K230432 · Hg Innovations, Ltd. · GEI · May 18, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230432
Device Name
Single Use Bipolar Forceps
Applicant
Hg Innovations, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Single-use, bipolar forceps are intended for bipolar coagulation to remove tissue and control bleeding by use of high frequency electrical current. The bipolar forceps must be operated with the following parameters: -Frequency range between 300 kHz-1,000 kHZ; -Maximum generator operating voltage 600Vp.
Device Story
Single-use bipolar forceps designed for grasping, manipulating, and coagulating soft tissue during open surgery. Device consists of forceps with pre-attached cables; tines act as electrodes. Operated by physician; connects to standard electrosurgical generator (ESU) via bipolar cable. Current passes between tines to desiccate and coagulate tissue. Single-use design eliminates reuse-related risks. Output is thermal coagulation of tissue; aids in bleeding control. Device provided sterile (EO).
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity) and pyrogenicity (USP 151) passed. Electrical safety testing per IEC 60601-2-2 (HiPot, continuity, dielectric strength) passed. Mechanical performance verified via drop testing.
Indicated for bipolar coagulation to remove tissue and control bleeding in general open surgery. For use by physicians familiar with electrosurgery. Requires bipolar coagulation current (300-1,000 kHz; max 600Vp).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 18, 2023
HG Innovations Ltd % Wondwossen Tekolla Associate Regulatory Consultant Medical Device Academy, Inc. 345 Lincoln Hill Road Shrewsbury, Vermont 05738
Re: K230432
Trade/Device Name: Single Use Bipolar Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 17, 2023 Received: February 17, 2023
Dear Wondwossen Tekolla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.05.18<br>11:13:23 -04'00' |
|---------------------|---------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------|
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230432
Device Name Single Use Bipolar Forceps
## Indications for Use (Describe)
Single-use, bipolar forceps are intended for bipolar coagulation to remove tissue and control bleeding by use of high frequency electrical current.
The bipolar forceps must be operated with the following parameters:
- -Frequency range between 300 kHz-1,000 kHZ;
-Maximum generator operating voltage 600Vp.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY- K230432
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
| SUBMITTER: | HG Innovations Ltd.<br>5 Elder Court,<br>Lions Drive Shadsworth Business Park,<br>Blackburn Lancashire, BB1 2EQ UK |
|------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Contact Person:<br>Phone No: | Dr. M. Umran Rafiq<br>+44 7702195623 |
| Date Prepared: | May 17 2023 |
#### DEVICE ll.
| Device Trade Name: | Single Use Bipolar Forceps |
|------------------------------|-----------------------------------------------------------------|
| Classification Name: | Electrosurgical Cutting & Coagulation Device and<br>Accessories |
| Regulation: | 21 CFR §878.4400 |
| Regulatory Class: | Class II |
| Device Panel: | General & Plastic Surgery |
| Product Classification Code: | GEI |
#### . PREDICATE DEVICE
| Predicate Manufacturer: | Titan Manufacturing Inc. |
|-------------------------|-------------------------------------|
| Predicate Trade Name: | Titan Manufacturing Bipolar Forceps |
| Predicate 510(k): | K213386 |
No reference devices were used in this submission.
#### DEVICE DESCRIPTION IV.
HG Innovations Ltd.'s Electrical Surqical Instruments encompass a wide variety of instruments that are widely used in the medical industry. Electrosurgical forceps and electrodes have been used in surgery for many years. The single-use, bipolar forceps (various lengths and designs), with preattached cables are designed to grasp, manipulate and coagulate soft tissues and are intended for use by a physician familiar with electrosurgery in bipolar coagulation for general open surgery where coagulation of soft tissue is needed. The blood vessel or tissue is grasped between the forceps tines, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue. Bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation current. The bipolar forceps must be operated with the following parameters:
-Frequency range between 300 kHz-1,000 kHZ;
{5}------------------------------------------------
-Maximum generator operating voltage 600Vp.
