← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K233351

# Single Use Non-Stick McPherson Bipolar Forceps, 110mm, 0.5mm Tip with Cable  (HNSAG-5110M); Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 0.5mm Tip with Cable  (HNSAG-5115J); Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 1.0mm Tip with Cable  (HNSAG-1115J); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 1.0mm Tip with Cable  (HNSAG-1120A); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 0.5mm Tip with Cable  (HNSAG-5120A); Single Use Non-Stick Adson Bipolar Force (K233351)

_Hg Innovations, Ltd. · GEI · Nov 20, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K233351

## Device Facts

- **Applicant:** Hg Innovations, Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Nov 20, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

McPherson Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are intended for use by a physician familiar with electrosurgery for bipolar coagulation and irrigation of tissue for general open surgery where coagulation of soft tissue is needed. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must be operated within the following parameters: -Frequency range between 300 kHz-1,000 kHZ; -Maximum generator operating voltage 600Vp. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures. The types of surgery intended include: -ENT -Gynecology -Urology -General Surgery -Neurosurgery -Laryngeal Surgery -Orthopedic Surgery -Thoracic Surgery Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must only be used with a bipolar coagulation current. HG Innovations, as a manufacturer, does not recommend a specific procedure. Use of bipolar techniques may be desirable in order to avoid unwanted tissue damage for surgical procedures where HF current could flow through relatively small cross-sectional area of body.

## Device Story

Single-use, non-stick bipolar forceps (various designs/sizes) with pre-attached cables; used to grasp, manipulate, coagulate, and irrigate soft tissue. Operates by connecting to the bipolar output of an electrosurgical generator; tissue grasped between tines acts as electrodes; current passes to desiccate/coagulate. Irrigating models feature a drain along the tines connected via Luer-Lock to an irrigation pump. Used in OR/clinic settings by physicians. Benefits include reduced unwanted tissue damage via bipolar technique and non-stick properties. Single-use design eliminates reprocessing risks.

## Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility (ISO 10993-1, 5, 10, 11, USP 151), electrical safety (IEC 60601-1, 60601-2-2), and mechanical testing demonstrated performance equivalent to predicates.

## Technological Characteristics

Materials: Ag800 (80% silver) tips, AISI 420 stainless steel arms, Polypropylene/PVC components. Energy: Bipolar electrosurgical current (300-1000 kHz, max 600Vp). Form factor: Various lengths (110-240mm) and tip configurations (Bayonet, Straight, McPherson, Jeweller, Adson). Connectivity: Wired to electrosurgical generator. Sterilization: Ethylene Oxide (EO).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Adeor Medical nxt™ Non-stick Bipolar Forceps / Adeor Medical nxt™ Single-Use Non-stick Bipolar Forceps ([K191847](/device/K191847.md))
- Single Use Non-Stick Bipolar Forceps sterile/non sterile, Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non sterile ([K182773](/device/K182773.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 20, 2023

HG Innovations Ltd % Wondwossem Tekolla Associate Regulatory Consultant Medical Device Academy 345 Lincoln Hill Road Shrewsbury, Vermont 05738

Re: K233351

Trade/Device Name: Single Use Non-Stick McPherson Bipolar Forceps, 110mm, 0.5mm Tip with Cable (HNSAG-5110M); Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 0.5mm Tip with Cable (HNSAG-5115J); Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 1.0mm Tip with Cable (HNSAG-1115J); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 1.0mm Tip with Cable (HNSAG-1120A); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 0.5mm Tip with Cable (HNSAG-5120A); Single Use Non-Stick Adson Bipolar Forceps, 150mm, 1.0mm Tip with Cable (HNSAG-1150A); Single Use Non-Stick Straight Bipolar Forceps, 150mm, 0.5mm Tip with Cable (HNSAG-5150S); Single Use Non-Stick Straight Bipolar Forceps, 150mm, 1.0mm Tip with Cable (HNSAG-1150S); Single Use Non-Stick Straight Bipolar Forceps, 160mm, 0.5mm Tip with Cable (HNSAG-5160S); Single Use Non-Stick Straight Bipolar Forceps, 160mm, 1.0mm Tip with Cable (HNSAG-1160S); Single Use Non-Stick Straight Bipolar Forceps, 180mm, 0.5mm Tip with Cable (HNSAG-5180S); Single Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 29, 2023 Received: September 29, 2023

Dear Wondwossem Tekolla:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some

{1}------------------------------------------------

cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See 

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.11.20
13:04:01 -05'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

Irrigation Tubing (HNSAG-2220S/IRS);

