← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K223033

# Tixel® 2 System (K223033)

_Novoxel , Ltd. · GEI · Jun 21, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K223033

## Device Facts

- **Applicant:** Novoxel , Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jun 21, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Tixel® 2 System is intended for dermatological procedures requiring ablation and resurfacing of the skin, and for treatment of periorbital wrinkles.

## Device Story

Tixel 2 is a thermo-mechanical fractional skin treatment device. It uses a handpiece with a tip containing 81 or 24 biocompatible titanium pyramids on a gold-plated copper base, heated by a ceramic element. The clinician operates the device via a console to control pulse parameters, speed, and contact distance. The pyramids contact the skin to transfer heat via direct conduction, creating a matrix of coagulation sites and thermal necrosis. Used in clinical settings by trained professionals. The device features an upgraded GUI, dual handpiece connectivity, and an improved air blower. The output is a controlled thermal injury to the skin, intended to improve skin texture and reduce periorbital wrinkles. Benefits include effective skin resurfacing and wrinkle reduction with a demonstrated safety profile.

## Clinical Evidence

Prospective, single-arm, blinded study (n=51, 48 completed) evaluated periorbital wrinkle treatment. Primary endpoints included safety and efficacy (FWCS and GAIS scores). Results showed >1 grade improvement in FWCS. Mean GAIS score was 3.54 at first follow-up and 3.52 at 3-month follow-up. 83.3% of subjects were responders. Safety profile was favorable with no SAEs; 3.92% AE rate (erythema, unrelated back pain). VAS pain scores were low.

## Technological Characteristics

Thermo-mechanical fractional skin treatment device. Patient-contacting materials: biocompatible titanium pyramids on gold-plated copper base. Energy source: ceramic heating element. Connectivity: console supports up to two handpieces. Software: GUI-controlled parameters. Standards: ISO 20857 (disinfection), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Novoxel Tixel® System ([K202988](/device/K202988.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

June 21, 2023

Novoxel Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103

Re: K223033

Trade/Device Name: Tixel® 2 System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: Mav 25, 2023 Received: May 25, 2023

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.06.21
13:17:13 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name Tixel® 2 System

Indications for Use (Describe)

The Tixel® 2 System is intended for dermatological procedures requiring ablation and resurfacing of the skin, and for treatment of periorbital wrinkles.

Type of Use (Select one or both, as applicable)

× | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) SUMMARY

#### Novoxel's Tixel® 2 System

#### Submitter's Name, Address, Telephone Number and Contact Person

Novoxel, Ltd. 43 Hamelaha Street P.O.B 8539 Netanya, Israel 4250573 Phone: +972 9 7739074 Facsimile: +972 9 793 9953 Contact Person: Omri Findler, Dir. Clin. QA/RA

#### Date Prepared

June 19th, 2023

#### Name of Device

Tixel® 2 System

#### Common or Usual Name

Thermo-mechanical fractional skin treatment device

#### Classification Name

21 CFR 878.4400, Class II, Product Code: GEI

# Predicate Device

Novoxel Tixel® System (K202988)

#### Intended Use / Indications for Use

The Tixel® 2 System is intended for dermatological procedures requiring ablation and resurfacing of the skin, and for treatment of periorbital wrinkles.

#### Device Description

The Tixel® 2 System is a thermo-mechanical fractional skin treatment device that is designed to perform ablative fractional skin treatment is achieved by transfer of energy in the form of heat to the skin to create coagulation sites. The treatment is applied through an operating Tip that consists of 81 (standard Tip) or 24 (small Tip) biocompatible titanium, square pyramidal shapes assembled over a gold-plated copper base that are heated by an underlying flat ceramic heating element. The desired skin treatment is achieved by defining the speed and distance at which the Tip contacts and pushes the skin, and the number of pulses performed. Based on the

{4}------------------------------------------------

treatment parameters selected, the pyramids contact the skin surface in 81 (or 24 for the small Tip) discrete, non-overlapping areas and by the transfer of heat, a matrix of coagulation sites and thermal necrosis is generated.

Tixel 2 transfers heat to the tissue by direct conduction to target tissue in a localized manner via the discrete non-overlapping pyramids. The complete Tixel 2 System contains two Handpieces, which are attached to the Tixel 2 Console by an umbilical tube. The Console's touch screen is used to control the system parameters.

# Technoloqical Characteristics

The Tixel 2 enables connection of up to two Handpieces in comparison to the Tixel, which connects to only one. Both the Tixel 2 and its predicate device transfer energy through a Handpiece attached to a system console, where the output is controlled by the clinician operating the device to achieve the desired effect. The Tixel 2 and the predicate device are identical in all aspects. All patient contacting materials for both the Tixel 2 and the predicate device are identical in all aspects and biocompatible. The Tixel 2 and its predicate device deliver non-invasive fractional skin treatments through the exact same method of application (same Handpiece and same Tip design applied to the area with identical treatment parameters controlled through the console).

The following differences exist between the subject and predicate devices:

- Addition of periorbital use
- . Up to two Handpieces can be plugged to the Tixel 2 in comparison to one in Tixel
- Increase in air blower capacity .
- Tip and HP use-life extended .
- Upgrade to the Graphic User Interface (GUI)
- Change in external casing design ●
- Shorter disinfection time ●
- Addition of repeat pulse mode ●
- Improvement to displayed Tip temperature during cooling .

The differences between the devices do not result in different types of safety or effectiveness questions because the delivered treatment is unchanged. Both devices are intended to deliver a fractional thermal treatment with limited heat delivered to adjacent tissue. The energy transfer results in local heating of the skin to cause thermally induced tissue coagulation and ablation.

# Performance Data

The following tests were performed to establish equivalence:

- Disinfection validation per ISO 20857
- Use-life validation testing .
- . Software and cybersecurity validation per FDA guidance
- Electrical safety per IEC 60601-1 and 60601-1-6 0
- EMC per IEC 60601-1-2 .
- Comparative animal tissue histology studies (In-Vivo and Ex-Vivo)

{5}------------------------------------------------

# Clinical Data

A prospective, blinded (pre- and post-treatment), single-arm clinical study was performed in two clinics (US, Israel). The study is designed to evaluate the safety and efficacy of the Tixel 2 in treatment of periorbital wrinkles and was conducted with 51 patients. Forty-eight patients have completed the study. Analgesic materials were not applied; only forced air cooling was used.

The results support the safety of the device when used for periorbital wrinkle treatment. No SAEs or significant related AEs were noted. 2 AEs were reported in two (n=2) subjects (3.92%); Erythema (probably related, resolved within two days) and back pain (unrelated). In both events, medications were administered to treat the event.

FWCS score was performed by blinded raters for baseline and the three months follow up visits and demonstrated improvement > 1 grade. GAIS assessment was performed at the follow up visit compared to baseline. The mean score at Visit 5 (1st follow up) was 3.54±0.68. The mean score remained almost the same, 3.52±0.58 at Visit 6 (3-month FU). Most of the subjects achieved the highest improvement grade (grade 4, 75-100 %). At least 2 out of the 3 raters were in agreement for grading 83.3% of subjects as responders.

Mean procedure-associated VAS pain scores reported by the subjects at all treatment visits were low.

Conclusions: Patient comfort, satisfaction, safety, and the efficacy of the Tixel 2 in the treatment of periorbital wrinkles have been adequately demonstrated.

# Conclusions

The Tixel 2 has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Tixel 2 and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Tixel 2 is as safe and effective as the predicate device. Therefore, the Tixel 2 is substantially equivalent to its predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K223033](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K223033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