#### INDICATIONS FOR USE V.
Single-use, bipolar forceps are intended for bipolar coagulation to remove tissue and control bleeding by use of high frequency electrical current.
The bipolar forceps must be operated with the following parameters:
-Frequency range between 300 kHz-1,000 kHZ;
-Maximum generator operating voltage 600Vp.
{6}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE & PERFORMANCE DATA
| Feature | Subject Device | Predicate Device | Justification For Differences |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| | Single-use Bipolar<br>Forceps with cable<br>(Multiple models with<br>different designs and<br>sizes) | Titan Manufacturing<br>Bipolar Forceps –<br><i>K213386</i><br>(Multiple models<br>with different<br>designs and sizes) | |
| Manufacturer | HG Innovations Ltd. | Titan Manufacturing, Inc. | -------- |
| Regulation | 878.4400 | 878.4400 | Same |
| Classification 21 | Class II | Class II | Same |
| Product Code | GEI | GEI | Same |
| Electrode Type | Bipolar | Bipolar | Same |
| Cable included | Yes | No | Different |
| Principles of Operation | The blood vessel or<br>tissue is grasped<br>between the forceps<br>tines or tips, each of<br>which acts as an<br>electrode, and current<br>passes to desiccate<br>and coagulate the<br>tissue. Bipolar<br>forceps are<br>connected through a<br>suitable bipolar cable<br>with the bipolar<br>output of an<br>electrosurgical<br>generator. Bipolar<br>forceps must only be<br>used with bipolar<br>coagulation current. | Bipolar forceps are<br>designed to grasp,<br>manipulate, and<br>coagulate selected<br>tissue. They are to<br>be connected<br>through a bipolar<br>cable with the<br>bipolar output of an<br>electrosurgical<br>generator.<br>Coagulation is<br>achieved using<br>electrosurgical<br>energy generated<br>by the<br>electrosurgical<br>generator unit<br>(ESU) and activated<br>by a footswitch.<br><br>Bipolar cables, ESU<br>and footswitch are<br>not part of the<br>subject<br>device. | The wording for<br>principles of<br>operation are<br>different, but the<br>principles of<br>operation are<br>equivalent. |
| Indications for use | Single-use, bipolar<br>forceps are intended for<br>bipolar coagulation to<br>remove tissue and<br>control bleeding by use | Titan Manufacturing<br>Bipolar Forceps are<br>intended to remove<br>tissue and control<br>bleeding by use of | Subject device<br>includes “bipolar<br>coagulation” which<br>is how bleeding is<br>controlled. The |
| | electrical current. The bipolar forceps must be operated with the following parameters: -Frequency range between 300 kHz-1,000 kHZ; -Maximum generator operating voltage 600Vp. | high frequency electrical current. | operating parameters were also added to the indications of the subject device. |
| Rx/ OTC | Rx | Rx | Same |
| Design | Straight | Straight | Same |
| | Jeweler | N/A | Similar to predicate (materials, components) |
| | Angled | Angled | Same |
| | Adson | N/A | Similar to predicate (materials, components) |
| | McPherson | N/A | Similar to predicate (materials, components) |
| | McPherson, Angled | N/A | Similar to predicate (materials, components) |
| | Bayonet | N/A | Similar to predicate (materials, components) |
| | N/A | Curved | N/A |
| | N/A | Coaptation | N/A |
| | Insulated | Insulated/noninsulated | Same |
| | N/A | Insulating/ Irrigating | N/A |
| | N/A | Non-insulating/Irrigating | N/A |
| Energy Source | Generator | Generator | Same |
| Single Use | Yes | Reusable | Disposable use of subject device eliminates risk of reuse. |
| Tip Sizes | 0.5-2mm | 0.2mm-2mm | Range of tip sizes falls within range of predicate device. |
| Lengths | 102-240mm | 89mm-273mm | Range of lengths falls within range of predicate device. |