### K233351

Device Name

Single Use Non-Stick McPherson Bipolar Forceps, 110mm, 0.5mm Tip with Cable (HNSAG-5110M): Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 0.5mm Tip with Cable (HNSAG-5115J); Single Use Non-Stick Jeweller Bipolar Forceps, 115mm, 1.0mm Tip with Cable (HNSAG-1115J); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 1.0mm Tip with Cable (HNSAG-1120A); Single Use Non-Stick Adson Bipolar Forceps, 120mm, 0.5mm Tip with Cable (HNSAG-5120A); Single Use Non-Stick Adson Bipolar Forceps, 150mm, 1.0mm Tip with Cable (HNSAG-1150A); Single Use Non-Stick Straight Bipolar Forceps, 150mm, 0.5mm Tip with Cable (HNSAG-5150S); Single Use Non-Stick Straight Bipolar Forceps, 150mm, 1.0mm Tip with Cable (HNSAG-1150S); Single Use Non-Stick Straight Bipolar Forceps, 160mm, 0.5mm Tip with Cable (HNSAG-5160S); Single Use Non-Stick Straight Bipolar Forceps, 160mm, 1.0mm Tip with Cable (HNSAG-1160S); Single Use Non-Stick Straight Bipolar Forceps, 180mm, 0.5mm Tip with Cable (HNSAG-5180S); Single Use Non-Stick Straight Bipolar Forceps, 180mm, 1.0mm Tip with Cable (HNSAG-1180S); Single Use Non-Stick Straight Bipolar Forceps, 200mm, 1.0mm Tip with Cable (HNSAG-1200S); Single Use Non-Stick Straight Bipolar Forceps, 200mm, Angled 1.0mm Tip with Cable (HNSAG-1200SA): Single Use Non-Stick Bayonet Bipolar Forceps, 165mm, 1.0mm Tip with Cable (HNSAG-1165B); Single Use Non-Stick Bayonet Bipolar Forceps, 200mm, 1.0mm Tip with Cable (HNSAG-1200B); Single Use Non-Stick Bayonet Bipolar Forceps, 220mm, 1.0mm Tip with Cable (HNSAG-1220B); Single Use Non-Stick Bayonet Bipolar Forceps, 240mm, 1.0mm Tip with Cable (HNSAG-1240B); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 150mm, 0.5mm Tip with Cable & Irrigation Tubing (HNSAG-5150S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 150mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1150S/IRS): Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 160mm, 0.5mm Tip with Cable & Irrigation Tubing (HNSAG-5160S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 160mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1160S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 180mm, 0.5mm Tip with Cable & Irrigation Tubing (HNSAG-5180S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 180mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1180S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 200mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1200S/IRS): Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 200mm, 2.0mm Tip with Cable & Irrigation Tubing (HNSAG-2200S/IRS): Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 200mm, Angled 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1200SA/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 200mm, Angled 2.0mm Tip with Cable & Irrigation Tubing (HNSAG-2200SA/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 220mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1220S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 220mm, 2.0mm Tip with Cable &

{4}------------------------------------------------

Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 220mm, Angled 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1220SA/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 220mm, Angled 2.0mm Tip with Cable & Irrigation Tubing (HNSAG-2220SA/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 240mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1240S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 240mm, 2.0mm Tip with Cable & Irrigation Tubing (HNSAG-2240S/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 240mm, Angled 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1240SA/IRS); Single Use Non-Stick, Irrigating, Straight Bipolar Forceps, 240mm, Angled 2.0mm Tip with Cable & Irrigation Tubing (HNSAG-2240SA/IRS); Single Use Non-Stick, Irrigating, Bayonet Bipolar Forceps, 165mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1165B/IRS): Single Use Non-Stick, Irrigating, Bayonet Bipolar Forceps, 200mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1200B/IRS): Single Use Non-Stick, Irrigating, Bayonet Bipolar Forceps, 220mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1220B/IRS); Single Use Non-Stick, Irrigating, Bayonet Bipolar Forceps, 240mm, 1.0mm Tip with Cable & Irrigation Tubing (HNSAG-1240B/IRS) ndications for Use (Describe)

McPherson Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are intended for use by a physician familiar with electrosurgery for bipolar coagulation and irrigation of tissue for general open surgery where coagulation of soft tissue is needed. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must be operated within the following parameters:

-Frequency range between 300 kHz-1,000 kHZ;

-Maximum generator operating voltage 600Vp.

Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures. The types of surgery intended include:

-ENT

-Gynecology

- -Urology
- -General Surgery
- -Neurosurgery
- -Laryngeal Surgery
- -Orthopedic Surgery
- -Thoracic Surgery

Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must only be used with a bipolar coagulation current.

HG Innovations, as a manufacturer, does not recommend a specific procedure.

Use of bipolar techniques may be desirable in order to avoid unwanted tissue damage for surgical procedures where HF current could flow through relatively small cross-sectional area of body.

