← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K240017

# InMode System with the Morpheus8 90 Applicator (K240017)

_InMode , Ltd. · GEI · Jun 13, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240017

## Device Facts

- **Applicant:** InMode , Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jun 13, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The InMode System with the Morpheus8 90 Applicator is intended for use in dermatological procedures where coaqulation/contraction of soft tissue or hemostasis is needed. At higher energy levels greater than 62 mJ/pin, the use of the Morpheus8 90 Applicator is limited to Skin Types I-IV.

## Device Story

InMode System with Morpheus8 90 Applicator is an RF-based electrosurgical device for dermatological procedures; provides soft tissue coagulation, contraction, and hemostasis. System comprises AC/DC power supply, RF generator, controller, and LCD touch screen; connects to a handheld applicator with a disposable, sterilized 24-pin tip. Physician operates device via foot switch to deliver fractional RF energy to skin. Device monitors RF parameters to enhance safety and minimize side effects. Output is thermal effect on tissue; used in clinical settings. Benefits include controlled coagulation necrosis for dermatological applications.

## Clinical Evidence

No clinical data submitted. Evidence consists of bench testing and ex-vivo tissue testing. Performance testing validated fractional coagulation necrosis patterns at depths up to 7mm, demonstrating equivalence to predicates K180189 and K200947. System meets requirements for electrical/mechanical safety and electromagnetic compatibility per IEC 60601 standards.

## Technological Characteristics

RF-based electrosurgical device; 1 MHz frequency; 65W max output power. Features fractional multi-electrode pin array. Console dimensions: 46cm x 46cm x 100cm; weight 30kg. Input: 100-240 VAC, 50-60 Hz. Patient contact materials identical to predicate K180189. Sterilization: disposable, single-use tip. Software-controlled energy delivery.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- InMode System with Fractora3D/3D-90 Applicators ([K180189](/device/K180189.md))
- InMode System with the Morpheus8 Applicators ([K200947](/device/K200947.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

June 13, 2024

InMode Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street. Floor 23 Philadelphia, Pennsylvania 19103

Re: K240017

Trade/Device Name: InMode System with the Morpheus8 90 Applicator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: NA Received: May 16, 2024

### Dear Janice Hogan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long H. Chen -- S Digitally signed by Long H.Chen -S
Long H. Chen -- S Date: 2024.06.13 07:46:52 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K240017

Device Name

InMode System with the Morpheus8 90 Applicator

Indications for Use (Describe)

The InMode System with the Morpheus8 90 Applicator is intended for use in dermatological procedures where coaqulation/contraction of soft tissue or hemostasis is needed. At higher energy levels greater than 62 mJ/pin, the use of the Morpheus8 90 Applicator is limited to Skin Types I-IV.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) #: | K240017 | 510(k) Summary | Prepared on: 2024-06-11 |
|-----------|---------|----------------|-------------------------|
|-----------|---------|----------------|-------------------------|

Contact Details

|                             | 21 CFR 807.92(a)(1)                                               |
|-----------------------------|-------------------------------------------------------------------|
| Applicant Name              | InMode Ltd.                                                       |
| Applicant Address           | Tabor Building, Shaar Yokneam POB 44 Yokneam Illit 2069200 Israel |
| Applicant Contact Telephone | +972-4-9097470                                                    |
| Applicant Contact           | Mrs. Suhair Francis                                               |
| Applicant Contact Email     | Suhair.Francis-Najjar@inmodemd.com                                |

Device Name

|                     | 21 CFR 807.92(a)(2)                                            |
|---------------------|----------------------------------------------------------------|
| Device Trade Name   | InMode System with the Morpheus8 90 Applicator                 |
| Common Name         | Electrosurgical cutting and coagulation device and accessories |
| Classification Name | Electrosurgical, Cutting & Coagulation & Accessories           |
| Regulation Number   | 878.4400                                                       |
| Product Code        | GEI                                                            |

Legally Marketed Predicate Devices

|             | 21 CFR 807.92(a)(3)                                      |              |
|-------------|----------------------------------------------------------|--------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K180189     | InMode System with Fractora3D/3D-90 Applicators          | GEI          |
| K200947     | InMode System with the Morpheus8 Applicators             | GEI          |

Device Description Summary

|  | 21 CFR 807.92(a)(4) |
|--|---------------------|
|--|---------------------|

The InMode System with the Morpheus8 90 Applicator is a RF technology based device intended for dermatological applications which require coagulation/contraction of soft tissue or hemostasis.