| Component Materials | | | |
| Forceps' Tip(s) | Stainless Steel AISI 420 | Stainless Steel or titanium | Subject device is equivalent to the stainless steel |
| | | | model of the |
| | | | predicate. |
| Arm Material | Stainless Steel AISI<br>420 | Stainless Steel or<br>titanium | Subject device is<br>equivalent to the<br>stainless steel<br>model of the<br>predicate. |
| Adhesive | None | Epoxy | N/A. Subject device<br>legs are over-<br>molded to foot<br>instead of using<br>adhesives. |
| Outer Cap | Polypropylene,<br>SABIC® PP<br>107M90T + Master<br>Color, Black | N/A | Used for shipping<br>purposes only,<br>because subject<br>device is shipped<br>sterile. |
| Cable | Polyvinyl Chloride<br>(PVC), 2/16 core,<br>Copper Wires | Not included | Cable wires used in<br>subject device is<br>made of equivalent<br>materials to<br>predicate wires. |
| Solder | Tin | N/A | N/A |
| Powder Coating | Polyamide 11 | Polyvinylidene<br>Fluoride (PVDF) | Equivalent, both are<br>thermoplastics. |
| Colorant | Pigment Blue 15:3<br>UN8632 | Kynar (blue) | Both colorants are<br>non-patient<br>contacting. |
| Forging Blank | Stainless Steel AISI<br>420 | N/A | N/A |
| Inner Cap | Polypropylene,<br>SABIC® PP<br>107M90T + Master<br>Color, Black | N/A | N/A |
| Banana Pin | Chrome-plated, Gold-<br>plated spring, Brass | N/A | N/A |
| Internal Plug Body | Polypropylene,<br>SABIC® PP<br>107M90T + Master<br>Color, Black | N/A | N/A |
| Outer Plug Body | Polyvinyl Chloride<br>(PVC) | N/A | Both are made of<br>plastic |
| Sterilization Method &<br>SAL | Ethylene Oxide (EO),<br>provided sterile to<br>user<br>SAL= 10-6 | Provided non-sterile<br>to user for pre-<br>vacuum sterilization<br>(steam).<br>SAL= 10-6 | Equivalent. Both<br>meet requirement<br>for sterile products. |
| Shelf-Life Testing | Real-time aging study shows product shelf life up to 3 years | N/A | N/A |
| Packaging | Paper/film (Tyvek/Film) pouch | Bipolar forcep in a propionate clear hard plastic tube with foam insets at both ends packaged in an instrument sturdy box with packaging material to protect the instrument (bipolar forcep) | Equivalent. Both meet requirements for package integrity. |
| Performance Testing<br>Biocompatibility (ISO 10993-1) | | | |
| Cytotoxicity (ISO 10993-5) | Non-Cytotoxic | Non-Cytotoxic | Same |
| Irritation (ISO 10993-23) | Non-Irritating | Non-Irritating | Same |
| Sensitization (ISO 10993-10) | Non-Sensitizing | Non-Sensitizing | Same |
| Pyrogenicity (USP151) | <10 EU/ sample (Pass) | Pass | Same |
| Systemic Toxicity (ISO 10993-11) | Non-Toxic | Non-Toxic | Same |
| Selection of Tests for Interactions with Blood (ISO 10993-4) | N/A | Pass | Not required for intended use. |
| Electrical Safety IEC-60601-2-2 | | | |
| HiPot & Continuity Test Dielectric strength insulation; insulated safety plug; | Pass | Pass | Same |
| Compatible with general electrosurgical generators with 4mm outlets | Pass | Pass | Same |
| Drop Test | Pass | Pass | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## VII. CONCLUSION
HG Innovation's bipolar forceps are identical to the predicate device in their intended use/indications for use and construction material of the tips. The tip-size range of the subject device also completely overlaps with that of the predicate device. There is a difference in the specific arm/body coatings used between the subject device and the predicate device.
Based on the indications for use, comparison of technological characteristics with the predicate device, and results of performance testing, the subject device is substantially equivalent to the predicate for the proposed indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.