{5}------------------------------------------------

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

# 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

| SUBMITTER:           | Mr. Wondwossen Tekolla      |
|----------------------|-----------------------------|
| Company Name:        | Medical Device Academy Inc. |
| Address:             | 345 Lincoln Hill Rd         |
| City, State, Zip USA | Shrewsbury, VT 05738        |
| Tel:                 | 919-903-0194                |
| Fax:                 | N/A                         |

| Contact Person: | Dr. M Umran Rafiq |
|-----------------|-------------------|
| Date Prepared:  | August 23, 2023   |

#### DEVICE ll.

| Device Trade Name:           | Heinrich Single Use Non-Stick Bipolar Forceps/ Heinrich<br>Single Use Irrigating Non-Stick Bipolar Forceps |
|------------------------------|------------------------------------------------------------------------------------------------------------|
| Classification Name:         | Electrosurgical Cutting & Coagulation Device and<br>Accessories                                            |
| Requlation:                  | 21 CFR §878.4400                                                                                           |
| Regulatory Class:            | Class II                                                                                                   |
| Device Panel:                | General & Plastic Surgery                                                                                  |
| Product Classification Code: | GEI                                                                                                        |

#### PREDICATE DEVICE lll.

| A) Predicate 1: Irrigating forceps. |                                                                                                         |
|-------------------------------------|---------------------------------------------------------------------------------------------------------|
| Predicate Manufacturer:             | Adeor Medical AG                                                                                        |
| Predicate Trade Name:               | Adeor Medical nxt™ Non-stick Bipolar Forceps<br>Adeor Medical nxt™ Single-Use Non-stick Bipolar Forceps |
| Predicate 510(k):                   | K191847                                                                                                 |

### B) Predicate 2: Non-irrigating forceps

| Predicate Manufacturer: | Faulhaber by Pinzetten OHG                                                                                                           |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Trade Name:   | Single Use Non-Stick Bipolar Forceps sterile/non sterile,<br>Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non<br>sterile |
| Predicate 510(k):       | K182773                                                                                                                              |

No reference devices were used in this submission.

### IV. DEVICE DESCRIPTION

The single-use, non-stick bipolar forceps and single-use, non-stick irrigating bipolar forceps (various sizes, designs and tip configurations), with preattached cables are designed to grasp, manipulate, coagulate and irrigate soft tissues and are intended for 

{7}------------------------------------------------

use by a physician familiar with electrosurgery in bipolar coagulation for general open surgery where coagulation of soft tissue is needed. The blood vessel or tissue is grasped between the forceps tines, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue. Bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coaqulation current. The bipolar forceps must be operated with the following parameters:

-Frequency range between 300 kHz-1,000 kHZ: -Maximum generator operating voltage 600Vp.

#### V. INDICATIONS FOR USE

Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are intended for use by a physician familiar with electrosurgery for bipolar coagulation and irrigation of tissue for general open surgery where coagulation of soft tissue is needed. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must be operated within the following parameters:

-Frequency range between 300 kHz-1,000 kHZ;

-Maximum generator operating voltage 600Vp.

Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures. The types of surgery intended include:

-ENT

- -Gynecology
- -Urology
- -General Surgery
- -Neurosurgery
- -Laryngeal Surgery
- -Orthopedic Surgery
- -Thoracic Surgery

Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must only be used with a bipolar coagulation current.

HG Innovations, as a manufacturer, does not recommend a specific procedure.

Use of bipolar techniques may be desirable in order to avoid unwanted tissue damage for surgical procedures where HF current could flow through relatively small crosssectional area of body.

{8}------------------------------------------------

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE & PERFORMANCE DATA

Table 1: Comparison of subject device, Heinrich Single Use Non-Stick Bipolar Forceps with Cable – and Predicate Device, Adeor Medical nxt™ Non-stick Bipolar Forceps Adeor Medical nxt™ Single-Use Non-stick Bipolar Forceps- K191847