The device platform is identical to the one in FDA-Cleared InMode System with the Fractora3D 90 Applicator (K180189), inclusive of minor software modifications. The InMode System with the Morpheus8 90 Applicator employs fractional RF multi-electrode technology for procedures requiring electrocoaqulation and hemostasis. The Morpheus8 90 Applicator is deliver radiofrequency energy to the skin in a fractional manner, via an array of multi-electrode pins. The device provides enhanced safety while minimizing possible side effects by monitoring RF parameters.

The InMode System with the Morpheus8 90 Applicator consists of an AC/DC power supply unit, RF generator, controller and user interface including LCD touch screen. The Morpheus8 90 Applicator is connected to the console via a cable and a foot switch activates the energy delivery to the applicator. The Morpheus a handle and a detachable, sterilized, disposable, singleuse 24 pin tip head accessory.

Following are the InMode System with the Morpheus8 90 Applicator specifications:

| RF Max Output Power: | 65 Watt                     |
|----------------------|-----------------------------|
| RF Output Frequency: | 1 MHz                       |
| Dimension:           | 46cm W x 46cm D x 100cm H   |
|                      | (18.2" W x 18.2" D x 40" H) |

{4}------------------------------------------------

Weight: 30 Kg (70.4 Ibs.)

Main Line Frequency: 50-60 Hz

Input Voltage (nominal): 100-240 VAC

# Intended Use/Indications for Use

The InMode System with the Morpheus8 90 Applicator is in dermatological procedures where coagulation/contraction of soft tissue or hemostasis is needed.

At higher energy levels greater than 62 ml/pin, the use of the Morpheus8 90 Applicator is limited to Skin Types I-V.

# Indications for Use Comparison

The device has similar indications for use as the predicate. Specifically, An additional change for this 510(k) submission consists of modified indications for use; the mention of "coagulation of soft tissue" is added to the indications. This change in the wording of the indications does not refer to a different intended use nor a different performance claim.

# Technological Comparison

The design and components in the InMode System, including the console (with power supply, RF generator, controller and display panel) and the Applicator (with cable, connole, handle and tip) are identical to the design and components found in the primary predicate device (K180189) except for minor design changes to the handle to improve ergonomics and manufacturing process. The safety features and compliance with safety standards in the Morpheus8 90 Applicator are identical to the safety features and compliance with safety standards found in the primary predicate device. (K180189). Patient contact materials are also identical.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The current system is a modification to InMode's system with the Fractora 3D 90 Applicator cleared for ocedures for electrocoagulation and hemostasis (K180189).

The subject of the current 510(k) is to clear the Fractora 3D 90 for a treatment depth of up to 7mm. This change is implemented via software only.

No clinical data was submitted.

Performance testing which supported the primary predicate device (K180189), remains applicable, including software validation testing, electrical and mechanical safety testing to EC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 and IEC 60601-2-2, bench testing, and ex-vivo tissue testing to evaluate the fractional coagulation necrosis pattern of target tissues formed by thermal effect. These performance tests demonstrated that the device's performance and specifications meet system requirements.

Further, ex-vivo testing was performed with the secondary predicate device K200947 up to a treatment depth of 7 mm. This testing is equally relevant to the current submission. Additional testing was provided in this submission comparing the thermal effect of the subject device to the primary predicate up to a treatment depth of 7 mm. Thus, the data and information provided demonstrates the subject device is substantially equivalent to and is as safe and as effective as the predicate device.

## 21 CFR 807.92(a)(5)

## 21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240017](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K240017)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
