is limited to Skin Types I-IV.

## 6. Comparison of Technological Characteristics with the Predicate Devices

Overall, the proposed and predicate devices are based on the following similar basic technological elements:

- 1. Device contains a central console that connects each of the main components to facilitate dermatologic procedures and facilitates the selection of fractional RF treatment modes via the graphic user interface
- 2. Device contains an applicator to allow energy delivery and application on to the area of interest on the skin
- 3. Device contains similar components between the subject and predicate devices (i.e., Applicator VO and disposable tips)
- 4. Device requires continuous direct control by the user to operate the Applicator
- 5. Device mode operation is only at the command of the user

Comparison of the technological characteristics with the predicate devices for each of the Legend X Applicator VO disposable tip components are described below:

{7}------------------------------------------------

| Proposed Device                                                                                       | Primary Predicate                                                                                                              | Secondary Predicate                                                                        | Technological Characteristics<br>Comparison                                                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Legend X<br>Applicator VO                                                                             | Legend X Platform<br>(K232903)                                                                                                 | InMode System<br>(K200947)                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Components:<br>- Applicator VO<br>(connected to a<br>cleared Legend X<br>Console) with<br>accessories | System Components:<br>- Console<br>- Patient-controlled<br>manual switch<br>- Applicators (1, 2, 3,<br>and VO)<br>- Footswitch | System Components:<br>- Console<br>- Applicator<br>- Footswitch                            | Identical Characteristics:<br>- The Legend X Applicator<br>VO is identical to the<br>primary predicate Legend<br>X Platform in<br>functionality, design, and<br>specifications.                                                                                                                                                                                                                                                      |
|                                                                                                       |                                                                                                                                |                                                                                            | Differences<br>- The subject Legend X<br>Applicator VO is<br>compatible with six<br>disposable tip accessories,<br>whereas the primary<br>predicate is compatible<br>with five disposable tip<br>accessories.                                                                                                                                                                                                                        |
| Applicator VO<br>Disposable Tips:<br>- gen12<br>- gen36<br>- gen36L<br>- gen100<br>- H7x7<br>- genXL  | Applicator VO<br>Disposable Tips:<br>- gen12<br>- gen36<br>- gen36L<br>- H7x7<br>- gen100                                      | Accessories<br>- T tip head<br>- 12-pin tip head<br>- 24-pin tip head<br>- 40-pin tip head | Identical Characteristics:<br>- Functionality: To provide<br>delivery of fractional RF<br>energy<br>- Similar materials,<br>specifications, and intended<br>functionality<br>Differences<br>- The Applicator VO genXL<br>disposable tip is an<br>additional disposable tip<br>with insertion depth<br>options of up to 4 mm,<br>which is within the range<br>of the secondary<br>predicate's disposable tips<br>that are up to 7 mm. |

# 7. Performance Data

{8}------------------------------------------------

The Legend X Applicator VO was assessed for performance in accordance with internal design specification with the applicable performance standards to demonstrate safety and effectiveness.

The testing identified no new issues of safety or effectiveness. The testing performed are summarized below:

### Summary of Performance Testing:

The reprocessing, sterilization, and shelf life for the subject device are identical to those evaluated for the primary predicate. The Legend X. Applicator VO disposable tips are EO sterilized for single use, and the Legend X Applicator VO is reusable.

Biocompatibility: The final finished form of the subject device has been used for the biocompatibility evaluation. Biocompatibility for patient-contacting components has been evaluated and validated in accordance with the provisions of the following guidance document and standards:

- . FDA Guidance for Industry and Food and Drug Administration Staff – Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
- ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing ● within a risk management process
- ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for in vitro ● cvtotoxicitv
- ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for irritation . and skin sensitization
- . ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- ISO 10993-23 Biological evaluation of medical devices Part 23: Tests for irritation ●

Electrical Safety and Electromagnetic Compatibility: The subject device has been fully evaluated for electrical safety and EMC compliance and is compliant with IEC 60601-1 standards.

Software: Software was developed, tested, and verified per Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

Results of verification and validation (V&V) testing confirm that the proposed device conforms to design specifications and meets the needs of the intended users.

V & V Testing: Performance testing was executed to verify the overall functionality of the proposed device to operate as specified by the design input requirements, including applicator and controls, various functional safety features, and other general functionality. Requirements

{9}------------------------------------------------

for safety and effectiveness of the device, including the adherence to regulatory standards, were verified. Results of verification testing confirm that the subject device conforms to design specifications and requirements and meets the needs of the intended users.

#### 8. Conclusion

Based on the same Indications for Use and similar technological characteristics as the secondary predicate and performance testing, the Legend X Applicator VO with genXL tip raises no new questions of safety and effectiveness and is substantially equivalent to the secondary predicate device in terms of safety, effectiveness, and performance.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240999](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240999)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