| Feature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Justification For<br>Differences                                                                                                                                                                                                           |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Heinrich Single Use<br>Non-Stick Bipolar<br>Forceps with Cable –<br>(Multiple models with<br>different sizes)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Adeor Medical nxt™<br>Non-stick Bipolar<br>Forceps/ Adeor<br>Medical nxt™<br>Single-Use Non-<br>stick Bipolar<br>Forceps K191847                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                            |
| Manufacturer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HG Innovations Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Adeor Medical AG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ----------                                                                                                                                                                                                                                 |
| Regulation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                       |
| Classification 21                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                       |
| Product Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | GEI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | GEI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                       |
| Device Description                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The product family<br>“Single Use Bipolar<br>Forceps”, including<br>Heinrich Single Use,<br>Non-Stick Bipolar<br>Forceps (various<br>sizes, designs and tip<br>configurations), with<br>Pre-Attached Cable<br>are designed to<br>grasp, manipulate,<br>coagulate and irrigate<br>soft tissues and are<br>intended for use by a<br>physician familiar with<br>electrosurgery in<br>bipolar coagulation<br>for general open<br>surgery where<br>coagulation of soft<br>tissue is needed. The<br>blood vessel or tissue<br>is grasped between<br>the forceps tines,<br>each of which acts as<br>an electrode, and<br>current passes to<br>desiccate and<br>coagulate the tissue.<br>Bipolar forceps are<br>coppected through a                                           | The Adeor Medical<br>AG Bipolar Forceps<br>are electrosurgical<br>instruments used to<br>grasp, manipulate,<br>cut or<br>coagulate tissue.<br>Bipolar forceps have<br>various lengths and<br>tip configurations, as<br>well as irrigation and<br>suction<br>technologies. Both<br>reusable and single-<br>use forceps are<br>available, with flat<br>plug or two pin plug<br>configurations.<br>Bipolar forceps are<br>connected through a<br>suitable bipolar<br>cable with the<br>bipolar output of a<br>high frequency<br>generator and may<br>be used only with<br>bipolar coagulation<br>current. Adeor<br>bipolar forceps<br>must be operated<br>with the following | Similar                                                                                                                                                                                                                                    |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                            |
| suitable bipolar cable<br>with the bipolar output<br>of an electrosurgical<br>generator. Bipolar<br>forceps must only be<br>used with bipolar<br>coagulation current.<br>The Single Use, Non-<br>Stick Bipolar Forceps<br>are single use<br>products and must<br>not be reused.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | parameters:<br>Frequency range<br>between 300 kHz<br>and 1,000 kHz:<br>maximum generator<br>operating voltage<br>600Vp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                            |
| <i>Indications for use</i>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                            |
| Heinrich Single-use<br>non-stick bipolar<br>forceps are intended<br>for use by a physician<br>familiar with<br>electrosurgery for<br>bipolar coagulation<br>and irrigation of tissue<br>for general open<br>surgery where<br>coagulation of soft<br>tissue is needed.<br>Single-use non-stick<br>bipolar forceps must<br>be operated within<br>the following<br>parameters:<br>-Frequency range<br>between 300 kHz-<br>1,000 kHZ;<br>-Maximum generator<br>operating voltage<br>600Vp.<br><br>Single-use non-stick<br>bipolar forceps have<br>not been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization<br>procedures and<br>should not be used<br>for these procedures.<br>The types of surgery<br>intended include: | The Adeor Medical<br>Non-stick Bipolar<br>Forceps are<br>intended for use by<br>a physician familiar<br>with<br>electrosurgery in<br>bipolar coagulation<br>for general open<br>surgery where<br>coagulation of soft<br>tissue is<br>needed. Adeor<br>bipolar forceps must<br>be operated with the<br>following<br>parameters:<br>Frequency range<br>between 300 kHz<br>and 1,000 kHZ;<br>maximum generator<br>operating voltage<br>600Vp.<br><br>The Adeor Medical<br>Non-stick Bipolar<br>Forceps have not<br>been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization<br>procedures and<br>should not be used<br>for these<br>procedures.<br>The types of surgery                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | -ENT<br>-Gynecology<br>-Urology<br>-General Surgery<br>-Neurosurgery<br>-Laryngeal Surgery<br>-Orthopedic Surgery<br>-Thoracic Surgery<br>Single-use non-stick<br>bipolar forceps and<br>are connected<br>through a suitable<br>bipolar cable with the<br>bipolar output of an<br>electrosurgical<br>generator. Single-use<br>non-stick bipolar<br>forceps must only be<br>used with a bipolar<br>coagulation current.<br>HG Innovations, as a<br>manufacturer, does<br>not recommend a<br>specific procedure.<br>Use of bipolar<br>techniques may be<br>desirable in order to<br>avoid unwanted<br>tissue damage for<br>surgical procedures<br>where HF current<br>could flow through<br>relatively small cross-<br>sectional area of<br>body. - Thoracic<br>Surgery | -ENT<br>-Gynecology (except<br>for use in female<br>sterilization)<br>-Urology<br>-General Surgery<br>-Neurosurgery<br>-Laryngeal Surgery<br>-Orthopedic Surgery<br>-Thoracic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                            |
| Rx/ OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                       |
| Design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Bayonet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bayonet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Mcpherson                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Jeweller                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Adson                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Straight                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Irrigating Bayonet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Irrigating Straight                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                        |
| Energy Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                       |
| Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Both single use and<br>multiuse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Subject device is<br>single use,<br>predicate device<br>has both single use<br>and multiuse<br>devices. Disposable<br>use of subject<br>device reduces risk.                                                                               |
| Maximum Peak Voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 600 Vp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 600Vp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                       |
| Electrode Type                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same…

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K233351](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K233351)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
